TURBIDITY (0 to 4400 FAU)
Method 8237
For water, wastewater, and seawater
Attenuated Radiation Method (direct reading)*
2, Rotate the
‘wavelength dial until the
small display shows:
860 am
1. Enter the stored
program number for
turbidity.
Press: 75 0 ENTER
‘The display will show:
1am to 860
When the correct
wavelength is dialed in,
the display will quickly
Note: The Powr-Thru show
Cell cannot be used with a
this procedure, ‘Zero Sample
then:
FAU TURBIDITY
0
6. Pour 25 mL of
sample into another
sample cell.
Immediately place this
sample cell into the cell
holder. Close the light
shield,
5, Press: ZERO
‘The display will show:
‘Leroing. ...
then:
0. FAU TURBIDITY
Note: The sample must
‘be well mixed before
transferring ito the
sample cell.
“Adapted from ISO Method 7027.
3. Pour 25 mL. of
deionized water (the
4, Place the blank into
the cell holder. Close the
blank) into a sample cell. light shield,
Note: For highly colored
samples, afitered portion
of the sample is used in
lace ofthe deionized
water
7. Press: READ
‘The display will show:
Reading...
then the result in
Formazin Attenuation
Units (FAU) will be
displayed.
853TURBIDITY, continued
‘Sampling and Storage
Collect samples in clean plastic or glass bottles. Analyze samples as soon
as possible after collection but can be stored 48 hours by cooling to 4 °C
(39 °F). Warm samples to room temperature before analyzing,
‘Accuracy Check
Standard Solution Method
Prepare a 200 FAU standard solution by pipetting 5.00 mL. of a 4000 NTU
Formazin stock solution into a 100-mL Class A Volumetric flask. Dilute
to the mark with deionized water. Prepare this solution daily
Precision
Ina single laboratory, using a standard solution of 1000 FAU with the
DR/2010, a single operator obtained a standard deviation of £49 FAU.
Summary of Method
This test measures turbidity, which is an optical property of the sample
that results from scattering and absorption of light by particles in the
sample. The amount of turbidity measured depends on the size, shape,
color and refractive properties of the particles.
Formazin standards are used for calibration and readings are taken using
Formazin Attenuation Units (EAU). A 4000 NTU Formazin stock
standard is also defined as 4000 FAU. The optical measurement method
for FAUs is very different than the NTU method. Color interference is
minimized by taking measurements at 860 nm.
‘This test cannot be used for USEPA reporting purposes, but it may
be used for daily in-plant monitoring and is best suited for measuring
turbidity levels greater than 20 NTU. A turbidimeter should be used for
accurately monitoring low levels of turbidity and for reporting purposes.TURBIDITY, continued
REQUIRED REAGENTS
Description
Water, deionized.
REQUIRED APPARATUS.
Sample Cells, 25 mL, matched pair
OPTIONAL REAGENTS
Formazin Stock Solution, 4000 NTU......
StabICal Solution, <0.1 NTU.
StabICal Solution, 20 NTU.
StablCal Solution, 200 NTU
StablCal Solution, 1000 NTU.......
OPTIONAL APPARATUS
Bottle, wash, 250 mL,
Flask, volumetric, Class A, 100 mL.
Flak, filter, 500 mL.
Filter Holder..
Filter Membrane, 47 mm, 0.45 microns..
Filter Pump, aspirator...
Pipet Filler, safety BUDD... cnn
Pipet, volumetric, Class A, 5.0 mL...
‘Stopper, rubber, one-hole,
‘Tubing, rubber, 5/16” 1
Tweezers, plastic...
14574-42
1546-49
Related Products
For the most precise turbidity measurements, the Hach Model
2100P Portable Turbidimeter, o Model 2100AN Turbidimeter are
recommended. These instruments are true nephelometers that meet all
the performance requirements for turbidity measurements described in
APHA Standard Methods, the EPA Methods of Chemical Analysis of
Water and Wastes and the Federal Register. For international use,
Hach recommends the 2100 AN IS, a true nephelomter that meets,
ISO 7027 requirements.
For technical support and ordering information, see Section V.