You are on page 1of 6

Customer Service Department Fast Fax

To: FAROArm User From: Customer Service Department

6 Pages total including this page. Contact us immediately if pages are missing or not legible.

Procedures for Probe Calibration

Revised: June 29, 2011 LM-06REF176-109.doc


© 2011 FARO Page 1 of 6
Revised: June 29, 2011 LM-06REF176-109.doc
© 2011 FARO Page 2 of 6
Revised: June 29, 2011 LM-06REF176-109.doc
© 2011 FARO Page 3 of 6
Revised: June 29, 2011 LM-06REF176-109.doc
© 2011 FARO Page 4 of 6
Revised: June 29, 2011 LM-06REF176-109.doc
© 2011 FARO Page 5 of 6
NOTE:
Max error displays the displacement of the point with the largest displacement. Two sigma displays the
deviation of all the points taken during the calibration. In order to pass calibration, this value has to be less than
the single point accuracy specification of the arm

The arm is factory calibrated to the last joint of the arm. The probe calibration is a performed operation by the
user that defines the position of the probe tip relative to the last joint of the arm. Any movement of the
unrelated joint arm induces systemic error and adds uncertainty into the probe calibration. Hand tighten Probe
(Do not over tighten).

Practice and experience is required to correctly calibrate the probe. Guidance should always be used when
performing any of these three operations listed.
 When testing the accuracy of the equipment.
 On high accuracy projects
 When experiencing issues during troubleshooting phase.

If you are still experiencing difficulties, please contact the customer service department by telephone at
1-800-736-2771 or by e-mail at support@faro.com.

FAX QUESTIONS TO: (407) 333-8056

Revised: June 29, 2011 LM-06REF176-109.doc


© 2011 FARO Page 6 of 6

You might also like