You are on page 1of 7

Designation: D 7091 – 05

Standard Practice for


Nondestructive Measurement of Dry Film Thickness of
Nonmagnetic Coatings Applied to Ferrous Metals and
Nonmagnetic, Nonconductive Coatings Applied to Non-
Ferrous Metals1
This standard is issued under the fixed designation D 7091; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope Even then, the thickness of the measured layer is the cumula-
1.1 This practice describes the nondestructive measurement tive thickness of that layer and all layers beneath it, down to the
of the dry film thickness of nonmagnetic coatings applied to the base metal.
surface of ferrous metals using magnetic gages and the 1.5 Gages which measure coating thickness using an ultra-
nondestructive measurement of the dry film thickness of sonic principle may also be able to measure the thickness of
electrically nonconductive, nonmagnetic coatings applied to coatings applied to metal surfaces. This practice does not
the surface of nonferrous metals using eddy current gages. This address the use of ultrasonic coating thickness gages. Test
practice is intended to supplement the manufacturers’ instruc- Method D 6132 should be referenced for this application.
tions for the manual operation of the gages and is not intended 1.6 Coating thickness can vary widely across a surface. As
to replace them. It includes definitions of key terms, reference a result, obtaining single-point measurements may not accu-
documents, the significance and use of the practice, the rately represent the actual coating system thickness. This
advantages and limitations of coating thickness gages, and a practice provides guidance for the frequency of coating thick-
description of test specimens. It describes the methods and ness measurements, based on large areas of coated surface, on
recommended frequency for verifying the accuracy of gages test panels and on small parts/components. The governing
and for adjusting (optimizing) the equipment, describes a specification is responsible for providing the user with the
frequency for measuring the thickness of the coating(s) and minimum and the maximum coating thickness for each layer,
lists the reporting recommendations. and for the total coating system.
1.2 These procedures are not applicable to coatings that will 1.7 This standard does not purport to address all of the
be readily deformed under the load of the measuring gages/ safety concerns, if any, associated with its use. It is the
probes, as the gage probe must be placed directly on the responsibility of the user of this standard to establish appro-
coating surface to obtain a reading. Provisions for measuring priate safety and health practices and determine the applica-
on soft or tacky coatings are described in 5.6. bility of regulatory limitations prior to use.
1.3 Use of film thickness measuring gages on metal-filled
2. Referenced Documents
coatings may produce erroneous results, depending on the type
and amount of metal in the coating film. The user should 2.1 ASTM Standards: 2
consult the manufacturers instructions regarding the use of D 609 Practice for Preparation of Cold-Rolled Steel Panels
coating thickness gages on these types of coatings. for Testing Paint, Varnish, Conversion Coatings, and
1.4 Coating thickness can be measured using a variety of Related Coating Products
gages. These gages are categorized as “magnetic pull-off” and D 823 Practices for Producing Films of Uniform Thickness
“electronic.” They use a sensing probe or magnet to measure of Paint, Varnish, and Related Products on Test Panels
the gap (distance) between the base metal and the probe. This D 1730 Practices for Preparation of Aluminum and
measured distance is displayed as coating thickness by the Aluminum-Alloy Surfaces for Painting
gages. Accordingly, these gages cannot distinguish the thick- D 6132 Test Method for Nondestructive Measurement of
ness of individual layers after they have all been applied. The Dry Film Thickness of Applied Organic Coatings Using an
thickness of each layer must be measured after it is applied. Ultrasonic Gage
2.2 SSPC Standards:

1 2
This practice is under the jurisdiction of ASTM Committee D01 on Paint and For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Related Coatings, Materials and Applications and is the direct responsibility of contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Subcommittee D01.23 on Physical Properties of Applied Paint Films. Standards volume information, refer to the standard’s Document Summary page on
Current edition approved Aug. 1, 2005. Published September 2005. the ASTM website.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
D 7091 – 05
SSPC-PA2 Paint Application Standard No. 2 Measurement 3.1.6 certification, n—documentation of the state of condi-
of Dry Coating Thickness with Magnetic Gauges3 tion of the gage, which can (but not required by definition) be
accompanied by corrective action (such as adjustment or
3. Terminology calibration, or both, or the replacement of components) neces-
3.1 Definitions of Terms Specific to This Standard: sary to correct any out-of-tolerance conditions.
3.1.1 accuracy, n—the measure of the magnitude of error 3.1.7 coating thickness standard (test block), n—a smooth,
between the result of a measurement and the true thickness of metal substrate with a coating of known thickness that is
the item being measured. traceable to a national standard.
3.1.1.1 Discussion—An accuracy statement predicts the 3.1.8 compensation value, n—generating a verifiable value,
ability of a coating thickness gage to measure the true thickness which is deducted from a measured value read from the gage,
of a coating to be measured. Accuracy statements provide the to correct for any surface conditions (that is, base metal effect).
performance capability across the full functional measurement 3.1.9 dry film thickness, n—the thickness of a coating (or
range of the gage. Accuracy statements frequently include a coating layers) as measured from the surface of the substrate.
fixed portion that remains constant across the measurement 3.1.9.1 Discussion—If the surface is roughened, the dry
range, plus a variable portion that is related to the measurement film thickness is considered the thickness of the coating or
result for a particular thickness. coating layers above the peaks of a surface profile.
3.1.2 adjustment, n—the physical act of aligning a gage’s 3.1.10 ferrous, n—a magnetic material such as carbon steel.
thickness readings to match those of a known thickness sample 3.1.10.1 Discussion—Also known as ferro-magnetic.
(removal of bias), in order to improve the accuracy of the gage 3.1.11 gage, n—an instrument for measuring quantity, or an
on a specific surface or within a specific portion of its instrument for testing.
measurement range. 3.1.11.1 Discussion—In this practice, the term “gage” refers
3.1.2.1 Discussion—An adjustment will affect the outcome to an instrument for quantifying coating thickness.
of subsequent readings. Also known as “optimization.” 3.1.12 manufacturer’s specifications, n—a statement or set
3.1.3 area measurement, n—the average of several spot of statements that describes the performance characteristics of
measurements obtained over the coated surface. the gage under a given set of conditions.
3.1.3.1 Discussion—For large surfaces, an area measure- 3.1.12.1 Discussion—Manufacturer’s specifications typi-
ment is commonly the average of at least five spot measure- cally includes the range of measurement, accuracy statement,
ments obtained in approximately 9 m2 (100 ft2) of surface operating temperature range, power source, dimensions and
area. For small test panels, it is commonly the average of three weight, and conformance to industry standards.
spot measurements on each test face (one from the top third, 3.1.13 measurement, n—the value obtained when placing
one from the middle third and one from the bottom third). the probe of a thickness gage in contact with a surface.
3.1.4 base metal reading (BMR), n—a measurement ob-
3.1.14 micrometer (micron), n—one one-thousands of a
tained on the uncoated substrate using a coating thickness
millimeter (0.001 mm); 25.4 microns = 1 mil.
gage.
3.1.15 mil, n—an imperial unit of measure; one one-
3.1.4.1 Discussion—The BMR is the measured effect of
thousands of an inch (0.001 in.); 1 mil = 25.4 microns.
substrate roughness on a coating thickness gage that is caused
by the manufacturing process (for example, castings) or 3.1.16 non-conductive, n— a material that is unable to
surface profile (roughness)-producing operations (for example, conduct electricity.
power tool cleaning, abrasive blast cleaning, etc.). Non- 3.1.17 non-ferrous metal, n—a non-magnetic metal such as
compensation for the base metal effect can result in an aluminum.
overstatement of the true thickness of the coating. The base 3.1.18 reference standard, n—a specimen of known thick-
metal reading is measured, recorded and deducted from the ness used to verify the accuracy of a coating thickness
thickness of each coat, in order to correctly state the thickness measuring gage.
of the coating over the surface roughness. 3.1.18.1 Discussion—A reference standard may or may not
3.1.5 calibration, n—the high-level, controlled and docu- be traceable to a National or International registry. References
mented process of obtaining measurements on traceable cali- should include notations stating the accuracy of the standard.
bration standards over the full operating range of the gage, then Reference Standards may be coated or plated metal plates, or
making the necessary gage adjustments (as required) to correct may be shims (with samples of substrate material). Subject to
any out-of-tolerance conditions. agreement between the buyer and the seller, a piece of coated
3.1.5.1 Discussion—Calibration of coating thickness gages substrate may also be used as a project-specific reference
is performed by the equipment manufacturer, an authorized standard. Traceability may not be required for reference
agent, or by an authorized, trained calibration laboratory in a standards used on a day-to-day basis. However, the reference
controlled environment using a documented process. The standard should contain the stated value and the degree of
outcome of the calibration process is to restore/realign the gage accuracy.
to meet/exceed the manufacturer’s stated accuracy. 3.1.19 shims, n—strips of flat, non-metallic sheet (typically
polyester, with the thickness stated or referenced in some form)
used to adjust (optimize) a coating thickness gage in the
3
Available from Society for Protective Coatings (SSPC), 40 24th St., 6th Floor, intended range of use, over the surface of the representative
Pittsburgh, PA 15222-4656. (see www.sspc.org) substrate material.

2
D 7091 – 05
3.1.19.1 Discussion—Other uses include: placement over suring gages, as well as the proper methods for obtaining
soft coatings to obtain thickness measurements without the coating thickness measurements on both ferrous and non-
gage probe depressing the coating film, and verification of gage ferrous metal substrates.
operation. Also known as “foils.”
3.1.20 spot measurement, n—the average of at least three 5. Principles, Advantages, and Limitations of Gages
measurements made within a 12 mm (1⁄2 in.) diameter circle. 5.1 Type 1 Magnetic Pull-Off gages are mechanical instru-
3.1.21 substrate, n—the base material, the type of surface, ments that measure the force required to pull a permanent
or the component that is being coated. magnet from a coated ferrous metal substrate. The magnetic
force of attraction to the steel substrate beneath the coating is
NOTE 1—This practice addresses only metal substrates.
opposed by a spring or coil. Tension is applied to the
3.1.22 surface profile, n—surface roughness generated dur- spring/coil until the magnetic attraction to the steel is over-
ing the manufacturing process (for example, casting), or the come. The gage must be placed directly on the coated surface
average peak-to-valley depth generated by some power tools to obtain a measurement. The force holding the permanent
and by abrasive blast cleaning operations. magnet to the ferrous base is inversely proportional to the
3.1.23 Type 1 gage, n—a magnetic pull-off instrument that thickness of the coating layer(s) between the magnet and the
measures the dry film thickness of non-magnetic coatings over ferrous substrate. For example, a thin coating applied to a
a ferrous metal base. ferrous substrate will require greater spring tension to pull the
3.1.23.1 Discussion—For Type 1 gages, a permanent mag- magnet off than will a thicker coating, since the magnet is
net is brought into direct contact with the coated surface. The closer to the ferrous substrate with the thinner coating. This
force necessary to pull the magnet from the surface is measured inverse relationship is reflected on the nonlinear gage scale.
and interpreted as the coating thickness value on a scale or Most Type 1 magnetic pull-off gages do not require a power
display on the gage. Less force is required to remove the source (for example, batteries). The manufacturer’s stated
magnet from a thick coating. The scale is nonlinear. Also accuracy is typically 5 to 10 % of the reading.
known as “pull-off gage.” 5.2 Type 1 magnetic pull-off gages are susceptible to
3.1.24 Type 2 gage, n—an electronic instrument that uses vibrations, which may cause the magnet to release from the
electronic circuitry and (but not limited to) the magnetic coated substrate prematurely, yielding a false high value. The
induction, Hall-effect or the eddy current principles, or both, to manually operated gages may be susceptible to human error
convert a reference signal into a coating thickness reading. caused by inadvertently turning the dial wheel past the point at
3.1.24.1 Discussion—The probe of a Type 2 gage remains which the magnet pulls from the surface, yielding a false low
on the surface during the measurement process. Also known as measurement. Type 1 gages should not be used on soft or tacky
“constant pressure probe” or “electronic” gages. coatings, as the magnet may adhere to the coating causing false
3.1.25 verification of accuracy, n—obtaining measurements low measurements, or coating materials may dry on the magnet
on a reference standard prior to gage use for the purpose of causing false high measurements. The exposed magnet may
determining the ability of the coating thickness gage to produce attract metal filings, which can contaminate the magnet and
reliable values, compared to the combined gage manufacturer’s cause false high measurements. Type 1 gages cannot be used to
stated accuracy and the stated accuracy of the reference measure the thickness of coatings applied to non-ferrous metal
standard. substrates. The manufacturer’s specifications will contain a
temperature operating range. Use of the gage outside of this
4. Significance and Use range may generate false coating thickness measurements and
4.1 Most specifications for commercial and industrial coat- may damage the instrument.
ings projects stipulate a minimum and a maximum dry film 5.3 Type 2 gages are instruments that employ a measuring
thickness for each layer in a coating system. Additionally, most probe and the magnetic induction, Hall-effect or eddy-current
manufacturers of high performance coatings will warranty measurement principle in conjunction with electronic micro-
coating systems based upon, in part, achieving the proper processors to produce a coating thickness measurement. The
thickness of each layer and the total coating system. Even if a gage probe must be placed directly (in a perpendicular posi-
project specification is not provided, the coating manufactur- tion) on the coated surface to obtain a measurement. For gages
er’s recommendations published on product data sheets can measuring on ferrous substrates, the magnetic induction or
become the governing document(s). Equipment manufacturers Hall-effect principles determine the effect on the magnetic field
produce non-destructive coating thickness testing gages that generated by the probe due to the presence of the substrate. For
are used to measure the cumulative thickness of the coating gages measuring on non-ferrous metals, the gage probe is
layers, after they are dry. The manufacturers provide informa- energized by alternating current that induces eddy currents in
tion for the adjustment and use of these gages, normally in the the metal substrate. Note that gages/probes for measuring
form of operating instructions. The user of this equipment must coating thickness on non-ferrous metals should not be used to
be knowledgeable in the proper operation of these devices, measure coating thickness on ferrous surfaces, even though a
including methods for verifying the accuracy of the equipment reading may be displayed.
prior to, during and after use as well as measurement proce- 5.4 Type 2 gages are available with integral, separate or
dures. remote (unconnected) probes, and they can be used to measure
4.2 This practice was prepared to describe the proper coating thickness on ferrous or non-ferrous metal substrates, or
methods for verifying the accuracy of coating thickness mea- both, depending on the probes supported by the particular gage

3
D 7091 – 05
platform. The thickness of the coating is displayed digitally. In tion showing traceability to a national standard can be issued.
general, access to tight areas is easier with Type 2 gages, There is no standard time interval for re-calibration, nor is one
especially those equipped with separate or remote probes. Type absolutely required, but a calibration interval can be estab-
2 gages are available with memory, measurement batching, lished based on experience and the work environment. A
statistical analysis packages and data download/print-out. The one-year calibration interval is a typical frequency suggested
manufacturer’s stated accuracy is typically 1 to 3 % of the by many gage manufacturers. The type of reference standard
reading. used to verify gage accuracy should be predicated upon the
5.5 Most electronic coating thickness-measuring gages can type that provides the best and most appropriate method of
be verified for accuracy using either traceable reference stan- verification, considering the type of gage, the sample surface
dards or measured non-metallic shims. Gage operation should geometry, and prevailing contract requirements.
be verified on prepared, uncoated substrate having the same 7.2 Type 1 (magnetic pull-off) and Type 2 (electronic)
composition, shape and surface roughness to which the coating coating thickness gages should be verified for accuracy (and
will be applied to, for the intended range of use. If necessary, adjusted if required and when possible) prior to, during and
the gage should be adjusted as described in 7.6 or 7.7. Gages after each period of use. Type 1 and Type 2 gages can be
that cannot be adjusted by the user should be returned to the verified for accuracy using coated/plated reference standards or
manufacturer for calibration if the readings obtained on the shims (see Note 5 and Appendix X1). The gage should be
reference standards or shims are outside of the combined verified for accuracy in the intended range of use. Also, the
accuracy of the standard/shim and the manufacturer’s stated probe should be examined for cleanliness before verifying the
gage accuracy. accuracy and before obtaining coating thickness measure-
5.6 Type 2 gages should not be used on soft or tacky ments.
coatings, as the pressure on the probe can indent the coating 7.3 Verification of Accuracy of Type 1 Gages Using Coated/
yielding false low measurements, or coating materials may Plated Reference Standards—Type 1 coating thickness gages
contaminate the probe yielding false high measurements. A can be verified for accuracy using traceable reference standards
shim (of known thickness) can be placed on top of the (for example, NIST4 Coating Thickness Reference Standards)
soft/tacky coating film and a measurement of the coating or proprietary reference standards produced by coating thick-
thickness obtained by subtracting the shim thickness from the ness gage manufacturers. If the gage reading is outside of the
total measurement of the shim and the coating. Note that some combined accuracy of the coating thickness standard (test
Type 2 gages can be programmed to automatically deduct the block) and the manufacturer’s stated gage accuracy, the instru-
shim thickness (known as “zero offset”). Type 2 gages may be ment should be returned to the manufacturer or authorized
sensitive (to some degree) to substrate effects including, but agency for calibration. Type 1 coating thickness gages should
not limited to edges, corners and holes in the substrate, as well never be adjusted by the user.
as substrate thickness, curvature or conductivity, or both. In 7.3.1 For example, if the gage accuracy is 65 % and the
general, the user should remain a minimum of 25 mm (1 in.) standards accuracy is 65 %, then the combined accuracy of the
from edges when measuring coating thickness, unless the gage and the standard will be 67 % as given by the sum of the
probe specification indicates otherwise. The manufacturer’s squares formula:
specifications will contain a temperature operating range. Use
of the gage or the probe outside of this range may generate =52 1 52 5 7.071 or approximately 7 % (1)
false coating thickness measurements and may damage the 7.3.2 For the gage to be in agreement with the standard, the
instrument. average thickness measured by the gage must be within 67 %
of the standard’s thickness. If the average thickness measured
6. Test Specimen on a 254 µm (10 mil) reference standard is between 236 µm
6.1 The test specimen can be the coated structure or (9.3 mils) and 272 µm (10.7 mils), the gage is properly
component/part on which the thickness is to be evaluated, or adjusted. The minimum value of 236 µm is calculated as 254
can be test panels of similar roughness, shape, thickness, µm minus 7 % of 254 µm (9.3 mils is 10 mils minus 7 % of 10
composition and magnetic properties on which it is desired to mils); the maximum of 272 µm is 254 µm plus 7 % of 254 µm
measure the coating thickness. (10.7 mils is 10 mils plus 7 % of 10 mils). Otherwise the
NOTE 2—Applicable test panel description and surface preparation reliability of the gage is suspect.
methods are given in Practices D 609 and D 1730. 7.4 Verification of Accuracy of Type 1 Gages Using Shims—
NOTE 3—Coatings should be applied in accordance with Practices Type 1 coating thickness gages may be verified for accuracy
D 823 or as agreed upon between the contracting parties. using shims (see Note 5 and Appendix X1). If the user elects to
NOTE 4—Test panels may be fabricated from thin gage materials and use shims to verify the accuracy of Type 1 gages, the shim(s)
special consideration for calibration adjustment may be required.
representing the intended range of use should be placed onto a
7. Frequency and Methods for Verifying the Accuracy smooth steel surface (zero plate), and a measurement taken. If
and for Adjusting a Coating Thickness Gage the gage reading is outside of the combined accuracy of the
7.1 Calibration of coating thickness gages is performed by
the equipment manufacturer, an authorized agent, or by an
authorized, trained calibration laboratory in a controlled envi- 4
Available from National Institute of Standards and Technology (NIST), 100
ronment using a documented process. A Certificate of Calibra- Bureau Dr., Stop 3460, Gaithersburg, MD 20899-3460.

4
D 7091 – 05
shim and the manufacturer’s stated gage accuracy, the instru- be verified for accuracy during the measurement procedure and
ment should be returned to the manufacturer or authorized after the coating thickness measurements have been acquired.
agency for calibration. See 7.3 for example of calculating the 7.6.3 Since the gage is verified for accuracy using smooth-
combined gage and calibration shim accuracy. Otherwise the surfaced reference standards, a compensation value may be
reliability of the gage is suspect. Type 1 coating thickness required if the substrate to be coated is roughened from the
gages should never be adjusted by the user. manufacturing process (for example, castings) or from abrasive
7.5 Since the gage is verified for accuracy using smooth- blast cleaning. This is known as a Base Metal Reading or
surfaced reference standards (or using a smooth zero plate), a BMR. The BMR is the effect of substrate roughness on a
compensation value may be required if the substrate to be coating thickness-measuring gage. The user obtains a mini-
coated is roughened from the manufacturing process (for mum of ten (10) readings on the prepared, uncoated substrate.
example, castings) or from abrasive blast cleaning. This is The arithmetic mean of these values becomes the Base Metal
known as a Base Metal Reading or BMR. The BMR is the Reading. The BMR is deducted from the coating thickness
effect of substrate roughness on a coating thickness gage. The values, in order to report the thickness of the coating layer(s)
user obtains a minimum of ten (10) readings on the prepared, over the surface profile.
uncoated substrate. The arithmetic mean of these values 7.7 Verification of Accuracy of Type 2 Gages Using Shims—
becomes the Base Metal Reading. The BMR is deducted from Type 2 coating thickness gages can be verified for accuracy
the coating thickness values, in order to report the thickness of using shims. The shim(s) representing the intended range of
the coating layer(s) over the surface profile. use should be placed onto the prepared, uncoated surface, and
a measurement taken. Most Type 2 gages can be adjusted using
NOTE 5—SSPC-PA2 prohibits the use of shims to verify the accuracy of either a one-point or a two-point procedure. Gages that cannot
Type 1 gages. See Appendix X1 for more information. be adjusted by the user should be returned to the manufacturer
7.6 Verification of Accuracy of Type 2 Gages Using Coated/ or authorized agency for calibration if the readings obtained on
Plated Reference Standards—Type 2 coating thickness gages the shim(s) are outside of the combined accuracy of the shim
can be verified for accuracy using traceable reference standards and the manufacturer’s stated gage accuracy. See 7.3 for
(for example, NIST Coating Thickness Reference Standards) example of calculating the combined gage and calibration shim
or proprietary reference standards produced by coating thick- accuracy. Otherwise the reliability of the gage is suspect.
ness gage manufacturers. Most Type 2 gages can be adjusted 7.7.1 Adjustment Using a One-Point Procedure—If the user
using either a one-point or a two-point procedure. Gages that elects to perform a one-point adjustment procedure, a shim
cannot be adjusted by the user should be returned to the representing the target range of gage use should be selected and
manufacturer or authorized agency for calibration if the read- placed onto the prepared, uncoated surface, and a measurement
ings obtained on the reference standards are outside of the taken. If the value displayed by the gage is within the
combined accuracy of the standard and the manufacturer’s combined accuracy of the shim and the manufacturer’s stated
stated gage accuracy. See 7.3 for example of calculating the gage accuracy (see 7.3), then no adjustment is necessary and
combined gage and calibration standard accuracy. Otherwise the gage is considered “optimized” for that targeted thickness.
the reliability of the gage is suspect. If the value is not within the combined accuracy of the shim
and manufacturer’s stated gage accuracy (see 7.3), then the
7.6.1 Adjustment Using a One-Point Procedure—If the user user should carefully follow the gage manufacturer’s instruc-
elects to perform a one-point adjustment procedure, a reference tions for proper adjustment, as the actual step-by-step proce-
standard representing the target range of gage use should be dures vary widely. Note that it may be necessary to “zero” the
selected and a measurement taken. If the value displayed by the gage on the uncoated substrate prior to conducting a one-point
gage is with the combined accuracy of the coating thickness adjustment procedure. It is recommended that the gage be
standard (test block) and the manufacturer’s stated gage verified for accuracy during the measurement procedure and
accuracy (see 7.3), then no adjustment is necessary and the after the coating thickness measurements have been acquired.
gage is considered “optimized” for that targeted thickness. If 7.7.2 Adjustment Using a Two-Point Procedure—If the user
the gage reading is outside of the combined accuracy of coating elects to perform a two-point adjustment procedure, two shims
thickness standard (test block) and the manufacturer’s stated (representing a thickness below and above the targeted range of
gage accuracy (see 7.3), then the user should carefully follow gage use) should be selected and placed onto the prepared,
the gage manufacturer’s instructions for proper adjustment, as uncoated surface. The user should carefully follow the gage
the actual step-by-step procedures vary widely. It is recom- manufacturer’s instructions for performing a two-point adjust-
mended that the gage be verified for accuracy during the ment, as the actual step-by-step procedures vary widely. Note
measurement procedure and after the coating thickness mea- that it may be necessary to “zero” the gage on the uncoated
surements have been acquired. substrate prior to conducting a two-point adjustment proce-
7.6.2 Adjustment Using a Two-Point Procedure—If the user dure. It is recommended that the gage be verified for accuracy
elects to perform a two-point adjustment procedure, two during the measurement procedure and after the coating
reference standards (representing a thickness below and above thickness measurements have been acquired.
the targeted range of gage use) should be selected. The user
should carefully follow the gage manufacturer’s instructions 8. Frequency for Measurement of Coating Thickness
for performing a two-point adjustment, as the actual step-by- 8.1 The thickness of a coating or a coating system can vary
step procedures vary widely. It is recommended that the gage from area to area on a structure or part. Accordingly, it is

5
D 7091 – 05
recommended that a number of measurements be obtained and during that work shift, in order to identify the deficient areas
the arithmetic mean calculated to determine the high, low and for subsequent rework.
average coating thickness in a given area. 8.3 For test panels having minimum dimensions of 3 by 5
8.2 For larger structures, five spot measurements (each spot in., three spot measurements (each spot measurement being the
measurement being the average of three individual gage average of three individual gage measurements acquired within
measurements acquired within a 12 mm (1⁄2 in.) diameter a 12 mm (1⁄2 in.) diameter circle) are recommended on each test
circle) should be obtained in every 9 m2 (100 ft2) area of face (one from the top third, one from the middle third and one
coated surface. The number of areas to measure is dependent from the bottom third).
on the size of the structure or part. The following general 8.4 For small parts or components, the number of coating
guidance is provided for measuring coating thickness on larger thickness measurements is typically based on the criticality of
surfaces. the application, and should be as agreed upon between the
purchaser and seller.
8.2.1 For surfaces totaling up to 28 m2 (300 ft2), measure
8.5 For mass quantities of manufactured products, the
and record the thickness as described in 8.2 in each 9 m2 (100
frequency of coating thickness measurement is dictated by the
ft2) area.
volume produced and should be based on statistical process
8.2.2 For surfaces totaling between 28 and 93 m2 (300 and control (SPC) calcualtions for sample size selection.
1000 ft2), measure and record the thickness as described in 8.2
in three random 9 m2 (100-ft2) areas. If thickness measure- 9. Report
ments outside of the specified range are observed, then more 9.1 The following items should be reported:
frequent measurements may be required. For example, the 9.1.1 Gage type (manufacturer, model and serial no.),
thickness may need to be measured in each 9 m2 (100-ft2) area 9.1.2 Date of calibration,
(in accordance with 8.2.1) coated during that work shift, in 9.1.3 Method of accuracy verification,
order to identify the deficient areas for subsequent rework. 9.1.4 Base Metal Reading (if appropriate),
8.2.3 For surfaces totaling greater than 93 m2 (1000 ft2), 9.1.5 Spot and area coating thickness measurements,
measure and record the thickness as described in 8.2 in three 9.1.6 Minimum and maximum thickness for each area,
random 9 m2 (100 ft2) areas for the first 93 m2 (1000 ft2) and 9.1.7 Depending upon the application, the individual gage
in one additional 9 m2 (100-ft2) area for each additional 93 readings may be recorded, and
m2 (1000 ft2). Example: for a structure comprised of 2323 9.1.8 Gage operator’s name.
m2 (25 000 ft2), a total of 27 — 9 m2 (100-ft2) areas should be
measured. If thickness measurements outside of the specified 10. Keywords
range are observed, then more frequent measurements may be 10.1 coating thickness; dry film thickness; eddy current
required. For example, the thickness may need to be measured thickness gages; magnetic gages; nondestructive thickness;
in each 9 m2 (100-ft2) area (in accordance with 8.2.1) coated paint thickness

APPENDIX

(Nonmandatory Information)

X1. PRECAUTIONS REGARDING VERIFICATION OF GAGE ACCURACY

X1.1 When selecting shims to verify the accuracy of Type standards include (but are not limited to):
1 coating thickness gages, it is necessary to be aware of X1.3.1 Permanent creases in the shim due to folding,
additional characteristics that can affect the measured values. X1.3.2 Air entrapment between the shim and substrate,
Shim stock typically has an inherent curve. This curve can act X1.3.3 Distortion due to environmental conditions, such as
as a leaf spring and cause a magnetic pull-off gage to be temperature, and
“pushed” off the surface prematurely, resulting in an incorrect X1.3.4 Shim thickness inconsistency (due to the pressure of
reading. the probe tip) that may be a permanent “dimple” in the shim.
X1.2 With some materials and thickness, it is possible that
the shim will not lie flat, which will also cause an erroneous X1.4 Even with these factors, in many applications, veri-
reading. Various techniques exist to minimize this effect, such fication of gage accuracy using shims directly on the sample to
as mounting the shim in a holder that maintains tension on the be measured can be more appropriate than using plated or
shim to eliminate the tendency of the shim to curve. Special coated standards. Some gage manufacturers produce certified
tensioned ring shims are also available. shims.
NOTE X1.1—Data produced using Type 1 (magnetic pull-off) gages and
X1.3 Other factors experienced with plastic shims (appli- shims placed onto an abrasive blast cleaned steel surface indicate that the
cable to both Type 1 and Type 2 coating thickness gages), surface texture beneath the shim has an adverse effect on the gage reading.
which are not usually present with coated or plated calibration Since Type 1 gages cannot be field adjusted, the use of shims to verify

6
D 7091 – 05
gage accuracy on abrasive blast cleaned surfaces is not recommended. If coated steel standards it is important to be aware of the effect
shims are selected to verify gage accuracy, they should be placed on a of the coating on some types of magnetic gages. For best
smooth steel plate (zero-plate). A Base Metal Reading (BMR) is obtained accuracy when measuring with magnetic induction principle
and subtracted from the coating thickness (to compensate for the effect of
roughness), as described in 7.5.
gages, consider the following:
X1.6.1 Verify gage accuracy on metal plated (conductive
X1.5 Independent of what standard is employed, they coating) standards when measuring conductive coatings (for
should be periodically verified to ensure the assigned value is example, chrome and zinc); verify gage accuracy on epoxy
correct. Even coated metal plates can wear or be damaged to an coated (non-conductive) standards when measuring non-
extent that gage readings are affected. conductive coatings (for example, paint).
X1.6.2 Gages that use the Hall-effect principle are not
X1.6 When verifying the accuracy of magnetic gages on effected by the conductive nature of the coating.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

You might also like