You are on page 1of 8

Food Control 17 (2006) 719–726

www.elsevier.com/locate/foodcont

Quality assurance of volumetric glassware for the determination


of vitamins in food
a,*
I. Castanheira , E. Batista b, A. Valente a, G. Dias a, M. Mora b, L. Pinto b, H.S. Costa a

a
Department of Food Safety and Nutrition, National Institute of Health Dr. Ricardo Jorge, Av. Padre Cruz, 1649-016 Lisboa, Portugal
b
Central Metrology Laboratory, Portuguese Institute for Quality Rua António Gião, 2 Caparica, Portugal

Received 2 December 2004; received in revised form 7 April 2005; accepted 11 April 2005

Abstract

Volumetric measurements are currently used in the analytical process to determine the vitamins content in foods. The variations
in stability and integrity of volumetric apparatus, although relatively small when compared with other sources of error, should be
monitored. The main purpose of this study was to investigate and identify a number of variables which, in daily use, can affect the
life cycle of volumetric glassware. The role of in-house metrological hierarchy, the operatorÕs performance and the effect of temper-
ature was evaluated for single volume flasks, single- and multi-mark pipettes. Biased nominal capacity was applied in the calculation
of water-soluble vitamins (B1, B2, B6, niacin) and fat-soluble vitamins (A, D, E) content. Persistent bias due to volumetric glass-
ware, can reach 11%, in the results obtained for different vitamins content. The largest bias was observed for vitamin B1, perhaps
due to the high amount of volumetric glassware used in the analytical procedure. This is an indication that the cost–benefit of a
calibration plan should be evaluated in a rational way. The regular checking of volumetric glassware for stability and integrity is
discussed on the light of the latest revision of ISO 17025.
Ó 2005 Elsevier Ltd. All rights reserved.

Keywords: Volumetric glassware; Bias; Vitamins; Food; Metrology in chemistry

1. Introduction now explicitly requested in guidelines dealing with food


measurements (Codex, 2002; EURACHEM, 2000).
Quality assurance of volumetric glassware is a rele- The quality criteria of volumetric glassware has been
vant step of the food analytical process (Anglov, Peter- studied in recent decades and granted through the use of
sen, & Kristiansen, 1999; Wood, Nilsson, & Wallin, metrological procedures, supported on traceability of
1998), and there are cases where volumetric methods nominal value and its confidence interval. ISO (ISO
are adopted as CEN methods: Foodstuffs––Determina- 835/1; ISO 4787) in the early 1980s published a series
tion of vitamins by high performance liquid chromatog- of standards where the concept of tolerance limits based
raphy (HPLC), (EN 12821; EN 12822; EN 14164; EN on repeatability of the calibration procedure was intro-
14152; EN 12823-1; EN 14122; CEN/TC 275/WG 9). duced. Studies on stability of volumetric glassware were
Calibrated glassware and traceability to SI units is described later in the 1980s by Burfield and Hefter
(1987). These concepts have recently been reviewed in
the literature (EURACHEM, 2000; Kadis, 2004): The
*
tolerance limits are calculated in compliance with the
Corresponding author. Tel.: +351 21 751 93 17; fax: +351 21 752
6400.
Guide ‘‘Quantifying Uncertainty in Analytical Measure-
E-mail address: isabel.castanheira@insa.min-saude.pt (I. Castan- ment’’, issued by EURACHEM/CITAC (2000); volu-
heira). metric glassware is considered appropriate if the

0956-7135/$ - see front matter Ó 2005 Elsevier Ltd. All rights reserved.
doi:10.1016/j.foodcont.2005.04.010
720 I. Castanheira et al. / Food Control 17 (2006) 719–726

nominal capacity value and its associated uncertainty re- calibrated against reference standards at a Food Analy-
main within acceptable limits. In the case of volumetric sis Laboratory (FAL) using an appropriate protocol.
glassware used in daily measurement practice, the expiry The nominal value of volumetric glassware, routinely
date of the calibration result, calibration plans and used in food analysis, was checked against the assigned
calibration intervals are defined by the laboratory and value of the working standard. A set of three or five
usually operation conditions are not mentioned. 20 ml and 50 ml flasks and 1 ml, 2 ml and 5 ml pipettes
However, this procedure tends to change in the future currently used in vitamin content determinations were
due to improved actions introduced by accreditation submitted to several tests described below.
bodies, as consensus exists, in the revision of ISO/IEC
17025 to include regular checking of equipment for 2.1.2. Metrological criteria
stability and integrity (FDIS ISO 17025, 2004). Class A volumetric glassware was calibrated at the
Confidence in the whole life cycle of the measurement NMI and calibration results issued by the NMI were
equipment, from its design to use, framed by metrolog- analysed in terms of the assigned value and associated
ical parameters is a thematic area that has received uncertainty. The acceptance criterion was defined
much attention in the field of legal metrology. Substan- according to ISO 4787 (1984). Compliance of the as-
tial information on this area exists in OIML D 20 signed value with the acceptance criteria is defined by
(1988), OIML D 27 (2001), Vaz, Castanheira, Fino, Eq. (1) and it was evaluated on the basis of accepted
and Van der Veen (2000), Sommer, Chappell, and or rejected.
Kochsiek (2001). The equipment is considered appropri- jxn  xr j
ate during the time in which deviation from the true qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi < TL ð1Þ
value remains under acceptable limits. The checking U 2n þ U 2a
intervals for the equipment in place are designated as
a periodic verification, and the decisions on equipment where xn is the nominal value, xr is the reference value,
performance are based on maximum permissible errors, Un is the uncertainty of the nominal value, Ua is the
as described in OIML recommendation. uncertainty of the assigned value and TL tolerance
This work is a three-step procedure dealing with limits defined in ISO 1042 and ISO 835.
monitoring of volumetric glassware used in the measure- Volumetric glassware was accepted as reference stan-
ment of vitamins content in food in compliance with the dard, when the assigned value and associated uncertainty
latest revision of ISO 17025 (FDIS ISO 17025, 2004). In were within the acceptance limits. The glassware was des-
the first step, an approach to establish in-house metro- ignated as WSVG when the assigned value plus or minus
logical hierarchy is discussed. The identification of crit- the associated uncertainty lies outside one of the accep-
ical parameters in the calibration procedure is presented tance limits. The glassware was rejected as RSVG, when
in the second step. Finally, in the third step, the impact the assigned value and associated uncertainty did not fall
of systematic error due to volumetric glassware in the in the previous judgement of compliance. In order to
measurement of vitamins B1, B2, B6, niacin, A, D and provide confidence of measurements, a calibration pro-
E content was estimated. gramme was developed and RSVG were calibrated once
a year by the NMI. The calibration results for time zero,
years two and four were compared with the maximum
2. Materials and methods acceptable deviation defined by Eq. (1), and used to
establish the external calibration intervals.
2.1. Metrological qualification of volumetric glassware
2.2. Calibration procedure
2.1.1. Metrological hierarchy for volumetric glassware
Laboratory glassware was classified according to its 2.2.1. Traceability
metrological grade. This was based on an established The in-house calibration hierarchy used to achieve
hierarchy of traceability chain to SI units through na- traceability to SI units was composed of: RSVG, WSVG
tional standards of measurements and framed by stan- and volumetric glassware. The reference value of capac-
dard definitions adopted by the International ity was obtained by a gravimetric process at the NMI,
Vocabulary of Metrology (VIM) (BIPM, 1993). The fol- according to ISO 4787 (1984), ISO 835 (1981) and ISO
lowing Reference Standards Volumetric Glassware 1042 (1998). This was achieved applying a weighing pro-
(RSVG) were calibrated at the National Metrology cedure consisting of 10 weighing cycles, where in each
Institute (NMI): burettes, pipettes, flasks and cylinders cycle, the mass corresponding to volume capacity was
with capacities in the range of 5–500 ml, and stored obtained by the difference between the filled vessel with
under appropriate conditions. distilled water and the empty vessel. The equation of
Working Standards Volumetric Glassware (WSVG) proportionality that converts the average mass to
(pipettes of 2 ml and 5 ml; flasks of 20 ml and 50 ml) were volume measurement is described as:
I. Castanheira et al. / Food Control 17 (2006) 719–726 721
   
1 q scribed above. The ‘‘blind’’ test was carried out by an
V 20 ¼ ðI L  I E Þ   1 A
qw  qA qB analyst with less experience in the calibration procedure,
 ð1  cðt  20ÞÞ ð2Þ and consisted on the evaluation of results at the gradu-
ation line under study, taking into account three param-
where V20 is the volume at 20 °C (reference tempera- eters: (a) reported value and associated uncertainty; (b)
ture), in ml; IL and IE, the weight of the vessel with water precision and (c) meniscus reading. The ‘‘blind’’ results
and of the empty vessel, respectively, in g; qw, the den- were benchmarked to the ones given by a NMI operator
sity of water, at calibration temperature, in g ml1; qA and adequate statistical tests were applied. The criteria
and qB, the density of air and of the mass pieces, in used to classify the operatorÕs performance were: poor
g ml1 (reference values of 0.0012 g ml1 and 8.0 g ml1, and good performance. This was derived from the sum
respectively); c, the coefficient of cubic thermal expan- of the overall results. Each analyst result was consid-
sion of the material, in °C1 and t, the temperature of ered: satisfactory; questionable or unsatisfactory, if the
the water used in the calibration, in °C. value was respectively, below, equal or above the tabu-
lated value of the statistical test. The operatorÕs perfor-
2.2.2. Measurement uncertainty mance was considered as good, if 75% of the values of
The uncertainty of the glassware calibration process the overall result were satisfactory. The remaining situ-
is defined as a parameter associated with the result of ations were considered as poor performance. This proce-
a measurement that characterizes the dispersion of the dure was applied to two types of volumetric glassware:
values that could be attributed to the measurand, pipettes and flasks of different capacities.
according to EURACHEM Guide (2000). The expres-
sion for the combined standard uncertainty (u(V20)) 2.4. Temperature conditions
was developed from Eq. (2) using the error propagation
formula. The major contributions to the combined stan- The effect of temperature on the nominal capacity
dard uncertainty were estimated through a function of volumetric glassware was carried out at NMI follow-
model, presented in Eq. (3), and considering seven build- ing a three step experimental protocol: first, the glass-
ing blocks: repeatability mass; meniscus; water density; ware was calibrated according to the calibration
temperature; air density; density of mass species and procedure described above; secondly, the glassware
thermal expansion of the material. The expanded uncer- was submitted to a chosen temperature during a period
tainty (U) defined ‘‘as an interval within which the value of time and in the third step, the glassware was equili-
of the measurand is believed to lie with a particular level brated at standard reference temperature and then sub-
of confidence’’ EURACHEM (2000) was calculated mitted to the calibration procedure again. The effect of
using a coverage factor of k = 2, considering 95% as temperature is given by the difference between the vol-
approximate level of confidence. ume at graduation line obtained in the first and third
" 2  2 step.
oV 20 oV 20 Pipettes and flasks were withdrawn from daily rou-
uðV 20 Þ ¼ u2 ðmÞ þ u2 ðqw Þ
om oqw tine laboratory work, divided into three different groups
 2  2 and submitted to this procedure. In group I, the glass-
oV 20 oV 20
þ u2 ðqA Þ þ u2 ðqB Þ ware was submitted daily to a temperature of 4 °C dur-
oqA oqB ing five days (120 h). In group II, the temperature was
 2  2 #12
oV 20 oV 20 set daily (24 h) to 50 °C during five days and the calibra-
2 2 2
þ u ðcÞ þ u ðtÞ þ u ðdV men Þ tion procedure was performed at day one, three and five.
oc ot
In group III, the glassware was maintained during five
ð3Þ days at 50 °C and submitted to the calibration proce-
dure at day five.
where u(m) is the uncertainty of the weighing; u(qw), the
uncertainty of water density; u(qA), the uncertainty of air 2.5. Testing for persistent bias
density; u(qB), the uncertainty of the mass pieces; u(c),
the uncertainty of the coefficient of cubic thermal expan- 2.5.1. Determination of vitamin content
sion of the material; u(t), the uncertainty of the temper- Vitamin determination was done by conventional
ature and u(dVmen), the uncertainty of the meniscus. methodology. This consisted on saponification for fat-
soluble vitamins and hydrolysis for water-soluble vita-
2.3. Operator’s performance mins, followed by liquid–liquid extraction and HPLC
analysis according to the methods described in the liter-
The influence of the operatorÕs performance on the ature: Vitamins A (EN 12823-1), D (EN 12821), E (EN
capacity value of WSVG was evaluated through a cali- 12822) and vitamins B1 (EN 14122), B2 (EN 14152), B6
bration test, using the same experimental procedure de- (EN 14164), niacin (CEN/TC 275/WG 9).
722 I. Castanheira et al. / Food Control 17 (2006) 719–726

2.5.2. Bias due to volumetric glassware


The potential bias of the analytical procedure, due to
laboratory volumetric glassware was studied using two
protocols (A) and (B), and assuming that no other
variation occurred in the measuring system during the
determination of vitamin content. In protocol (A), mea-
surement results were obtained according to standard
operation procedure. In protocol (B) and for each step
of the analytical procedure, the ‘‘nominal value’’ of
volumetric glassware was estimated by adding to the
reference value, the worst value within experimental
variability for the operator and temperature tests. Bias
was considered significant, if the relative difference
between protocol (A) and (B) was larger than the
expanded uncertainty of this difference.

2.6. Statistical analysis


Fig. 1. Calibration results of volumetric glassware as reference
HPLC response characteristics were estimated standards.
through calibration functions obtained by least-squares
regression analysis. The following parameters were con-
sidered: polynomial coefficients and their uncertainties, Reference standard
residuals and regression coefficients. Mean and standard Working standard
deviations of analytical data for vitamins content were UAL
calculated for a significance level of P < 0.05.
The following statistical tests were applied for data
analysis: F-test was used to compare variances of the
replicate analysis, En-score was used to compare two
independent means with the same number of observa- xr
tions and single outlier selection was done by DixonÕs
test (Miller & Miller, 1988).

3. Results and discussion LAL

An overview of calibration results for RSVG is given


in Fig. 1. It can be seen that for different types of
volumetric glassware, the percentage of glassware cali- 0 2 4
brated that failed the acceptance criteria varied from Time (years)
0% up to 40%. Cylinders are in agreement with the
acceptance criteria defined above; 10% of the flasks cal- Fig. 2. Examples of calibration results with reference value and
ibrated do not fulfill this criterion; 20% of burettes can- associated uncertainty: xr, reference value; UAL, upper acceptable
limit; LAL, lower acceptable limit.
not be labeled as reference standards and for pipettes,
40% of the candidate standards were not compliant with
the specifications of the acceptance criteria as reference An indirect calibration procedure (Rodrı́guez, Gra-
standards (BIPM, 1993). Calibration results issued for cia, López, & Sánchez, 2001) which consists on the con-
pipettes as RSVG are presented graphically in Fig. 2, version of the mass into volume and comparison of the
for time zero, year two and four, revealing that appro- deliverable water of RSVG and WSVG, on a periodi-
priate care of volumetric glassware was taken. The same cally time schedule of six months and under the assump-
graphic profile was obtained for flasks. These results, tion that measurement uncertainty remains constant,
indicating that nominal capacity remained constant was also used.
overtime is probably due to consistent storing and clean- The establishment of in-house metrological hierarchy
ing conditions as described by accreditation bodies based on an indirect calibration procedure of WSVG,
(UKAS, 2002). For working standards more than 90% with more rigorous specifications than the manufacture
of pipettes and flasks passed the defined acceptance tolerances, allows regular checking of volumetric glass-
criteria. ware for stability and integrity of daily used glassware
I. Castanheira et al. / Food Control 17 (2006) 719–726 723

without lost of traceability links. This procedure is more


advantageous than to refer all volumetric glassware to
an external link (European Commission, 1998; EURA-
CHEM, 2003).
Evaluation of the measurement uncertainty, in the
calibration procedure of reference standards, was based
on a model which consists on the contribution of several
standard uncertainties, quantified in terms of experi-
mental standard deviation times the sensitivity coeffi-
cient of each source, as described in Eq. 3. This
empirical model was evaluated by the Ishikawa (1976)
diagram and based on type A (experimental) and type
B (estimated) uncertainties inputs which combines
intralaboratory repeatability and reproducibility with
parameters (e.g.: water density; thermal expansion coef-
ficient) whose contributions derive from tabulated
values published in the literature (Spieweck & Bettin, Fig. 3. Schematic representation of sources and main components of
1992). All the values obtained for the expanded uncer- measurement uncertainty, for the calibration procedure, performed
against national standards.
tainty were below the predicted uncertainties of 0.1%;
0.2%; 0.3% (UKAS, 2002). The values of reference
standards were well within the variation claimed by
the manufacture (Noack, 1998) and in agreement with provided by AFAL are different from the assigned value
the results of Euromet intercomparisons (Euromet, (Armishaw, King, & Millar, 2003). The analystÕs preci-
2004). The uncertainty budget provides a mean to com- sion (ANMI versus AFAL) was evaluated using F-test
pare different contributions to the combined standard and the results were compared with the tabulated values
uncertainty (Fig. 3). The largest contribution came from (Miller & Miller, 1988) demonstrating that AFAL has
experimental sources: mass determination, meniscus been working with a different level of precision. The
reading, temperature, precision of each determination evaluation of operatorÕs performance was grounded on
which includes the operatorÕs performance and which a measurement theory (Mari, 2003), implemented with
was found to be the highest source. Therefore, the oper- two kinds of tests: pair comparison and magnitude esti-
atorÕs performance was accounted for and was subject to mation using a set of reference values in all points of
a more detailed study. scale (Rossi, Crenna, & Codda, 2003). The judgement
The acceptable operatorÕs performance was evaluated of evaluation occurs in three steps: (a) single measure-
applying appropriate statistical tests to identify the main ment through outliers tests; (b) error propagation due
causes of error. Table 1 summarizes the results obtained to statistical evaluation of mean of results (c) estimation
by analysts from the National Metrology Institute degree of inaccuracy by null hypothesis. This approach
(ANMI) and from a Food Analysis Laboratory (AFAL) seems to be very useful to evaluate the analyst skills,
for flasks and pipettes, and for the same calibration pro- to identify the cause of error and to decide on further
cedure. Outliers were not found when DixonÕs test was training programmes. This is helpful not only during
applied to ANMI or AFAL results. checking and calibration procedure, but also in the over-
Significant differences between the results given by all analytical procedure, since a large number of vita-
the two analysts ANMI and AFAL were detected for both mins analysis measurements are based on volumetric
types of volumetric glassware. jEn-scorej > 1 was ob- measurements (Klejdus et al., 2004; Sharpless, Margolis,
served for flasks and pipettes indicating that the results & Thomas, 2000).

Table 1
Overview of statistical tests, applied to evaluate operatorÕs results
Flask En-score F-test Pipette En-score F-test
20 ml 50 ml 20 ml 50 ml 2 ml 5 ml 2 ml 5 ml
#1 22.5 26.4 9.00 4.00 #1 1.3 6.7 1.00 4.00
#2 14.8 15.6 1.00 1.00 #2 2.5 2.4 9.00 1.00
#3 3.1 7.4 4.00 1.00 #3 4.2 3.1 4.00 2.25
#4 22.6 10.7 1.00 4.00 #4 – 2.8 – 1.00
#5 4.5 9.9 4.00 1.00 #5 – 0.6 – 9.00
Criteria used to classify operatorÕs results: jEn-scorej > 1, unsatisfactory; jEn-scorej = 1, questionable; jEn-scorej < 1, satisfactory; F0.95(9,9) = 3.18.
724 I. Castanheira et al. / Food Control 17 (2006) 719–726

Deviation from the reference value due to tempera- glassware. Burfield and Hefter (1987) has concluded that
ture and exposure intervals is presented in Fig. 4. The for oven drying borosilicate glassware, the calibration
nominal capacity did not change when volumetric glass- value is not significantly affected. Our results are not
ware was maintained at 4 °C. These results are expected in agreement with BurfieldÕs conclusion possibly because
for borosilicate material (Burfield & Hefter, 1987). But the calibration methods and the uncertainty calculations
at 50 °C, significant effects were noticed, depending on applied here are different. In this study, uncertainty cal-
the type of volumetric glassware, capacity and exposure culations and the calibration process were determined
period. The highest deviation from reference value was through a metrological approach which is more rigorous
observed at day five for both types of volumetric glass- to investigate differences between or within measure-
ware and the largest effect was observed for pipettes of ments results (De Bièvre & Taylor, 1997).
1 ml (Fig. 5). The results in the present study show that In daily use, machine washing and oven or air drying
temperature effect is different according to temperature of glassware at moderate temperatures (around 60 °C)
value, period of time, type and capacity of volumetric are unavoidable, for volumetric glassware used with
complex food matrices. Therefore, the glassware should
be submitted to a qualification performance test regu-
0.006 larly (Bedson & Sargent, 1996; UKAS, 2002).
4 ºC Fig. 6 shows the mean bias due to volumetric glass-
ware in the determination of vitamins content. Persistent
Deviation from reference value (%)

50 ºC
0.002 bias (Taverniers, De Loose, & Bockstaele, 2004), is de-
fined as the difference between the expectation of the test
results (protocol B) and the reference value (protocol
-0.002
A). Potential bias due to volumetric glassware was
investigated in all steps of the analytical procedure for
the determination of vitamins content and the largest
-0.006
contribution to bias was observed for the calibration
curve and extraction step.
-0.010 The operatorÕs performance and temperature effects
on volumetric glassware are the factors that influence
the most the nominal capacity. This enabled us to
-0.014 demonstrate that lack of accuracy can overestimate the
1 3 5 value of vitamins content, in the range of 1–11%, except
Time (days)
for Vitamin D where the results are unbiased. This
pattern of results strongly suggests the influence of
Fig. 4. Results of temperature tests for graduated pipettes as a
function of time.

0.4
Deviation from reference value (%)

0.3

0.2

0.1

0
1 2 5

Capacity of pipette (ml)


Fig. 6. Influence of bias due to volumetric glassware on reported value
Fig. 5. Temperature effect on nominal capacity of graduated pipettes of vitamin concentration. The largest source comes from low capacity
after five days at 50 °C. glassware.
I. Castanheira et al. / Food Control 17 (2006) 719–726 725

systematic errors in the preparation of the calibration Furthermore, collaboration between food analysis
curve. Vitamin D was quantified using an internal stan- and metrological laboratories, reveals to be very useful:
dard and for the other vitamins, the content was calcu- (a) to provide support for education and training of per-
lated by regression analysis using a calibration curve sonnel; (b) to improve the traceability of working level
with five external standards, which were prepared by a by increasing the quality of link to SI units of assigned
dilution procedure. The largest bias was observed for values and associated uncertainties; (c) to support the
vitamin B1, perhaps due to the high amount of volumet- organization of inter-laboratory comparisons, a tool to
ric glassware used in the analytical procedure, especially assess the ability of food laboratories to generate data
during the preparation of calibration curve. For vita- quality that may be utilized in aims such as establish-
mins A, E and B2, bias is introduced by low capacity ment of food/disease-health-status using different
volumetric glassware (pipettes and flasks) and for vita- sources of analytical data.
min B6 bias is due to pipettes of 2 ml and 5 ml. The
results obtained when a small number of flasks and pip-
ettes are used did not differ significantly in both proto- Acknowledgments
cols, while those obtained when a high number of low
capacity pipettes are used were systematically biased. The authors would like to thank Dr. Maria Antónia
From this study, we conclude that a cumulative effect Calhau, Head of Department-INSA and to Engo Antó-
of error in the dilution steps for the preparation of the nio Cruz, Director of IPQ/LCM for encouragement of
calibration curve is probably associated with the lack this work, to Dr. Luı́sa Oliveira and Enga Eduarda Fil-
of fit of linearity and can contribute significantly to less ipe, MD for providing laboratory facilities and to Hel-
accurate results. This study has implications concerning ena Bártolo for vitamins analysis. This work was
routine work in the determination of vitamins content in partially supported by PIDDAC/IPQ/LCM/Universid-
food and was motivated by the need of traceability and ades (1997–2004) and POCTI /CTA/95/2001.
comparability of results. Differences ranging from 10%
to 40% for the value of vitamin A, E, B1, B2 and B6
content have been reported by food analysis laborato-
ries and are largely connected with the expertise level References
of participants (Part 1: Hollman et al., 1993a; Part 2:
Anglov, T., Petersen, I. M., & Kristiansen, J. (1999). Uncertainty of
Hollman et al., 1993b; Mathiasson et al., 2002). In the nitrogen determination by the Kjeldahl method. Accreditation
determination of Vitamin A, in serum, the same conclu- Quality Assurance, 7(4), 504–510.
sion was reported (Hulshof, Brouwer, Burema, & West, Armishaw, P., King, B., & Millar, R. G. (2003). Achieving traceability
2002). Although variations in stability and integrity of in chemical measurement—a metrological approach to proficiency
testing. Accreditation and Quality Assurance: Journal for Quality,
volumetric glassware can be relatively small when com-
Comparability and Reliability in Chemical Measurement, 8(5),
pared to other sources of error, they need to be properly 184–190.
evaluated as defined by item no. 5.9.1, in the new Bedson, P., & Sargent, M. (1996). The development and application of
revision of ISO 17025 (FDIS ISO 17025, 2004), and guidance on equipment qualification of analytical instruments.
are necessary to improve collaborative tests results and Accreditation Quality Assurance, 1(6), 265–274.
BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML (1993). International
inter-laboratory comparability. This work raises the
vocabulary of basic and general terms in metrology (2nd ed.). ISO,
problem of correlating the nominal capacity variation Geneva, Switzerland.
which occurs in daily used volumetric glassware with Burfield, D. R., & Hefter, G. (1987). Oven drying of volumetric
the overestimation content of vitamins. Therefore, our glassware. Journal Chemical Education, 64(12), 1054.
results can contribute to improve the evaluation of CEN (European Committee for Standardization) (2000a). EN 12821
foodstuffs––Determination of vitamin D by high performance liquid
laboratory performances and comparability of results
chromatography––Measurement of cholecalciferol (D3) and ergo-
as recommended by Codex Alimentarius (Codex, 2002). calciferol (D2). CEN, Brussels, Belgium.
CEN (European Committee for Standardization) (2000b). EN 12822
foodstuffs––Determination of vitamin E by high performance liquid
4. Conclusions chromatography––Measurement of a-, b-, c- and d-tocopherols.
CEN, Brussels, Belgium.
CEN (European Committee for Standardization) (2000c). EN 12823-1
The present methodology evaluates the factors affect- foodstuffs––Determination of vitamin A by high performance liquid
ing the accuracy of volumetric glassware. This approach chromatography––Part 1: Measurement of all-trans-retinol and 13-
is robust and feasible, thus, enabling its adoption in cis-retinol. CEN, Brussels, Belgium.
food analysis laboratories. CEN (European Committee for Standardization) (2002). EN 14164
foodstuffs––Determination of vitamin B6 by high performance liquid
This study highlights the critical control points of lab-
chromatography. CEN, Brussels, Belgium.
oratory volumetric glassware addressing the need of CEN (European Committee for Standardization) (2003a). EN 14152
periodic check intervals and calibration in the measure- foodstuffs––Determination of vitamin B2 by high performance liquid
ment plan. chromatographys. CEN, Brussels, Belgium.
726 I. Castanheira et al. / Food Control 17 (2006) 719–726

CEN (European Committee for Standardization) (2003b). CEN/TC Kadis, R. (2004). Evaluation of measurement uncertainty in volumet-
275/WG 9 determination of niacin by high performance liquid ric operations: The tolerance-based approach and the actual
chromatography. CEN, Brussels, Belgium. performance-based approach. Talanta, 64(1), 167–173.
CEN (European Committee for Standardization) (2003c). EN 14122 Klejdus, B., Petrlová, J., Potěšil, D., Adam, V., Mikelová, R., Vacek,
foodstuffs––Determination of vitamin B1 by high performance liquid J., Kizek, R., & Kubáň, V. (2004). Simultaneous determination of
chromatography. CEN, Brussels, Belgium. water- and fat-soluble vitamins in pharmaceutical preparations by
CX/MAS/02/13, Codex Alimentarius Commission (2002). Codex high-performance liquid chromatography coupled with diode array
committee on methods of analysis and sampling (FAO/WHO). detection. Analytica Chimica Acta, 520(1–2), 57–67.
The use of analytical results: Sampling, relationship between the Mari, L. (2003). Epistemology of measurement. Measurement, 34(1),
analytical results, the measurement uncertainty, recovery factors 17–30.
and the provisions in Codex Standards. Agenda Item 9 of the 24 Th Mathiasson, L., Turner, C., Berg, H., Dahlberg, L., Theobald, A.,
Sessions, Budapest, Hungary, 18–22 November 2002. Anklam, E., Ginn, R., Sharman, M., Ulberth, F., & Gabernig, R.
De Bièvre, P., & Taylor, P. D. P. (1997). Traceability to the SI of (2002). Development of methods for the determination of vitamins
amount-of-substance measurements: From ignoring to realizing, a A, E and beta-carotene in processed foods based on supercritical
chemistÕs view. Metrologia, 34(1), 67–75. fluid extraction: A collaborative study. Food Additives Contami-
EURACHEM (2000). EURACHEM/CITAC. Quantifying uncertainty nants, 19(7), 632–646.
in analytical measurement (2nd ed.). EURACHEM/CITAC Guide, Miller, J. C., & Miller, J. N. (1988). Statistics for analytical chemistry
EURACHEM/CITAC, Teddington, UK. (2nd ed.). Chichester, UK: Ellis Horwood.
EURACHEM (2003). EURACHEM/CITAC Guide: Traceability in Noack, S. (1998). New tools: Expert systems for uncertainty budgets.
chemical measurement. A guide to achieving comparable results in Accreditation Quality Assurance, 3(11), 436–443.
chemicals measurement. Joint EURACHEM/CITAC working OIML (International Organization of Legal Metrology) (1988). Initial
group on measurement uncertainty and traceability. Available and subsequent verification of measuring instruments and process
from http://www.eurachem.ul.pt. (OIML D 20).
EUROMET––European Collaboration in Measurement Standards OIML (International Organization of Legal Metrology) (2001). Initial
(2004). Homepage. Available from http://www.euromet.org/pro- verification of measuring instruments using the manufactures quality
jects.html. Cited 15 October 2004. management system (OIML D 27).
European Commission (1998). Metrology in chemistry and biology: A Rodrı́guez, C. L., Gracia, G. L., López, A. E., & Sánchez, L. J. (2001).
practical approach report. EUR 18045 EN Luxembourg. Calibration in chemical measurement processes. I. A metrological
Hollman, P. C. H., Slangen, J. H., Wagstaffe, P. J., Faure, U., approach. TrAC Trends in Analytical Chemistry, 20(4), 195–
Southgate, D. A. T., & Finglas, P. M. (1993a). Intercomparison of 206.
methods for the determination of vitamins in foods. Part 1. Fat- Rossi, G., Crenna, F., & Codda, M. (2003). Measurement of quantities
soluble vitamins. The Analyst, 118(5), 475–480. depending upon perception by jury-test methods. Measurement,
Hollman, P. C. H., Slangen, J. H., Wagstaffe, P. J., Faure, U., 34(1), 57–66.
Southgate, D. A. T., & Finglas, P. M. (1993b). Intercomparison of Sharpless, K. E., Margolis, S., & Thomas, J. B. (2000). Determination
methods for the determination of vitamins in foods. Part 2. Water- of vitamins in food-matrix Standard Reference Materials. Journal
soluble vitamins. The Analyst, 118(5), 481–488. of Chromatography A, 881(1-2), 171–181.
Hulshof, P. J. M., Brouwer, J. T., Burema, J., & West, C. E. (2002). Spieweck, F., & Bettin, H. (1992). Review: Solid liquid density
Bias and random error in retinol measurements of laboratories in determination. Tm-Technishes Messen, 59, 237–292.
countries with populations with mild to severe vitamin A Sommer, K. D., Chappell, S. E., & Kochsiek, M. (2001). Calibration
deficiency. Clinical Chemistry, 48(11), 2061–2063. and verification: Two procedures having comparable objectives
Ishikawa, K. (1976). Guide to quality control. Tokyo: Asian Produc- and results. OIML Bulletin, XLII, 1, 5–12.
tivity Organization. Taverniers, I., De Loose, M., & Bockstaele, E. (2004). Trends in
ISO (International Organization for Standardization) (1998). ISO 1042 quality in the analytical laboratory. I. Traceability and measure-
laboratory glassware––One-mark volumetric flasks. ISO, Geneva, ment uncertainty of analytical results. Trends in Analytical Chem-
Switzerland. istry, 23(7), 480–490.
ISO (International Organization for Standardization) (2004). TC UKAS-Lab 15(2002). Traceability volumetric apparatus. United King-
CASCO/SC/WG 2525 ISO/IEC FDIS 17025: 2004––5.9.1 (f)––on dom Accreditation Service. Available from http://www.ukas.com/
general requirements for the competence of calibration and testing downloads/publications/Lab 15. Cited 15th October.
laboratories. ISO Geneva, Switzerland. Vaz, D., Castanheira, I., Fino, H., & Van der Veen, A. M. H. (2000).
ISO (International Organization for Standardization) (1981). ISO 835/ Uncertainty budgets and mpes in refractometry: A project study.
1 laboratory glassware––Graduated pipettes––Part 1: General OIML Bulletin, XLI, 4, 9–12.
requirements. ISO, Geneva, Switzerland. Wood, R., Nilsson, A., & Wallin, H. (1998). Quality in the food
ISO (International Organization for Standardization) (1984). ISO 4787 analysis laboratory. In P. S. Belton (Ed.), RSC food analysis
laboratory glassware––Volumetric glassware––Methods for use and monographs. Cambridge, UK: The Royal Society of Chemistry, pp.
testing of capacity. ISO, Geneva, Switzerland. 16–35, 235–236.

You might also like