Professional Documents
Culture Documents
Ident-Nr. 48 001
This Issue 06, 11/2007 applies to the following software versions:
1 Introduction ........................................ 1-1 1.3 The USM 35X family ................................. 1-8
The different instrument versions ................. 1-8
1.1 Safety information .................................... 1-2
Special features ........................................... 1-9
Batteries ...................................................... 1-2
Software ...................................................... 1-2 1.4 How to use this manual ......................... 1-10
Defects/errors and exceptional stresses ..... 1-3 1.5 Layout and presentation in this
manual ..................................................... 1-11
1.2 Important information on
ultrasonic testing ...................................... 1-3 Attention and Note symbols ...................... 1-11
Preconditions for testing with ultrasonic Listings ..................................................... 1-11
test equipment ............................................ 1-3 Operating steps ......................................... 1-11
Operator training .......................................... 1-4
Technical test requirements ........................ 1-4 2 Standard package and
Limits of testing ........................................... 1-5 accessories ........................................ 2-1
Ultrasonic wall thickness measurement ...... 1-5
2.1 Standard package .................................... 2-3
Effect of the test object’s material ............... 1-5
Effect of temperature variations ................... 1-6 2.2 Recommended accessories ..................... 2-5
Measurement of remaining wall thickness .... 1-6
Ultrasonic evaluation of flaws ...................... 1-6
Flaw boundary method ................................ 1-6
Echo display comparison method ............... 1-7
3 Initial start-up ...................................... 3-1 4.3 Keys and rotary knobs .............................. 4-6
Function keys ............................................. 4-6
3.1 Power supply ............................................ 3-2
On/Off key ................................................... 4-6
Operation using the power supply unit ........ 3-2
Special keys ................................................ 4-7
Operation using batteries ............................ 3-3
Rotary knobs ............................................... 4-8
Charging the batteries ................................. 3-5
4.4 Operational concept ................................. 4-8
3.2 Connecting a probe .................................. 3-7
Setting the functions ................................... 4-9
3.3 Starting the USM 35X ............................... 3-8 Alternative operation without rotary
Switching on ................................................ 3-8 knobs ........................................................ 4-10
Basic initialization ....................................... 3-8 4.5 Important basic settings ......................... 4-12
Information lines in the startup screen ......... 3-8 Selecting the language .............................. 4-12
Selecting units .......................................... 4-13
4 Principles of operation ...................... 4-1 Setting the date ......................................... 4-14
4.1 Operator’s controls ................................... 4-2 Setting the time ......................................... 4-15
4.2 Screen display .......................................... 4-3 4.6 Basic settings of the display .................. 4-16
Functions on the display ............................. 4-4 Selecting the color scheme ....................... 4-16
Other displays ............................................. 4-5 Setting the lighting .................................... 4-16
5.7 Calibrating the USM 35X ........................ 5-17 5.11 Calculation of flaw position
Calibrating the display range ..................... 5-17 (function group TRIG) ............................. 5-31
Choosing the measuring point ................... 5-17 ANGLE (Angle of incidence) ...................... 5-32
Calibration with straight- and X-VALUE (X-value of the probe) ................. 5-32
angle-beam probes .................................... 5-18 COLOR ..................................................... 5-33
Calibration with dual-element (TR) THICKNE (Material thickness) ................... 5-33
probes ....................................................... 5-21 DIAMET (Outside diameter of the test
object) ....................................................... 5-33
5.8 Measuring ................................................ 5-23
General notes ............................................ 5-23 5.12 Data saving
(function group MEM) ............................. 5-34
5.9 Measurement of dB difference
Storing a data set ...................................... 5-35
(function group REF) ............................... 5-25
Deleting a data set .................................... 5-35
Recording a reference echo ....................... 5-26
Deleting all data set .................................. 5-36
Deleting a reference echo .......................... 5-26
Recalling a stored data set ....................... 5-36
Echo comparison ...................................... 5-27
5.13 Dataset management
5.10 Classification of welds
(function group DATA) ............................ 5-38
(function group AWS) ............................. 5-28
TESTINF (Storing additional information) .... 5-39
Rating of welds according to AWS ............ 5-28
PREVIEW (Dataset preview) ..................... 5-41
DIR (Dataset directory) .............................. 5-42
SETTING (Function list) ............................ 5-42
5.14 Configuring the USM 35X for a test TIME/DATE (Setting the time and date) .... 5-56
application .............................................. 5-43 ANAMOD .................................................. 5-57
TOF (Selecting the measuring point) ......... 5-44 HORN ........................................................ 5-58
S-DISP (Zoomed display of reading) .......... 5-46 EVAMOD (Echo evaluation) ...................... 5-58
MAGNIFY (Gate spreading) ...................... 5-48
5.16 Other functions with special keys ......... 5-59
A-Scan (Setting the A-scan) ...................... 5-48
Freeze ....................................................... 5-59
Configuring the measurement line ............. 5-49
Zooming the echo display ......................... 5-59
Setting the display .................................... 5-50
FILLED (Echo display mode) .................... 5-51 The key ................................................ 5-59
VGA .......................................................... 5-51 5.17 Status symbols and LEDs ....................... 5-60
SCHEME .................................................. 5-51 Status symbols ......................................... 5-60
LIGHT (LCD backlight) ............................... 5-52 LEDs ......................................................... 5-60
SCALE (Configuring the measurement
line) ........................................................... 5-52 5.18 Distance-amplitude curve
(only USM 35X DAC and USM 35S) ....... 5-61
5.15 General configuration ............................ 5-53 DACMOD (Activating DAC/TCG) ............... 5-62
DIALOG (Selecting the language) ............. 5-53 DACECHO (Recording reference curve) .... 5-63
UNIT (Selecting units of measurement) ..... 5-54 T-CORR (Sensitivity correction) ................. 5-64
BAUD-R (Baud rate for transmission) ........ 5-55 OFFSET (Distance of multiple DAC) ......... 5-65
PRINTER (Printer for test report) ............... 5-55 Echo evaluation with DAC ......................... 5-65
COPYMOD (Assignment of the key) ... 5-55
that the required functions operate perfectly in the in- 1.2 Important information on
tended combination.
ultrasonic testing
If you have any questions about the use of your test
equipment, please contact your nearest representative Please read the following information before using your
of GE Inspection Technologies. USM 35X. It is important that you understand and ob-
serve this information to avoid any operator errors that
Defects/errors and exceptional stresses might lead to false test results. This could result in per-
sonal injuries or damages to property.
If you have reason to believe that a safe operation of
your USM 35X is no longer possible, you have to dis-
connect the instrument and secure it against uninten- Preconditions for testing with ultrasonic
tional reconnection. Remove the batteries if necessary. test equipment
A safe operation is e.g. no longer possible This operating manual contains essential information on
how to operate your test equipment. In addition, there
• if the instrument shows visible damages, are a number of factors which affect the test results. A
• if the instrument no longer operates perfectly, description of these factors would go beyond the scope
of an operating manual. The following list therefore only
• after prolonged storage under adverse conditions mentions the three most important conditions for a safe
(e.g. exceptional temperatures and/or especially high and reliable ultrasonic inspection:
air humidity, or corrosive environmental conditions),
• the operator training
• after being subjected to heavy stresses during trans-
portation. • the knowledge of special technical test requirements
and limits
• the choice of appropriate test equipment.
A proper training comprises for example adequate • the definition of the scope of inspection
knowledge of: • the choice of the appropriate test method
• the theory of sound propagation • the consideration of material properties
• the effects of sound velocity in the test material • the determination of limits for recording and
• the behavior of the sound wave at interfaces be- evaluation.
tween different materials It is the task of those with overall responsibility for test-
• the propagation of the sound beam ing to ensure that the inspector is fully informed about
these requirements. The best basis for such information
• the influence of sound attenuation in the test object is experience with identical test objects. It is also es-
and the influence of surface quality of the test ob- sential that the relevant test specifications be clearly
ject. and completely understood by the inspector.
Lack of such knowledge could lead to false test results
with unforeseeable consequences. You can contact for GE Inspection Technologies regularly holds specialized
example NDT societies or organizations in your country training courses in the field of ultrasonic testing. The
(DGZfP in Germany; ASNT in the USA), or also GE scheduled dates for these courses will be given to you
Inspection Technologies, for information concerning on request.
existing possibilities for the training of ultrasonic in-
spectors as well as on the qualifications and certifi-
cates that can finally be obtained.
Echo display comparison method The ultrasonic wave is attenuated in any material. This
sound attenuation is very low, e.g. in parts made of
The echo from a small, natural flaw is usually smaller fine-grained steel, likewise in many small parts made of
than the echo from an artificial comparison flaw, e.g. other materials. However, if the sound wave travels
circular disc flaw of the same size. This is due, for in- larger distances through the material, a high cumulative
stance, to the roughness of the surface of a natural sound attenuation can result even with small attenua-
flaw, or to the fact that the beam does not impinge on it tion coefficients. There is then a danger that echoes
at right angles. from natural flaws appear too small. For this reason, an
If this fact is not taken into account when evaluating estimate must always be made of the effects of attenu-
natural flaws, there is a danger of underestimating their ation on the evaluation result and taken into account if
magnitude. applicable.
In the case of very jagged or fissured flaws, e.g. shrink If the test object has a rough surface, part of the inci-
holes in castings, it may be that the sound scattering dent sound energy will be scattered at its surface and
occurring at the boundary surface of the flaw is so is not available for the test. The larger this initial scat-
strong that no echo at all is produced. In such cases, a tering, the smaller the flaw echoes appear, and the
different evaluation method should be chosen, e.g. use more errors occur in the evaluation result.
of the backwall echo attenuation in the evaluation. It is therefore important to take the effect of the test
The distance sensitivity of the flaw echo plays an im- object’s surfaces on the height of the echo into account
portant part when testing large components. Attention (transfer correction).
should be paid here to choosing artificial comparison
flaws which are as far as possible governed by the
same ”distance laws” as the natural flaws to be evalu-
ated.
H Note:
Note contains e.g. references to other chapters or spe-
cial recommendations for a function.
It describes
• accessories included in the standard package,
• recommended accessories.
or
or
You can connect the USM 35X to the mains supply – Push the Lemo plug of the power supply unit into the
system even if it carries batteries. A discharged battery plug receptacle until it snaps into place with a clearly
is charged in this case, viz. parallel to the instrument audible click.
operation.
– When pulling off the Lemo plug, withdraw the metal – Lift the lid off upward. To the right in the open battery
bushing on the plug first in order to release the lock. compartment, you will see two springs and several
connection pins.
The power supply unit is automatically set to any nomi-
nal voltage between 90 VAC and 240 VAC.
– Press the button PUSH at the front side of the battery. Charging the batteries
Four LEDs indicate the charge level of battery.
You can charge the lithium-ion battery either directly in
the instrument or by means of an external battery
H Note: charger. You always need an external battery charger to
You can also check the battery charge even if the battery charge standard C-cells.
is located in the battery compartment of the instrument.
Internal charging
Analog battery charge indicator Requirement:
The USM 35X is equipped with an analog battery charge • Lithium-ion battery, order number 102 208
indicator which enables you to estimate the remaining
• Power supply/charger unit, order number 102 163
battery life. In the measurement line to the very right a
battery symbol displays the corresponding charging If a battery is located in the instrument, the charging
level.The battery symbol indicates the charge in steps process is started automatically when you connect the
of 10 %. plug-in power supply unit. You can carry out ultrasonic
inspections and charge a battery at the same time.
H Note: The charging time is 10 hours with a simultaneous ultra-
If the battery charge indicator displays a low battery sonic inspection. If the instrument is not being used for
charge, you should urgently close your test job and ultrasonic inspections, the charging time is 8 hours. This
exchange the battery. You should take a second battery charging time applies to ambient temperatures from 25 to
or replacement batteries with you if you cannot connect 30 °C. Please take into consideration that the batteries are
the instrument to mains. not charged to their full capacity at higher temperatures.
External charging
Lithium-ion batteries can be charged by means of an
external battery charger. We recommend the battery
charger with the order number 35 297. To charge single
NiCad or NiMH cells, you need the external desktop
battery charger with the order number 101 729.
Switching on A Attention:
To start the USM 35X, press the switch-on key . All saved data are deleted.
The start display of the USM 35X appears; here you will – Press the and the key simultaneously.
also see the current software version of the instrument.
The instrument is reset to factory setting (cold start).
The instrument carries out a self-check and then
switches over to stand-by mode.
The settings of all function values and the basic set- In the start-up screen the message “Basic Initialization”
tings (language and units) are the same as before appears, and the unit will bei initialized and reset to its
switching-on of the instrument. basic setup (dialog language: English, for more details
on how to select the language, please refer to chapter 4).
Basic initialization
In case the instrument can no longer be operated, or Information lines in the startup screen
you need to make a basic initialization (factory setting), You can enter two lines (each with up to 39 characters) for
you have two options. information purposes in the startup screen. For this use
the remote function (codes I1 and I2, refer to chapter 8).
To keep the saved data:
H Note:
The screen display always shows the gain and the ad-
justed dB step value. All other functions are locked in
zoom mode.
Other displays
The measurement line below the screen display shows
values of settings, measured values, and status indica-
tions. As an alternative, a scale can be shown here,
giving an overview of the echo positions.
H Note:
Every measurement value can also be shown in an en-
larged display at the top right corner of the A-scan (set-
ting in the function group MEAS, function S-DISP).
Example of a measurement
line
USM 35X Issue 06, 11/2007 4-5
Principles of operation Keys and rotary knobs
Function keys
For changing between operation levels (below),
On/Off key
For turning the device on or off.
Special keys
To directly activate individual instrument functions:
H Note:
As an alternative, you can operate the USM 35X via a Second operating level
special menu line without using the rotary knobs.A de-
scription can be found on page 4-10.
Setting the functions The following functions offer a choice between coarse
and fine adjustment:
Shown below the A-scan are five function groups that
you can directly select using the corresponding key. Function Function group
The selected function group is highlighted and the corre-
sponding four functions are displayed next to the RANGE BASE
A-scan on the right. You can likewise directly select the MTLVEL BASE
individual functions using the corresponding keys. D-DELAY BASE
aSTART aGAT
Functions with double assignments aWIDTH aGAT
bSTART bGAT
Some functions have double assignments. You will rec- bWIDTH bGAT
ognize the functions with double assignments by an cSTART cGAT
arrow (icon >) after the function name. cWIDTH cGAT
S-REF1 CAL
Toggle between the two functions by repeatedly press-
S-REF2 CAL
ing the corresponding key .
ANGLE TRIG
Coarse and fine adjustment of functions THICKNE TRIG
DIAMET TRIG
You can choose between coarse and fine adjustment for
some functions. You can toggle between these two For more details on the adjustment possibilities, please
adjustment modes by pressing the corresponding read from page 5-5 onward.
key several times. The fine adjustment is identified
by an asterisk preceding the function value.
The operation mode of the USM 35X without rotary dB– reduces the instrument gain by the selected step
knobs is intended for all applications where the ultra- dB+ increases the instrument gain by the selected step
sonic instrument is used with a dustproof or watertight
cover, e.g. in contaminated areas of nuclear power
plants or in dusty environments. H Note:
The key remains active, i.e. you can select another
– Browse through the menus using the key to
select the required menu. dB step size if required.
– Press the key and keep it pressed for one second Val– reduces the value of the currently chosen
to switch to the operation mode without rotary knobs. function, e.g. RANGE
The instrument still displays the last chosen menu but Val+ increases the value of the currently chosen
changes the menu key assignment to the following function
menu line: – Press the key next to one the four functions to
select it for editing. With another press of the
respective key you can choose between coarse and
fine setting if applicable for this function.
H Note:
The rotary knobs also work if the instrument is operated
with keys only.
H Note:
Double assignment of the function DIALOG/UNIT
(icon >). Toggle between the two functions by repeat-
edly pressing the corresponding key .
Selecting units
In the function UNIT (function group CFG1) you can
choose your favorite units between mm or inch.
H Note:
– If necessary, go to the third operating level.
Double assignment of the function DIALOG/UNIT
– In the function group CFG1 select the function UNIT. (icon >). Toggle between the two functions by repeat-
edly pressing the corresponding key .
A Attention:
Select your units immediately when you start working
with the USM 35X because if you change the unit, all
the current settings are deleted, and the basic setup is
loaded again.
USM 35X Issue 06, 11/2007 4-13
Principles of operation Important basic settings
In order not to delete anything by accident, a safety – If necessary, change to the third operation level.
prompt is displayed in the measurement line.
– Select the function DATE in the function group CFG2.
– If you are sure that you want to change the unit,
press the key belonging to the function UNIT one
more time.
A Attention:
For a correct documentation you should always make
sure that you are using the correct time settings.
4.6 Basic settings of the display – If necessary, change to the third operation level.
All color schemes are suitable for indoor use. For out- – Use the right-hand rotary knob to set the required
door use, we recommend the color schemes 3 and 4. lighting.
5.1 Overview of the functions Each operating level contains five function groups. You
will recognize your currently active operating level by
the number on the separation line between the first and
The functions of the USM 35X are combined to form
the second function group.
function groups on three operating levels.
First operating level
If the instrument has the option Data Logger, there is an
additional fourth operation level.
– Press the key to select the function shown next Third operating level
to it. The setting of the selected function is carried
out via the right-hand rotary knob.
You can carry out important functions (switch on/off, If the instrument is equipped with the Data Logger op-
dB-step, freeze, zoom and report printout) by pressing tion, a fourth operating level is added. For this, refer to
the special keys (ref. chapter 4). the corresponding chapter Option Data Logger.
Function groups first operating level Function groups second operating level
BASE The functions that you find here are re- CAL This function group makes functions for the
quired for the basic adjustment of the semiautomatic calibration available to you.
screen displays.
REF This function group serves for measuring
PULS Combined in this group are the functions the dB difference between a reference
that serve for the adjustment of pulser. echo and the reflector echo.
or
MEM These functions serve for storing, loading LCD This is the function group where you can
and deleting of data sets. set the LCD contrast and backlight as well
as the echo display mode on the screen.
DATA The functions of this group serve for the
dataset management and documentation. CFG1 Functions for the configuration: unit, dialog
language, printer driver and assignment of
the key
H Note:
The setting 0.0 dB locks the gain in this way preventing
any accidental change of setting.
You can determine the step size of the sixth step using
the function dBSTEP in the function group RECV.
H Note:
The adjustment range for the display range depends on
the frequency range setting (function FREQU in function
group RECV).
5-6 Issue 06, 11/2007 USM 35X
Adjusting the display range (function group BASE) Operation
the test object, the value in D-DELAY must be set to 0. P-DELAY (Probe delay)
• Coarse adjustment Every probe has a delay line between the transducer
–10 mm ... 1024 mm/–0.3" ... 40" in even steps element and the coupling face. This means that the
initial pulse must first pass through this delay line be-
• Fine adjustment fore the sound wave can enter the test object. You can
up to 99.9 mm/9.999" in steps of 0.01 mm/0.001" compensate for this influence of the delay line in the
up to 1024 mm/10" in steps of 0.1 mm/0.001" function P-DELAY.
– Select the function D-DELAY.
H Note:
– If required, toggle between coarse and fine adjust-
ment. If the value for P-DELAY is not known, read the section
Calibrating the USM 35X, chapter 5.7, in order to deter-
– Adjust the value for the display starting point by mine this value.
means of the right-hand rotary knob.
– Select the function P-DELAY.
• high
This setting reduces the echo height but mostly also
produces narrow echoes with higher resolution.
The larger your workpiece, the smaller PRF values are – Select the function group RECV.
needed in order to avoid phantom echoes. In the case of
smaller PRF values, however, the A-scan update rate
becomes lower; for this reason, high values are required
if a workpiece should be scanned fast.
H Note:
Error alarms can be triggered unter certain circumstanc-
es. These are caused by intermediate conditions in
instrument operation occuring when the instrument is
used, i.e. when function parameters are changed. Pos- • off – Evaluation logic off
sible alarms occuring during instrument operation The alarm and measurement capability are switched
(setting of functions) are to be ignored. off. The gate is not visible.
• pos – Coincidence
Display of gates
The alarm (LED A) is on if the preset response
To make the assignment easier, the gates are displayed threshold of the gate is exceeded within the dis-
in different colors. You cannot vary the colors of gates played range.
because they are fixed as follows:
• neg – Anticoincidence
• Gate A – red The alarm (LED A) is on if the preset response
threshold of the gate is not reached within the
• Gate B – green displayed range.
• Gate C – blue • a trig – Triggering by interface echo
When using gate A as echo-start gate (setting of the
evaluation logic for gate B)
aLOGIC/bLOGIC
(Evaluation logic of the gates) – Select the function aLOGIC or bLOGIC.
This function allows you to choose the method for trig- – Set the required alarm logic by means of the right-
gering the gate alarm. The alarm is output to the LED A hand rotary knob.
on the front panel of the USM 35X. There are four set-
ting options available: H Note:
The alarm and measurement function of the gates is
only active within the display range.
– Set the known material velocity in MTLVEL (function – Set RANGE to 100 mm/5".
group BASE).
– Set the known material velocity of 5920 m/s
– Couple the probe to the calibration block. (233 "/ms) in MTLVEL.
– Set the required display range in RANGE (function – Set the gate so that it is positioned on the first
group BASE). The calibration echo must be dis- calibration echo (from 25 mm/1").
played on the screen.
– Read the sound path in the measurement line. If this
– Position the gate on one of the calibration echoes value is not equal to 25 mm/1", change the adjust-
until the sound path of the echo is indicated in the ment for the function P-DELAY until it is at 25 mm/1".
measurement line.
This completes the calibration of the USM 35X to the
– After this, change the adjustment of the function material velocity of 5920 m/s (233 "/ms) with a calibra-
P-DELAY (function group BASE) until the correct tion range of 100 mm/5" for the probe used.
sound path for the selected calibration echo is
indicated in the measurement line.
The distances between 2 calibration echoes must be – The recording of the first calibration echo is confirmed
entered as default data. The USM 35X will then carry by the message “Echo is recorded”, and the func-
out a plausibility check, calculate the material velocity tion CAL indicates the value 1.
and the probe delay, and automatically set the parame-
ters. – Move the gate to the second calibration echo.
H Note:
If the instrument is not able to carry out any valid cali-
bration on the basis of the input values and the echoes
recorded, a corresponding error message is displayed.
In that case, please check the values of your calibration
– Position the gate on the first calibration echo.
lines and repeat the process of recording the calibration
echoes. – Press to record the first calibration echo.
– Position the gate on the second calibration echo: Calibration with dual-element (TR) probes
Dual-element (TR) probes are especially used for wall
thickness measurement. The following peculiarities
must be taken into account when using these probes:
Echo flank
Most dual-element (TR) probes have a roof angle (trans-
ducer elements with inclined orientation toward the test
surface). This causes mode conversions both at beam
index (sound entry into the material) and at the reflec-
– Press . tion from the backwall, which can result in very jagged
The second echo is stored, the calibration is carried echoes.
out, and the CAL function is reset to 0. The valid cali-
bration is briefly confirmed and carried out. V-path error
Dual-element (TR) probes produce a v-shaped sound
If you select the function group BASE, you can read the path from the pulser via the reflection from the backwall
material velocity and probe delay. to the receiver element. This so-called “V-path error”
affects the measuring accuracy. You should therefore
choose two wall thicknesses that cover the expected
thickness measurement range for the calibration. In this
way, the V-path error can be corrected to the greatest
possible extent.
Higher material velocity – Set the pulser and receiver functions according to the
probe used and the test application.
Due to the V-path error, a higher material velocity than
that of the test material is given during calibration, espe- – Set the function TOF (function group MEAS) to
cially with small thicknesses. This is typical of dual- flank.
element (TR) probes and serves for compensation of the
V-path error. – Vary the gain so that the highest echo reaches
approximately the full screen height.
With small wall thicknesses, the above-described effect
leads to an echo amplitude drop which has to be espe- – Set the gate threshold to the required height for
cially taken into account with thicknesses less than measuring the sound paths at the echo flanks.
2 mm/0.08".
– Select the function group CAL.
A stepped reference block having different wall thick-
nesses is required for calibration. The wall thicknesses – Enter the distances of the two calibration echoes in
must be selected so that they cover the expected read- S-REF1 and S-REF2.
ings. – Position the gate (function aSTART) on the first
calibration echo.
Calibration process:
We recommend to use the semiautomatic calibration – Press to record the first calibration echo.
function for the calibration with T/R probes. – Couple the probe to the calibration block containing
– Set the required test range. the second calibration line, and set the height so that
it’s about as high as the first calibration echo.
– Increase the probe delay (P-DELAY) until the two
calibration lines are displayed within the range. – Move the gate to the second calibration echo.
H Note:
The point of amplitude measurement is marked with a
small upward triangle on the corresponding gate bar.
The point of distance measurement is marked with a
small downward triangle.
Gate threshold at 80 %
measured sound path: 24.91 mm
Gate threshold at 20 %
measured sound path: 24.44 mm
5-24 Issue 06, 11/2007 USM 35X
Measurement of dB difference (function group REF) Operation
A Attention:
When recording a reference echo, an already stored
reference echo is overwritten after a corresponding
warning.
Echo comparison
You can compare the echo of any reflector of your
choice with the reference echo. The displayed result is
the dB difference of the two echoes.
H Note:
The dB difference is independent of any possible gain
variation.
– If necessary, change to the second operation level. Depending on the setting in the function EVAMOD
(function group CFG2), one of the function groups REF,
– Select the function group AWS. DAC, JDAC, or DGS may also be displayed at this
point. Please also refer to chapter 5.15 General configu-
ration.
D=A–B–C H Notes:
with: Make sure that all instrument options for the special
test are calibrated before starting with the rating accord-
• A = Indication (in dB) ing to AWS.
Absolute instrument gain with which the maximum
flaw echo is at 50 % (±5 %) echo height Pay attention to peaking an echo with an amplitude
between 45 % and 55 % screen height. A rating is not
• B = Reference (in dB) possible with other amplitudes.
Absolute instrument gain with which the maximum
reference echo (1.5 mm side-drilled hole from the – Apply couplant, and couple the probe to the refer-
reference block 1) is at 50 % (±5 %) echo height ence block 1. Peak the echo from the 1.5 mm side-
drilled hole.
• C = Attenuation (in dB)
This value is calculated according to the formula – Select the function aSTART, and set up the A gate
C = 0.079 dB/mm (s – 25.4 mm). With s = sound on the reference echo.
path of the flaw echo.
– Vary the gain so that the reference echo is displayed
The sound attenuation correction is automatically at 50 % screen height.
calculated and displayed by the instrument. For
– Choose the function REFRNCE, and confirm the
sound paths smaller than or equal to 25.4 mm
choice in order to save the reference gain.
(1 inch), the value is set to zero.
• Depth d:
Distance between flaw position and surface
When using angle-beam probes, the instrument can X-VALUE (X-value of the probe)
additionally calculate the sound path section or so-
called leg L up to the next reflection point. This sound The function X-VALUE enables you to set the X-value
path section or leg can be displayed as the measured (distance between the probe’s leading face and probe
value La, Lb, or Lc. index/sound exit point) of the probe used. This value is
required for the automatic calculation of the reduced
projection distance.
5.12 Data saving A data set contains all instrument settings as well as
the A-scan. This means that whenever you recall a
(function group MEM) stored data set, your instrument is again set up exactly
the same as it was at the moment when the data set
You will find all functions for storing, recalling and delet- was stored. This makes each one of your tests repro-
ing complete data sets in the function group MEM. ducible.
– If required, go to the second operating level. You will find the following functions:
– Select the function group MEM. SET-# selecting number of a data set
Storing a data set All active entries in the information table (TESTINF) are
automatically allocated to the data set being stored
You can save your current setup to a data set. (see chapter 5.13 Dataset management).
– Select the function SET-#.
Deleting a data set
– Use the right-hand rotary knob to set the number
where you would want to store the current data set An occupied data set is marked with an asterisk (*)
(1 to 800). before the data set number. You can delete these data
sets if you no longer need them.
– Select the function STORE.
– Select the function SET-#.
– Use the right-hand rotary knob to set it to on.
– Use the right-hand rotary knob to set the number of
The USM 35X stores the current data set. When the the data set that you want to delete.
storage process is completed, the function STORE is
automatically reset to off. – Select the function DELETE.
Deleting all data set – Use the right-hand rotary knob to set it to on. The
measurement line will then prompt: Delete all data
You can delete all data sets if you no longer need them. sets?
– Select the function DELETE. – Confirm by pressing the corresponding key one
more time (all other keys would abort the process).
H Note: A Attention:
Double assignment of the function DELETE/DELALL.
If a saved data set is loaded, the current instrument
Press the corresponding key repeatedly to toggle
setup is lost. If necessary, save the current instrument
between the functions.
setup to a new data set before loading a saved data set.
H Note:
The gate for surveying the echo can be moved in the
recalled A-scan. However, as the evaluation is made in
the frozen A-scan, the measurement resolution is only
0.5 % of the adjusted calibration range.
5.13 Dataset management The functions in the function group DATA enable you to
easily manage the data sets stored in the USM 35X.
(function group DATA)
The following functions are available:
The USM 35X offers comprehensive functions for an
TESTINF You can save a lot of additional information
easy dataset management.
for every data set, e.g. data on the test
– If required, go to the second operating level. object, on the flaw detected, or comments.
– Select the function group DATA. PREVIEW In this dataset preview you will see the
A-scan, the dataset name and the storage
date of each data set.
H Note:
You cannot edit the field SET-#. The number of the cur-
rent data set is displayed here.
– Use the right-hand rotary knob to set the function to – Use the right-hand rotary knob to set the function to
on. on. The list of the currently set functions is displayed.
The directory list of the stored data sets is displayed – Turn the right-hand rotary knob to have other lines
(dataset numbers and names). The display shows displayed. The list is advanced by one line each.
12 data sets at a time. Occupied data sets are
marked with an asterisk (*). – If necessary, press one of the keys , or to
go back to the currently active A-scan.
– Turn the right-hand rotary knob to have other data
sets displayed. The list always advances by one line
each.
5.14 Configuring the USM 35X for – If required, go to the third operating level.
a test application – Select the function group MEAS.
TOF (Selecting the measuring point) When adjusted to flank or jflank the sound path mea-
surement is made at the point of intersection of the
The sound path measurement in the calibration process monitor gate with the rising flank of the first echo in the
or in the subsequent echo evaluation process depends gate.
on the selected measuring point which can be adjusted
either to flank, to peak or to jflank in the USM 35X.
H Note:
While DAC, TCG or JISDAC is active you can change
the TOF mode from peak to flank.
A Attention:
The highest echo in the gate does not have to be the
echo for which the sound path has been measured. This
may lead to false echo evaluation!
In order to identify the points of measurement and to In TOF mode jflank the sound path measurement is
avoid misinterpretation two indicators were introduced made at the point of intersection of the monitor gate
per gate: The first triangle pointing downwards indicates with the rising flank of the first echo in the gate. The
the position of the measured sound path (distance), amplitude is measured at the maximum of the first echo
whereas the triangle pointing upwards marks the posi- in the gate even if there are further signals with higher
tion of the measured amplitude. amplitudes in the gate.
In principle, the peak measurement should be preferred S-DISP (Zoomed display of reading)
because the measured distances do not depend on the
echo height in that case. However, there are application You can have a selected reading zoomed in the A-scan
cases in which the flank measurement is either specified, display. The following readings can be selected for the
or it must be applied for technical reasons, e.g. in many zoomed display (in the second column the indication of
tests using dual-element (TR) probes. the readings in the measurement line):
Sa Sa Sound path for gate A
A Attention: Sb Sb Sound path for gate B
In any case, the adjustment of the measuring point for Sb-a ba Difference of single measurements for
calibration and for the subsequent test use must al- sound path gate B – gate A
ways be identical. Otherwise measurement errors may Ha % Ha Echo height gate A in % screen height
occur. Hb % Hb Echo height gate B in % screen height
– Select the TOF function. Ha dB ha Echo height gate A in dB
Hb dB hb Echo height gate B in dB
– Use the right-hand rotary knob to select the required R-start Rs Range start
setting.
R-end Re Range end
La La Number of legs in gate A
Lb Lb Number of legs in gate B
Lc Lc Number of legs in gate C
Only for flaw position calculation:
Da Da Depth for gate A
All measured values which have also been described for Setting the display
the zoomed display of the function S-DISP are availabe
to you at each position. In the function group LCD, you will find setting options
for the display screen itself and for the echo display.
H Note:
Double assignment of the function FILLED/VGA. Toggle
between the two functions by repeatedly pressing the
corresponding key .
H Note:
Double assignment of the functions DIALOG/UNIT and
DATE/TIME. Toggle between the two functions by re-
peatedly pressing the corresponding key .
• meas P1 H Note:
The measured value given at position 1 in the
Please also refer to chapter 6 Documentation.
measurement line
• pardump – Select the function COPYMOD.
All instrument functions with the current settings
• PCX – Use the right-hand rotary knob to set the required
Screen contents as a PCX-format file. To transfer the assignment for the key.
data to the PC, you will need a terminal program.
• store TIME/DATE (Setting the time and date)
The current instrument setting is stored to the
selected (free) data set, and the data set number You have to check the current date and time and, if
(DAT-#) is automatically increased. required, set them so that these data are correctly
saved together with the test results.
• datalog (only with Data Logger option)
The selected job is printed out as a report including
all measured values.
• off
The key is deactivated.
• special
as setting “hardcpy”. After printout of the screen
contents no form feed, every press on the key prints
out the next hardcopy on the same page (three or
four hardcopies depending on the printer).
H Note: ANAMOD
Double assignment of the function DATE/TIME (icon >). You can output results of measurements at the analog
Toggle between the two functions by repeatedly press- output for external further processing. Use the function
ing the corresponding key . ANAMOD to configure the analog output in case there
is no echo in the evaluation gate and the analog voltage
A Attention: has been selected for the sound path at the output.
For a correct documentation always make sure that you You have the following setting options:
are using correctly set time and date values. Be aware
that the USM 35X displays the year as a two digit num- • lo volt
ber! The analog output supplies 0 volt.
– Use the right-hand rotary knob to change the high- – Use the right-hand rotary knob to choose the required
lighted value. value.
• AWS
H Note:
Rating of welds according to AWS D1.1
Error alarms can be triggered unter certain circumstanc-
es. These are caused by intermediate conditions in • DAC (only USM 35X DAC and USM 35X S)
instrument operation occuring when the instrument is Evaluation using the Distance-Amplitude Curve
used, i.e. when function parameters are changed. Pos-
• JISDAC (only USM 35X DAC and USM 35X S)
sible alarms occuring during instrument operation
Evaluation using the Distance-Amplitude Curve
(setting of functions) are to be ignored.
according to JIS Z3060-2002
5.18 Distance-amplitude curve You will find the functions for the distance-amplitude
curve in the function group DAC. If required, select the
(only USM 35X DAC and setting DAC in the function group EVAMOD first.
USM 35S)
– If required, go to the third operating level.
H Note: – Select the function group CFG2.
The DAC function is available as a fixed function in the – Switch the function EVA-MOD over to the setting
second operating level on the USM 35X DAC. With the DAC.
USM 35X S, the DAC function can be additionally
switched over to DGS evaluation mode. – Go to the second operating level.
Due to the angle of the sound beam spread and to the – Select the function group DAC.
sound attenuation in the material the echo height of
reflectors of equal size depends on the distance to the
probe.
H Note: H Note:
Double assignment of the function T-CORR/OFFSET. No reference echoes can be recorded with
Toggle between the two functions by repeatedly press- DACMODE = TCG. TCG can only be activated if the
ing the corresponding key . reference echoes recorded are situated within a dynam-
ic range of 40 dB. Otherwise an error message is out-
put. If the TCG setting should be nevertheless be used
DACMOD (Activating DAC/TCG) in this case, then the DAC must be reduced (by delet-
You can use this function to activate the DAC. The fol- ing the last reference points) until TCG can be switched
lowing settings are available: on.
DACECHO (Recording reference curve) – Press to record the first reference echo. The
instrument gain will automatically change until the
A Attention: DAC echo in gate A reaches 80 % screen height
(+/–0,3 dB). The function DACECHO is set to 1 to
Before starting to record a reference curve, the instru- indicate that the first reference echo has been
ment must be correctly calibrated (ref. section successfully recorded. Simultaneously the status
5.7 Calibrating the USM 35X). symbol “R” appears (= reference echo stored).
The moment a new curve is recorded, a possibly al- – Peak the next reference echo, and repeat the record-
ready existing curve must be deleted. If necessary, ing process for other curve points. The number in the
make sure that the old curve has been stored in a free function DACECHO is increased by 1 with each
data set before starting to record a new curve! recording.
– Select the function DACMOD.
H Note:
– Use the right-hand rotary knob to set the function to
DAC. The function DACECHO is set to 0 since there If the message “Echo is not valid” appears, the refer-
is no previously recorded echo. ence point could not be recorded. Check the gate posi-
tion as well as the height of the reference echo and
– Couple the probe to the reference block, and peak repeat the recording.
the first reference echo. Use the left-hand rotary knob
to bring the echo to an amplitude between 70 % and As soon as you have recorded at least two curve refer-
100 % screen height. ence points. Your DAC is already active (please see
previous section). You can record a maximum of
– Select the function aSTART, and then move the gate 10 curve reference points.
so that the selected echo is the highest of the echo
sequence within the gate range.
In flank mode the TOF reads the first echo in the gate,
A Attention: but the amplitude will be measured at the highest peak
A recorded DAC will no longer be valid, in case you in the gate which possibly might belong to another sig-
change the probe delay to a larger value, e.g. by adding nal. For a better identification of the TOF and amplitude
an additional delay line to the probe after the DAC has measurement the instrument's SW has been improved
been recorded without the delay line in place. and indicates the TOF measurement point with the gate
cursor ∇, and the amplitude with the second gate cur-
The same applies for immersion testing: the DAC re- sor ∆.
cording must be performed when the final water delay is
set.
– Select the function group JDAC. DACMOD (Activating DAC according to JIS)
You can use this function to activate the DAC. The fol-
lowing settings are available:
• off
No DAC is active.
• DAC
DAC according to JIS with 6 curves. The first 4 curves
are identified with the letters L, M, H and U dedicated
to these curves, and therefore move with any gain
change.
– Use the right-hand rotary knob to select the DAC DACECHO (Recording reference curve)
setting.
– Couple the probe to the reference block, and peak The function DACECHO displays the number 1.
the first reference echo. Use the left-hand rotary knob
to bring the echo to an amplitude between 70 % and
100 % screen height.
Deleting reference points or the complete DAC BOLDLI (Choice of a registration curve)
You can delete the reference point which was recorded One of the four curves marked with a character (L, M,
last in each case, or the complete DAC. H, U) can be selected as the registration curve. This
registration curve will then be displayed as a bold line
– Select the function DACECHO.
and an echo amplitude evaluation (dB-to-curve) will be
– Turn the right-hand rotary knob downward (counter- made for this curve.
clockwise). The message “Do you want to delete
– Select the function BOLDLI.
the DAC echo?” appears in the measurement line.
– Use the right-hand rotary knob to select the required
– Press the key in order to delete the last echo, or
setting.
press another key in order to cancel the process of
deleting.
T-CORR (Sensitivity correction)
– In order to delete the complete DAC, turn the right-
hand rotary knob upward (clockwise). The message This function enables you to compensate for the trans-
“Do you want to delete all DAC echoes?” appears fer losses in the material under test. This correction is
in the measurement line. necessary if test object and reference block have differ-
ent surface qualities.
– Press the key in order to delete all echoes, or
press another key in order to cancel the process of You have to find out the adjustment value for the com-
deleting. pensation of transfer losses by experiments. The gain is
varied accordingly in this connection, the curve line
In this way, you can record one or several new reference remains the same.
points.
In flank mode the TOF reads the first echo in the gate,
A Attention: but the amplitude will be measured at the highest peak
A recorded DAC will no longer be valid, in case you in the gate which possibly might belong to another sig-
change the probe delay to a larger value, e.g. by adding nal. For a better identification of the TOF and amplitude
an additional delay line to the probe after the DAC has measurement the instrument's SW has been improved
been recorded without the delay line in place. and indicates the TOF measurement point with the gate
cursor ∇, and the amplitude with the second gate cur-
The same applies for immersion testing: the DAC re- sor ∆.
cording must be performed when the final water delay is
set.
A Attention: A Attention:
Switching to flank mode when DAC/TCG is active may You are comparing the reflecting power of a natural flaw
lead to the following effects: with that of a theoretical flaw. No definite conclusions
• Sound path readings may be affected by an error due may be drawn on the natural flaw (roughness, inclined
to the fact that the instrument had previously been position, etc.).
calibrated in peak mode. The so-called DGS diagram forms the basis for this
• In case more than one echo is in the gate, TOF (e.g. comparison of the reflecting power. This diagram con-
sound path) and amplitude reading may no longer sists of a set of curves showing the correlation of three
belong to the same echo. influencing variables:
• Distance D between the probe and circular disk- The following probe parameters affect the curve shape:
shaped equivalent reflector
• element or crystal diameter
• Difference in gain G between various large circular
disk-shaped equivalent reflectors and an infinitely • frequency
large backwall • delay length
• Size S of the circular disk-shaped equivalent reflec-
• delay velocity
tor. The influencing variable S always remains con-
stant for one curve of the set of curves You can adjust these parameters on the USM 35X S in
such a way that you can use the DGS method with
The advantage of the DGS method lies in the fact that
many different probes and on different materials.
you can carry out reproducible evaluations of small dis-
continuities. The reproducibility is most of all important,
for example, whenever you aim to carry out an accep- H Note:
tance test. Before setting the DGS function, the instrument must
Apart from the influencing variables already mentioned, first be calibrated because all functions affecting the
there are other factors determining the curve shape: DGS evaluation mode (MTLVEL, P-DELAY, DAMPING,
POWER, FINE G, FREQU, RECTIFY) can no longer be
• sound attenuation changed after the reference echo has been recorded.
• transfer losses For T/R probes, MTLVEL can only be set from 5350 to
6500 m/s.
• amplitude correction value
Please also refer to chapter 5.7 Calibrating the
• probe. USM 35X on this subject.
Validity of the DGS method • Echo amplitudes for reflector distances smaller than
half of the probe’s near-field length are subject to
Echo amplitude evaluations according to the DGS heavy variation – for physical reasons due to interfer-
method are only reliable and reproducible in cases ence phenomena effecting the area. Thus, evaluation
when: results may fluctuate more as the usually permissi-
• On materials having non-negligible attenuation ble ±2 dB. An evaluation according to the DGS
losses, the compensation coefficients will have to be method is possible but not recommended for this
defined. In this case, it is necessary to calibrate the cases.
equipment on test blocks containing test reflectors of
known size at different distances to make a correct Change of probe delay in DGS mode
compensation for the attenuation losses on the In general changing the delay line automatically influ-
actual test object. After matching the results of the ences the shape of the sound beam, and this would
test reflectors to the diagram or built-in curves this theoretically require a new DGS setup. However, a
setting is valid for all flaw sizes and for all distances small change of the probe delay, which typically occurs
covered by the DGS curves. with normal wear during inspection, has no significant
influence on the programmed distance law.
• The reference echo is received from the test object if
possible. If this is not possible, it should be ensured
that the reference block is made of the same material A Attention:
as the test object. An existing DGS setup will no longer be valid, in case
you change the probe delay to a larger value, e.g. by
• The evaluation is carried out using the same probe
adding an additional delay line to the probe after the
which was also used for recording the reference
DGS reference echo has been recorded without the
echo. Another probe of the same type can be used
delay line in place.
after recording a new reference echo.
In flank mode the TOF reads the first echo in the gate,
A Attention:
but the amplitude will be measured at the highest peak Switching to flank mode when DGS is active may lead
in the gate which possibly might belong to another sig- to the following effects:
nal. For a better identification of the TOF and amplitude
• Sound path readings may be affected by an error due
measurement the instrument's SW has been improved
to the fact that the instrument had previously been
and indicates the TOF measurement point with the gate
calibrated in peak mode.
cursor ∇, and the amplitude with the second gate cur-
sor ∆. • In case more than one echo is in the gate, TOF (e.g.
sound path) and amplitude reading may no longer
belong to the same echo.
MB 2 S 3.0 4.5 8 12
MB 4 S 1.5 2.3 15 23
MB 5 S 1.2 1.8 20 30
MSEB 4 1.5 10 ±2 You can only carry out a DGS evaluation with one of the
available dual element probes, if the material velocity is
MSEB 4 0° 1.5 18 ±4 in between 5330 m/s and 6500 m/s.
MSEB 5 1.2 10 ±2
Locks, error messages
SEB 1 5.9 20 ±4
As long as a valid reference echo is stored, no functions
SEB 2 KF 5 3.0 6 ±2 can be changed which could cause an incorrect DGS
evaluation. If an attempt is made to change such a func-
SEB 4 KF 8 1.5 6 ±2 tion, the corresponding error message will appear, e.g.
SEB 2 3.0 15 ±3 “FREQUEN blocked by DGS-REF = on”
SEB 4 1.5 12 ±2 The DGS evaluation must likewise be switched off and
the reference echo deleted in the case of selecting a
new probe, e.g. for a new test application.
H Note:
The setting PCX generates a PCX-format file which you
can transfer to a PC by means of a suitable program
capable of receiving and storing data.
Printing
If you have connected, prepared and activated the
printer, just press the key.
• water,
A Attention:
Do not use any methyl alcohol, solvents, or dye pen-
etrant cleaners!
The plastic parts can be damaged or embrittled by this.
• after a storage time of 3 months or longer – Start by fully discharging the batteries. You can use
the discharging function of the charger for this. For
• after frequent partial discharge more details, please read the notes on the operation
of the battery charger.
Charging the batteries – The batteries are automatically charged after that.
You can charge the lithium-ion battery either directly in
Charging of exhausted NiCd batteries
the instrument or by means of the external battery
charger DR36 (order number 35 297) recommended by If batteries are exhausted, e.g. after a prolonged stor-
us. You always need an external battery charger to age time in empty state, they often reach their full ca-
charge standard C-cells. In this regard, please pay at- pacity only after repeated discharge/charge cycles.
tention to the information on the operation of the battery
charger.
The charger identifies defective batteries. In that case, How to handle alkaline batteries
replace the batteries by a new set. Otherwise there is
the danger that individual cells have different capacities – Please remove the batteries from the instrument if it
so that you will no longer obtain the normal operating has not been operated for a longer time.
time with the instrument in battery operation.
A Attention:
Leaking batteries may cause severe damages to the
instrument! You should always only use leak-proof
batteries and remove them from the instrument after
turning it off.
H Note:
Used batteries are special waste and have to be
disposed of according to legal requirements!
7.3 Maintenance
The USM 35X requires basically no maintenance.
A Attention:
Repair work may only be carried out by members of
authorized Service staff of GE Inspection Technologies.
7.4 Recycling
Total 2.20
* Materials/components, which cannot be separated into mono materials by destructive mechanical processes
8.1 Interfaces
The USM 35X presents different interfaces for the con-
nection of external units and for the data exchange. All
interfaces are located at the instrument front. The fol-
lowing figure gives an overview of the position of inter-
faces.
3 RS 232
serial interface, 9-way Sub-D socket
4 I/O
analog interface, 8-way LEMO-1-B socket
5 RGB-OUT
VGA interface, 10-way LEMO-1-B socket
6 12V DC
Mains connection socket, 4-way LEMO-0-B socket
H Note:
View of the 8-way LEMO-1-B socket
You are able to externally process the alarm condition
with the alarm output, e.g. for sorting and other control
purposes. Error alarms can be triggered under certain
circumstances. These are caused by intermediate con-
ditions in instrument operation occuring when the in-
strument is used, i.e. when function parameters are
changed. Possible alarms occuring during instru-
ment operation (setting of functions) are to be ig-
nored.
H Note:
Switch off the instrument before connecting a cable to
the RS 232 socket or before withdrawing any plugs.
8.6 Remote control • Entry of a new value or state of a function using the
command structure:
You can use a connected PC for the remote control of
<ESC> <COMMAND> <SPACE> <VALUE> <RE TURN>
the USM 35X.
All values are entered or transmitted by the USM 35X
The data transfer is carried out by means of a remote without a comma or a point. The resolution of the func-
control program and the corresponding remote control tion should therefore be observed with all values. The
commands. These commands represent instructions resolution of a function applies to the entire value
referring to the individual functions of the USM 35X. range of that function.
The Windows program HyperTerminal can for example A resolution of 0.01 means:
be used as remote control program.
The USM 35X transmits the value of a function multi-
After the remote control program has been started and plied by the factor of 100. The entry of a value must be
the program interface has been configured, the com- done multiplied by the factor of 100.
mands are input via the keyboard of the computer. In
this connection, please observe the following differen- Examples
tiation:
• Setting of the display delay to 72.39 mm:
• Interrogation of a value or state of a USM 35X
function using the command structure: <ESC>dd 7239 <RETURN>
The USM 35X transmits the value of the current <ESC>dw 7230 <RETURN>
setting.
Example:
A resolution of 1 means:
Example:
<ESC>at 41 <RETURN>
With:
Transmission timing
• As soon as the instrument has received the ESC
command, is will return the * which then will be
displayed on PC screen.
Example:
Request RANGE value from the USM 35X
Example:
Request reading at position 2 in the measurement line
Alphanumerical entries
Key in [ESC] DN [Space] Weld inspection B 45/2
[CR] in order to enter the dataset name (DATNAME)
“Weld inspection B 45/2”. All alphanumerical entries
may have a maximum length of 24 characters. In case
the string length exceeds 24 characters, it will auto-
matically be cut to 24 characters.
Transfer of datasets
A total of 800 datasets (complete instrument setup
including A-scan) can be stored in the instrument. The
stored datasets including the actual setting (dataset # 0)
can be transferred to the PC in compressed binary
format for archiving purposes. If required, the datasets
may be downloaded back to the insturment for re-use
or echo comparison. This bi-directional dataset trans-
fer is part of the software UltraDOC.
Presettings are in bold-face type. You will find a brief aLOGIC AM 0 = off 1
description of all functions in chapter 9.1 Function 1 = pos
2 = neg
directory.
If not otherwise stated, all values refer to steel, AMPLCOR** AC -25 to +25 dB / 0 0.1
C = 5920 m/s. ANAMODE AQ 0 = 0 volt
1 = 5 volts
The functions marked with * are only available in the
USM 35X DAC and USM 35X S (DAC evaluation), the ANGLE PA 0 - 90 / 0 0.1
functions marked with ** are only available in the
USM 35X S (DGS evaluation). A-SCAN AS 0 = stndard 1
1 = compare
Functions which are only availabe in connection with 2 = envelop
the Data Logger option are marked with *** (please 3 = peak b
4 = afreeze
also refer to the section Remote control in chapter
5 = bfreeze
Option Data Logger on this subject). 6 = cfreeze***
aTHRSH AT 10 - 90 % / 40 1
-90 - -10 % additionally with rf
DGSMENU T5 0 = off
DATE DE numerical input
1 = on
e.g. 26-01-99
DGSMOD** DS 0 = off 1
DATNAME DN alphanumerical input 1 = on
POWER PI 0 = low 1
HORN HO 0 = off 1 1 = high
1 = on
PRBFREQ** XF 0.5 - 10.0 MHz 0.04
INDICAR BA -1101 - 1101 dB 0.1
PRBNAME** PN alphanumerical input
LIGHT LT 0 = off 1 PREVIEW T2 0 = off
1 = on 1 = on
2 - 20 / 0.8 - 8 MHz:
REJECT RJ 0 - 80 % / 0 1
0.5 - 1420 mm / 250 0.01
0.02 - 56" / 10 0.001
SCALE SE 0 = measval
1 = snd-pth.
RATING R
BD -1101 - 1101 db 0.1
2 = div.
HD Editable header data from report transferred as UR Read data set as binary data
ASCII-format data
TF Freeze on/off:
0 = off
1 = on
Control codes for the rotary knobs/ Function Key Code Range
function keys
ENTER R off / on
Function Key Code Range
Right-hand K+ increment
rotary knob K- decrement
dB-STEP P 0 = 0.0
1 = 0.5
2 = 1.0
3 = 2.0
4 = 6.0
5 = 6.5 – 20
FREEZE F off / on
ZOOM Z off / on
COPY C off / on
BASE 5 MEAS 5
PULS 6 MSEL 6
RECV 7 LCD 7
aGAT 8 CFG1 8
bGAT 9 CFG2 9
CAL 5
Function Key Code
REF/DAC/ 6
DGS/JDAC
first 1
TRIG 7
second 2
MEM 8
third 3
DATA 9
fourth 4
COPYMOD CFG1 Assignment of the key DGS-CRV** DGS Recording curve for DGS
evaluation mode
DACECHO* DAC/JDAC Recording of a reference echo
for the DAC DGSMENU** DGS Activating/Deactivating the DGS
menu table for selecting the
DACMODE* DAC/JDAC Activating/Deactivating the DAC probe and further DGS settings
DELALL MEM Deleting all stored data sets DIR DATA Dataset directory
DELETE MEM Deleting a stored data set DUAL PULS Separation of pulser and
receiver
DEL-VEL** DGS Material velocity for probe delay
EVAMOD CFG2 Switchover REF – DAC – DGS
– JDAC
FILLED LCD Selecting the echo display mode MEAS-P1 MSEL Selection of measured values
(filled or normal) MEAS-P2 at four positions of the
MEAS-P3 measurment line
FINE G RECV Fine adjustment of gain within a MEAS-P4
range of approx. 4 dB in 40 steps
MTLVEL BASE Setting of the material sound
FLAWLEN DATA Flawlength velocity
FREQU RECV Selecting the frequency range OBJECT DATA Object description
for the connected probe
OFFSET* DAC/JDAC Offset for multiple DAC
GAIN left-hand Setting of the gain
rotary knob OPERAT DATA Name of the operator
HORN CFG2 Switching the acoustic alarm P-DELAY BASE Compensating for the probe
signal on/off delay line
INDICA AWS Flaw gain in dB POWER PULS Setting the power of the initial
for AWS evaluation pulse
PRF-MOD PULS Setting the pulse repetition REFRNCE AWS Reference gain in dB
frequency for AWS evaluation
PROBE-#** DGS Probe number REFSIZE** DGS Size of the reference reflector
PRINTER CFG1 Selecting the printer for the test REJECT RECV Suppression of unwanted or
report spurious echo indications
RANGE BASE Setting of the range in which the SCALE LCD Choice of display mode
measurement is made. for the measurement line
RATING AWS Flaw rating as dB value SCHEME LCD Choice of a color scheme for
the screen display
RECALL MEM Retrieving a stored data set
S-DISP MEAS Zoomed display of a selected
RECTIFY RECV Selection of rectification parameter
REFECHO REF For storing a reference echo SET-# MEM Number of the data set
for the measurement of
dB difference SETTING DATA Display of a function list
REFECHO** DGS Type of the reference reflextor S-REF1 CAL Reference echo 1 for calibration
used S-REF2 Reference echo 2 for calibration
REFMOD REF Activation of echo comparison STO-INF DATA Saving the current additional
information
STORE MEM Saving the data set X-VALUE TRIG Entry of the distance between
probe index (sound exit point)
SURFACE DATA Condition of surface and leading face of the
angle-beam probe
T-CORR* DAC/AVG/ Sensitivity correction, e.g. to
JDAC compensate for transfer losses Y-POS DATA Y-position coordinate
• EN 61 000-6-2:2005
• EN 61 000-6-4:2001
• EN 61 010-1:2001
9.3 Manufacturer/ Keep the shipping container for any repairs possibly
required which cannot be made on the spot.
Service addresses
If there is anything special that you would like to know
The USM 35X is manufactured by: about the use, handling, operation and specifications of
the instruments, please contact your nearest GE
GE Inspection Technologies GmbH Inspection Technologies representative or directly:
Robert-Bosch-Str. 3
50354 Hürth GE Inspection Technologies GmbH
GERMANY
Service-Center
Phone +49 (0) 22 33 - 601 111 Robert-Bosch-Str. 3
Fax +49 (0) 22 33 - 601 402 50354 Hürth
GERMANY
The USM 35X is manufactured according to state-of-the-
art methods using high-quality components. Thorough or:
in-process inspections or intermediate tests and a quality
Postfach 1363
management system certified to DIN EN ISO 9001
50330 Hürth
ensure an optimum quality of conformance of the
GERMANY
instrument.
Phone +49 (0) 22 33 - 601 111
Should you nevertheless detect an error on your
Fax +49 (0) 22 33 - 601 402
instrument, switch the instrument off and remove the
batteries. Inform your local GE Inspection Technologies
Service indicating the error and describing it.
France USA
Great Britain
GE Inspection Technologies
892 Charter Avenue
Canley
Coventry CV4 8AF
GREAT BRITAIN
4
33
24
35
25
34
50
10
11
37
38
54
2
12
53
30
3 6
1
21
27
51
20
26
28 7
9
36
52 8
22
29
USM 35X
(contains Pos. 11, 6 and 30)
3 1 Battery cover, complete 36165-3.190
Spare parts list
08.March.05 / Ant
9-13
Appendix
9-14 Issue 06, 11/2007 USM 35X
Changes 10
B bTHRSH
Threshold of gate B ............................................. 5-16
Backlight ................................................................ 5-52
bWIDTH
BASE (function group) .............................................. 5-6 Width of gate B ................................................... 5-16
Basic initialization .................................................... 3-8
Basic settings ........................................................ 4-12 C
Batteries ................................................................... 1-2
CAL
Care ...................................................................... 7-3
Semiautomatic calibration ......................... 5-19, 5-22
Charging the NiCd batteries .................................. 7-3
CAL (function group) ............................................... 5-19
Baud rate for transmission ...................................... 5-55
Calibration .............................................................. 5-17
BAUD-R (Baud rate for transmission) ..................... 5-55
Dual-element (TR) probes ................................... 5-21
bGAT (function group) ............................................. 5-14 Straight-beam probes .......................................... 5-18
With unknown materials ...................................... 5-19
bLOGIC
Evaluation logic gate B ....................................... 5-15 CFG1 (function group) ............................................ 5-53
BOLDLI (Choice of a registration curve) .................. 5-71 CFG2 (function group) ............................................ 5-53
bSTART Changes ................................................................. 10-1
Starting point of gate B ....................................... 5-16
Changing a function value ....................................... 8-14
Codes DACECHO
Codes for function keys ...................................... 8-26 Recording reference curve ................................... 5-69
Function codes ................................................... 8-17
DACECHO (Recording reference curve) .................. 5-63
Remote control ..................................................... 8-9
DACMOD
Coincidence ............................................................ 5-15
Activating DAC according to JIS ......................... 5-68
COLOR ................................................................... 5-33
DACMOD (Activating DAC) ................. 5-61, 5-68, 5-78
Configuration ........................................................... 5-53
DACMOD (Activating DAC/TCG) ............................ 5-62
For test applications ........................................... 5-43
DAMPING (Probe matching) ..................................... 5-9
Configuring the measurement line .......................... 5-52
DATA (function group) ............................................. 5-38
Connecting a probe ................................................... 3-7
Data exchange ......................................................... 8-7
Controls .................................................................... 4-2
Data saving ............................................................. 5-34
COPY key
Assignment ......................................................... 5-55 Data set
Deleting ............................................................... 5-35
COPYMOD ....................................................... 6-3, 8-8
Deleting all .......................................................... 5-36
Assignment of the COPY key ............................. 5-55
Management ....................................................... 5-38
Recalling ............................................................. 5-36
Storing ................................................................. 5-35
D
Dataset
D-DELAY (Display starting point) .............................. 5-7 Directory ............................................................. 5-42
DAC (function group) .............................................. 5-61 Preview ............................................................... 5-41
F G
FILLED Gain .......................................................................... 5-5
Echo display ....................................................... 5-51 Fine adjustment .................................................. 5-12
Incrementation ...................................................... 5-5
FINE G (Fine adjustment of gain) ................. 5-11, 5-12
Gate setting ............................................................ 5-14
Flaw evaluation ......................................................... 1-6
Gate spreading ....................................................... 5-48
Flaw position calculation ........................................ 5-31
Gates ...................................................................... 5-14
Freeze function ....................................................... 5-59
Starting point ....................................................... 5-16
FREQU (Frequency range) ..................................... 5-13 Width .................................................................. 5-16
Frequency range ..................................................... 5-13
Function directory ..................................................... 9-2 H
Function groups ........................................ 4-4, 4-8, 5-3
HORN
Functions .................................................................. 4-4 Acoustic alarm .................................................... 5-58
Function directory ................................................. 9-2
Overview ................................................................ 5-2
Setting the functions ............................................. 4-9 I
Special keys ....................................................... 5-59
Instrument
Care of the instrument ........................................... 7-2
Instrument versions ................................................... 1-8
J M
JDAC (function group) ............................................. 5-67
MAGNIFY
Gate spreading ................................................... 5-48
K Maintenance ............................................................. 7-5
OFFSET (Distance of multiple DAC) ............ 5-65, 5-72 PRINTER (Printer for test report) ............................ 5-55
THICKNE V
Material thickness ............................................... 5-32
VGA ....................................................................... 5-51
THICKNE (Material thickness) ................................ 5-33
Threshold of the gates ............................................ 5-16
TIME
W
Setting the time .................................................. 5-56 Wall thickness .......................................................... 1-6
TOF Waste disposal ......................................................... 7-6
Selecting the measuring point ............................. 5-44
Width of the gates .................................................. 5-16
Transfer correction (DGS) ....................................... 5-83
Transfer of datasets ................................................ 8-15
X
Transmission timing ............................................... 8-12
TRIG (function group) .............................................. 5-31 X-position ................................................................ 5-40
X-VALUE
Of the probe ........................................................ 5-32
U
Ultrasonic testing ...................................................... 1-3 Z
UNIT
Units of measurement ......................................... 5-54 Zoom mode ............................................................... 4-3