You are on page 1of 10

Received: 15 November 2018 Revised: 14 December 2018 Accepted: 27 December 2018

DOI: 10.1111/jerd.12457

RESEARCH ARTICLE

Influence of preparation designs on marginal adaptation and


failure load of full-coverage occlusal veneers after
thermomechanical aging simulation
Daniele Angerame MD, DDS1 | Matteo De Biasi DDS, MS, PhD1 |
Martina Agostinetto DDS2 | Andrea Franzò DDS3 | Giulio Marchesi DDS, MS, PhD1

1
University Clinical Department of Medical,
Surgical, and Health Sciences, University of Abstract
Trieste, Trieste, Italy Objectives: To evaluate the fracture resistance and marginal quality of maxillary molars restored
2
Private Practice, Jesolo, Italy using lithium disilicate glass-ceramic (LDG) occlusal veneers with two preparation designs.
3
Private Practice, Gorizia, Italy Methods: Sixteen extracted maxillary molars were assigned to two groups (n = 8). In group
Correspondence 1 (G1), the teeth received a preparation for a conservative full-coverage occlusal veneer restora-
Daniele Angerame, University Clinical
tion with a 90 rounded shoulder margin. In group 2 (G2), the teeth underwent a 1-mm cusp
Department of Medical, Surgical, and Health
Sciences, Piazza Ospedale 1, Trieste, I-34125, reduction with a marginal chamfer. LDG restorations (IPS e.max CAD) were obtained with the
Italy. Cerec 3 CAD/CAM system and luted with Variolink II cement. After thermomechanical aging
Email: d.angerame@fmc.units.it (1 250 000 cycles), the specimens were loaded to fracture. A semiquantitative marginal seal
evaluation was performed observing resin replicas of the specimens at the scanning electron
microscope. Cement thickness was assessed at the stereomicroscope on sectioned specimens.
Collected data were statistically analyzed by parametric and nonparametric tests.
Results: The maximum load to fracture was 2395.01 ± 150.96 N in G1 and 2408.39 ± 112.66 N
in G2. Most of the observed specimens exhibited restorable fractures and continuous margins.
Cement thickness was 132 ± 38 μm in G1 and 150 ± 41 μm in G2. No differences between the
groups emerged.
Conclusion: This study demonstrated similar satisfactory performance of the two considered
preparations designs for occlusal veneer with LDG.
Clinical Significance: A new minimally invasive occlusal veneer preparation with marginal cham-
fer exhibited promising fracture resistance and marginal adaptation that were comparable to
those of a standard conservative preparation for the restoration of molars with CAD/CAM lith-
ium disilicate occlusal veneers.

KEYWORDS

CAD/CAM, cement thickness, lithium disilicate, marginal adaptation, occlusal veneers

1 | I N T RO D UC T I O N lead to improved preservation of the residual tooth structures, espe-


cially for the treatment of the single tooth.1,2
Obtaining excellent esthetic results without renouncing to the preser- Lithium disilicate glass-ceramic (LDG) is a recently introduced
vation of the biological structures is one of the main goals of modern ceramic material composed of quartz, lithium dioxide, phosphor oxide,
restorative dentistry. In both the clinical and laboratory setting, the alumina, potassium oxide and other components, characterized by
restorative protocols have become simpler, predictable, and capable flexural strength up to 440 MPa.3 The mechanical stability of LDG
of providing satisfactory esthetic and functional results. The use of ceramic is assured by the embedment of lithium disilicate crystals
new generation all-ceramic restorations and adhesive systems can (SiO2-Li2O) into a matrix of glass that minimizes microcrack

J Esthet Restor Dent. 2019;1–10. wileyonlinelibrary.com/journal/jerd © 2019 Wiley Periodicals, Inc. 1


2 ANGERAME ET AL.

propagation.4,5 The first available and most tested commercial product treatment option to restore the shape and anatomy of posterior teeth
belonging to the class of LDG materials is IPS e.max lithium disilicate, affected by occlusal wear and/or advanced erosive lesions.20,21,27–29
introduced in 2005 by Ivoclar Vivadent. LDG restorations can be man- Minimally invasive designs or the “no-preparation” approach have
ufactured making use of either lost-wax hot-pressing techniques or been advocated for teeth where a significant amount of dental tissue
computer-aided-design/computer-aided manufacturing (CAD/CAM) has already been lost by wear and erosion, because further tooth
milling procedures. CAD/CAM LDG ceramic blocks, which are avail- preparation may be counterproductive in these cases.21 Although
6
able for both the dental laboratory and office milling machines, are there is some evidence dissuading from the use of LDG single crowns
fabricated at the industrial level under controlled conditions, that min- with feather edge preparation in frontal teeth due to marginal crack
imize the formation of voids and volume defects within the material.7 formation,30 little is known about the performance of LDG occlusal
Among the several CAD/CAM ceramic materials, IPS e.max CAD veneers with reduced marginal thickness.
exhibited superior mechanical properties and better internal fit than The purpose of this study was to evaluate the influence of two
that of other ceramic blocks used for all-ceramic restorations.8 The different occlusal veneer preparations on the fracture load and mar-
clinical effectiveness of monolithic LDG CAD/CAM crowns have been ginal quality in maxillary molars restored using monolithic lithium disi-
already documented by a considerable amount of studies, attesting licate after chewing simulation. A rounded 90 shoulder was
the capability of LDG crowns to maintain their structural integrity compared to a new minimally invasive chamfer design. The following
after 2 years of clinical service,9 and to reach a 10-year survival rate hypotheses were tested: (1) there is no difference in the load to frac-
equal to 96.7%.10 Laboratory studies on LDG crowns have shown
ture, fracture pattern, internal fit, and marginal quality between the
their superior reliability after fatigue in comparison with porcelain-
two preparation techniques and (2) the thermomechanical aging simu-
layered Yttrium-tetragonal zirconia polycrystal-based crowns.11,12
lation has no influence on the marginal adaptation of occlusal veneers.
Molar crowns made of IPS e.max CAD can withstand mean fracture
loads up to 2500 N, which is approximately 3 to 4 times the maximum
force that can be exerted by the masticatory muscles in the posterior 2 | M A T E R I A L S A N D M ET H O D S
area.11 The use of monolithic LDG restorations has been advocated
for the rehabilitation of conditions of erosion or abrasion where the
2.1 | Sample selection
replacement of the damaged enamel is indicated through a “re-enam-
eling” process.13 The sample size was calculated on the basis of data gathered in prelimi-
Different factors play a significant role in determining the fracture nary studies (α = 0.05; β = 0.20; δ = 250.0; σ = 160.0). Sixteen extracted
resistance of all-ceramic restorations. The luting technique has an noncarious human maxillary molars were collected after obtaining the
impact in this regard as adhesively bonded all-ceramic restorations are patients' informed consent to their use for research purposes under a
characterized by greater fracture resistance than traditionally cemen- protocol approved by the institutional review board of the University of
ted restorations. 14,15
Depending on the adhesive system being used, Trieste. The teeth were stored at 4 C in 0.5% chloramine-T solution for
adhesive restorations bonded making use of total-etch techniques can no more than 1 month before use. With the aim of selecting teeth of
16
reach a bond strength up to 28 MPa within enamel and 13 to comparable size, the teeth to be included in the present experimentation
17
20 MPa within dentin. Immediate sealing of the freshly cut dentin were measured with a digital caliper and discarded if exceeding the fol-
has been suggested to improve bond strength.18 lowing ranges: crown mesiodistal width, 10.5 ± 1.0 mm; crown buccolin-
The increased thickness of the ceramic restoration is surely gual width, 11.5 ± 1.0 mm; overall length, 20.0 ± 1.0 mm.
another factor positively affecting its fracture resistance,19 but the The remnants of periodontal ligament and calculus were removed
modern minimally invasive concepts of restorative and prosthetic den- with a manual scaler. Then, the teeth were immersed into a 5%
tistry often impose the use of thin occlusal restorations. Scientific evi- sodium hypochlorite solution for 15 minutes and stored in saline.
dence on the minimum thickness allowed for LDG ceramic partial The teeth were randomly assigned to two test groups of eight
crowns or occlusal veneers is still scarce. Bonded occlusal veneers elements each according to the preparation technique to receive a
made of e.max CAD were found to resist to forces of up to 800 N full-coverage lithium disilicate occlusal veneer. The teeth were pre-
and 1000 N, when their thickness was 0.6-1.0 mm or 1.2-1.8 mm, pared either with a rounded 90 shoulder finish line in group 1 (G1),
respectively20,21; these values are only slightly inferior to those of or a minimally invasive chamfer in group 2 (G2), as described below.
monolithic full crowns of comparable thickness.22 The use of ultrathin
LDG occlusal veneers (0.6 mm) has been reported to be a promising
2.2 | Tooth preparation
and viable restorative procedure.23,24 Further, Guess et al.25 reported
no significant effect on the fracture resistance of pressable LDG A single experienced prosthodontist performed all the operative
ceramic onlay restorations on premolars reducing the preparation phases for sample preparation. The occlusal reduction for both groups
depth to 1.00 and 0.5 mm. was set at 1 mm, following the indications about the minimum occlu-
The preparation design is another factor potentially capable of sal thickness for the final restoration given by the manufacturer of the
influencing the fracture resistance of all-ceramic restorations. Ideally, lithium disilicate CAD/CAM blocks used in the present study (IPS e.
the preparation surface of a tooth receiving an all-ceramic restoration max, Ivoclar Vivadent, Schaan, Liechtenstein). For each tooth, two sili-
26
should have no sharp angles and be rounded as much as possible. con indexes were obtained to check the occlusal reduction in mesio-
Nonretentive full-coverage occlusal veneers are becoming a popular distal and buccolingual direction.
ANGERAME ET AL. 3

FIGURE 1 Detail of the occlusal veneer preparation criteria in the two experimental groups: 90 rounded shoulder (A) and minimally invasive chamfer (B)

The teeth were prepared with a simple nonretentive full-coverage dentin sealing.31–33 The adhesive primer was applied on the exposed den-
geometry with blunt internal edges. Specifically, the step-by-step tine with a light brushing motion for 20 seconds and gently air-dried for
preparation protocol for each group was the following: 5 seconds. Afterward, the adhesive resin was applied only on dentin and
light-cured with a LED lamp at 1500 mW/cm2 (Radii Plus, SDI Limited,
• G1, rounded 90 shoulder (Figure 1A). Depth grooves were placed Bayswater, Australia) for 20 seconds. Excess resin accumulated on the
all over the occlusal surface of the tooth following the original anat- enamel margins was carefully removed with the aid of an abrasive rubber
omy with a cylindrical coarse diamond bur (835 KRM 314008, point (9608 314 030, Komet) activated at low-speed. The prepared speci-
Komet, Lemgo, Germany) with a 0.8 mm diameter. The guiding mens were embedded into self-curing resin blocks (Protemp 4, 3M ESPE,
grooves were connected by removing the enamel septa between St. Paul, MN) up to a depth of 3 mm apical to the cementoenamel junc-
the grooves with a conical bur (846 KR 314 016, Komet). The occlu- tion after the periodontal ligament simulation was obtained by applying a
sal preparation was finished with a fine-grit diamond bur (8846 KR layer of regular-viscosity vinyl polysiloxane (Express Standard, 3M ESPE)
314 016, Komet). A wider round-ended cylindrical diamond bur onto the root portion of each tooth following a previously described pro-
(836 KR 314 018, Komet) was used to prepare the axial walls creat- cedure.34 Polytetrafluoroethylene cylinders having an internal diameter of

ing a 90 shoulder finish line, which was finished with a fine-grit bur 15 mm were used as mold. The specimens were stored in distilled water
(88836 KR 314 018, Komet). The prepared surfaces were then for 24 hours at 37 C before the restorative procedures.
polished with abrasive rubber points (9608 314 030, Komet).
• G2, minimally invasive chamfer technique (Figure 1B). The occlusal
reduction was carried out as described in G1. The marginal chamfer 2.4 | Restorations design and manufacturing
was prepared on the top of the axial wall with the aid of a specific
Table 1 reports the details of the tested materials. The whole process
tapered bur with a noncutting guide pin at its tip (6856P 314 018,
of occlusal veneer design and manufacturing was carried out with the
Komet) and finished with a fine-grit bur (8856 P 314 018, Komet;
Cerec 3 CAD/CAM system (Cerec software v. 3.03, Sirona Dental
Figure 2). The prepared surfaces were polished as described above.
Systems GmbH, Bensheim, Germany). The specimens were subjected
to optical impression to replicate the clinical setting. A stainless-steel
ball with a diameter of 5 mm with similar dimensions of the head of
2.3 | Immediate dentin sealing the stylus used in the fracture resistance test was also scanned. This
The dentin surfaces exposed by the tooth preparation process were approach allowed to plan the occlusal anatomy of the restoration with
immediately treated with a self-etch adhesive system (Clearfil SE Bond the aim of providing an ideal tripod between the central fossa of the
2, Kuraray, Osaka, Japan) by following the principles of the immediate occlusal veneer and the stylus.

FIGURE 2 Preparation of the minimally invasive chamfer with coarse (A) and fine grit (B) diamond burs. Note the noncutting pin at the burs' tip
4 ANGERAME ET AL.

TABLE 1 Chemical compositions and application procedures of the tested materials

Material Application procedures Composition


Clearfil SE Bond 2, Etch-and-dry acid primer 20 seconds, Primer: 10-methacryloyloxydecyl dihydrogen phosphate,
Kuraray, Osaka, Japan air blow s, bonding for 20 seconds 2-hydroxyethyl methacrylate, hydrophilic aliphatic dimethacrylate,
and light curing for 20 s dl-camphorquinone, water
Bonding: 10-methacryloyloxydecyl dihydrogen phosphate, bisphenol
A diglycidylmethacrylate, 2-hydroxyethyl methacrylate,
hydrophobic aliphatic dimethacrylate, dl-camphorquinone,
initiators, accelerators, silanated colloidal silica.
Fill Etch, Dental World, Application on enamel for Phosphoric acid, purified water, thickener, colorant.
Molfetta, Italy 30 seconds
IPS e.max CAD, Ivoclar-Vivadent, SiO2, Li2O, K2O, P2O5, ZrO2, ZnO, other and coloring oxides.
Shaan, Liechtenstein
Variolink II, Ivoclar-Vivadent, Bisphenol A diglycidylmethacrylate, urethane dimethacrylate,
Shaan, Liechtenstein triethylene glycol dimethacrylate, barium glass, ytterbium
trifluoride, Ba-Al-fluorosilicate glass, spheroid mixed oxide,
catalysts, stabilizers, pigments (particle size: 0.04-3 μm; mean filler
size: 0.7 μm; base: 46.7 vol%, low viscosity catalyst: 43.6 vol%).

Sixteen occlusal veneers were milled from an equal number of St. Paul, Minnesota). The enamel margins were covered with an etchant
lithium disilicate blocks (IPS e.max CAD) following the same principles gel containing 37% orthophosphoric acid (Fill Etch, Dental World), which
already described by Magne et al.20 After milling, each restoration was was rinsed after 30 seconds. Both the internal surface of the restoration
carefully examined and no cracks were observed. After manufacturing, and the prepared tooth were covered with adhesive resin (Clearfil SE
the internal surface of each restoration was acid-etched with viscous Bond 2, Kuraray). The two components of a dual-cure resin luting
buffered 9% hydrofluoric acid (HF; Porcelain etch, Ultradent Products, cement (Variolink II, shade A3, Ivoclar-Vivadent) were mixed in a 1:1
Inc., South Jordan, Utah) for 20 seconds, which was then rinsed with ratio and placed onto the treated veneer surface. Seating pressure of
water spray for 20 seconds. Postetching cleaning was performed with 20 g/mm2 was applied with a plunger loaded by a box filled with lead
37% orthophosphoric acid (Fill Etch, Dental World, Molfetta, Italy). pellets for 5 minutes to the specimens.35. The excess luting cement
After 60 seconds, the gel was rinsed with water spray for the same extruding at the adhesive interface was removed with a manual instru-
amount of time and the restoration immersed in an ultrasonic bath ment and each surface of the restored tooth was exposed to three light-
containing absolute ethanol for 3 minutes. Once dried, a silane primer curing cycles of 20 s each with a LED lamp at 1500 mW/cm2 (Radii
(Monobond Plus, Ivoclar Vivadent) was applied thrice onto the inner Plus, SDI Limited). The margins were polished with rotary rubber points
surface of each veneer and gently dried with air. (Identoflex Composite Polishers, Kerr, Orange, California) (Figure 3).

2.5 | Adhesive placement 2.6 | Thermomechanical aging


The prepared tooth surfaces were treated with an intraoral sandblaster Specimens were incubated in distilled water at 37 C for 24 hours and
and 35 μm aluminum oxide particles at 30 psi (Cojet System, 3M, then cleaned for 10 minutes by sonication.

FIGURE 3 Representative specimens of the two experimental groups after preparation and occlusal veneer cementation: rounded 90 shoulder
(A and C) and minimally invasive chamfer (B and D)
ANGERAME ET AL. 5

The restored samples underwent thermomechanical aging by


means of a chewing simulator (CS-4.4, Willytech, Munich, Germany).
A steatite ceramic sphere with a 5-mm diameter (Hoechst Ceram
Tec, Wunsiedel, Germany) was applied with a downward speed of
16 mm/s exerting a force of 50 N, at a frequency of 1 Hz for
1 250 000 cycles.36 At the same time, the device subjected the speci-
mens to thermocycles of 80 seconds each in distilled water at
5 ± 3 C e 55 ± 3 C. Correct thermo-mechanical loading was checked
every 10 000 cycles by monitoring the mechanical action and water
temperature within the chewing chambers.

2.7 | Load-to-fracture test and failure analysis


The resistance to fracture of the specimens was tested with a univer-
sal testing machine (Quasar, Galdabini, Cardano al Campo, Italy),
which was set to exert an increasing force with a round 5 mm wide
stainless steel stylus onto the restored tooth at a speed of FIGURE 5 Transillumination of a specimen after testing. Note
1 mm/min.25 The tip of the stylus was positioned over the central “mosaiclike” cracks as a result of force propagation through a brittle,
fossa to achieve tripodization of contacts along the cuspal inclines crystalline, solid material. This type of restorable fracture was the
(Figure 4). The maximum load to fracture was registered and collected most frequent under the conditions of the present study

in a dedicated spreadsheet.
(EpoThin Epoxy Resin and Hardener, Buehler, Lake Bluff, Illinois) and
The type of fracture was qualitatively appraised by two indepen-
then poured inside the impressions. After 24 hours, the positive resin
dent trained examiners, who classified the failure mode distinguishing
replicas of the restored teeth were removed from the impressions and
between failures above or below the cementoenamel junction and
sputter-coated with gold (Sputter Coater K550X, Fei Company, Hills-
between fractures and cracks (fracture without fragment detachment).37
boro, Nebraska) in order to be observed at the scanning electron
microscope (Quanta 250, Fei Company). A low-magnification image
2.8 | Marginal quality and internal fit analysis
(50×) of each of the four faces of the restored tooth (mesial, buccal,
Making use of custom box-shaped trays, two-phase two-stage preci- distal, and lingual) was acquired for the semiquantitative evaluation of
sion impressions (Aquasil Putty and Light, Dentsply Caulk, Milford, marginal seal, which was carried out according to a previously
Delaware) of the restored teeth were taken immediately after the described protocol38 that involves the following categories: continu-
occlusal veneer cementation, at the end of thermomechanical aging ous margin, gap/irregularity, or not assessable/artifact.
and after the load-to-fracture test. All the specimens were longitudinally sectioned trice with a
Epoxy resin was prepared by mixing the two components of a microtome (Micromet, Remet, Bologna, Italy; Figure 5), thus creating
commercial product in the 5:2 ratio suggested by the manufacturer four surfaces to be analyzed. Each section was inspected and photo-
graphed at the stereomicroscope (MZ16, Leica, Wetzlar, Germany), to
measure the thickness of the cement layer with the aid of the same
software mentioned above. The cement thickness was measured at
20 equidistant points along the adhesive interface.

2.9 | Statistical analysis


All the data sets were handled with statistical software (Statistical
Package for Social Sciences v.15, SPSS Inc., Chicago, Illinois). Scalar
data were tested for the existence of the assumptions for the use of
parametric tests (Shapiro-Wilk and Levene tests). The differences in
maximum load to fracture between the two groups were carried out
with an independent sample t test. A Chi-squared test served to com-
pare the groups in terms of fracture pattern after the load-to-fracture
test and in terms of margin quality. For the measurement of internal
fit, the average value of the readings obtained for each specimen con-
stituted the statistical unit and the comparisons between groups were
performed by means of independent sample t tests. The value of α
FIGURE 4 Load to fracture test in the universal testing machine was set at 5%.
6 ANGERAME ET AL.

TABLE 2 Failure types and numbers (crack implies fracture without isolated fragment)

Group Intact specimen Fracture above CEJ Fracture below CEJ


Crack Fragment Crack Fragment
Chamfer (n = 8) 0 6 0 0 2
Shoulder (n = 8) 0 5 1 0 2

3 | RESULTS conservation of coronal structure, often avoiding the root canal treat-
ment. For this reason, the present investigation was conducted on teeth
The means of the maximum load to fracture were 2395.01 ± 150.96 N that intentionally were not endodontically treated. This kind of restora-
and 2408.39 ± 112.66 N in shoulder and chamfer groups, respectively; tions can be fabricated at a minimal thickness (0.6 mm) even with CAD/-
the difference was not statistically significant (P = 0.844). The fracture CAM techniques, such as the Cerec system.21
pattern, summarized in Table 2, was also similar in the two groups, with After thermomechanical loading, all ceramic restorations survived
cracks above the CEJ being the most predominant type of restoration without fracture or chipping. Masticatory forces during normal func-
failure (Figure 6). In both groups, only two specimens were characterized tion range from 50 N to 250 N and 500 to 800 N in bruxism in the
by fracture below the CEJ. posterior region.39,40 Mean fracture loads for both groups (~2300 N)
As to margin quality assessment, the vast majority of the observed exceeded those values. Comparing our results to other investigations
specimens exhibited continuous margins (87.5%-100%), irrespective of on the fracture resistance of LDG restorations, resistance to load was
experimental phase or group (P > 0.05) (Table 3 and Figure 7). significantly higher than that reported in a study on premolars
The internal fit measurement revealed good adaptation of the restored with LDG overlays (maximum mean value 1361 ± 333 N).25
occlusal veneer restoration, with cement thickness values equal to Such difference is possibly attributed to the different tooth type. On
132 ± 38 μm in G1 and 150 ± 41 μm in G2 (Figure 8). The difference the other hand, a recent work on LDG occlusal veneers considering
in internal fit was not statistically significant (P = 0.393). different molar tooth preparations showed a wide range of fracture
loads among the considered experimental groups (median value
610 N-3390 N),27 which were slightly below or greatly above the
4 | DISCUSSION recommended minimum fracture strength for posterior restorations of
500 N-700 N.41 The authors obtained the best results with 0.7 to
The null hypotheses that the tested types of preparation technique do
1.0 mm-thick restoration in terms of the survival rate after dynamic
not affect load to fracture, fracture pattern, internal fit, and marginal
loading and fracture resistance regardless of the bonding substrate
quality, as well as that thermomechanical loading has no influence on
(only enamel or enamel and dentin), whereas the performance of thin-
the marginal adaptation were both accepted.
ner restorations depended on the bonding surface, with 0.3 to
Occlusal veneers have been demonstrated as a conservative alter-
0.6 mm-thick restorations bonded only to enamel showing the worst
native to traditional onlays and complete coverage crowns for the treat-
results.27 In the present study, the restorations were mainly bonded
ment of severe abrasive/erosive lesions. Occlusal veneers are essentially
to enamel, but a certain degree of dentin exposure on the occlusal
thin overlay restorations with a nonretentive design that allow the
surface was unavoidable in all teeth (Figure 8); nonetheless, the occlu-
sal veneers were 1.0 mm thick (with the sole exception of the margin
in the chamfer group) and thus the possible effect of bonding to den-
tin was likely decreased. The findings of the present study corroborate
the evidence in favor of the treatment of occlusal abrasion or malposi-
tioned teeth with nonretentive lithium disilicate ceramic occlusal
veneers with a marginal thickness even inferior to 1 mm.
In our study two finishing lines were chosen, a 90 shoulder and a
minimally invasive chamfer, but the marginal preparation design
showed no significant influence on the fracture resistance (P = 0.844).
Clausen et al. evaluated the influence of different preparation designs
(chamfer finishing line or straight-beveled finishing line) on fracture
resistance of adhesively luted nonretentive full-coverage molar resto-
rations, made of lithium or leucite-reinforced glass ceramic, and found
no statistically significant differences on the fracture resistance.42
These considerations suggest that the finishing line seems not to play
a primary role in the fracture resistance of partial indirect restoration,
provided that the other variables are kept constant. Probably the
results of our study might be explained by the location of point of
FIGURE 6 Specimen after microtome sectioning before internal fit loading during the fracture test, which was not near the finishing line.
assessment Enamel and dentin are unique tissues with highly specialized function.
ANGERAME ET AL. 7

TABLE 3 Results for marginal quality of the tested specimens

Number of Continuos Gap/ Not judgeable/ % Continuos


Groupa Margin Time analized surfaces margin irregularity artifact margina
Group 1 Shoulder Baseline 32 31 1 0 96.9%
After aging 32 31 1 0 96.9%
Group 2 Chamfer Baseline 32 32 0 0 100%
After aging 32 30 2 0 93.8%
a
No statically significant differences were found among all groups.

When restoring a tooth one should consider not only the restorative glassy matrix of the specimens to the depth of a few microns,
materials that best emulates enamel and dentin, but also the simula- increased the surface area and facilitated the penetration and reten-
tion of the dentin-enamel junction through the restoration-dentin tion of resin cement into the microretentions of the treated surface.46
43
bond, which can be considered a true composite structure. Using The ideal HF concentration to maximize the ceramic surface rough-
extremely thin “enamel-like” restorations, the bonding strategy ness and avoid the potential detrimental effects of over-etching on
becomes yet more important.27 the strength of the LDG restorations is still debated. Some authors
The clinical success of a ceramic restoration depends on the qual- have advocated the use of 5% HF acid to reduce the risk of defect
ity and durability of the bond between the ceramic and the resin formation on the ceramic surface and their propagation in the bulk
cement.44 The quality of this bond is determined by the bonding structure of the LDG restoration.47 Some others reported no signifi-
mechanism, which is controlled in part by specific surface treatment cant differences between 5% and 10% HF etching in terms of bond
used to promote micromechanical or chemical retention to the strength to LDG ceramic48 and, most importantly, of cyclic load-to-
45
ceramic substrate. In the present study, the ceramic restorations failure of lithium disilicate crowns.49 Therefore, making use of differ-
were etched with 9% HF for 20 seconds and covered with silane. ent combinations of HF concentrations and etching times appears
Etching the ceramic surface with 9% HF promoted dissolution in the reasonable, provided that over-etching is avoided. As to alternative

FIGURE 7 Representative scanning electron microscope images of the margin quality before and after thermomechanical loading in the 90
rounded shoulder group (A and B) and in the minimally invasive chamfer group (C and D)
8 ANGERAME ET AL.

have shown that oral fluids degrade the resin matrix, and it has been
shown that thermomechanical loading causes the marginal adaptation
of adhesive restorations to break down.58 In this study, no differences
in the margin with two different preparations designs were observed
at time zero and after thermo-mechanical loading. Thus, satisfactory
adhesion to enamel was evident in all groups. This investigation is in
agreement with other studies in which enamel margins demonstrated
a good seal in contrast to the dentin margins because bonding to
enamel produces the smallest degree of microleakage.59
The internal fit measurement revealed good adaptation of the
occlusal veneer restoration, with cement thickness values equal to
0.132 ± 0.038 μm in G1 and 0.150 ± 0.041 μm in G2 (P = 0.393).
These findings attest that the registered internal adaptation is clini-
cally acceptable. To the best of the Authors' knowledge, no previous
study has investigated the internal fit of LDG occlusal veneers. Never-
theless, it is conceivable that the internal fit of such kind of restoration
would not substantially vary from that obtainable with full crowns at
the occlusal level, given the standardization of the scanning and mill-
ing processes. In accordance with this speculation, a recent study
reported internal fit values at the occlusal space of e.max LDG crowns
that were very similar to that found in the present study (median
values 139-166 μm, depending on the intraoral camera used for the
scanning process).60

5 | CONC LU SIONS

FIGURE 8 Appearance of two representative sections observed In conclusion, the results of the present study support the use of dif-
during the internal fit assessment: 90 rounded shoulder (A) and ferent preparations lines for occlusal veneer with LDG. Further in vivo
minimally invasive chamfer (B)
studies are needed to validate the clinical performance of the more
conservative preparation design.
surface treatments, several studies demonstrated that there is a
decrease in shear bond strength when the ceramic surface was air
abraded with Al2O3 when compared with a ceramic surface etched ACKNOWLEDGMENTS
with 9%-10% HF.46,50,51 Differently, Panah et al. showed no signifi-
The authors thank the staff of the dental laboratory Momic, Via
cant differences between a ceramic surface etched with 10 HF% and
Nicolò degli Aldegardi 18, Trieste, Italy, for providing their essential
one air abraded with Al2O3.52
contribution for the occlusal veneer fabrication. The authors declare
Specimens were thermo-mechanically cycled with a specific load-
that they do not have any financial interest in the companies whose
ing machine to simulate in vitro the physical stress to which restored
materials are included in this article.
teeth are exposed in the oral environment. Tensions and thermo-
mechanical stress occur at the adhesive interface due to the different
ORCID
thermal expansion coefficients of the materials and are able to induce
physicochemical alterations in dental materials.53,54 In the present Daniele Angerame https://orcid.org/0000-0003-4345-3790
 
study, the temperature ranged from 5 C to 55 C in 1 250 000 cycles, Matteo De Biasi https://orcid.org/0000-0003-2261-0400

because thermal records in the oral environment ranged between 0 C


and 67 C.55 RE FE RE NC ES
The coefficient of thermal expansion is defined as a change in 1. Edelhoff D, Sorensen JA. Tooth structure removal associated with var-
length per unit of the original length when the temperature is changed ious preparation designs for posterior teeth. Int J Periodontics Restor-
by 1 K. It is known that the coefficients of thermal expansion of ative Dent. 2002;22:241-249.
2. Valenti M, Valenti A. Retrospective survival analysis of 261 lithium dis-
enamel and dentin are approximately 17 × 10−6/ C and 11 × 10−6/
ilicate crowns in a private general practice. Quintessence Int. 2009;40:
C respectively and the coefficients for adhesive and resin cement 573-579.
range between 20 × 10−6/ C and 80 × 10−6/ C.56 For the restored 3. Tysowsky GW. The science behind lithium disilicate: a metal-free
alternative. Dent Today. 2009;28:112-113.
tooth, the problem is amplified by the use of composites, because
4. Demir N, Ozturk AN, Malkoc MA. Evaluation of the marginal fit of full
some of these materials may exhibit high thermal expansion coeffi- ceramic crowns by the microcomputed tomography (micro-CT) tech-
cients when compared with porcelain or ceramics.57 in vitro studies nique. Eur J Dent. 2014;8:437-444.
ANGERAME ET AL. 9

5. Silva NR, Thompson VP, Valverde GB, et al. Comparative reliability 27. Sasse M, Krummel A, Klosa K, Kern M. Influence of restoration thickness
analyses of zirconium oxide and lithium disilicate restorations in vitro and dental bonding surface on the fracture resistance of full-coverage
and in vivo. J Am Dent Assoc. 2011;142(suppl 2):4S-9S. occlusal veneers made from lithium disilicate ceramic. Dent Mater. 2015;
6. Anadioti E, Aquilino SA, Gratton DG, et al. 3D and 2D marginal fit of 31:907-915.
pressed and CAD/CAM lithium disilicate crowns made from digital 28. Magne P, Stanley K, Schlichting LH. Modeling of ultrathin occlusal
and conventional impressions. J Prosthodont. 2014;23:610-617. veneers. Dent Mater. 2012;28:777-782.
7. Giordano R. Materials for chairside CAD/CAM-produced restorations. 29. Schlichting LH, Resende TH, Reis KR, Magne P. Simplified treatment
J Am Dent Assoc. 2006;137(suppl 1):14S-21S. of severe dental erosion with ultrathin CAD-CAM composite occlusal
8. Goujat A, Abouelleil H, Colon P, et al. Mechanical properties and inter- veneers and anterior bilaminar veneers. J Prosthet Dent. 2016;116:
nal fit of 4 CAD-CAM block materials. J Prosthet Dent. 2018;119: 474-482.
384-389. 30. Fuzzi M, Tricarico MG, Ferrari Cagidiaco E, et al. Nanoleakage and
9. Fasbinder DJ, Dennison JB, Heys D, Neiva G. A clinical evaluation of internal adaptation of zirconia and lithium disilicate single crowns with
chairside lithium disilicate CAD/CAM crowns: a two-year report. J Am feather edge preparation. J Osseointegr. 2017;9:250-262.
Dent Assoc. 2010;141(suppl 2):10S-14S. 31. Magne P. Immediate dentin sealing: a fundamental procedure for indi-
10. Pieger S, Salman A, Bidra AS. Clinical outcomes of lithium disilicate rect bonded restorations. J Esthet Restor Dent. 2005;17:144-154. dis-
single crowns and partial fixed dental prostheses: a systematic review. cussion 155.
J Prosthet Dent. 2014;112:22-30. 32. Magne P, So WS, Cascione D. Immediate dentin sealing supports
11. Guess PC, Zavanelli RA, Silva NR, Bonfante EA, Coelho PG, delayed restoration placement. J Prosthet Dent. 2007;98:166-174.
Thompson VP. Monolithic CAD/CAM lithium disilicate versus 33. Oliveira L, Mota EG, Borges GA, Burnett LH Jr, Spohr AM. Influence
veneered Y-TZP crowns: comparison of failure modes and reliability of immediate dentin sealing techniques on cuspal deflection and frac-
after fatigue. Int J Prosthodont. 2010;23:434-442. ture resistance of teeth restored with composite resin inlays. Oper
12. Silva NR, Bonfante EA, Martins LM, et al. Reliability of reduced- Dent. 2014;39:72-80.
thickness and thinly veneered lithium disilicate crowns. J Dent Res. 34. Wafaie RA, Ibrahim Ali A, Mahmoud SH. Fracture resistance of pre-
2012;91:305-310. pared premolars restored with bonded new lab composite and all-
13. Cortellini D, Parvizi A. Rehabilitation of severely eroded dentition uti- ceramic inlay/onlay restorations: Laboratory study. J Esthet Restor
lizing all-ceramic restorations. Pract Proced Aesthet Dent. 2003;15: Dent. 2018;30:229-239.
275-282. quiz 284. 35. Goracci C, Cury AH, Cantoro A, Papacchini F, Tay FR, Ferrari M.
14. Piwowarczyk A, Bender R, Ottl P, Lauer HC. Long-term bond between Microtensile bond strength and interfacial properties of self-etching
dual-polymerizing cementing agents and human hard dental tissue. and self-adhesive resin cements used to lute composite onlays under
Dent Mater. 2007;23:211-217. different seating forces. J Adhes Dent. 2006;8:327-335.
15. Klosa K, Wolfart S, Lehmann F, Wenz HJ, Kern M. The effect of stor- 36. Kern M, Strub JR, Lu XY. Wear of composite resin veneering materials
age conditions, contamination modes and cleaning procedures on the in a dual-axis chewing simulator. J Oral Rehabil. 1999;26:372-378.
resin bond strength to lithium disilicate ceramic. J Adhes Dent. 2009; 37. Magne P, Knezevic A. Thickness of CAD-CAM composite resin over-
11:127-135. lays influences fatigue resistance of endodontically treated premolars.
16. Fortin D, Swift EJ Jr, Denehy GE, Reinhardt JW. Bond strength and Dent Mater. 2009;25:1264-1268.
microleakage of current dentin adhesives. Dent Mater. 1994;10: 38. Spreafico R, Marchesi G, Turco G, et al. Evaluation of the in vitro
253-258. effects of cervical marginal relocation using composite resins on the
17. Sarr M, Mine A, De Munck J, et al. Immediate bonding effectiveness marginal quality of CAD/CAM crowns. J Adhes Dent. 2016;18:
of contemporary composite cements to dentin. Clin Oral Investig. 355-362.
2010;14:569-577. 39. Kampe T. Function and dysfunction of the masticatory system in indi-
18. Magne P. IDS: Immediate Dentin Sealing (IDS) for tooth preparations. viduals with intact and restored dentitions. A clinical, psychological
J Adhes Dent. 2014;16:594. and physiological study. Swed Dent J Suppl. 1987;42:1-68.
19. Lima JM, Souza AC, Anami LC, et al. Effects of thickness, processing 40. Kiliaridis S, Kjellberg H, Wenneberg B, Engstrom C. The relationship
technique, and cooling rate protocol on the flexural strength of a between maximal bite force, bite force endurance, and facial morphol-
bilayer ceramic system. Dent Mater. 2013;29:1063-1072. ogy during growth. A cross-sectional study. Acta Odontol Scand. 1993;
20. Magne P, Schlichting LH, Maia HP, Baratieri LN. In vitro fatigue resis- 51:323-331.
tance of CAD/CAM composite resin and ceramic posterior occlusal 41. Körber KH, Ludwig K. Maximum biteforces as a basis for the design of
veneers. J Prosthet Dent. 2010;104:149-157. dental restorations. Dent Labor. 1983;31:55-60.
21. Schlichting LH, Maia HP, Baratieri LN, Magne P. Novel-design ultra- 42. Clausen JO, Abou Tara M, Kern M. Dynamic fatigue and fracture resis-
thin CAD/CAM composite resin and ceramic occlusal veneers for the tance of non-retentive all-ceramic full-coverage molar restorations.
treatment of severe dental erosion. J Prosthet Dent. 2011;105: Influence of ceramic material and preparation design. Dent Mater.
217-226. 2010;26:533-538.
22. Sorrentino R, Nagasawa Y, Infelise M, et al. In vitro analysis of the 43. Magne P, Knezevic A. Influence of overlay restorative materials and
fracture resistance of CAD-CAM monolithic lithium disilicate molar load cusps on the fatigue resistance of endodontically treated molars.
crowns with different occlusal thickness. J Osseointegr. 2018;10: Quintessence Int. 2009;40:729-737.
11-17. 44. Soares CJ, Soares PV, Pereira JC, Fonseca RB. Surface treatment pro-
23. Al-Akhali M, Chaar MS, Elsayed A, et al. Fracture resistance of ceramic tocols in the cementation process of ceramic and laboratory-
and polymer-based occlusal veneer restorations. J Mech Behav Biomed processed composite restorations: a literature review. J Esthet Restor
Mater. 2017;74:245-250. Dent. 2005;17:224-235.
24. Andrade JP, Stona D, Bittencourt HR, Borges GA, Burnett LH Júnior, 45. Della Bona A, Shen C, Anusavice KJ. Work of adhesion of resin on
Spohr AM. Effect of different computer-aided design/computer-aided treated lithia disilicate-based ceramic. Dent Mater. 2004;20:
manufacturing (CAD/CAM) materials and thicknesses on the fracture 338-344.
resistance of occlusal veneers. Oper Dent. 2018;43:539-548. 46. Spohr AM, Sobrinho LC, Consani S, et al. Influence of surface condi-
25. Guess PC, Schultheis S, Wolkewitz M, Zhang Y, Strub JR. Influence of tions and silane agent on the bond of resin to IPS Empress 2 ceramic.
preparation design and ceramic thicknesses on fracture resistance and Int J Prosthodont. 2003;16:277-282.
failure modes of premolar partial coverage restorations. J Prosthet 47. Murillo-Gomez F, Palma-Dibb RG, De Goes MF. Effect of acid etching
Dent. 2013;110:264-273. on tridimensional microstructure of etchable CAD/CAM materials.
26. Federlin M, Krifka S, Herpich M, Hiller KA, Schmalz G. Partial ceramic Dent Mater. 2018;34:944-955.
crowns: influence of ceramic thickness, preparation design and luting 48. Puppin-Rontani J, Sundfeld D, Costa AR, et al. Effect of hydrofluoric
material on fracture resistance and marginal integrity in vitro. Oper acid concentration and etching time on bond strength to lithium disili-
Dent. 2007;32:251-260. cate glass ceramic. Oper Dent. 2017;42:606-615.
10 ANGERAME ET AL.

49. Prochnow C, Venturini AB, Guilardi LF, et al. Hydrofluoric acid concen- 57. Versluis A, Douglas WH, Sakaguchi RL. Thermal expansion coefficient
trations: Effect on the cyclic load-to-failure of machined lithium disilicate of dental composites measured with strain gauges. Dent Mater. 1996;
restorations. Dent Mater. 2018;34:e255-e263. 12:290-294.
50. Salvio LA, Correr-Sobrinho L, Consani S, Sinhoreti MAC, de Goes MF, 58. Roulet JF, Walti C. Influence of oral fluid on composite resin and
Knowles JC. Effect of water storage and surface treatments on the glass-ionomer cement. J Prosthet Dent. 1984;52:182-189.
tensile bond strength of IPS Empress 2 ceramic. J Prosthodont. 2007; 59. Uludag B, Ozturk O, Ozturk AN. Microleakage of ceramic inlays luted
16:192-199. with different resin cements and dentin adhesives. J Prosthet Dent.
51. Ayad MF, Fahmy NZ, Rosenstiel SF. Effect of surface treatment on 2009;102:235-241.
roughness and bond strength of a heat-pressed ceramic. J Prosthet 60. de Paula Silveira AC, Chaves SB, Hilgert LA, Ribeiro AP. Marginal and
Dent. 2008;99:123-130. internal fit of CAD-CAM-fabricated composite resin and ceramic
52. Panah FG, Rezai SM, Ahmadian L. The influence of ceramic surface crowns scanned by 2 intraoral cameras. J Prosthet Dent. 2017;117:
treatments on the micro-shear bond strength of composite resin to 386-392.
IPS Empress 2. J Prosthodont. 2008;17:409-414.
53. DeLong R, Douglas WH. Development of an artificial oral environ-
ment for the testing of dental restoratives: bi-axial force and move-
ment control. J Dent Res. 1983;62:32-36. How to cite this article: Angerame D, De Biasi M,
54. Oyafuso DK, Ozcan M, Bottino MA, Itinoche MK. Influence of thermal Agostinetto M, Franzò A, Marchesi G. Influence of preparation
and mechanical cycling on the flexural strength of ceramics with tita-
nium or gold alloy frameworks. Dent Mater. 2008;24:351-356.
designs on marginal adaptation and failure load of full-
55. Palmer DS, Barco MT, Billy EJ. Temperature extremes produced orally coverage occlusal veneers after thermomechanical aging simu-
by hot and cold liquids. J Prosthet Dent. 1992;67:325-327. lation. J Esthet Restor Dent. 2019;1–10. https://doi.org/10.
56. Anusavice KJ, Brantley WA. Physical properties of dental materials. In:
1111/jerd.12457
Anusavice KJ, ed. Phillips' Science of Dental Materials. 11th ed. St.
Louis, MO: Saunders; 2005:52-53.

You might also like