You are on page 1of 7

Hemodynamic Monitoring: Principles to Practice – M. L.

Cheatham, MD, FACS, FCCM

SUMMARY

• Fluid-
Fluid-filled catheters are commonly utilized in the
ICU to measure a variety of physiologic parameters
– Systemic blood pressures
– Pulmonary blood pressures
– Intra
Intra--abdominal pressure
HEMODYNAMIC MONITORING:
MONITORING
– Intracranial pressure
FROM PRINCIPLES TO PRACTICE
• To appropriately assess and utilize the data
provided by such monitoring devices, certain basic
Michael L. Cheatham, MD, FACS, FCCM concepts must be understood
Director, Surgical Intensive Care Units
Orlando Regional Medical Center
Orlando, Florida

INTRODUCTION INTRODUCTION

• Invasive pressure monitoring in the critically ill • To improve patient outcome, accurate measurement
provides valuable information of physiologic variables requires:
– Most accurate method for determining blood – Constant vigilance to ensure accuracy
pressure – Thorough understanding of monitoring
– Allows continuous p physiologic
y g monitoring g principles
p p and pitfalls
p
– Identifies physiologic abnormalities – Appropriate application of information gained
– Can be used to guide appropriate resuscitative
therapies
• WARNING: Physics lies up ahead! If you developed
– Waveform interpretation provides valuable PTSD following college physics, you may want to
information on the patient’s cardiac contractility skip the remainder of this lecture (relax, it’s actually
and heart valve competency pretty cool and explains why you needed to take
physics after all)

G-I-G-O MEASURING PRESSURE VARIABLES

• Each fluid
fluid--based pressure monitoring system has
• “Garbage In…Garbage Out”
the following components
– Erroneous physiologic measurements can result
– Intravascular catheter
in inappropriate patient therapy
– Connecting tubing and stopcocks Hydraulic
– You should always ask “Is this data valid?”
– Pressure transducer system
– Example
E l
– Continuous flush device
» Diebel et al. found that 52% of PAOP
measurements in surgical patients are – Amplifier
inaccurate and misleading as a result of – Oscilloscope / Digital display Electronic
monitoring artifacts – Processor system
» Reliance upon erroneous PAOP and CVP values – Recorder
to resuscitate critically ill patients may lead to
under--resuscitation and inappropriate therapy
under

Revised 01/13/2009 1
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM

MEASURING PRESSURE VARIABLES HYDRAULIC SYSTEM COMPONENTS

• The hydraulic system is much more subject to • Intravascular catheter


potential errors and artifacts than is the electronic – Tubing to access the desired blood vessel or
system compartment
– Learning to troubleshoot the hydraulic portion of a – Detects the pressure waves generated by cardiac
invasive pressure monitoring system is essential contraction
» Confirms the validity of the data
• Connecting tubing and stopcocks
» Helps avoid inappropriate therapeutic
– Transmits pressure wave from patient to
interventions based upon erroneous data
measuring device
– Allows control of blood vessel to avoid
hemorrhage, introduction of air, etc…

PRESSURE TRANSDUCER WHEATSTONE BRIDGE

• Converts the mechanical impulse of a pressure wave • An electrical circuit for


into an electrical signal through movement of a “strain comparison of resistances
gauge”
• Consists of a power source and
– Modern transducers utilize a displaceable silicon four resistors, three of which
sensing diaphragm and a “Wheatstone Bridge” have a known resistance
• To determine the unknown
Transducer Dome Sensing Diaphragm
resistance, the resistance of the
other three are balanced until the
current passing through both
sides of the parallel “bridge”
decreases to zero

Pressure Tubing Pressure Wave

WHEATSTONE BRIDGE HYDRAULIC SYSTEM COMPONENTS

• The Wheatstone Bridge is used to measure the • Continuous flush device


resistance change in a strain gauge – Flushes the tubing with fluid
– The resistance change is proportional to the at a rate of 1-
1-3 mL/hr and
changing physiologic variable (i.e. pressure) or helps prevent blood from
mechanical strain applied to the transducer clotting off the catheter
– “Zeroing” a transducer is simply determining the – The “fast flush” feature
value of the unknown resistance at rest increases the flow to 30
» “Balancing the bridge” mL/min and can be used to
– As the physiologic variable changes, the test the system’s compliance
resistance varies proportionally, a current is
induced across the bridge, and the voltmeter
value is converted to a pressure measurement

Revised 01/13/2009 2
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM

ELECTRONIC SYSTEM COMPONENTS MEASURING PRESSURE VARIABLES

• Amplifier
– Increases the low voltage signal from the pressure
transducer to a signal that can be displayed
– Most include electronic filters to remove unwanted SIGNAL
PRESSURE AMPLIFIER & SIGNAL ANALOG TO DIGITAL
TRANSDUCER CONDITIONING CONVERTER
physiologic
p y g “noise”
• Oscilloscope / Digital display
– Used to display waveforms and numerical data STRIP RECORDER BEDSIDE MONITOR MICROPROCESSOR

• Processor
– Used to calculate various hemodynamic parameters
• Recorder
– A printer, strip chart recorder, or other device

TYPICAL RADIAL ARTERY PRESSURE


MONITORING SYSTEM CLINICAL CORRELATION

• What do these measurements mean clinically?


– Systolic pressure
» The pressure exerted on the artery walls due to
left ventricular contraction (i.e., contractility)
– Diastolic pressure
» The pressure exerted during left ventricular
relaxation (i.e., vascular resistance)
– Pulse pressure
» The difference between peak systolic and
diastolic pressures (i.e., perfusion)

CLINICAL CORRELATION PHYSICS OF PRESSURE MONITORING

• What do these measurements mean clinically? • The typical catheter


catheter--transducer system in the ICU is
– Dicrotic notch pressure considered to be a “second-
“second-order dynamic system”
» The pressure that – The pressure waveform dampens over time
reflects aortic valve – Determined by two factors
closure » “Natural
Natural frequency”
frequency
– Heart rate – The number of oscillations per unit time that
» The frequency of occurs without any damping
contraction measured » “Damping coefficient”
in beats per minute – The time taken to dampen the waveform
– Ejection time
» The time of left
ventricular ejection

Revised 01/13/2009 3
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM

PHYSICS OF PRESSURE MONITORING PHYSICS OF PRESSURE MONITORING

• For example, when dropped onto a hard floor, a ball • However, if the same ball is dropped onto soft earth,
bounces several times before coming to rest the ball will not bounce as high, resulting in a
– With each successive bounce, it does not rise as decreased frequency, and will come to rest sooner,
high as on the previous bounce reflecting an increased damping coefficient
– Each bounce has a characteristic frequency
q y ((the – This can be expressed mathematically BUT you
number of oscillations per unit time) and don’t need to know how to calculate it ☺
damping coefficient (time that it takes the ball to P2 1
come to a rest) =
P1 f2 f
+ 2 jξ +1
Frequency
− Fn2 Fn
P1, P2 are output and input signals of the pressure transducer
respectively, f is an arbitrary frequency, Fn is the natural frequency,
Damping Coefficient ξ is the damping coefficient, and j is the complex number

DETERMINING THE NATURAL FREQUENCY


PHYSICS OF PRESSURE MONITORING AND DAMPING COEFFICIENT

• The accuracy of a second


second--order system is subject to
• This can also be expressed mathematically BUT
three mechanical factors
you don’t need to know how to calculate it ☺
1. Compliance
• The stiffness of the fluid
fluid--filled system (tubing)
2. Fluid inertia 1 R C
• The pressure required to move fluid (blood) Fn = ξ=
through the system 2π IC 2 I
3. Fluid resistance
• The viscosity of the fluid moving through the
system (resistance due to friction) Fn is the natural frequency, ξ is the damping coefficient,
C = compliance, I = inertia, and R = resistance

PHYSICS OF PRESSURE MONITORING PHYSICS OF PRESSURE MONITORING

• The complex pressure wave generated with each • If the waveform is reflected off a transducer or other
beat of the heart is not unlike the bouncing ball obstruction within the catheter
catheter--tubing system, a
– A pressure waveform is propagated at a given wave reflection or “harmonic” is generated
frequency (beats per minute) – These harmonic waveforms are additive and can
– The vascular resistance acts as the damping p g introduce error into pressure measurements
coefficient and diminishes the waveform’s
energy and magnitude over time
• The resulting arterial sine wave, occurring at the
rate of the patient’s pulse, is called the first
harmonic or fundamental

Revised 01/13/2009 4
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM

PHYSICS OF PRESSURE MONITORING UNDERDAMPED WAVEFORM

• Without some degree of damping within a system, • Note the characteristic


pressure waves reverberate within the catheter and narrow, peaked waveform
tubing leading to the formation of harmonics and – Overestimates systolic
overestimation of true blood pressure and underestimates
– An “underdamped
“underdamped”” system diastolic blood pressure
• With too much damping, the frictional forces impede – Mean arterial pressure
the arterial waveform such that it loses energy remains unchanged!
leading to underestimation of blood pressure • Causes
– An “overdamped
“overdamped”” system – Long stiff tubing,
increased vascular
resistance

OVERDAMPED WAVEFORM OPTIMAL DAMPING

• Note the characteristic widened


• Some damping is essential
and slurred waveform
to avoid harmonics
– Underestimates systolic
– The “optimal” amount of
and overestimates diastolic
damping is crucial to
blood pressure
accurate measurement
– Mean
M arterial
i l pressure of physiologic pressures
remains unchanged!
• A catheter-
catheter-transducer
• Causes system accurately measures
– Air bubbles, compliant pressure only if its natural
tubing, catheter kinks, frequency and damping
blood clots / fibrin, coefficient are appropriate
stopcocks, no fluid or low
flush bag pressure

FREQUENCY RESPONSE DYNAMIC RESPONSE ARTIFACTS

• A pressure monitoring system should be able to • Underdamped and overdamped waveforms are
detect changes quickly (known as the “frequency encountered in the ICU on a daily basis
response”) – Look for them at the bedside!
• Damping will tend to decrease frequency response • The ability to recognize when these potential
– Important if changes are occurring rapidly such sources of error or “dynamic
dynamic response artifacts
artifacts” are
as with tachycardia or a hyperdynamic heart present is essential to the…
• The ideal monitoring system has a high “natural” or – Appropriate use of hemodynamic measurements
“undamped
undamped” ” frequency – Prevention of inappropriate therapy based upon
– The frequency that would occur in the absence of erroneous data
any frictional forces or damping
– Allows accurate measurement of fast heart rates

Revised 01/13/2009 5
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM

DYNAMIC RESPONSE ARTIFACTS TROUBLESHOOTING

• Because dynamic response artifacts are commonly • OK, now let’s get practical…
encountered during patient monitoring, titration of
• The simpler the pressure monitoring system, the
medications should ALWAYS be based upon mean
higher its fidelity, the less it is subject to dynamic
arterial pressure (MAP)
response artifacts, and the less likely it will be to
– This variable is less subject to measurement produce erroneous data
p
error due to under
under-- or overdamping
• There are a number of steps that should be
• Systolic and diastolic blood pressure should NOT undertaken whenever setting up or troubleshooting
be used to titrate therapy! a catheter
catheter--transducer system

TROUBLESHOOTING TROUBLESHOOTING

• Remove multiple stopcocks, multiple injection ports, • Remove all air bubbles from the system
and long lengths of tubing whenever possible – Perhaps the single most important step in
– The optimal length of pressure tubing is 4 feet optimizing dynamic response
» Longer lengths of tubing promote harmonic » Air acts as a “shock absorber”
amplification
p and underdamping
p g – Bubbles as small as 1 mm in diameter can
– Ensure that arterial pressure tubing is being used cause substantial waveform distortion
» Overly compliant tubing leads to overdamping » Leads to overdamping and flattened waveforms
– Avoid large diameter tubing – Ensure that all connections are tight and
» Prevents achievement of optimal damping periodically flush all tubing and stopcocks to
remove air bubbles

TROUBLESHOOTING TROUBLESHOOTING

• Whenever you are evaluating a patient’s changing • Zero the transducer


hemodynamics – The accuracy of invasive pressure measurements
– Check all transducers, stopcocks, tubing, and is dependent upon a proper reference point
injection ports for air » The “midaxillary line” or “phlebostatic axis” is
– Gentlyy tap
p the tubing
g and stopcocks
p as the commonlyy utilized
continuous flush valve is opened to dislodge any – Each transducer should be zeroed at least once
bubbles each day and whenever data is considered suspect
» This will usually clear the system and restore
measurement accuracy
» Flushing a few small bubbles through the
catheter is OK; if more air is present, aspirate it
from the tubing

Revised 01/13/2009 6
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM

TROUBLESHOOTING TROUBLESHOOTING

– Transducers may be attached to the patient or to a • The “fast


“fast--flush” or “square wave” test
pole at the head of the bed – Performed by opening the continuous flush valve
– Changes in bed positioning generally require re-
re- for several seconds creating a square wave
zeroing the pressure transducer » A system with appropriate dynamic response
» If the transducer is below the p
phlebostatic axis, characteristics will return to the baseline
the resulting arterial pressure will be waveform within one to two oscillations
erroneously high » If dynamic response characteristics are
» If the transducer is above the phlebostatic axis, inadequate, troubleshoot the system until
the resulting arterial pressure will be acceptable dynamic response is achieved
erroneously low

CONCLUSIONS CONCLUSIONS

• Direct pressure monitoring is essential for • All hemodynamic data should be considered
determining immediate changes in blood pressure erroneous until you are satisfied that the dynamic
response characteristics of the monitoring system
• The arterial waveforms provide valuable diagnostic
are appropriate
and treatment information
• Learn to troubleshoot each monitoringg system
y so
• If not accurate,
acc rate the erroneo
erroneous
s data within
ithin these
that you can ensure the accuracy of your patient’s
waveforms can potentially lead to detrimental
data
treatment
• Physics is important after all ☺
• Look at both the bedside monitor waveforms AND
the numerical data
– The waveforms can tell you a great deal

Revised 01/13/2009 7

You might also like