Error messages of the master Er OXX
Er 004
“Communication master - AEC faulty”
Displayed on PC: Communication master <-> AEC dturbed or AEC not ready
Description
ifthe master processor (D702) cannot activate the AEC processor (D701) via the serial
interface any more, or f any important data from the control deck Is missing, the above
‘mentioned error is displayed
Fault elimination
First you must chieck it PC board D701 is plugged in correclly and i the connections to the
power supply and to the serial Interface (see wiring diagram) function correctly
{ino fault ean be found, PC board D701 or D702 should be replaces.
Er 005
“Flash in AEC is defective or not correctly initiated.”
Deplayedon PC: Flashin AECis detectve ornstconecty nt
Description
‘During initialization, the AEC processor (D701) checks the data stored in its EPROM by
‘means of stored check sums, If these data are incorrect or if the E°PROM falls, the AEC
proceseor informs the master processor via the serial interface and the above mentioned
terror is displayed.
‘The master also géneratos this error if, dvi
the AEC has been established yet.
exposure release, no communication with,
NOTICE | As opposed to this, error 004 appears if communication started
with the master is disturbed.
Fault elimination
Reprogram E2PROM or replace E2PROM or PC board D701. Reprogram with the service
PC, menu Backup,
Er 008
“Stand not ready”
Deplayed on PG: Communication master <-> stand disturbed or stand not eady
Description
‘This error means that the communication between master processor and unit processor
Vie the serial interface Is faully or interrupted.
Fault elimination
Check ifthe power supply and the serial interface (see witing diagram) fo the unit function
correctly,
Siemens Elem AB RegaierS SPHT DRED Page? o130 _, MiIBOOBOOTSOO0 Nova
Solna, Swedon Fev.03 06.00 SPSUD Serve ProgramError messages
Erott
“Exposure has been stopped"
Displayed cn PC: Expoaute aborted hy user
Description
the exposure buttons ara released prematurely the exposure is aborted immediately,
(On the control deck, the LIMIT LED (+0) lights up, an acoustic signal sounds and the
above-mentioned error is displayed,
Fault elimination
This error indication is only for information and requires no further action. Shouldit appear
frequently, however, check the contacts of the exposure release buttons. Replace expo-
sure roleaso button(s), if necessary.
Er 012
“Time limit reached”
Displayed en PC: Tine ima resched
Description
In order to prevent the tube assombly from overioad, a timer runs during exposure. This is
sot with the time fimit before exposure. If his timer runs down, the exposure is aborted
immediately, and the above error message is dsplayed. Depending on the mode of oper=
ation, the time Hit is calculated as follows:
‘AEC mode
The lime limits the power-dependent load time plus a tolerance value (approx. 400 ms)
mAs mode
Tho time limit is the calculated exposure time plus a cortain reserve. ti this period is
shorter than 2 8, 2¢ are assumed. Ifit exceeds 2 s, a reserve of 400 ms is added.
‘The time limit can be read off in ms with the service PC in “normal mode* under itam "AEC.
ata’. The value shown hero doss not include the extra 400 ms.
Fault elimination
‘The tube current and the filament current must be measured again with the osciloscope
(lest points log and I, on D705), The nominal and actual valve of the tube current can be
‘ead off withthe service PC in normal mode. I the actual tube current is too low, you must
fist ensure that the dynamic learn filament curront (DLF switch) is switched on. The tube
‘must be readjusted in any case, Should diffcutties stil ocour, the tube-current actual value
‘acquisition must be checked (see wiring diagram).Er 013
Limit mAs during AEC exposure’
Dielayedon PC: mae init eachad in AEC exposure
Description
For safety reasons, with AEC exposures, the accumulated mAs is integrated via the timer
independently of the AEC. Ifa focus, tube assembly and kV-dapendent ime limit (see
table) is obtained without the AEC being switched off, the exposure is aborted and the
‘above-mentioned error message is displayed.
Tube aw ‘aOR SRT
assenbly
PAOMORIOOS FL | 166mAs | 183mAs | WO mAs
PaoManeTO0GF2 | comms | GO0mAs | 428 mAs
Paowari-iooGFs | 2eemas | 19BmAe | _17OmAS i
PeOMaN-TODGFA | 7S2mAs | Ga7mAS | _SS7mAS
Fault elimination
The dose signal must be checked from the detector or the chamber to the hardware of the
AEC (see witing diagram).
Iino error can be detected, PC board D701 or D702 must be replaced.
Er 020
“PC mess bad length”
Displayedion PC: PC message hed bad length
Description
‘The length of the message, sent from the service PC to the MAMMOMAT, is too long.
Fault elimination
‘Chock that the right version of the service program is used.
Er 021
“PC dsr lost”
Displayedon PC: PC data sot ready lost
Description
This error message is not enabled today.
Fault elimination
‘Simone Fiore 5 Rossers SPETEROIIIGS Faye 30.3) MiCOOOOGSODD Nove
Soin, Sueden Fev.08 05.00 Sexvive Progeam3-4
Error messages
Er022
“PC rxerror”
Displayed on PC: Roosive message fom PC taled.
Description
‘An ertor occurred when the MAMMOMAT was receiving data from the PC or printer.
Note: May occur ifthe printer is switched on/off while the rest ofthe system Is powered.
Fault elimination
‘Chock the cables and conneetors to the PC or printer. Board 702 defective. PC/Printer
defective.
Er 023
“PC tx error”
Displayed on PC: Transmitmessage to PC tailed
Description
‘An error occurted when the MAMMOMIAT tried to send data to the service PC or printer
Fault elimination
Check the cables and connectors to the PC or printer. Board 702 defective, PCIPrinior
defective,
Er 024
“PC time-out”
Displayecien PC: PC, ID camera or workstation hae ot sent ACK/NAK within vo seconds,
Thee res are made.
Description
Tho service PC didn't answer.
Fautt elimination
Check the cables and connectors to the PC. PC defective.
Er 025
“PC NAK”
Displayed on PC: PC, 1D camera or workstation sonds NAK toe mee as response te the
sane message.
Deseription
‘The Service PC did not receive information from Master. Time-out occured and Service
PC responds with NAK.
Fautt elimination
1. Reset error on panel and try again,
2. Reset MAMMOMAT end PC.Error messages of the panel Er 1XX
Er 105
“Incorrect check sum in the E7PROM”
Displayed on PC: Checksum enorin instalation deta stored n panel EEPROM
Description
Each timo the MAMMOMAT 3000 Nova is switched on, the installation area in the
PROM on the control deck are checked with a shadow area also stored in tha
EPROM. If any discrepancy was detected, the above mentioned error is displayed.
Fault elimination
‘The installation data can be corrected by means ofthe service PC. It this error appears
‘more frequently, exchange the deck PC board D740 and make a now instalation of the
parameters in the control panal. The opdose parameters have to be programmed manu
ally via the contro! deck. For tho rest ofthe paramoters, usa the sorvice PC, menu
Backup.
Er 408
“*PROM program area error”
Displayedon PC: EEPROM program area ator
Description
When data is stored in the E2PROM (D740 112} of the control panel during programming
of the opdose parameters, this is read back once mote to check whether it has bean
stored correctly. If eny diferences occur, the above-mentioned error is displayed,
Fault elimination
‘The E2PROM (D740 112) in the panel must be replaced. If the error still occurs, the whole
10740 must be replaced. In both cases a new installation of the parametars in tho control
panel has to be performed. The opdose parameters have tobe programmed manually via
the control deck. For the rest of the parameters, use the sorvice PC, menu Backup,
Er 107
“Deck in STAND ALONE mode on exposure release”
Displayed an PC: No comrurizaion panel e» master
Description
tno data transmission takes place between the deck and the master during the first 20 s
after power on, the control deck switches to the so-called “stand alone” mode. The key-
board can be operated normally. Ifthe exposure release bution is operated in this mode,
the above mentioned error message is shown on the display.
Fault elimination
‘The cause of this error can be an interruption on the serial interface to the master. For this
reason, this connection must be checked exactly.
Siemens-Elema AS RassierS ‘SPB7-260.11403 “Pago 5 130 MIOEOIROONSGDO Nowa
Sone, Sweden Rov.03 05.00 SPSUD Senice ProgramError messages
Error messages of the filament Er 3XX_
Er 302
“Filament current monitoring lyimin has responded.”
Displayed onPC: ng Hament cent oo low at stench
Description
During *Stand-by’, the master checks whether there is any filament current. ifthe filement
urrent doas not reach a predetermined value, tite above-mentioned error is displayed,
Fault elimination
(Check fuse F3 on T2 (see wiring diagram).
Check leads.
Filament of tube assembly defective.
PC board D705 detective.Error messages of the AEC Er 4XX
Er 401
“Error during configuration of PLD.”
DipeyedonPC: Ero duingcoguaion of PLD
Description
Configuration of PLD device in AEC falled.
Fault elimination
‘Switch off the MAMMOMAT. Wait approx. one minute, Start up the MAMMOMAT. if he
error occurs again change D701 circuit board.
Er 402
“Self-test of PLD fail
Diplayecon PC: Seltest of PLD alos
Description
Gain-tost of circuit board D701 falled.
Fault elimination
‘Switch off the MAMMOMAT. Wait approx. one minute. Start up the MAMMOMAT. If the
error oceurs again change D701 circuit board
Er 403
““Selfest time-out, Fatal error in PLD’s internal program.”
Daplayedion PC: Setttoat time-out. Fetal eon PLO’ toma progran.
Description
Time-out occurred in the PLD device in AEC during gairtest of circuit board D701.
Fautt elimination
Switch off the MANMOMAT, Wait epprox. one minute. Tum on the MAMMOMAT. It the
error occurs again change D701 circuit board.
"PLD cannot perform requested operation because it is busy selt-testin;
Digplayod on PC: PLD cannot perf requested operation because its busy seeing,
Description
PLD in AEC cannot perform requested operation because itis busy sett-tasting.
Fault elimination
Acknowledge displayed error and wait approx. one minute.
‘Siemens Bema AB Rogeios ‘SFBT 28011408 Page 7 of 30 MiG007BO00I3000 Nove
Solna, Sweden Rev.08 05.00 SPEUD Serves Program3-8
Error messages
Er 407
“Time-out error while erasing AEC flash memory” ~.
Displayedon PC: Tano-out err while erasing AEG fash memory
Description
‘The D701 contains a memory device called "flash Memory”. This memory is used for stor
ing AEC parameters and correction tables. Before writing data into the memory it is first
necessary to erase the sector in which data is to be stored. The erasing procedure is con-
trolled by a timer. I anything goes wrong while erasing AEC flash memory and time-out
‘occurs, this will result in an error .
Fault elimination
‘Switch off tho MAMMOMAT, Wait approx. one minute, Start up the MAMMORAT. If the
error occurs again change D701 circuit board.
Er 408
“Time-out error while writing to AEC flash memory (time-out in waiting for
OK indication).”
Displayed on PC: Tme-0u err whe weting to REC fash memory line-out in wai for
ck indection),
Description
‘The D701 contains a memory device callad “flash memory”. This memory is usod for stor-
ing AEC parameters and correction tables. The procedure of writing this data is controlled
by a timer. if anything goes wrong while writing to the AEC flash memory and time-out
‘occurs, this will result in an error.
Fault elimination
‘Switch off the MAMMOMAT. Wait approx. one minute, Start up the MAMMOMAT. ifthe
error oceurs again change D701 circult board.
Er 409
“AEC flash test has locked up (time-out of 15 seconds exceeded).”
Displayed on PC: AEC fash estas looked up (time-out of 15 seconds excooded)
Description
During start-up ofthe unit the AEC performs a number of tests, 0.g. testing ofthe flash
memory device on D701. This testing is controlled by a timer and if something tums out to
be wrong a time-out is generated,
Fault elimination
‘Switch off the MAMMOMAT. Wait approx. one minute. Start up the MAMMOMAT. I the
error occurs again change D701 circuit board.Er 4i0
“Attempt to start an exposure before AEC flash memory has delivered cor-
rection values.”
Displayedon FC: Altomels stat an exposure before AEC fash memory has devered coe
rection vals,
Bofore an AEG exposure fs released, and based on the exposure settings on the panel,
the AEC fetches an estimated dose from a correction table stored in the flash memory. If
the exposure buttons are pressed during reading from flash, an error wil occur.
Fault elimination
Acknowtedge displayed error. Walt 15 seconds. Try to pertarm an exposure.
Er4i2
“Attempt made to start an exposure while self-testing of the PLD in AEC was
in progress”
Dieplayedon PC: Attomot mado to start an exposure while seltestng oe PLO in AEC
wasn proses
Description
Attempt mado to start an exposure while sa-testing of the PLD in AEC was in progress.
Fault elimination
Acknowledge displayed err. Walt 15 seconds. Try o perform an exposure.
Er ais :
“Exposure too short, correction value could not be loaded in time.”
Denlayecion FC: Excocwro too shor, ccrecton valve cout nots loaded in imo.
Description
‘An estimated dose is pre-calculated before an exposure is released. The minimum esti-
mated dose is dofinad for a minimum object thickness. For a normal objact thickness
{25 mm) the AEC will calculate @ necessary correction to add to the estimated dose dur-
ing the exposure. This procedure takes a while, and for very thin objects itis possibie that
the AEC will not manage to, within the time limit, decide if a correction is necessary or not.
This wil result in a errr, due to small object thickness.
Fault elimination
Acknowiedge dispiayed error. Change exposure parameters (e.9. lower kV).
‘Slamane-Eiema 0B Begowrs SPaTaaGs Page v oF) MITOONTRNOOAGON Nowe
‘Soin, Sweden Fev.08 0500 SPSUD ‘Sorviva Program3-10
Error messages
raid
“Attempt to release an exposure after the PLD in AEC has failed self-test ~
configuration.” .
Displayed on PC: Atomptio recone an exposure afer fhe PLD in AE has fled cottets
ceatiguration,
Description
Caused by previously generated errors 401, 402, 403, 409 and not restarted MAMMO-
MAT.
Fault elimination
‘Switch off the MAMMOMAT. Wait approx. one minute. Start up the MAMMOMAT. If the
error occurs again change D701 circuit board.
Er 416
“Estimated dose out of range.”
Displayed en PC: Estimated dose ct of ango.
Description
Estimated dose out of range, which can be caused by
~ xtreme setting of sensitivity
+ extreme setting of senstivity correction
+ extreme sotting of density correction
‘This moans that the exposure will be eliminated by the max. mAs limit.
Fault elimination
Acknowledge displayed error. Make sure the correct AEC Correction tables are installed,
‘Check settings of sensitivity, sansitivity correction and density correction.
Eraie
“Attempt to make offset compensation test while in AEC mode.”
Displayed on PO: Atlempt io make ofisot compansaton test while in AEC-mode.
Description
During an offset compensation test the AEC board is set in a special test mode and can-
‘not perform an AEC exposure, Because of this, the test exposures have to be performed
inmAs mode during this test.
Fault elimination
‘Acknowledge displayed error. Change to mAs and follow the instructions for offset com-
pensation test correctly,Er 419
“Attempt to write to registers of PLD not available during exposure.”
Dieplayedan PC: Atemptto wit to rgistors of PLD notevatebe dung expose.
Description
This is an intemal error not interfering with an exposure and not shown on the control
panel.
Fault elimination
If this error is discovered during reading of the error buffer through service PC, ploase
report this immediately.
Er 420
“Dose or time monitor overflow in PLD.”
Dieplayod on PC: Doe0 or imo mento overfowin PLD.
Description
This is an intemal esror not interfering with an exposure and not shown on the control
panel. Time and dose values reported by the service PC are not vali.
Fault elimination
ltthis error is discovered during reading ofthe error butler through service PC, please
report this immediately,
Er 42
“Wrong test mode request received by PLD.”
Displayed oc PC: Wong et rod equotrceived by PLD.
Description
This is an intemal error not interfering with an exposure and not shown on the control
panel.
Fault elimination
I this errors discovered curing reading ofthe error butter through service PO, please
report this immediataly.
Er 422
“Attempt to read registers of PLD not available during exposure.”
Dieplayodon PC: Atemptto read rege of PLD not avalable ding expose
Description
This is an internal error not interfering with an exposure and not shown on the control
panel. :
Fault elimination
I this error is discovered during reading of the error butler through service PC, please
report this immediately.
‘Slemene Eloma AB Ragen SPOTDRITIADS Page TAS) MTOOORIOO R000 Nova
Solna, Seeder Rov.03 05.00 ‘SPS-UD ‘Serves Program3-12
Error messages
Er 423
“Counters for estimated doses in PLD not initialized in time.”
Displayed on PC: Gourters or estimated dose in PLO ot iaed ina.
Description
mAs mode: Internal error not interfring with an exposure and nat shown on the control
panel
IWAEC mode: Internal error interfering wth an exposure and shown on the control panel
Fault elimination
If mAs mode: If this eror is discovered during reacing of the errr buffer through service
PC, please report this immediately,
ITAEC mode: Acknowledge displayed error. Walt 15 seconds. Try to perform an exposure,
Er 450
“Exposure aborted by dose monitoring,”
Displayeden PC: Exposure aborted by dose moni.
Description
An AEC exposure could not be performed due to too low detected dose rate.
Fault elimination
Acknowledge displayed error. Change exposure parameter saltings (e.g. higher kV).Error messages of the power pack Er 6XX
Er 601
“Rotation not reached”
Deplyeden PC: Acton speed ot reached win 3 3
Description
During staring of the rotating anode, the master counts the control pulses and compares
them witha tube assembly and intermediate circuit voltage”-dependent table value, If this
is not reached within 3 s, Er 601 Is displayed.
Fault elimination
BS signal missing (see wiring diagram).
Actuation of WRANST 2 fails (see wiring diagram).
‘Check oscilation current (see witing diagram).
Rotating anode cable or stator defective,
Er 602
“Tube voltage < 17 kV”
Deplayad en PC: qq imum tbo vollge 17KV ret resco
Description
1f17 KV Is not reached within 250 ms after the KVE signal, the above-mentioned error is
displayed (LED V39 on D705 lights up).
Fautt elimination
Intermediate circuit voltage too low (see wiring diagram).
‘Tube current or power too high.
‘Check cscitation currant (50 wiring diagram).
Actuation of WRANST 1 or WRANST 2 fails (see wiring diagram).
$6 relay not pulled up (see witing diagram).
“RV poi" 100 HOW,
Board 0702, D705 or D710 defective.
Er 603
“Tube voltage > 50 KV”
Displayed on PC: MVay= be veiage gear than SOR
Description
AA threshold switch in the kV controller watches during exposure whether the high vollage
exceeds 50 ky. If this is the case, the exposure is aborted immediately via the KVA leas.
‘The error is indicated by LED "> 50 kVsq" (D705.V41) lighting up and the above men-
tioned error message.
Fault elimination
Chock kV nominal value. Check tube current and power. Actual value acquisition defec-
tive (see wiring diagram). Replace PC board D705,
Slamone-Elama AB Faget SPAT IOTIAOT Tots) MIODUETVGOOG Nova
Solna, Sweden Rey03 06.00, SPUD ‘Sonvoe Program,3-14
Error messages
Er 604
“WR blank is applied”
Displayedon PC: WR_AUSTAST--inveier short cicult Blanking ignal rervains
Description
When Ina in the inverter is excoeded, the control is intorrupted for 200 ms. Tho software
thon checks whetner the blanking signal is still applied. It this is the casa, the above men-
‘toned error is displayed.
Fautt elimination
Replace PC board D705,
Er 605
“Module current too high during rotary anode acceleration’
Displayed on PC: Inverter shot cuit dung arode rotation boost
Description
|f the Irae Monitoring responds during acceleration of he rotating anode, the above men
tioned error is displayed,
ses
NOTICE | If this occurs, the system has to be switched off and resiaried in
order to get out of the error state.
Fault elimination
Transistor module in Inverter defactive,
Stator short-0, OV->BV or 5V->OV) will take place. the flank, lies in the middle of the mov-
ing aroa of the colimator package, and between the two postions for Molly Large focus
‘and Wolfram Large focus.
Aer having reached the ‘lank, i will output 2 certain number of stepping motor steps in
the direction of the wishad focus position.
The exor is displayed if the lank cannot be sensed within a cenain number of stepping
motor steps.
Fault elimination, trouble shooting
Switch between W and Mo anode materials withthe anode/fiter combinations buttons on
the contro! panel and check i the coll mator package moves at all,
Ifitdoes not move:
~ check if the CPU sends out pulses to the wing board, FOCUS_ST test point on
‘D805, if net, chock cables and connectors between D&C and D805 and maybe
replace CPU board (0801) or wing board (D805).
Check voltages on wing board, SV and 14V.
+ chock STEP ENABLE test point on D806, it should be low (OV) when pulses are
‘output to the motor.
~ _checkif itis at an end position, and the direction is into the end. Direction test
Point FOCUS _D al D805 should be low (OV) when moving collimator package
up (towards the tube).
| something is wrang, check cables and connectors.
Itit moves:
~ check if the collimator package position ight switch works as in the description
above,
~ check iit moves just slowiy, then something may be in the way that slows tho
‘spe6¢ down, or the motor may be faulty. n the lastease, try changing the motor
~ ifthe motor stops etter ust a short time either the motor orthe wing board (D805)
fs faulty, Try replacing one of them.Er 850°
“Incorrect checksum for stand insiallation menu in stand EEPROM”
Dieplayec on PC: Wite/ead back eure for nb 1h stand EEPROM
Description
Each time the MAMMOMAT 3000 Nova is switched on, the adjustment data in the stand
EEPROM (D801.134) are checked with a checksum. Each menu under “Stand contig” has
its own checksum. Ia discrepancy was detected in the stand installation manu, the above
mentioned errors displayed.
Fault etimination
Install all stand parameters using the service PC “Back-up” menu.
Er a5t
“incorrect checksum for calibrate compression menu in stand EEPROM”
Displayed on PC: WitfRead back fare for table 2 in sland EEPROM
Description
Each time the MAMMOMAT 3000 Nova is éwitched on, the adjustment data in the stand
EEPROM (D807.134) are checked with a checksum. Each menu under “Stand config’ has
its ewm checksum. Ifa discrepancy is detected in the calibrate compression menu, the
above mentioned error is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Er 862
“Incorrect checksum for rotation calibration menu in stand EEPROM”
Displayed on PC: Whie/Read back faire for table in siand EEPROM
Description
Each time the MAMMOMAT 3000 Nova is switched on, the adjustment data in the stand
EEPROM (801.134) are checked with a checksum. Each menu under “Stand config’ has
its own checksum. Ifa discrepancy was detected in the rotation calibration menu, the
above mentioned error is displayed,
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
‘Siomens-Flema A Repiters ‘SPE7-TOTTEOa Page 27 6130 WTOONSCOTEGDO Nova
Solna, Sweden Rev.03 06.00 SPSUD ‘Sones ProgramError messages
Er 853
“Incorrect checksum for rotation calibration menu in stand EEPROM”
Dieplayedon PC: WiteRead back fale for tae 4.in stand EEPROM
Description
Each time the MAMIMOMAT 3000 Nova is switched on, the adjustment data in the stand
EEPROM (080i 134) are checked with a checksum. Each menu under “Stand config’ has
lis own checksum. If discrepancy was detected in the rotation callbration menu, the
Fault elimination
Install all stand parameters using the service PC “Back-up" menu.
Er 854
“Incorrect checksum for lift parameters menu in stand EEPROM”
Displayed on PC: WiiteRtead back laure for tblo Sin stand EEPROM
Description
Each time the MAMMOMAT 3000 Nova is switched on, the acjustment data in the stand
EEPROM (D801.134) are checked with a checksum. Each menu under “Stand cortigr has
‘ts own checksum. If a discrepancy was detected in the lit parameters menu, the above
mentioned errors displayed.
Fault elimination
‘nstall all stand parameters using the service PC “Back-up” menu.
Er 855
“Incorrect checksum for best compression menu in stand EEPROM”
Displayed on PC: Write/ead back lure for able 8 in land EEPROM
Description
Each time the MAMMOMAT 3000 Nova is switched on, the adjustment data in the stand
EEPROM (D801 94) are checked with a checksum. Each menu under*Stand config’ has
its own checksum. Ifa discrepancy was detected in the bost compression menu, tho
‘above mentioned error is displayed.
Fault elimination
Install all stand parameters using the service PC “Back-up" menu,Er 856"
“Incorrect checksum for rotation parameters menu in stand EEPROM”
Deplayed on PC: Wiitoad back flue for tabe 7 in stand EEPROM
Description
Each time the MAMMOMAT 3000 Nova is switched on, the adjustment data in the stand
EEPROM (D801 134) are checkad with a checksum, Each menu under “Stand config” has
lis own checksum. Ifa discrepancy was detested in the rotation parameters menu, the
above mentioned errors cisplayed.
Fault elimination
Install al stand parameters using the service PC “Back-up” menu,
Er 857
“Incorrect checksum for beam limiting device menu in stand EEPROM”
Displayed on PC: Wirtefeed back fare for table 8 stand EEPROM
Description
Each time the MAMMOMAT 3000 Nova is svitchad on, tho adjustment data in the stand
EEPROM (D801 194) are checked with a checksum. Each menu under “Stand config” has
lts own checksum. fa discrepancy was detected in the beam limiting device menu, the
above mentioned error is cisplayed,
Fault elimination
Install all stand parameters using the service PC “Back-up” menu.
Er 858
“incorrect checksum for rotation parameters menu in stand EEPROM”
Displayed on: —WhtePead back Sale fortable 9 in stand EEPROM
Description
Each time the MAMMOMAT 2000 Nova is switched on, the adjustment data in the etand
EEPROM (D801.134) are checked with a checksum. Each menu under “Stand config" has
its own checksum. Ifa discrepancy was detected in the rotation parameters menu, the
above mentioned error is cisplayed.
Fault elimination
Install all stand parameters using the service PC "Back-up" menu
Sieméne-Eioma AS ogerars ‘SPBT ROTTS Page 29 of WICOR/SOO0G000 Nowa
Solna, Sneden ev.03 05.00 SPS-UD ‘Servibo ProgramError messages
Er 859
“Incorrect checksum for rotation calibration menu in stand EEPROM”
Digpiaysd on PC: White/Reed back flue for table 10 in stand EEPROM.
Description
Each time the MAMMIOMAT 3000 Nova is switched on, the adjustment data in the stand
EEPHOM (D801 194) are checked with a checksum, Each menu under “Stand contig” has
lis own checksum. Ia discrepancy was detected in the rotation calibration menu, the
‘above mentioned error is displayed
Fault elimination
Insta all stand parameters using the service PC "Back-up" menuChap. 0 Front page and table of contents new generated
Page 2-3 ‘Text change in picture
Pago2-4 ‘Text change in picture
Page 2-5 Notice is new
Page 2-6 Text change in picture and in text
Page 2-7 ‘This section is new
Page 2-10 ‘Text change
Pago 2-14 Text change
Page 2-19 Text change in notice
Pago 2-20 Text change in text, picture and in notice
Page 2-23 Text change in picture
Page 2-47 ‘Text change in picture and in text
Page 2-€0 Notice is new
Semone Bea 55 FaakioS SPETAIOTAOS Page O12 MTOUEODUUOO Nova
‘Soin, Swocen Rev.03 05.00 SPS. ‘Senice ProgiamnThis page intentionally left blank,
Changes to previous version