You are on page 1of 4

Acta Pædiatrica ISSN 0803-5253

REGULAR ARTICLE

Accuracy of tympanic and forehead thermometers in private paediatric


practice
J Teller1*, M Ragazzi2*, GD Simonetti (giacomo.simonetti@insel.ch)2, SAG Lava2,3
1.Private Paediatric Practice, Langnau i.E., Switzerland
2.Division of Paediatric Nephrology, University Children’s Hospital Bern, University of Bern, Bern, Switzerland
3.Department of Paediatrics, Ospedale Regionale Bellinzona e Valli, Bellinzona and University of Bern, Bern, Switzerland

Keywords ABSTRACT
Children, Fever, Infrared forehead, Infrared tympanic, Aim: To compare infrared tympanic and infrared contact forehead thermometer
Thermometer
measurements with traditional rectal digital thermometers.
Correspondence
Methods: A total of 254 children (137 girls) aged one to 24 months (median 7 months)
GD Simonetti, MD, Division of Paediatric
Nephrology, University Children’s Hospital, consulting a private paediatric practice because of fever were prospectively recruited. Body
Inselspital and University of Bern, Freiburgstrasse temperature was measured using the three different devices.
3010 Bern, Switzerland.
Tel: +41 31 632 95 12 |
Results: The median and interquartile range for rectal, tympanic and forehead
Fax: +41 31 632 94 20 | thermometers were 37.6 (37.1–38.4)°C, 37.5 (37.0–38.1)°C and 37.5 (37.1–37.9)°C,
Email: giacomo.simonetti@insel.ch respectively (p < 0.01). The limits of agreement in the Bland-Altman plots were 0.73 to
Received +1.04°C for the tympanic thermometer and 1.18 to +1.64°C for the forehead
24 July 2013; revised 7 September 2013; thermometer. The specificity of both the tympanic and forehead thermometers for
accepted 10 October 2013.
detecting fever above 38°C was good, but sensitivity was low. Forehead measurements
DOI:10.1111/apa.12464 were susceptible to the use of a radiant warmer.
*These authors contributed equally to this work. Conclusion: Both the tympanic and forehead devices recorded lower temperatures than
the rectal thermometers. The limits of agreement were particularly wide for the forehead
thermometer and considerable for the tympanic thermometer. In the absence of valid
alternatives, because of the ease to use and little degree of discomfort, tympanic
thermometers can still be used with some reservations. Forehead thermometers should
not be used in paediatric practice.

INTRODUCTION study, performed more than 20 years ago (6), investigated


Fever is a frequent symptom in childhood and one of the tympanic thermometers in the setting of a private paediatric
most common reasons for taking a child to the doctor (1). practice. More recently, infrared cutaneous contact ther-
Ideally, temperature measurement techniques should be mometers that can assess forehead temperature have
safe, easy to perform, noninvasive, cost-effective and time- appeared on the market. So far, only a few published
efficient and should accurately reflect core body tempera- studies have examined the validity of these thermometers
ture (2,3). Traditionally, body temperature has typically and none have been performed in a private paediatric
been taken orally in older children and adults and rectally practice (2,3,7).
in infants and young children. Rectal measurement, the gold
standard in young children (3), is reliable and accurately
reflects the body’s core temperature. However, the proce- Key notes
dure is invasive and not well accepted among children and  Our study of 254 children presenting with fever in a
parents. Therefore, alternative measurement methods are private paediatric practice compared readings from
needed (2,4). infrared tympanic, infrared contact forehead and tradi-
In infants younger than 4 weeks, axillary measurement is tional rectal digital thermometers.
extremely accurate (2,4). In older children and adults,  We found that tympanic and forehead thermometers
however, this method, although easy to perform and recorded lower temperatures than digital rectal ther-
generally well tolerated, showed variable sensitivity. Fur- mometers, and their limits of agreement were partic-
thermore, although an axillary reading is generally 0.5°C ularly wide.
lower than a rectal reading, an exact conversion factor  The findings suggest that tympanic thermometers can
cannot be established (2). Tympanic measurements using be used with some reservations, but forehead ther-
infrared thermometers have also been repeatedly investi- mometers should not be used in paediatric practice.
gated, with conflicting results (5). Furthermore, only one

e80 ©2013 Foundation Acta Pædiatrica. Published by John Wiley & Sons Ltd 2014 103, pp. e80–e83
Teller et al. Accuracy of thermometers in paediatric practice

The aim of this noncommercially sponsored study was to was examined using Bland–Altman plots, which display the
compare infrared tympanic and infrared contact forehead mean differences and 95% limits of agreement (10). Statis-
thermometer measurements with those from traditional tical significance was assigned at p < 0.05.
rectal digital thermometers within a large sample of infants
and toddlers presenting at a private paediatric practice.
RESULTS
Our final sample included 254 children (117 boys and 137
PATIENTS AND METHODS girls) ranging in age from one to 24 months, with a median
In this prospective study, infants and toddlers between one of 7.0 (5.0–11.0) months.
and 24 months of age with no underlying chronic illness or Figure 1 shows the temperature readings from the three
known pathology and presenting to the private paediatric types of thermometers investigated. The medians and IQR
practice because of fever were eligible to participate. for the rectal, tympanic and forehead thermometers were
Participants were consecutively recruited between Novem- 37.6 (37.1–38.4)°C, 37.5 (37.0–38.1)°C and 37.5 (37.1–
ber 1, 2011 and April 2, 2012. 37.9)°C, respectively. The tympanic and forehead measure-
Age and gender were recorded for each patient, and ments were 0.1°C ( 0.1–0.4, p < 0.01) and 0.15°C ( 0.3–
temperature measurements were collected as part of the 0.7; p < 0.05) lower than the rectal measurements, respec-
routine clinical diagnostic procedure. After explaining the tively (Fig. 1). However, tympanic and forehead measure-
purpose of temperature measurement, informed consent ments were not significantly different from each other
was obtained from the parents. No data are available on (Fig. 1). Using Bland–Altman plots, the 95% limits of
those who declined to consent. After having created a quiet agreement were 0.73 to +1.04°C (Fig. 2, upper panel) for
atmosphere, during the physical examination, taking care of the tympanic measurements and 1.18 to +1.64°C (Fig. 2,
avoiding any haste, a licensed paediatrician on the project lower panel) for the forehead measurements.
(JT) took the three temperature measurements, each of While the specificity in detecting temperatures above
which was performed at room temperature (22–24°C). All 38°C was good, the sensitivity was low when using either
three measurements were taken within 3 min. the tympanic or the forehead thermometer (Table 1).
Each patient’s rectal temperature was measured using the Similarly, the positive predictive value was good (ranging
Microlife MT1961TM (Microlife, Widnau, Switzerland) con- from 88% to 95%), and the negative predictive value was
ventional digital thermometer. Infrared tympanic tempera- adequate (ranging from 71% to 84%, Table 1). Calculating
ture was measured using the ThermoScan 6022TM (Braun, the ROC curve, we found that the best threshold for
Kronberg, Germany) thermometer, and infrared skin fore- detecting rectal temperature ≥38°C was 37.55°C for both
head temperature was measured using the VoiceThermo the tympanic and the forehead thermometers. In particular,
wdc 6603 BTM (Wellness-design-company, Du € sseldorf, for a tympanic temperature of 37.55°C sensitivity was
Germany) thermometer. All the thermometers (always the 93.1% (95% CI: 86.4–97.2%) and specificity 83.5% (95% CI:
same device for each type of thermometer) were used 76.7–89.0%) with an area under the curve of 0.94 (95% CI:
strictly according to the manufacturers’ manuals.
The use of an infrared radiant warmer was sometimes
necessary. Its use was annotated, and its influence on
measurements was evaluated.
Power calculation was performed to determine sample
size. At least 198 children were needed to achieve a study
power of 80%, with a error = 0.05, estimating a potential
difference between the two methods of 0.2°C and a
standard deviation of 1°C (8). Normal distribution of
variables was tested using the D’Agostino–Pearson omnibus
test for normality. Not normally distributed data are
presented as median and interquartile range (IQR). Vari-
ables were compared by the nonparametric Friedman
analysis of variance test, and the Dunn’s post-test procedure
was used to compare the different measurements. The
Fisher exact test was used to compare dichotomous
variables; sensitivity, specificity, positive predictive value
and negative predictive value were calculated, using rectal
temperature of 38.0°C, as the generally used cut-off point
Figure 1 Box plot of temperatures measured with the 3 different devices. The
for rectally measured fever (3,9). Moreover, receiver oper-
boundary of the box closest to zero indicates the 25th percentile, the line within
ating characteristics (ROC) curve analysis was performed to the box marks the median, and the boundary of the box farthest from zero
establish the best threshold in determining the presence of indicates the 75th percentile. Whiskers (error bars) above and below the box
rectally measured fever (≥38°C) for tympanic and forehead indicate the 90th and 10th percentiles. (ns = not significant).
thermometers. Concordance among the different methods

©2013 Foundation Acta Pædiatrica. Published by John Wiley & Sons Ltd 2014 103, pp. e80–e83 e81
Accuracy of thermometers in paediatric practice Teller et al.

DISCUSSION
This study demonstrates that both tympanic and forehead
thermometers devices record lower temperatures than
rectal thermometers. This can frequently underestimate
the core temperature. Although both devices showed
excellent specificity and positive predictive values, sensitiv-
ity was low, and the negative predictive values were
suboptimal. With a lower threshold of 37.55°C for both
the tympanic and forehead thermometers, derived from a
ROC curve, sensitivity slightly improved, particularly for the
tympanic thermometer.
The forehead thermometer showed clear inconsistency,
with limits of agreement encompassing a range of almost
3°C. While the tympanic thermometer showed a slightly
better degree of agreement, its range of almost 2°C is
clinically inappropriate (11,12). These results are consistent
with previous data (13–15). More importantly, a well-
performed systematic review of studies comparing rectal
and tympanic thermometers found acceptable mean differ-
ences but similarly wide limits of agreement (5).
In addition, the present results indicate that forehead
measurements were susceptible to the use of a radiant
warmer, explaining at least in part their insufficient accu-
racy, as well as supporting the conclusion that these devices
are not appropriate for use in a private paediatric practice.
Figure 2 Bland–Altman plots, which display the difference between two
There are two innovative aspects of this study. First, the
measurements against their average. The data plotted should ideally scatter measurements were performed in the real-world setting of a
around a difference of zero across the range of means: the smaller the scatter is private paediatric practice using modern devices. Children
away from the mean (solid line), the greater the agreement. The scatter is presenting to a private paediatric practice represent the vast
quantified by calculating the limits of agreement (dotted lines), which are majority of feverish children cared for by paediatricians.
estimated by the mean difference and 2 standard deviations of the difference. Nonetheless, only one study, performed more than 20 years
Upper panel: Bland–Altman plot of tympanic measurements. The mean
ago, analysed tympanic thermometer measurements in this
difference is 0.15°C, and the 95% limits of agreement range from 0.73 to
+1.04°C. Lower panel: Bland–Altman plot of the forehead thermometer. The setting (6). Second, we assessed the influence of external
mean difference is 0.23°C, and the 95% limits of agreement range from 1.18 factors on recorded temperatures, such as the use of a
to +1.64°C. radiant warmer. Although the negative effect of radiant
heaters on temperature measurement might appear evident,
our data demonstrate that rectal and tympanic readings are
not affected by the presence of a radiant warmer and can
Table 1 Sensitivity, specificity, positive (PPV) and negative predictive values (NPV) therefore be rated as robust. On the other side, forehead
with 95% confidence interval for detecting rectal fever (≥38.0°C) with the tympanic thermometers are extremely dependent on external vari-
and infrared skin thermometers ables and are therefore even more unreliable as already
Thermometer Sensitivity Specificity PPV NPV shown by their wide limits of agreement.
Another significant strength of the present study is the
Tympanic 0.72 0.97 0.95 0.84
(0.62–0.80) (0.93–0.99) (0.87–0.99) (0.77–0.89)
inclusion of children with a febrile illness. In contrast, most
Infrared skin 0.42 0.96 0.88 0.71 thermometer studies have utilised a mixed sample of
(0.32–0.52) (0.92–0.99) (0.75–0.95) (0.64–0.77) healthy and sick patients, without a clear selection based
on the pretest probability of finding a feverish temperature.
Moreover, infrared cutaneous contact thermometers have
only been commercially available for a few years, and data
0.91–0.97, p < 0.0001). On the other side, for a forehead on their use in a private paediatric practice are lacking.
temperature of 37.55°C sensitivity was 80.4% (95% CI: The present study has some limitations. First, only one
71.4–87.6%) and specificity 73.0% (95% CI: 65.2–79.9%) model for each type of thermometer was analysed; thus, our
with an area under the curve of 0.83 (95% CI: 0.78–0.88, results cannot be generalised to other models. Second, only
p < 0.0001). children 24 months of age or less were included. Thus, no
A radiant warmer was used in 117 children, but not conclusions for older children can be drawn. Thirdly,
documented in 42 children. Subgroup analyses indicated that sensitivity, specificity, positive predictive value and negative
the radiant warmer only influenced forehead measurements. predictive value are computed from children presenting

e82 ©2013 Foundation Acta Pædiatrica. Published by John Wiley & Sons Ltd 2014 103, pp. e80–e83
Teller et al. Accuracy of thermometers in paediatric practice

with fever; therefore, these data should be interpreted with of the Italian Pediatric Society guidelines. Clin Ther 2009; 31:
respect to the study design. Fourthly, we did not evaluate 1826–43.
the reproducibility of measurements performed by each 3. Paes BF, Vermeulen K, Brohet RM, van der Ploeg T, de Winter
JP. Accuracy of tympanic and infrared skin thermometers in
type of thermometer. However, the high number of partic-
children. Arch Dis Child 2010; 95: 974–8.
ipants included in the study should have minimised the 4. Richardson M, Lakhanpaul M. Assessment and initial
impact of measurement imprecision. Finally, our results management of feverish illness in children younger than 5
might not be automatically generalised to inpatients or to years: summary of NICE guidance. BMJ 2007; 334: 1163–4.
children presenting to a paediatric emergency department. 5. Craig JV, Lancaster GA, Taylor S, Williamson PR, Smyth RL.
In conclusion, according to the results of the present Infrared ear thermometry compared with rectal thermometry in
children: a systematic review. Lancet 2002; 360: 603–9.
study, a reliable and valid alternative to invasive rectal
6. Freed GL, Fraley JK. Lack of agreement of tympanic membrane
temperature measurement is still lacking. Forehead ther- temperature assessments with conventional methods in a
mometers similar to that investigated in the present study private practice setting. Pediatrics 1992; 89: 384–6.
should not be used in private paediatric practice because of 7. De Curtis M, Calzolari F, Marciano A, Cardilli V, Barba G.
their wide limits of agreement along with their vulnerability Comparison between rectal and infrared skin temperature
to the influence of external factors such as a radiant in the newborn. Arch Dis Child Fetal Neonatal Ed 2008; 93:
warmer. In the absence of valid alternatives, and consider- F55–7.
8. Chiappini E, Sollai S, Longhi R, Morandini L, Laghi A, Osio
ing that tympanic thermometers are easy to use, create little
CE, et al. Performance of non-contact infrared thermometer
discomfort and are the preferred method of physicians and for detecting febrile children in hospital and ambulatory
nurses at several institutions, such thermometers can be settings. J Clin Nurs 2011; 20: 1311–8.
employed but with some reservations. Their lack of sensi- 9. Avner JR, Baker MD. Management of fever in infants and
tivity and negative predictive value, together with their wide children. Emerg Med Clin North Am 2002; 20: 49–67.
limits of agreement, render them suboptimal as a screening 10. Bland JM, Altman DG. Agreed statistics - measurement method
tool, in contrast to some recently proposed data (16). comparison. Anesthesiology 2011; 116: 182–5.
11. Duncan AL, Bell AJ, Chu K, Greenslade JH. Can a non-contact
According to other studies, axillary measurement might
infrared thermometer be used interchangeably with other
represent a better screening tool (15). This notion is also thermometers in an adult Emergency Department? Australas
supported by the recently published update of the Italian Emerg Nurs J 2008; 11: 130–4.
Guidelines on Fever Management (17). Finally, very 12. Rajee M, Sultana RV. NexTemp thermometer can be used
recently, a temporal artery thermometer was also shown interchangeably with tympanic or mercury thermometers for
to perform very well, with limits of agreement in a range of emergency department use. Emerg Med Australas 2006; 18:
245–51.
approximately 0.5°C (18).
13. Brennan DF, Falk JL, Rothrock SG, Kerr RB. Reliability of
Patient history and a careful clinical examination remain infrared tympanic thermometry in the detection of rectal fever
the most important criteria for the evaluation of a feverish in children. Ann Emerg Med 1995; 25: 21–30.
child. The clinical assessment and suspicion should guide 14. Fadzil FM, Choon D, Arumugam K. A comparative study on
diagnostic and therapeutic decisions, rather than tempera- the accuracy of noninvasive thermometers. Aust Fam
ture readings (19). As already suggested more than 40 years Physician 2010; 39: 237–9.
ago by Smith, it is the child that should be treated, not the 15. Shann F, Mackenzie A. Comparison of rectal, axillary, and
forehead temperatures. Arch Pediatr Adolesc Med 1996; 150:
thermometer (20).
74–8.
16. Edelu BO, Ojinnaka NC, Ikefuna AN. Fever detection in under
5 children in a tertiary health facility using the infrared tympanic
CONFLICT OF INTEREST thermometer in the oral mode. Ital J Pediatr 2011; 37: 8.
None. 17. Chiappini E, Venturini E, Principi N, Longhi R, Tovo PA,
Becherucci P, et al. Update of the 2009 Italian Pediatric
Society Guidelines about management of fever in children. Clin
Ther 2012; 34: 1648–53 e3.
ACKNOWLEDGEMENTS
18. Batra P, Goyal S. Comparison of rectal, axillary, tympanic, and
No support from any organisation for the submitted work. temporal artery thermometry in the pediatric emergency room.
Pediatr Emerg Care 2013; 29: 63–6.
19. Lava SA, Simonetti GD, Ferrarini A, Ramelli GP, Bianchetti
References MG. Regional differences in symptomatic fever management
1. Diethelm M, Largo RH. Prospektive Studie u € ber die Ta
€tigkeit among paediatricians in Switzerland: the results of a
des praktizierenden Kinderarztes – II. Besuchsgru€ nde und cross-sectional Web-based survey. Br J Clin Pharmacol 2013;
Diagnosen. Helv Paediatr Acta 1986; 41: 29–52. 75: 236–43.
2. Chiappini E, Principi N, Longhi R, Tovo PA, Becherucci P, 20. Smith DS. Fever and the pediatrician. J Pediatr 1970; 77:
Bonsignori F, et al. Management of fever in children: summary 935–6.

©2013 Foundation Acta Pædiatrica. Published by John Wiley & Sons Ltd 2014 103, pp. e80–e83 e83

You might also like