You are on page 1of 7

MAPANCalibration

- Journal ofof Special Relative


Metrology Society Humidity and23,
of India, Vol. Temperature
No. 2, 2008;(RHT) ......................Measurement
pp. 115-121

Calibration of Special Relative Humidity and


Temperature (RHT) Sensors and Evaluation and
Expression of Uncertainty in the Measurement

BHIKHAM SINGH, HARI KISHAN* and Y. P. SINGH

National Physical Laboratory (NPLI), CSIR


New Delhi - 110 012, India
*e-mail : hkishan@mail.nplindia.ernet.in

[Received : 06.03.2007 ; Revised : 07.01.2008 ; Accepted : 29.05.2008]

Abstract
This paper reports the calibration of special type of relative humidity and temperature sensor, which
gives the output in terms of millivolts, measured by precision digital multimeter (DMM) to further
estimation of relative humidity in the range of 25 %RH to 95 %RH. The sensor has been calibrated
against the standard aspirated psychrometer having a pair of matched dry and wet bulb quartz
thermometers. The temperature of wet bulb thermometer is effectively depressed when a ventilated
airflow is maintained at a speed more than 3 metre/second. The relative humidity of reference standard
psychrometer is computed using psychrometric equation to relate with the output of the RH indicator
under comparison. The paper also describes the details of estimation and expression of expanded
uncertainty in the calibration of the sensor at measured relative humidity. A sufficient number of data
were undertaken to evaluate the uncertainty at a confidence-level of approximately 95% for a normal
probability distribution.

1. Introduction standard of humidity against which other methods


are compared. The National Institute of Standards
The measurement and control of relative humidity and Technology (NIST), USA, National Physical
plays a significant role in industry, including various Laboratory (NPL), UK, Physikalisch-Technische
sectors such as electrical, pharmaceutical, textile, food Bundesanstalt (PTB), Germany, National Metrology
and aerospace. Its need is increasing day by day with Institute of Japan (NMIJ), Japan and Korea Research
the modernization of industries and awareness of Institute of Standards and Science (KRISS), Republic
quality system in regard to ISO 9000. This awareness of Korea use gravimetric method [1-6]. Various other
has led to the industries for getting their RH indicators techniques such as two-pressure, two-temperature,
calibrated against the National Humidity Standard. two pressure-two temperature, mixed flow and
The gravimetric method of water vapour measurement aspirated psychrometer [7-11] are used as the
is considered the most precise and accurate. This secondary humidity standards. At NPL, India an
method permits the measurement of the amount of aspirated psychrometric technique has been
water vapour in a given mass of the moist gas sample. established [12-14] as the reference humidity
The quantities measured in this method are standard. No measurement is complete without
fundamental; it gives an absolute measure of humidity. assigning the value of uncertainty to it. The detailed
Hence the gravimetric method is used as a primary study of the uncertainty associated with the
© Metrology Society of India, All rights reserved. measurement has been evaluated and discussed

115
Bhikham Singh, Hari Kishan and Y. P. Singh

[15-20] as per the GUM/ISO/IEC: 17025 documents. control equipment comprises dry and wet bulb contact
thermometers together with an air heater, a boiler, a
2. Experimental Setup fan, window heater, and a spray cooler. This
environmental chamber is being used to generate RH
The following instruments are required to be used
of different levels while the quartz thermometers carry
for calibration of RHT Sensor:
out the measurements. The Quartz thermometers and
1. Digital indicator (Testo 781) with Quartz pressure indicator DPI 145 are hooked on to a suitable
thermometers (dry and wet bulb) and PC adapter compatible computer, which collects the data every
second/minute of the two thermometers and pressure
2. Environmental chamber Gallenkamp Make sensor in a file.
(model OVR-403-010L).
The RHT sensor is kept very near to the aspirated
3. Anemometer with probe model AM-4201, make psychrometer and the readings of the RHT sensor in
LT Lutron Sr.No.480014. millivolt/volt are recorded manually at a regular
interval of time at the calibration temperature 24 ºC ±
4. Mini/small high-speed fan 4.5" size operated at 1 ºC. The reference humidity standard (an aspirated
220-230 volts ac supply. psychrometer as described above) has an expanded
uncertainty of ±1% RH at a confidence level
5. Digital pressure indicator, Model Druck DPI-145 approximately 95 % coverage factor k=2 with normal
probability distribution [14]. Fig. 1 shows the
6. Temperature & Humidity Indicator, Make
schematic block diagram of the entire experimental
Indosaw.
set-up used for the calibration of RHT sensors/
7. Special Relative Humidity Sensor under hygrometers (dial and digital type). A program (in
calibration with digital multi-meter. C++ language) reads the data and computes mean
value of relative humidity (RH) and the standard
3. Calibration Procedure uncertainty with the help of the software. The
evaluation of uncertainty is also checked by manual
The calibration setup for RH measurements for calculation in order to validate the software used. The
computing the RH is based on measuring the dry & relative humidity of reference standard is evaluated
wet bulb temperatures with the help of two matched, by using the following psychrometric equations
precision quartz thermometers (Model TESTO 781). [11-14].
These thermometers, having a resolution of 0.001 C,
have been periodically calibrated against the standard (%) RH = 100 x [Pd/Psd] (1)
PRT of the National Physical Laboratory, New Delhi
and are found to be reproducible within ± 0.005 ºC. where
The expanded uncertainties of both the thermometers
Psd = Saturation vapour pressure of water vapour at
have been calculated and estimated to be within
the dry bulb temperature (td)
± 0.03 ºC for the coverage factor, k = 2, at a level of
confidence 95 % approximately. These two quartz Pd = Vapour pressure of water vapour at dry bulb
thermometers are supported on stands very near to temperature (td)
each other. One of these is provided with a clean wick,
immersed in distilled water in a Pyrex bottle. A small Pd = Psw - AP (td -tw) (2)
fan is placed near to the thermometers blowing air at
a speed of >3 metre/second. The use of new wick where
over the wet bulb thermometer ensures the effect on
radiation interchange between the bulbs due to td = dry-bulb temperature (ºC)
airflow. An anemometer measures the speed of air.
tw = wet-bulb temperature ( ºC)
The entire setup (i.e. the quartz thermometers along
with fan) is then placed inside the Gallenkamp (td - tw) = depression (ºC)
environmental chamber (Model OVR-403-010L). The
environmental chamber is a stainless steel cabinet Psw = Saturation vapour pressure of water vapour
with glazed door. The basic temperature and humidity at the wet bulb temperature (tw)

116
Calibration of Special Relative Humidity and Temperature (RHT) ......................Measurement

Fig. 1. Schematic diagram of the experimental setup for calibration of RHT sensor

A = Psychrometric constant, 0.000666 / ºC for 4.1 Mathematical Model


moving air
The following mathematical model is expressed
P = Total atmospheric pressure on the basis of various components of uncertainty for
RH measurement in % for the calibration.
Substituting the value of Pd from Eq. (2) in Eq. (1), we
get RH(meas) = RH(S) ± ΔRH (4)
(%) RH = 100 x [Psw - AP (td -tw)]/[ Psd ] (3) ΔRH = U(RH) (5)
U(RH) = k x uc(RH) (6)
4. Evaluation of Uncertainty
uc(RH) = √[u12 δRH(S)+ u22 δRH(C)+ u32 δRH(A)+u42
In our experiment, the uncertainty of measurement δRH(P)+ u52 δRH(D)+u62 δRH(R)+u72
associated with the input estimates is evaluated by δRH(RO)+ u82 δ RH(DVM)+u92 δRH(Rep)]
'Type-A' method of evaluation. The Type-A evaluation (7)
of standard uncertainty is performed by the statistical where
analysis of a series of observations. The Type-B
evaluation of standard uncertainty is the method of u1 [δRH(S)] = Standard uncertainty component
evaluating the uncertainty by means other than the of Reference Standard Aspirated
statistical analysis of a series of observations. In this psychrometer
case the evaluation of the standard uncertainty is
u2 [δRH(C)] = Standard Uncertainty due to
based on some other scientific knowledge. The various environmental Chamber
sources of uncertainty have been identified and
described in a mathematical model. The mathematical u3 [δRH(A)] = Standard Uncertainty due to
model has been formed to include the various error Anemometer
components affecting the measurement and contribute
u4 [δRH(P)] = Standard Uncertainty component
towards the uncertainty in the calibration. Some of
due to Pressure Indicator/Sensor
the components are already included in the
uncertainty due to standard reference RH u5 [δRH(D)] = Standard Uncertainty component
Standard. due to Drift of Reference Standard

117
Bhikham Singh, Hari Kishan and Y. P. Singh

u6 [δRH(R)] = Standard Uncertainty component Referring to the data in Table 1, calculation of


due to resolution of DUC uncertainty (Type-A) was estimated as;

u7 [δRH(RO)] = Standard Uncertainty component ua = 0.062272 % RH


due to rounding off significant
digit 4.3 Evaluation of Type 'B' Components of Uncertainty,
(uiB)
u8 [δRH(DVM)]= Standard Uncertainty component
due to DVM/DMM used 4.3.1 Uncertainty contribution due to reference
standard (an aspirated psychrometer)
u9 [δRH(Rep)] = Standard Uncertainty due to
Standard deviation of the Data on The reference standard is calibrated and traceable
DUC. to the national standard of temperature and
pressure. The expanded uncertainty of the
uc(RH) = √Σ[δf(RH)/δxi (RH)] 2ui2(RH) reference standard is ± 1% RH at approximately
95% confidence level with coverage factor k = 2
Sensitivity Coefficients 'c i' = δf(RH)/δxi (RH) =
(1x 102 %RH/V) for a normal probability distribution. Thus the
standard uncertainty, u1 is obtained by dividing
The components of uncertainty due to the expanded uncertainty by coverage factor,
anemometer and pressure indicator are included in therefore, u1 = 1 % RH/2 = 0.5 % RH, Sensitivity
the uncertainty contribution of the reference RH coefficient c1=1, Degree of freedom, ν1 = ∞.
standard.
4.3.2 Uncertainty contribution due to the stability
4.2 Evaluation of Type 'A' Component of Uncertainty and uniformity of the environmental chamber
(ua) at 50% RH
Based on our measurements of uniformity and
Besides observations of the reference humidity stability inside the humidity oven, the maximum
standard, we also note down simultaneously the variation in RH from one corner to the other is
readings of RHT sensor as shown in the digital multi- found to be 1.465 % giving a limiting value of ±
meter (DMM) display of the special sensor (which is 0.7325 % RH. Assuming rectangular probability
to be calibrated). Table 1 shows some of the typical distribution, the standard uncertainty, u2 = 0.7325
observations of RHT sensor in Volts/milli Volts in % /√3 ≅ 0.423 % RH, Sensitivity coefficient,
(DMM) under calibration. It is a set of 10 observations. c2=1,Degree of freedom, ν2 = ∞.

Table 1
Observation
S.No. RH % of DUC reading Deviation from (Vi-Vavg )2
Ref. Std Vi, in Volts average in V, (Vi-Vavg) in volts

1 50.951 0.510 3.1 x 10-3 9.61 x 10-6


2 50.862 0.509 2.1 x 10-3 4.41 x 10-6
3 50.861 0.507 1 x 10-4 1 x 10-8
4 50.922 0.508 1.1 x 10-3 1.21 x 10-6
5 50.892 0.507 1 x 10-4 1 x 10-8
6 50.928 0.507 1 x 10-4 1 x 10-8
7 51.010 0.507 1 x 10-4 1 x 10-8
8 50.650 0.506 -9 x 10-4 8.1 x 10-7
9 50.314 0.505 -1.9 x 10-3 3.61 x 10-6
10 50.240 0.503 -3.9 x 10-3 1.521 x 10-5
Mean 50.763 % RH Vavg = 0.5069 Σ = 3.49 x 10-5

118
Calibration of Special Relative Humidity and Temperature (RHT) ......................Measurement

4.3.3 Uncertainty contribution due to the resolution distribution = ± 0.07/√3 = 0.04041 % RH,
of the Device Under Calibration Sensitivity coefficient, c5= 1, Degree of freedom, ν5
= ∞, u5=0.04041% RH.
The resolution of RHT sensor in digital multi
meter (DMM) is 0.00001V. This gives a voltage 4.4 Combined Standard Uncertainty, uc
resolution limit of ± 5x10 -6 V. Assuming
rectangular probability distribution, the standard The combined uncertainty uc(y) in output estimate
uncertainty u3 is u3 = 5x10-6V /√3 = 2.8867513x is equal to root sum square (RSS) of all contributions
10-6 V, Sensitivity coefficient, c3 = 1x 102 %RH/V, discussed above
u3 = 2.8867513x10-4 % RH, Degree of freedom,
ν3 = ∞. uc 2 = Σci2ui2 = 0.6602498 %RH (8)

4.3.4 Uncertainty due to drift of the standard RH 4.5 Effective Degree of Freedom

Drift caused in the Reference Standard in the years The effective degree of freedom (νeff) of combined
2004-06 = 0.06686 % RH, standard. Uncertainty standard uncertainty is calculated using the well
limit = ± 0.06686/√3=0.0386%RH.Assuming known Welch-Satterthwaite equation given as
rectangular probability distribution. Sensitivity following:
coefficient, c4 = 1, Degree of freedom, ν4 = ∞, u4
= 0.0386 % RH. νeff = [uc(RH)]4/[ui(RH)4/νi] = 113738 ∞ (9)

4.3.5 Uncertainty due to rounding off significant digits/ 4.6 Expanded Uncertainty, U
decimal places
For effective degree of freedom, νeff ≅ ∞ at 95.45%,
Limiting value of RH for significant digits = level of confidence, the 't' factor from student's 't' table
± 0.07. Assuming rectangular probability is 2, so the coverage factor 'k' =2.

Uncertainty Budget
Source of Estimated Std. Probability Sensitivity Uncertainty Degree
Uncertainty Value Uncertainty Distribution Coefficient Contribution of
Divisor ci (% RH) Freedom
(a) (b) (c) (d) f = (b/c).d

u1 (Ref. Std.) 50.69 % RH ±1 Normal Type B, 2 1 0.5 ∞

u2 (Hum. Oven) - ±0.7325 Rectangular Type B, √3 1 0.423 ∞

u3 (Resolution) - 5x10-6 V Rectangular Type B, √3 1 x 102 % 2.8867513x10-4 ∞


RH /V
u4 (Drift) - 0.06686 Rectangular Type B, √3 1 0.0386 ∞

u5 (Round off) - 0.07 Rectangular Type B, √3 1 0.04041 ∞

(ua) Repeatability - 0.5069 V Normal Type A, √10 1 x 102% 0.062272 9


1.962074 x 10 -3
RH /V
(uc) Combined - uc = √Σui2 = 0.6602498 % RH 0.6602498 ∞
Uncertainty

(U) 50.69 % RH ≅ ± 1.32 %RH ∞


Expanded νeff = 113737
Uncertainty U = uc x k =1.3204996

119
Bhikham Singh, Hari Kishan and Y. P. Singh

Calibration Results

S. No. Value of Ref. Value measured in RHT Expanded Uncertainty of


Std. (RH %) measurement, % RH
V RH%

1. 25.25 0.2489 24.89 ± 1.32


2. 50.76 0.5069 50.69 ± 1.32
3. 95.47 0.9261 92.61 ± 1.32

The expanded uncertainty, U Acknowledgement


U = kxuc(RH) = 1.32% RH (10) The authors express their sincere gratitude to
Prof. Vikram Kumar, Director of the National Physical
5. Results and Discussion
Laboratory, New Delhi for his constant
The relative humidity & temperature Sensor encouragement. The help provided by Mr. Shiv Dutt
(RHT), which measures the (RH) output in terms of Sharma is greatly acknowledged.
the electrical potential was calibrated against the
References
reference RH standard generated in the environmental
chamber. This is an indirect comparison/calibration [1] A. Wexler and D.D. Raymond, Pressure
of RHT sensor whose output is further evaluated in Humidity Apparatus, J. Res. Natl. Bur. Std., 48
RH % using a sensitivity coefficient of 1 x 102% RH / (1952) 269.
V. Four numbers of similar sensors have been
[2] A. Wexler, Calibration of Humidity-Measuring
compared against the reference RH standard. A case
Instruments at the National Bureau of
study has been presented on calibration of one RHT
Standards, ISA Transaction, 7 (1968) 356-362.
sensor in the range 25 % RH to 95 % RH. The
measurements have been reported at 25 % RH, 50 % [3] T. Inamatsu, Establishment of Humidity
RH and 95 % RH and associated expanded Standard in Japan, Bull NRLM, 44 (1995)
uncertainty of calibration has also been evaluated and 107-113.
expressed under calibration results. [4] A.G. Forton and R.F. Pragnell, Development of
The reported expanded uncertainty of Primary Gravimetric Hygrometer for the UK
measurement is stated as the standard uncertainty National Humidity Standard Facility,
estimated at a confidence level of approximately 95%, Proceedings International Symposium on
k=2, for a normal probability distribution. Moisture and Humidity, Washington DC,
15-18 April 1985, 229-234.
6. Conclusion [5] A Guide to the Measurement of Humidity,
Published by The Institute of Measurement &
The RHT sensor undertaken for calibration was
Control, London, 1996.
a special sensor device, which gives an output in the
form of an electrical signal in volt to evaluate the [6] H.S. Nham, KRISS (Republic of Korea) Private
relative humidity (RH%). The sensor needs a high Communication.
stability environmental chamber to produce [7] A. Wexler and S. Hasegawa, Relative Humidity
effectively stable output voltage of the sensor. The -Temperature Relationships of Some Saturated
equipment utilized for comparison of such type of Salt Solutions in the Temperature Range 0 to
sensor was found to be significantly useful in the 50 C, J. Res. NBS, 53A (1954) 19.
calibration of specialized humidity sensors. These
[8] G. Scholz, PTB (Germany) Private
devices are used in the measurement of wind energy
Communication.
evaluation. These are frequently required to be
calibrated against the reference standard of humidity [9] M. Heinonen, Measurement, 17 (1996) 183-188.
established at NPL, New Delhi. [10] M. Stevens and S.A. Bell, The NPL Standard

120
Calibration of Special Relative Humidity and Temperature (RHT) ......................Measurement

Humidity Generator: an Analysis of [15] Bhikham Singh and Hari Kishan, Calibration
Uncertainty by Validation of Individual Proc.No.DP#1.03(Part-C)/Doc.3/CP#01,
Component Performances, Meas. Sci. Technol., Document Manual of Humidity Standards of
3 (1992) 943-952. National Physical Laboratory, New Delhi.
[11] H. Kishan and B. Singh, Mixed Flow Relative [16] Hari Kishan, Calibration of Relative Humidity
Humidity Generator, MAPAN- Journal of Indicators and Evaluation of Uncertainty,
Metrology Society of India, 23 (2008) 21-24. NABL News, 41 (2006) 27-32.
[12] Hari Kishan, Bhikham Singh, V.P. Wasan, R.S. [17] Guide to the Expression of Uncertainty in
Khandekar, S.S. Verma and R.G. Sharma, Measurements, International Organization for
Development of an Aspirated Psychrometer Standardization (ISO), First Edition (1995).
using Two Quartz Thermometers, MAPAN-
[18] Expression of the Uncertainty of Measurement
Journal of Metrology Society of India, 12 (1997)
119-122. in Calibration, Published by European
cooperation for Accreditation of Laboratories
[13] Hari Kishan, NPL (India) Humidity Standard, (EAL), Reference EAL-4/02, 1999.
Proceedings 3rd International Conference on
Metrology in New Millennium and Global [19] NABL-141, Guidelines for Estimation and
Trade (MMGT-2001), NPL, New Delhi, 8-10 Expression of Uncertainty in Measurements,
February 2001, 62-65. (2000).
[14] Hari Kishan and Bhikham Singh, [20] Y.P. Singh, Evaluation and Expression of
Establishment of the Indian National Uncertainty in the Calibration of Screw Type
Humidity Standard Facility, 4th International Industrial Lamps Using Photoelectric
Symposium on Humidity and Moisture, Taipei, Pyrometer, MAPAN- Journal of Metrology
Taiwan, 16-19 September 2002, 566-570. Society of India, 22 (2007) 235-245.

121

You might also like