Professional Documents
Culture Documents
Training Manual
Typically the flow system consists of an Inlet separator, filter holder, flow meter,
flow controller and a pump (Figure 1).
The BAM1020 also has a flow system, measuring system and a data system. The
advantage of the BAM1020 is in the automation of the measuring system and the
simplification of both the data and flow systems. Each system can be
independently audited to assure accurate data. The following sections give details
on verifying the three systems of the BAM1020.
A. Leak Check
Tools Required: BX-305 or BX-302
Silicone Lube
Step 1. If the BAM102 is installed in a shelter remove the PM10 head (if
necessary also the PM2.5 SCC). If the unit isn’t installed insert the BAM
inlet adapter from the BX-305/302 (Note: if you have a BX-305 you will
not have the Zero filter) into the inlet of the BAM1020. See Figure 4
Figure 4 BX-302/305 Installation
Step 2. Place the Inlet coupler onto the BAM1020 inlet. Press it firmly onto the
inlet until it both O-rings are completely sealed to the inlet. This step may
require a small amount of silicone lube on the inlet to allow proper
Troubleshooting
Step 1. Check the Zero on the Flow meter.
a. Remove the pump tubing from the quick disconnect fitting on the back of
the BAM.
b. From the main menu press TEST. In the TEST menu select PUMP and
turn the pump on.
c. Flow should be less then 0.2 LPM. If the flow is greater than this value do
a flow audit from page 56-57 from manual.
d. Retry the LEAK TEST PROCEDURE.
Step 2. Check all connections from the inlet adapter to the vacuum pump.
a. Remove the inlet adapter visually check for leaks and loose fittings and
replace.
b. Remove the inlet tube from the BAM1020. Visually inspect for leaks or
loose fittings and replace. Check the two orings in the BAM1020 inlet for
wear or splits.
c. Open the front door on the BAM1020. Manually lift and lower the nozzle
letting it slowly contact the filter paper. The nozzle should be free to move
up and down. If the Nozzle sticks contact the factory and request a BX-
308 kit.
d. Remove the tubing from the back of the BAM1020 and inspect for splits
or loose connections. If the tube is split, cut out the bad portion and
reinsert into the back of the BAM1020
e. Retry the LEAK TEST PROCEDURE.
Disconnect and
plug here.
If the leak is below the inline filter then the flow meter has a leak or is
faulty. Replace with a new flow meter. MOI part number 970608
An option for the BAM1020 is the RH controlled heater. This option uses an RH
sensor and a temperature sensor located below the filter paper. Testing has shown
that if the RH on the filter paper increases above 55% the particulate can absorb
water and the measured mass will increase. The RH controlled heater minimizes
this effect. Met One suggests setting the RH value to 45%. This is the
equilibration value for the FRM filters.
Secondly, the filter temperature and ambient temperature are measured and a
Delta-Temperature is computed from the difference. Volatile organic compounds
(VOCs) and Semi-VOCs can be vaporized and removed from the measured mass
if the filter temperature is greater than the ambient temperature by more than 5
degrees C. Met One suggest setting the Delta-T set point to 8 degrees Celsius.
Note: Delta-T is only available if an ambient temperature probe (BX-592) is
connected to channel 6.
Heater is enabled if the BAM-1020 is on and the sample pump is on. The heater
will turn off when RH is 1% below Setpoint or if the Delta-T Setpoint is reached.
Heater Setup
RH Control: YES
RH Setpoint: 45 %
Datalog RH: YES (Chan 4)
Delta-T Control: YES
Delta-T Setpoint: 5 C
Datalog Delta-T: YES (Chan 5)
Save Cancel
RH Control Options are YES or NO. If YES is selected the sample heater
will be turned on until the RH is within 1% of the RH Setpoint.
RH Setpoint The Setpoint can be a number from 1-99. This is the value of
RH that will be maintained at the filter.
Delta-T Control Options are YES or NO. If YES is selected the sample heater
will be turned off if the Delta-T Setpoint is exceeded. If the
Delta-T Setpoint is exceeded by 1 degree Celsius an error (N)
will be logged. See Section 4.16.
Datalog Delta-T Options are YES or NO. If YES is selected the Delta-T will be
logged on channel 5 of the BAM1020 Datalogger. Note: if
YES is selected external inputs on channel 5 are ignored.
A. Self Test
The BAM-1020 has a built-in self-test function. Press the TAPE key and then
press the SELF TEST key. The Self-Test Menu is shown in Figure 9.
02/08/1999 15:29:30
LATCH: OFF TAPE BREAK: OK
CAPSTAN: OK TAPE TENSION: OK
NOZZLE DN: OK SHUTTLE: OK
NOZZLE UP: OK REF EXTEND: OK
FLOW: OK REF WITHDRAW: OK
Status: SELF TEST PASSED
TENSION SELF TEST EXIT
Upon depressing the SELF TEST key the will automatically go through a series
of internal checks displaying, in sequence, the results of these checks. If the check
passed, the user will see the word OK displayed, or if it does not operate
correctly, the user will see the word FAIL displayed. With the large display it is
not necessary to watch each segment being tested, as all of the results will be
displayed upon completion of the test. The test will take approximately 4 minutes.
The results of a properly operating unit are shown below.
LATCH
Checks the position of the pinch roller latch, if the LATCH is OFF then the
capstan will engage the pinch roller. If the LATCH is ON the capstan will not be
able to shuttle the filter tape. If the LATCH is ON during normal operation the
error message will be shown on the display. See Figure 12 number 1 in the BAM
manual.
CAPSTAN
Capstan operation is tested by moving the paper forward and then back, this test is
keyed by proper operation of photo-interrupters. See Figure 12 number 6.
NOZZLE DN (DOWN)
Checks the nozzle movement and reports the proper operation of the ‘nozzle
down’ photo-interrupter sensor.
NOZZLE UP
FLOW
The pump is turned on and the flow is tested, to verify operation for operation
between 10-20 LPM.
TAPE BREAK
Detects broken tape by looking at the operation of the feed and take up motors
photo interrupters. The motors are un-tensioned and then tensioned and the
change in the photo interrupters.
TAPE TENSION
Checks the condition of the IDLER rollers, verifies operation of the photo
interrupters on the tension idlers.
SHUTTLE
Moves the shuttle left and right and verifies correct operation using the photo
interrupters.
B. Membrane Test
CALIBRATION MODE
REF MBRN: <
COUNT (Io): 634000
COUNT (I): 556234
CAL MASS M: 0.801 mg/m3
START STOP EXIT
COUNT (I0 )
Indicates the total count during a four-minute period of filter paper.
COUNT (I)
Indicates the total count during a four-minute period of filter paper with
the membrane installed.
START
Starts the calibration cycle, the counting will immediately begin. After 4
minutes the counting of I0 will stop the membrane will be installed and the
count I will begin. At the completion of the cycle the counting will stop
and the Calibration Mass of the membrane will be calculated.
A. Analog Voltage
The DAC test provides a simple way to verify the digital to analog conversion
electronics and to verify correct data logger setup. The front panel display shows
a keyboard-controllable voltage that the DAC is providing to the rear panel
terminal strip.
Note: Use cursor keys to alter the values. Up/Dn steps the voltage in millivolt
increments. Left cursor sets the output to zero. Right cursor sets the output to full
scale. The DAC Output Test Menu is shown in Figure 11.
0.000 VOLT
EXIT
Analog outputs of the BAM-1020 are selectable and may be set for isolated
voltage (0-1 Vdc or 0-10 Vdc) and isolated current (4-20 or 0-16 mA). Rear panel
dipswitch is used to make selection, analog signal is located on rear panel. The
full-scale value of both outputs is determined by the concentration range setting in
the INTERFACE SETUP screen.
B. Download Data
Printers, computers and modems may be connected to the RS-232 ports on the
rear panel.
Serial interface #1 handles data transfer and instrument operation status. This
interface is often used with modem for remote communications. For ease of
operation this terminal is supplied with a switch to select ‘modem’ or ‘computer’
operation. The switch and connections are located on the rear panel. Recorded
data may be retrieved using the RS-232 communications serial port when used
with a PC , Data Transfer Module, or Modem.
Serial interface #2 is output only and may be used with serial printer or computer.
Output may be set for date, time, data printout or may be set to one of three
diagnostic modes (Refer to Appendix D, command a of the BAM1020 manual).
Software packages include AutoMet Report, Air Plus and MicroMet Plus.
The BAM-1020 may also be used with simple terminal software to retrieve data
and setup station parameters. Specific communications features include; setting
of the Date, Time, erasing the BAM-1020 memory, binary downloading of data,
and the ability to view the current BAM-1020 setups. Any simple PC
communications or terminal program can also be used for monitoring of the
system or collection of recorded data.
The following gives details on how to download data using a terminal
program with a computer directly connected to a BAM.
1. Open the terminal software. If Windows HyperTerminal is used
go to START / PROGRAMS / ACCESSORIES /
COMMUNICATIONS / HyperTerminal.
2. Open HyperTerminal, enter a name and press OK. A Connect To
window will open like
*H
By typing one of the number keys, the associated function will be performed.
See Appendix C of the manual for command details.