Professional Documents
Culture Documents
net/publication/304908123
CITATIONS READS
2 960
4 authors, including:
Sunyoung Ahn
Yonsei University Hospital
2 PUBLICATIONS 5 CITATIONS
SEE PROFILE
All content following this page was uploaded by Sunyoung Ahn on 21 June 2017.
ORIGINAL ARTICLE
limits of agreement, and the number of outliers was are summarized in Table 2. The stability test revealed
limited (Fig. 1). Passing-Bablok regression analysis that many of the analytes results re-measured at 72 hour
revealed that there is definite linearity between two kinds after sampling showed statistically significant differences
of tubes and the values of R2 were near 1 for all cases when compared to the results at 0 hour. However, most
(Fig. 2). Also 95% confidence interval (CI) for intercept of the calculated differences according to the above-
of the regression included value zero and 95% CI of slope mentioned formula were located within the total allowable
included value one for the four analytes (data not shown error defined in prior to the analysis. There were two
in the figure), which means there are no constant or parameters which exceeded the total allowable error:
proportional differences between the results from two MCHC and MPV: -1.71% and -1.84% in MCHC, and
kinds of tubes. For the calculated difference (%) of these 8.06% and 8.24% in MPV, for BD Vacutainer EDTA tube
four analytes, they were all within the total allowable and Improvacuter EDTA tube, respectively.
error range defined in prior to the analysis. All the other
comparison results except for these four analytes showed 3. Differences
between K2 EDTA and K3 EDTA in Analysis
no statistically significant differences. of Hematologic Parameters
As for the red blood cell shrinkage effect, Hct from
2. Stability Test for BD Vacutainer EDTA Tube and Improvacuter K3 EDTA tube was 0.86% higher than
Improvacuter EDTA Tube from BD Vacutainer K2 EDTA tube with statistical
The statistic results of stability study for each BD significance (P<0.0001), in contrary to the previously
Vacutainer EDTA tube and Improvacuter EDTA tube reported studies [3] (Table 1). However, as mentioned
A B
Mean Mean
1.50 95% confidence interval 1.95 95% confidence interval
95% limits of agreement 95% limits of agreement
1.25 1.70
1.00 1.45
Improve 0 hr-BD 0 hr
Improve 0 hr-BD 0 hr
0.75 1.20
0.50 0.95
0.25 0.70
0 0.45
0.25 0.20
0.50 0.05
0.75 0.30
27.5 30.0 32.5 35.0 37.5 40.0 42.5 45.0 47.5 50.0 52.5 70 75 80 85 90 95 100 105 110
(BD 0 hr+improve 0 hr)/2 (BD 0 hr+improve 0 hr)/2
C D
Mean Mean
95% confidence interval 95% confidence interval
0.8 95% limits of agreement 95% limits of agreement
6
0.4
Improve 0 hr-BD 0 hr
Improve 0 hr-BD 0 hr
2
0
0.4 2
0.8
6
1.2
1.6 10
28 29 30 31 32 33 34 35 36 37 0 20 40 60
(BD 0 hr+improve 0 hr)/2 (BD 0 hr+improve 0 hr)/2
Fig. 1. Bland-Altman difference plot between BD Vacutainer and Improvacuter tubes for 4 analytes that showed statistical
significance (P <0.05) when compared using the paired t -test. (A) Hematocrit. (B) Mean corpuscular volume. (C) Mean
corpuscular hemoglobin concentration. (D) Erythrocyte sedimentation rate.
before, the difference was within the total allowable error differences between BD Vacutainer spray-dried K2 EDTA
and clinically insignificant. The estimated MCV from both tube and Improvacuter liquid K3 EDTA tube (P>0.05)
tubes also did not show clinically significant difference (Table 1), which means there was no dilution effect due to
(Table 1). When applying the above-mentioned reference the liquid form K3 EDTA anticoagulant.
range of MCV, only two cases out of total 100 subjects
(2%) were differently categorized due to the inconsistent DISCUSSION
MCV results of two tubes; one case was microcytic in BD
Vacutainer tube but normocytic in Improvacuter tube, A large number of articles have been published evaluating
the other case was normocytic in BD Vacutainer tube but many kinds of blood tubes for different analyzing purpose.
classified as macrocytic in Improvacuter tube (data not Particularly in Korea, many laboratories and hospitals
shown). have routinely used BD blood tubes. However, a variety of
For dilution effect, all the results of WBC count, RBC other brands of blood tubes have been introduced recently
count, and PLT count did not show statistically significant and related comparison and stability studies have been
A B
52.5 Passing-Bablok fit 110 Passing-Bablok fit
50.0 95% CI 105 95% CI
47.5 Linearity (95% CI) Linearity (95% CI)
100
45.0
Improve 0 hr
Improve 0 hr
42.5 95
40.0 90
37.5 85
35.0
80
32.5
y=0.9979x+0.3994 75 y=1.0151x 0.6145
30.0 2 2
R =0.9996 R =0.9996
27.5 70
27.5 30.0 32.5 35.0 37.5 40.0 42.5 45.0 47.5 50.0 52.5 70 75 80 85 90 95 100 105 110
BD 0 hr BD 0 hr
C D
Improve 0 hr
33
60
32
31 40
30
y=0.9837x+0.2164 20 y=0.9668x 0.1622
29 2 2
R =0.9955 R =0.9947
28 0
28 29 30 31 32 33 34 35 36 37 0 20 40 60 80 100 120
BD 0 hr BD 0 hr
Fig. 2. Passing-Bablok regression analysis: Comparison between BD Vacutainer and Improvacuter tubes for 4 analytes that
showed statistical significance (P <0.05) when compared using the paired t -test. (A) Hematocrit. (B) Mean corpuscular volume. (C)
Mean corpuscular hemoglobin concentration. (D) Erythrocyte sedimentation rate. Abbreviation: CI, confidence interval.
performed for implementation of new kinds of blood tubes the paired t-test showed statistically significant differences
[6,9,10]. for some analytes, all the differences were within the
This research performed the comparison study for the total allowable error. Also Bland-Altman difference plots
BD Vacutainer EDTA tube and Improvacuter EDTA tube demonstrated in most of the results were within the total
for 17 routine hematologic analytes and additional two allowable error, and Passing-Bablok regression analysis
analytes, HbA1c and ESR. The stability test for each tube revealed that the 95% CI of slope included value one and
and evaluation of the effect of different anticoagulants, K2 intercept included value zero, which suggests the two
EDTA and K3 EDTA, were also conducted. tubes are interchangeable in use.
The result of comparison study between the BD The stability test for 72 hour at 4oC also suggested
Vacutainer EDTA tube and Improvacuter EDTA tube favorable outcome for both tubes. Although the differences
suggested that there are no clinically significant differences from most of the analytes between the 0 hour result and
for all the hematologic analytes mentioned before and also 72 hour were statistically significant, they were within
for two additional analytes, HbA1c and ESR. Although the total allowable error and therefore not likely to be of
Hemoglobin A1c (%) 5.50±0.66 5.52±0.66 0.001 0.34 5.49±0.66 5.51±0.67 <0.0001 0.39 ±3%
Values are presented as mean±SD. Level of significance (P -value) was tested using the paired t -test.
83
Sunyoung Ahn et al • Comparison of Improvacuter and BD EDTA Tubes
clinical relevance except for MCHC and MPV. However, In addition, we have observed that the two different
these amounts of differences would be irrelevant for anticoagulants, K2 and K3 EDTA, did not make any
clinical practice. These results are more favorable than difference in hematologic analytes. The results of our
those reported before in that the both tubes used in this study are probably useful information for those planning
study showed more stable results. Some prior studies to implement Improvacuter EDTA tubes to the clinical
reported that WBC, MCV, MCHC, and basophil were laboratory instead of currently used BD Vacutainer EDTA
the analytes that mostly showed larger differences than tubes for hematologic analysis, or vice versa.
the total allowable error [6,11]. In conclusion, both tube Preferably further studies should progress to include
evaluated in this study showed satisfactory results in the analysis on insufficient blood volume effect associated
stability test. with K2 and K3 EDTA anticoagulant additives, which are
If we discuss the effect of EDTA with different salts, in not infrequent circumstances for the blood samples from
fact, all salts of EDTA are hyperosmolar, which causes the emergency room or pediatric wards. Also further
water to leave from inside of the cells and eventually studies about other factors besides of mentioned ones
results in cell shrinkage in high EDTA level. The higher here, such as vacuum function of the blood tube, will be
the concentration of EDTA, the greater the osmotic needed because that is directly related to the appropriate
withdrawal of water from cells. Therefore it is important blood sampling volume and the malfunction of vacuum
to ensure that the blood tubes are completely filled causes inconvenience for both patient and phlebotomist.
to the suggested volume so that the concentration of However, this factor is considered to be beyond the scope
anticoagulant in the blood sample would be appropriate. of this article.
However, some studies reported that even with the In conclusion, we have conducted comparability test
sufficiently collected blood volume, red cell volume shrank and demonstrated that Improvacuter EDTA tubes
and cell count also decreased [3]. In contrary to those performed equivalently to the currently used BD
reports, the comparison of K2 EDTA tube and K3 EDTA Vacutainer EDTA tube for routine hematologic analytes.
tube in this study suggests that the differences between Therefore, these tubes are interchangeable for routine
results obtained from each tube are minimal and unlikely laboratory hematologic analysis. For the stability test,
of any clinical significance, both for red cell shrinkage most of the results from both tubes did not show any
effect and dilution effect. clinically significant differences when comparing the
Our study had some limitations. First, the evaluation results of 0 hour to those of 72 hour. These results
about K2 and K3 EDTA were not based on the tubes of indicate that stability of both tubes is suitable for routine
the same company; the K2 EDTA tube was the product laboratory use. Also we were not able to find any clinically
of BD and the K3 EDTA tube was from Improve Medical. meaningful differences about the parameters related to
It could have affected the results although both tube were red cell shrinkage effect or dilution effect between the
made from plastic. Second, all the blood samples used for use of K2 and K3 EDTA coagulants, at least in optimal
evaluation had sufficient volume for routine hematological conditions such as in this study; when applied to the
analysis. Therefore, the dilution effect of liquid K3 EDTA sufficient blood sampling volume and analyzed without
with insufficient blood volume could not be assessed. time delay.
Up to date, several studies, which evaluated commercial
blood tube products for routine laboratory use, have been REFERENCES
reported. What makes this study different from previous
one is that we included sufficient number of 100 patient- 1. Wiwanitkit V. Types and frequency of preanalytical
subjects, which enables it to have strong statistical power mistakes in the first Thai ISO 9002:1994 certified clinical
and to be evaluated within wider reportable range. laboratory, a 6-month monitoring. BMC Clin Pathol
교신저자: 조선미
우)06273 서울시 강남구 언주로 211, 연세대학교 의과대학 강남세브란스병원 진단검사의학교실
Tel: 02)2019-3775, Fax: 02)2019-4822, E-mail: SM178@yuhs.ac