Professional Documents
Culture Documents
SUMMARY
• Fluid-
Fluid-filled catheters are commonly utilized in the
ICU to measure a variety of physiologic parameters
– Systemic blood pressures
– Pulmonary blood pressures
– Intra
Intra--abdominal pressure
HEMODYNAMIC MONITORING:
MONITORING
– Intracranial pressure
FROM PRINCIPLES TO PRACTICE
• To appropriately assess and utilize the data
provided by such monitoring devices, certain basic
Michael L. Cheatham, MD, FACS, FCCM concepts must be understood
Director, Surgical Intensive Care Units
Orlando Regional Medical Center
Orlando, Florida
INTRODUCTION INTRODUCTION
• Invasive pressure monitoring in the critically ill • To improve patient outcome, accurate measurement
provides valuable information of physiologic variables requires:
– Most accurate method for determining blood – Constant vigilance to ensure accuracy
pressure – Thorough understanding of monitoring
– Allows continuous p physiologic
y g monitoring g principles
p p and pitfalls
p
– Identifies physiologic abnormalities – Appropriate application of information gained
– Can be used to guide appropriate resuscitative
therapies
• WARNING: Physics lies up ahead! If you developed
– Waveform interpretation provides valuable PTSD following college physics, you may want to
information on the patient’s cardiac contractility skip the remainder of this lecture (relax, it’s actually
and heart valve competency pretty cool and explains why you needed to take
physics after all)
• Each fluid
fluid--based pressure monitoring system has
• “Garbage In…Garbage Out”
the following components
– Erroneous physiologic measurements can result
– Intravascular catheter
in inappropriate patient therapy
– Connecting tubing and stopcocks Hydraulic
– You should always ask “Is this data valid?”
– Pressure transducer system
– Example
E l
– Continuous flush device
» Diebel et al. found that 52% of PAOP
measurements in surgical patients are – Amplifier
inaccurate and misleading as a result of – Oscilloscope / Digital display Electronic
monitoring artifacts – Processor system
» Reliance upon erroneous PAOP and CVP values – Recorder
to resuscitate critically ill patients may lead to
under--resuscitation and inappropriate therapy
under
Revised 01/13/2009 1
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM
Revised 01/13/2009 2
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM
• Amplifier
– Increases the low voltage signal from the pressure
transducer to a signal that can be displayed
– Most include electronic filters to remove unwanted SIGNAL
PRESSURE AMPLIFIER & SIGNAL ANALOG TO DIGITAL
TRANSDUCER CONDITIONING CONVERTER
physiologic
p y g “noise”
• Oscilloscope / Digital display
– Used to display waveforms and numerical data STRIP RECORDER BEDSIDE MONITOR MICROPROCESSOR
• Processor
– Used to calculate various hemodynamic parameters
• Recorder
– A printer, strip chart recorder, or other device
Revised 01/13/2009 3
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM
• For example, when dropped onto a hard floor, a ball • However, if the same ball is dropped onto soft earth,
bounces several times before coming to rest the ball will not bounce as high, resulting in a
– With each successive bounce, it does not rise as decreased frequency, and will come to rest sooner,
high as on the previous bounce reflecting an increased damping coefficient
– Each bounce has a characteristic frequency
q y ((the – This can be expressed mathematically BUT you
number of oscillations per unit time) and don’t need to know how to calculate it ☺
damping coefficient (time that it takes the ball to P2 1
come to a rest) =
P1 f2 f
+ 2 jξ +1
Frequency
− Fn2 Fn
P1, P2 are output and input signals of the pressure transducer
respectively, f is an arbitrary frequency, Fn is the natural frequency,
Damping Coefficient ξ is the damping coefficient, and j is the complex number
• The complex pressure wave generated with each • If the waveform is reflected off a transducer or other
beat of the heart is not unlike the bouncing ball obstruction within the catheter
catheter--tubing system, a
– A pressure waveform is propagated at a given wave reflection or “harmonic” is generated
frequency (beats per minute) – These harmonic waveforms are additive and can
– The vascular resistance acts as the damping p g introduce error into pressure measurements
coefficient and diminishes the waveform’s
energy and magnitude over time
• The resulting arterial sine wave, occurring at the
rate of the patient’s pulse, is called the first
harmonic or fundamental
Revised 01/13/2009 4
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM
• A pressure monitoring system should be able to • Underdamped and overdamped waveforms are
detect changes quickly (known as the “frequency encountered in the ICU on a daily basis
response”) – Look for them at the bedside!
• Damping will tend to decrease frequency response • The ability to recognize when these potential
– Important if changes are occurring rapidly such sources of error or “dynamic
dynamic response artifacts
artifacts” are
as with tachycardia or a hyperdynamic heart present is essential to the…
• The ideal monitoring system has a high “natural” or – Appropriate use of hemodynamic measurements
“undamped
undamped” ” frequency – Prevention of inappropriate therapy based upon
– The frequency that would occur in the absence of erroneous data
any frictional forces or damping
– Allows accurate measurement of fast heart rates
Revised 01/13/2009 5
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM
• Because dynamic response artifacts are commonly • OK, now let’s get practical…
encountered during patient monitoring, titration of
• The simpler the pressure monitoring system, the
medications should ALWAYS be based upon mean
higher its fidelity, the less it is subject to dynamic
arterial pressure (MAP)
response artifacts, and the less likely it will be to
– This variable is less subject to measurement produce erroneous data
p
error due to under
under-- or overdamping
• There are a number of steps that should be
• Systolic and diastolic blood pressure should NOT undertaken whenever setting up or troubleshooting
be used to titrate therapy! a catheter
catheter--transducer system
TROUBLESHOOTING TROUBLESHOOTING
• Remove multiple stopcocks, multiple injection ports, • Remove all air bubbles from the system
and long lengths of tubing whenever possible – Perhaps the single most important step in
– The optimal length of pressure tubing is 4 feet optimizing dynamic response
» Longer lengths of tubing promote harmonic » Air acts as a “shock absorber”
amplification
p and underdamping
p g – Bubbles as small as 1 mm in diameter can
– Ensure that arterial pressure tubing is being used cause substantial waveform distortion
» Overly compliant tubing leads to overdamping » Leads to overdamping and flattened waveforms
– Avoid large diameter tubing – Ensure that all connections are tight and
» Prevents achievement of optimal damping periodically flush all tubing and stopcocks to
remove air bubbles
TROUBLESHOOTING TROUBLESHOOTING
Revised 01/13/2009 6
Hemodynamic Monitoring: Principles to Practice – M. L. Cheatham, MD, FACS, FCCM
TROUBLESHOOTING TROUBLESHOOTING
CONCLUSIONS CONCLUSIONS
• Direct pressure monitoring is essential for • All hemodynamic data should be considered
determining immediate changes in blood pressure erroneous until you are satisfied that the dynamic
response characteristics of the monitoring system
• The arterial waveforms provide valuable diagnostic
are appropriate
and treatment information
• Learn to troubleshoot each monitoringg system
y so
• If not accurate,
acc rate the erroneo
erroneous
s data within
ithin these
that you can ensure the accuracy of your patient’s
waveforms can potentially lead to detrimental
data
treatment
• Physics is important after all ☺
• Look at both the bedside monitor waveforms AND
the numerical data
– The waveforms can tell you a great deal
Revised 01/13/2009 7