Clean Dust of Light System - Chemray 240 Rayto
Clean Dust of Light System - Chemray 240 Rayto
(1) Power off the instrument. Remove back cover, side cover and front cover of this
instrument.
(2) Remove all screws which fix the work platform. Then remove the small cover
(3) Remove right part cover from work platform. Push sampling needle to reagent
(5) Push the sampling needle to the back word. When you push it, please do not let
(8) Do not remove the screws which fix the black cuvette holder, other wise the instrument can not
(9) Remove the lamp block. There are 3 screws fix the lamp block.
(10) Remove “Optical set” after remove the screw.
(11) Use ear washing bulb to blow two sides of all filters.
(12) Clean dust of the two lens. Do not remove black cuvette holder.
(13) After clean the light system, fix all the parts and covers.