You are on page 1of 2

Safety, Reliability,

and Flexibility
Aitecs 2016
Advanced Syringe Pump
Aitecs 2016 Advanced Syringe Pump

Designed for Safety,


Ease-of-Use, and Reliability

Specifications
Flow rate range 0.01 - 99.99 ml/h in 0.01 ml/h steps, 100.0 - 999.9 ml/h
in 0.1 ml/h steps, 1000 - 2200 ml/h in 1 ml/h steps

Mechanical accuracy ± 1%

Volumetric accuracy with ± 2% or ± 0.1 ml/h (whichever is greater)


approved syringes

Compatible syringes, ml 2/3, 5, 10, 20, 30 and 50/60 of all major brands

Occlusion pressure L1 (approx. 50 mmHg) - L10 (approx. 950 mmHg)


(adjustable)

Dose units* g, mg, mcg, ng, Units, kUnits, mUnits, mcUnits, mol, mmol,
mcmol, mEq, and other configured by the medical personnel

Purge volume 0.1 - 4.0 ml in 0.1 ml steps


The Aitecs 2016 offers enhanced safety,
reliability, ease of use, and flexibility. Bolus/purge rate 1 - 2200 ml/h in 0.1 ml/h steps

The Aitecs 2016 is extremely versatile Bolus volume 0.1 - 60.0 ml in 0.1 ml steps

and meets the needs of all patient Keep Vein Open (KVO) rate 0.1 ml/h - 5 ml/h

types from neonatal to adults. KVO volume 0.1% - 10% of the syringe volume

Volume to be infused (VTBI) 0.1 - 99.9 ml in steps 0f 0.01 ml, 100 - 999 ml in steps of
4 Accepts syringes 0.1 ml, 1000 - 9999 ml in steps of 1 ml or No Limit
from 2/3 to 60 ml
Post VTBI rate Stop, KVO, or continue with set rate
4 Customized drug
Standby time 1 min - 24 h
library (up to 300
drug labels) Alarms Visual and audio

4 Highly visible pump Dose error reduction Infugard® protocol (with safety limits)
status indicator
Interfaces USB (peripheral), IrDA (IR Comms port), Nurse Call - optional
4 Easy intuitive
programming Mounting Table top operation, Universal pole clamp, Draeger
(horizontal) bar, Infusion device rack
and operation
Power supply 100 - 240 VAC, 50/60 Hz, 50 VA or internal rechargeable battery
4 Automatic
capturing of Battery type Built-in NiMH 7.2 V / 2.5 Ah

syringe plunger Battery life 10 h @ 5 ml/h


4 Improved alarm Battery charging time 5 hours to 100% charge
signals
Event log > 2000 events
4 Graphic LCD
display History log > 500 events

4 Rate range from Key log > last 300 presses


0.01 to 2200 ml/h Protection Class II, defibrillation-proof type CF applied part, IPX1
4 Advanced communication (USB, IrDA interfaces)
CE marked Council Directive 93/42/EEC (14 June 1993) concerning medical devices

Electrical safety and EMC IEC 60601 - (1, 1-2, 1-6, 1-8, 2-24)
* available in Universal version
Weight 2.3 kg

Dimensions 120 x 320 x 137 mm

Accessories
Infusion device racks Centralized power distribution: IDR 16-03/06/08
Including communication capabilities: IDS (03-08)

Fixation Rotatable pole clamp

Software tools AIMS 16 infusion monitoring system

Distributed by:

UAB “Viltechmeda” tel +3705 2776745


125 Kalvariju St. LT-08221 fx +3705 2763867
Vilnius, Lithuania www.moog.com/medical
www.aitecs.com
Moog is a registered trademark of Moog, Inc. All other trademarks indicated herein are the
property of Moog, Inc. and/or its subsidiaries. © 2009 Moog, Inc. All rights reserved.