You are on page 1of 11

Clin Oral Impl Res 2000: 11: 465–475 Copyright C Munksgaard 2000

Printed in Denmark ¡ All rights reserved

ISSN 0905-7161

Magnitude and distribution of occlusal forces


on oral implants supporting fixed prostheses: an
in vivo study
Duyck J, Van Oosterwyck H, Vander Sloten J, De Cooman M, Puers R, Joke Duyck1,
Naert I. Magnitude and distribution of occlusal forces on oral implants Hans Van Oosterwyck2,
supporting fixed prostheses: an in vivo study. Jos Vander Sloten2,
Clin Oral Impl Res 2000: 11: 465–475. C Munksgaard 2000. Michel De Cooman3,
Robert Puers3, Ignace Naert1
1
Department of Prosthetic Dentistry/
BIOMAT research group, 2Department of
Since loading is increasingly believed to be a determining factor in the
Mechanics – Division of Biomechanics
treatment outcome with oral implants, there is a need to expand the and Engineering Design, 3Department of
knowledge related to the biomechanics of oral implants. The aim of this Electrotechnics – ESAT-MICAS, Catholic
study is to gain insight in the distribution and magnitude of occlusal forces University Leuven, Belgium
on oral implants carrying fixed prostheses. This is done by in vivo quanti-
fication and qualification of these forces, which implies that not only the
magnitude of the load but also its type (axial force or bending moment)
will be registered. A total of 13 patients with an implant supported fixed Key words: oral implants – fixed
full prosthesis were selected. Occlusal forces on the supporting implants prosthesis – biomechanics – strain
were quantified and qualified during controlled load application of 50 gauges – in vivo
N on several positions along the occlusal surface of the prostheses and
during maximal biting in maximal occlusion by use of strain gauged abut- Prof. I.E. Naert, Head, Department of
ments. The test was conducted when the prostheses were supported by all Prosthetic Dentistry, BIOMAT Research
(5 or 6) implants and was repeated when the prostheses were supported Group, Catholic University Leuven,
U.Z. St Raphaël, Kapucijnenvoer 33,
by 4 and by 3 implants only. Despite considerable inter-individual vari-
B-3000 Leuven, Belgium
ation, clear differences in implant loading between these test conditions Tel.: π32 16 332438
were seen. Loading of the extension parts of the prostheses caused a hin- Fax: π32 16 332309
ging effect which induced considerable compressive forces on the im- e-mail:
plants closest to the place of load application and lower compressive or Ignace.Naert/med.kuleuven.ac.be
tensile forces on other implants. On average, higher forces were observed
with a decreasing number of supporting implants. Bending moments were Accepted for publication 20 September
highest when 3 implants only were used. 1999

Since osseointegration was introduced as a pri- 1997; Barbier & Schepers 1997), and human
mary demand for oral implant success, fixed pros- (Glantz et al. 1993; Mericske-Stern et al. 1996)
theses supported by osseointegrated implants may studies, there is a growing belief that implant fail-
be regarded as a reliable treatment alternative (Ad- ure might be caused by a disequilibrium of forces
ell et al. 1981; Lindquist et al. 1987, 1996; Al- acting on the oral implants (Esposito et al. 1998).
brektsson 1988; Haraldson & Zarb 1988; Lund- Although an adaptive remodelling response is ob-
qvist & Haraldson 1992; Naert et al. 1992). De- served around oral implants, the biomechanical as-
spite the favourable long-term success rates, pects of these processes remain uncertain (Hoshaw
success cannot be guaranteed. Some animal studies et al. 1994). A first essential step towards the in-
(Becker et al. 1990; Lindhe et al. 1992) focused on sight in the forces which are transferred to the
peri-implant infection as a possible cause for im- bone and which therefore affect the latter, is to
plant failure. Based on in vitro (Smedberg et al. quantify and qualify the forces applied on oral im-
1996), animal (Hoshaw et al. 1994; Isidor 1996, plants supporting an oral prosthesis. In addition

465
Duyck et al.

of fracture of a resin facet. No further mechanical


complications occurred and none of the implants
became mobile. The mean marginal bone loss –
rated mesially and distally on retro-alveolar radio-
graphs that were placed parallel to the implant
body using a long-cone device – averaged 0.15 mm
(SEM: 0.22 mm), the first year included.

Measuring device
Standard 5.5- and 7-mm abutments were equipped
with three strain gauges (Type FLG-02–11 of
Fig. 1. Fixed full prosthesis supported by strain gauged abut- TML; Tokyo Sokki Kenkyujo Co., Ltd.; Resis-
ments. tance: 120∫0.3 W; Gauge factor: 2.05∫1%) each,
120æ apart. These strain gauges were adapted to
the abutment surface to register deformations of
to this, finite element analyses based on individual the latter. Each strain gauge was wired separately
CT-Scan data and to some extent animal experi- into a Wheatstone bridge and when sufficiently
mental studies, can provide information about the balanced, the output voltage of the bridge is pro-
way these forces will be transferred to and affect portional to the resistance change of the strain
the surrounding bone. By in vivo registration of the gauge, which is in turn proportional to the strain
axial forces and bending moments that are induced in the abutment. A computer program under Win-
on the implants during controlled load application dows has been developed to process the data and
and during clenching, this study contributes to the to transform these data into a numerical represen-
first step. tation of the axial force and 2 bending moments

Materials and methods


Patients
A total of 13 patients with an implant supported
fixed prosthesis in the fully edentulous mandible
(9) or maxilla (4) were selected. In all patients
Brånemark implants (Nobel Biocare, Sweden) were
installed and 5.5- or 7-mm standard abutments
were connected to the implants to support the
prostheses. All original abutments were replaced
by strain gauged abutments of the same dimen-
sions to allow load registration as described below
(Fig 1). All but one prostheses were supported by
6 oral implants. The remaining prosthesis was sup-
ported by 5 implants. Moreover, in 2 patients it
was only possible to measure the load on 5 of the 6
supporting implants because there was not enough
inter-implant space to replace all abutments by the Fig. 2. Schematical illustration of a fixed full prosthesis sup-
strain gauged abutments which are slightly broader ported by 6 oral implants. The numbers indicate the positions
due to the attached strain gauges. In the antagon- along the occlusal surfaces of the prostheses where 50 Newton
istic jaw 8 and 4 of the patients were provided with was applied: 1) above the distal extension, left quadrant; 2)
a fixed full implant supported prosthesis or a re- above the distal implant, left quadrant; 3) above the medial
implant, left quadrant; 4) above the mesial implant, left quad-
movable denture respectively, one had a natural rant; 5) in between the 2 mesial implants (frontal extension); 6)
dentition. All patients demonstrated a stable oc- above the mesial implant, right quadrant; 7) above the medial
clusion. The mean age of the patients was 58 years implant, right quadrant; 8) above the distal implant, right
(range: 41 to 75) and they were functioning suc- quadrant; and 9) above the distal extension, right quadrant.
The abutments in position 3 and 7 are left out when the pros-
cessfully with their oral prostheses since 4.9 years thesis is supported by only 4 implants. One additional mesial
on average (range: 0.5 to 12.75). Over this period, implant (in position 4 or 6) is excluded when the prosthesis is
3 fixed prostheses needed to be repaired because supported by only 3 implants.

466
Occlusal forces on oral implants supporting fixed prostheses

Fig. 3. (a) Axial force induced on each of the 6 implants supporting 1 particular prosthesis during controlled load application of
50 N on the 9 indicated places along the occlusal surface. (b) Bending moment induced on each of the 6 implants supporting 1
particular prosthesis during controlled load application of 50 N on the 9 indicated places along the occlusal surface.

around perpendicular axes. By use of the bending


moments around both perpendicular axes, the Test conditions
total bending moment on the abutments could be To create situations, which can be compared with
calculated (Duyck et al. 1999). To apply measur- one another, the miniature bite fork was used to
able loads at the level of the occlusion, a miniature apply controllable loads on several positions on
bite fork was used (Fløystrand et al. 1982). When the prostheses. The bite fork was placed at defined
the patient bit on the fork, the forces were regis- consecutive positions along the occlusal surface of
tered both on the occlusal and the abutment level. the prostheses. The positions are indicated in Fig
This allowed excluding any influence of the 2. Because there was not enough space to measure
existing occlusal arrangement and articulation be- between the 2 mesial implants in 2 of the pros-
tween patients on the load registration. theses, no measurements could be performed while

467
Duyck et al.

loading those prostheses on position 5. Except for


one, all prostheses had distal extensions of one
cantilever unit. Because these cantilever units were
lacking in one prosthesis, measurements when
loading on positions 1 and 9 were impossible. The
patients could control the applied force on the bite
fork by means of visual feedback and they were
asked to bite with a force of 50 N. The bite fork
forces and the forces on the abutments were re-
corded simultaneously. The registrations of the
forces on the abutments were collected at the same
Fig. 4. Average (SEM) axial force and bending moment per time point as when the patient bit approximately
abutment for the 3 test conditions (6/5, 4, and 3 supporting 50 N. Each strain gauge value was a mean of 20
implants).
consecutive registrations, representing 0.4 s of

Fig. 5. (a) Average (SEM) axial force on the abutments during loading of the ipsi-lateral distal extension with 50 N for the 3 test
conditions (6/5, 4, and 3 supporting implants). (b) Average (SEM) bending moment on the abutments during loading of the ipsi-
lateral distal extension with 50 N for the 3 test conditions (6/5, 4, and 3 supporting implants).

468
Occlusal forces on oral implants supporting fixed prostheses
registration time in total. A linear relationship be-
tween applied and transferred forces was assumed,
allowing for a recalculation of the forces on the
bite fork and on the abutments to exactly 50 N
to make comparisons possible. In addition, load
registrations were performed during maximal
biting in maximal occlusion (clenching). The used
strain gauge values are a mean of 10 consecutive
registrations, representing 0.2 s of registration
time.
All tests were performed when the prostheses
were supported by all 5/6 implants and were re-
peated when the prostheses were only supported
by 4 and 3 implants respectively. In case the pros-
theses were only supported by 4 implants, the two
abutments connected on the implants positioned
in the middle of each quadrant (medial implants)
were removed. In case the prostheses were only
supported by 3 implants, an additional mesial
abutment was removed (see Fig. 2).
To get an idea of the influence of the antagonistic
teeth or restoration on the load distribution among
the implants, a comparison was made between the
fixed prostheses in the mandible. Five out of the 9
mandibular implant supported prostheses antagon-
ized a natural dentition or a fixed restoration, while
the remaining 4 occluded with a maxillar removable
full denture. The load distribution during clenching Fig. 6. (a) Average (SEM) axial force on the abutments during
in both groups was compared. loading in between the 2 mesial implants (frontal extension)
with 50 N for the 3 test conditions (6/5, 4, and 3 supporting
implants). (b) Average (SEM) bending moment on the abut-
Statistical analysis ments during loading in between the 2 mesial implants (frontal
extension) with 50 N for the 3 test conditions (6/5, 4, and 3
The data were compared by means of a linear supporting implants).
mixed model (Verbeke and Molenberghs 1997)
which corrects for the fact that several dependent
measurements were done. Tukey’s method was used
to correct the P-values for the paired comparisons (P∞0.0001) or 3 (P∞0.002) implants. The differ-
in order to correct for multiple testing. The analy- ence in axial force between 4 and 3 supporting im-
ses were performed using the SAS-program (SAS/ plants is not statistically significant. The bending
STAT software: Changes and Enhancements moments on the other hand are significantly higher
trough Release 6.12, SAS Institute Inc., Cary, NC, with 3 supporting implants in comparison with 4
USA). The level of significance was set at 0.05 for or 5/6 supporting implants (P∞0.0001). The differ-
all tests. ence in bending moment between 4 and 5/6 sup-
porting implants is not statistically significant.
Fig. 5a,b depicts the average induced axial
Results forces and bending moments during loading of the
To get a better insight in the individual loading of ipsi-lateral distal extension with 50 N. These data
each of the implants, 1 patient, representative for comprise 2 tests per patient since each prosthesis
all others, is fully presented for the controlled load has 2 distal extensions. Regardless of the number
application (50 N). Fig. 3a,b represents the forces of supporting implants, the highest compressive
on all 6 supporting abutments for 1 particular pa- forces are recorded on the ipsi-distal implant
tient. The average axial forces and bending mo- (P∞0.0001) while tensile forces (negative axial
ments per abutment per test condition for the forces) are observed on the implants positioned
whole group during controlled load application of medially and mesially in each quadrant. The bend-
50 N are presented in Fig 4. The mean axial force ing moments on the ipsi-distal and ipsi-medial im-
is significantly lower when the prostheses are sup- plants do not differ significantly from each other,
ported by all implants in comparison with 4 but they are significantly higher than the bending

469
Duyck et al.

Fig. 7. (a) Average (SEM) axial force on the abutments during loading of the prostheses above the ipsi-medial implant with 50 N
for the 3 test conditions (6/5, 4, and 3 supporting implants). (b) Average (SEM) bending moment on the abutments during loading
of the prostheses above the ipsi-medial implant with 50 N for the 3 test conditions (6/5, 4, and 3 supporting implants).

moments on the other implants (P∞0.0008). The axial forces in that particular loading situation are
loading conditions of the implants are also investi- not significantly dependent on the number of sup-
gated when loading the prosthesis in the middle porting implants. The accompanying bending mo-
between the two mesial abutments (frontal exten- ments increase significantly (P∞0.0002) when the
sion). These data are presented in Fig. 6a,b. The prostheses are supported by only 3 implants in
axial forces that are recorded on the mesial im- comparison with 5/6 or 4 implants. Such a differ-
plants are significantly higher than on the medial ence is not observed between 5/6 or 4 supporting
(P∞0.02) and distal implants (P∞0.0001). Small implants. There is no significant difference in
compressive and tensile forces are observed on the bending moment between the different implants.
medial and distal implants. The difference in axial Fig. 7a,b summarizes the results when loading the
forces between the medial and mesial implants is prosthesis above the ipsi-medial implant. When all
significant (P∞0.002). Nevertheless, the observed implants support the prosthesis, the highest axial

470
Occlusal forces on oral implants supporting fixed prostheses

(P∞0.02) implants. The average (SEM) maximal


axial force on the distal implants in case of 5/6, 4
and 3 supporting implants is 106.2 N (23.3 N),
136.2 N (31.5 N), and 165.8 N (33.9 N) which is
significantly higher than the medial and the mesial
implants (P∞0.02). This was not statistically sig-
nificant for the maximal bending moments.
Peak axial forces and bending moments on the
implants supporting the mandibular prostheses
during clenching were selected. The prostheses ant-
agonizing a natural dentition or a fixed restoration
were compared with those antagonizing a remov-
able maxillar denture. The mean (SEM) peak axial
forces and bending moments were calculated for the
distally, medially and mesially positioned implants
and presented in Fig. 9a,b. Despite the lack of statis-

Fig. 8. (a) Average (SEM) maximal axial force registered on the


abutments during maximal biting in maximal occlusion for the
3 test conditions (6/5, 4, and 3 supporting implants). (b) Aver-
age (SEM) maximal bending moment registered on the abut-
ments during maximal biting in maximal occlusion for the 3
test conditions (6/5, 4, and 3 supporting implants).

force is seen on the ipsi-medial implant, although


the difference with the ipsi-mesial implant is not
statistically significant. Comparable bending mo-
ments are observed on the ipsi-distal and ipsi-me-
dial implants, which are significantly higher
(P∞0.02) than on the other implants. The neigh-
bouring implants are subjected to significantly
higher axial forces (P∞0.02) when this ipsi-medial
abutment is dropped. Only minor forces are re-
corded on the implants at the contra-lateral side.
Those data comprise again 2 tests per patient since
the patients have 2 medial implants. When the
prosthesis is only supported by 3 implants, Fig.
7a,b suggests that the mesial implant is always po-
sitioned at the ipsi-lateral side. Nevertheless, the
mesial implant remains the same when loading
both medial implants, which implies that the me-
sial implant is located on the ipsi-lateral side in
only half of the tests.
To get a better view on the load distribution dur-
ing maximal biting in maximal occlusion, the
maximal axial forces and the maximal bending
Fig. 9. (a) Average (SEM) peak axial forces registered on the
moments registered on the abutments during the distal, medial and mesial abutments supporting a mandibular
test are analysed. The average values are presented fixed prosthesis occluding with a natural dentition/a fixed res-
in Fig. 8a,b. The maximal axial forces are not sig- toration (fixed) or a maxillar removable denture (removable)
nificantly different when varying the number of during maximal biting in maximal occlusion. (b) Average
supporting implants. The maximal bending mo- (SEM) peak bending moments registered on the distal, medial
and mesial abutments supporting a mandibular fixed prosthesis
ments on the other hand are significantly higher occluding with a natural dentition/a fixed restoration (fixed) or
when the prostheses are supported by only 3 im- a maxillar removable denture (removable) during maximal
plants in comparison with 4 (P∞0.003) or 5/6 biting in maximal occlusion.

471
Duyck et al.
tically significant differences between both groups, An interesting view on the data is given in Figs
there is a tendency for higher axial forces as well as 5, 6, and 7 where the effect of a 50 N axial force
bending moments on the distally positioned im- applied on a specific position is evaluated. Fig.
plants of the dentures wearers whereas the loads are 5a,b indicates what happens with the supporting
more evenly distributed in the other group. implants when the distal extension pontic of the
prosthesis at the ipsi-lateral side is loaded with 50
N. The distal implant needs to bear high compres-
Discussion sive forces and bending moments which are on av-
Fig. 3a,b clearly shows the variability of forces on erage at least the double of the applied load due
each of the supporting implants when loading the to a cantilever effect. This is in accordance with
prosthesis on different positions. It is obvious that observations from previous studies where the
the load applied on one position is distributed to highest loads were observed on the distal implants
several supporting implants through the prosthesis. and their surrounding bone when loading a distal
Interpretation of the results should be handled care- cantilever (Skalak 1983; Haraldson & Zarb 1988;
fully because the distribution of forces seems to be Glantz et al. 1993; Benzing et al. 1995; Patterson
influenced by a variety of factors such as number, et al. 1995; Tashkandi et al. 1996; Assif et al. 1996).
spreading, and inclination of the implants, canti- Only minor forces are observed on the implants at
lever length, design and rigidity of the prosthetic the contra-lateral side. Assuming a hinging effect
superstructure, bone quality, and functional jaw de- of the prosthesis, the negative axial forces acting
formation (Skalak 1983; Weinberg 1993; Benzing et on the mesial implants – which can be interpreted
al. 1995; Patterson et al. 1995; van Zyl et al. 1995; as tensile forces – are a logic observation and were
McAlarney & Stavropoulos 1996; Sertgöz & Güv- also observed in previous studies (Patterson et al.
ener 1996; Hobkirk & Havthoulas 1998). It is for ex- 1995; Assif et al. 1996). The magnitude of the
ample interesting to observe that when load is ap- bending moments decreased with increasing dis-
plied at position 6 (above the mesial implant in the tance to the point of load application. Fig. 6a,b
right quadrant), most of the load is transferred to presents the load on the supporting implants when
the neighbouring implants instead of to the implant the frontal extension of the prosthesis (in the
lying underneath the place of load application. This middle between the two mesial implants) is loaded
can be explained by the fact that due to a certain with 50 N. When the prosthesis is supported by all
misfit of the superstructure, the prosthesis will not implants, again a hinging effect is observed with
be seated identical on all supporting abutments. If the highest compressive forces on the mesial im-
there is a larger vertical misfit between the prosthesis plant, less forces on the medial implants and small
and a particular abutment in comparison with the tensile or compressive forces on the distal im-
neighbouring abutments, the prosthesis will be plants. The loading conditions are rather similar
seated better on the neighbouring abutments. after the 2 medial implants are excluded, but when
Therefore, a load applied on the prosthesis will there is only 1 mesial supporting implant, the situ-
more likely be transferred to those abutments with ation seriously alters. Besides the compressive
the best fit and to those integrated in the bone with forces on the mesial implants, also the bending
highest density. moments on all supporting implants increased sig-
Loading the extension in the left quadrant (posi- nificantly when only 3 implants were present. The
tion 1) of this particular prosthesis seems to have foregoing suggests that spreading of implants as
a more distinct influence on the neighbouring im- well as the actual number of supporting implants
plants than the same load application on the ex- is of great importance. When loading the implants
tension in the right quadrant (position 9) does. It positioned in the middle of each quadrant, loads
is also striking that load application on position 8 are well distributed along the other implants in the
(above the distal implant in the right quadrant) ipsi-lateral quadrant whereas hardly any load is
seems to induce a lower overall load on the im- transferred to the contra-lateral quadrant. Fig.
plants than load application on other places does. 7a,b shows that when the prosthesis is carried by
These observations are supposed to be the result of 5/6 implants, most of the load is carried by the
a synergism and antagonism of several individual implant underneath the position of load appli-
factors as mentioned above. cation (ipsi–medial). When implants are excluded,
Fig. 4 indicates that the mean load per implant there is a distinct increase of forces applied on the
increases when the number of supporting implants neighbouring implants. The negative axial forces
decreases. Nevertheless, the particular loads on the on the distal implants in the contra-lateral quad-
individual implants should be taken into consider- rant suggest that a hinging-effect of the prosthesis
ation to gain insight in the way forces act on im- may occur.
plants supporting an oral prosthesis. As mentioned in the results, Fig. 8a,b presents

472
Occlusal forces on oral implants supporting fixed prostheses
the average maximal axial force and the maximal ure of the implants. More research is needed to
bending moment respectively, registered on the define more precisely the difference in loading con-
abutments during maximal biting in maximal oc- ditions according to the number of supporting im-
clusion. This gives an idea of which implants take plants. In addition to that, finite element analyses
the highest loads. The figure shows that the maxi- based on individualized CT-Scans derived from
mal axial loads registered on the distal implants humans are compulsory to further understand the
are higher than those on the medial and mesial aetiology of failure of implants, especially to load-
implants are. Despite the lack of statistical signifi- ing. Such studies are underway.
cance, Fig. 9a,b shows on average higher peak ax-
ial forces and bending moments on the distal im-
plants supporting a mandibular prosthesis antag- Acknowledgements
onizing a removable denture in comparison with This study was supported by the Research Fund of Research
those antagonizing a fixed restoration. The load Council K.U.Leuven (no. OT95/27) and the Fund for Scientific
Research, Flanders (no. G.0180.26). The authors would like to
seems to be better distributed when the prosthesis thank Kris Bogaerts for the statistical analyses.
is antagonizing a natural dentition or a fixed res-
toration instead of a removable denture. This
corroborates the findings of Falk et al. (1989/1990) Résumé
and Lundgren et al. (1989) who observed increas- Comme la charge s’avère de plus en plus être un facteur déter-
ing local closing and chewing forces bilaterally in minant dans le résultat du traitement par implant dentaire, il
the distal direction on mandibular implant-sup- est nécessaire d’étendre notre connaissance de leur biomécani-
que. Le but de cette étude a été d’élargir nos connaissances en
ported cross-arch prostheses occluding with maxil- ce qui concerne la répartition et la magnitude des forces occlu-
lary complete dentures. This was said to be caused sales des implants dentaires portant des prothèses fixes. Ceci a
by the behaviour of the maxillary complete den- été effectué par quantification et qualification de ces forces in
ture during function. Probably the larger anterior vivo ce qui signifie que non seulement la magnitude de la charge
resiliency of the oral maxillary mucosa causes a mais aussi son type (force axiale ou mouvement de pliage) ont
été enregistrés. Treize patients avec une prothèse complète fixe
hinging of the denture basis, thereby transferring sur implants ont été sélectionnés. Des forces occlusales sur les
more load onto the posterior area. implants ont été quantifiées et qualifiées durant les charges
Although Tashkandi et al. (1996) indicated contrôlées de 50 N sur différentes positions le long de la surface
higher loads on the distal implants with an increas- occlusale des prothèses et durant le serrage maximum en occlu-
sion maximale en utilisant des jauges de force. Le test a été
ing cantilever length, Naert et al. (1992) could not effectué lorsque les prothèses étaient supportées par tous (cinq
correlate marginal bone loss with the cantilever ou six) les implants et a été répété lorsque les prothèses n’étaient
length. Lindquist et al. (1988, 1996) observed even supportées que par quatre ou trois implants seulement. Malgré
more bone loss around mesially positioned im- une variation interindividuelle considérable, des différences
plants in comparison with distal implants carrying claires dans la charge imposée à l’implant entre ces différentes
conditions tests ont été mises en évidence. La charge des parties
a fixed full prosthesis. However, when only exten- en extension de ces prothèses causait un effet charnière qui ap-
sion contact was present, thus leaving the anterior portait des forces de compression considérables sur les implants
area without contact, more implant failures and les plus proches de l’endroit où se trouvait l’application de la
bone loss were observed especially at the distally charge, et des forces de compression ou de tension inférieures
sur les autres implants. En moyenne, des forces plus importan-
positioned implants (Quirynen et al. 1992). There tes ont été observées avec un nombre réduit d’implants de sup-
is a tendency for an increased average load with a port. Les moments de pliage les plus importants étaient pré-
decreasing number of implants which brings the sents lorsque seulement trois implants étaient utilisés.
situation on 3 implants only at a higher risk for
(mechanical) failure, keeping the implant dimen-
sions the same. Nevertheless, its clinical relevance Zusammenfassung
towards the bone surrounding the implants needs Seit immer mehr angenommen wird, dass die Belastung einen
entscheidenden Faktor beim Behandlungserfolg mit oralen Im-
to be further investigated. Since a study of Bråne- plantaten darstellt, besteht ein Bedürfnis, das Wissen bezüglich
mark et al. (Brånemark et al. 1995) showed similar Biomechanik bei oralen Implantaten zu erweitern. Das Ziel die-
survival rates for both implants and prostheses ser Studie ist es, Einsicht in die Verteilung und die Grösse von
when fixed prostheses were supported by 6 or 4 okklusalen Kräften bei oralen Implantaten, welche festsitzende
implants at the end of a 10-year observation Rekonstruktionen tragen, zu gewinnen. Dies wird getan durch
die in vivo Quantifikation und Qualifikation dieser Kräfte. Das
period, it is possible that the difference in loading bedeutet, dass nicht nur die Grösse der Belastung, sondern
conditions between 6 or 4 supporting implants has auch die Art der Kraft (Axiale Kraft oder Biegemoment) regi-
no clinical relevance at all. A recent paper (Bråne- striert wird. Es wurden 13 Patienten mit implantatgetragenen
mark et al. 1999) even reports remarkably good festsitzenden Totalprothesen ausgeẅahlt. Okklusale Kräfte bei
den Implantaten wurden während einer kontrollierten Kraftan-
results when fixed full prostheses are mounted on wendung von 50 N und beim maximalen Zusammenbeissen in
only 3 implants. In these cases, larger diameter im- der Zentrik an mehreren Stellen entlang der Okklusalflächen
plants (5 mm) were used to avoid mechanical fail- der Rekonstruktionen mit Hilfe von Dehnmessstreifen in den

473
Duyck et al.
Prothetikzylindern quantifiziert und qualifiziert. Der Versuch
wurde durchgeführt, wenn die Rekonstruktion auf allen (5 oder
6) Implantaten verankert waren und wurde wiederholt, wenn
die Rekonstruktionen nur auf 4 oder 3 Implantaten befestigt
waren. Nebst beträchtlichen Unterschieden zwischen den ein-
zelnen Individuen konnten klare Unterschiede in der Belastung
der Implantate bei den verschiedenen Testbedingungen gesehen
werden. Die Belastung der Extensionen der Rekonstruktionen
bewirkte einen Hebeleffekt, welcher beträchtliche Krompres-
sionskräfte bei den Implantaten in unmittelbarer Nähe der Be-
lastung zur Folge hatte, während es bei den anderen Implanta-
ten zu geringerer Kompression oder zu Zugkräften kam. Im
Durchschnitt konnten mit abnehmender Anzahl von Befesti-
gungsimplantaten zunehmende Kräfte beobachtet werden. Die
Biegemomente waren am grössten, wenn nur 3 Implantate ver-
wendet wurden.

Resumen
Desde que se considera cada vez mas que la carga es un factor
determinante en los resultados terapéuticos con implantes ora-
les, existe la necesidad de expandir los conocimientos relaciona-
dos con la biomecánica de los implantes orales. La intención References
de este estudio fue aumentar la comprensión en la distribución Adell, R., Lekholm, U., Rockler, B., Brånemark, P.-I. & Jemt,
y magnitud de las fuerzas oclusales en los implantes que sopor- T. (1981) A 15-year study of osseointegrated implants in the
tan prótesis fijas. Esto se lleva a cabo por cuantificación y cuali- treatment of the edentulous jaw. International Journal of
ficación in vivo de estas fuerzas, lo que implica que no solo se Oral Surgery 10: 387–416.
registrará la magnitud de la carga sino también su tipo (fuerza Albrektsson, T. (1988) A multicenter report of osseointegrated
axial o momento de torsión). Se seleccionaron 13 pacientes con oral implants. Journal of Prosthetic Dentistry 60: 75–84.
una prótesis completa fija implantosoportada. Se cuantificaron Assif, D., Marshak, B. & Horowitz, A. (1996) Analysis of load
y cualificaron las fuerzas oclusales en los implantes de soporte transfer and stress distribution by an implant-supported
durante la aplicación controlada de una carga de 50 N en dife- fixed partial denture. Journal of Prosthetic Dentistry 75:
rentes posiciones a lo largo de la superficie oclusal de la prótesis 285–291.
y durante la mordida máxima en máxima oclusión por medio Barbier, L. & Schepers, E. (1997) Adaptive bone remodelling
del uso de pilares calibrados para tensión. El test se Ilevó acabo around oral implants under axial and nonaxial loading con-
cuando la prótesis estaba soportada por totos los implantes (5 ditions in the dog mandible. International Journal of Oral
ó 6) y se repitió cuando la prótesis estaba soportada por 4 y and Maxillofacial Implants 12: 215–223.
por 3 implantes solamente. A pesar de la variación considerable Becker, W., Becker, B.E., Newman, M.G. & Nyman, S. (1990)
interindividual, se observaron clares diferencias en la carga del Clinical and microbiologic findings that may contribute to
implante entre estas condiciones de prueba. La carga de las dental implant failure. International Journal of Oral and
partes en extensión de la prótesis produjo un efecto de bisagra Maxillofacial Implants 5: 31–38.
que indujo fuerzas de compresión considerables en los implan- Benzing, U.R., Gall, H. & Weber, H. (1995) Biomechanical as-
tes más cercanos al lugar de aplicación de la carga y menores pects of two different implant-prosthetic concepts for eden-
fuerzas compresivas o tensiones en otros implantes. De media tulous maxillae. International Journal of Oral and Maxillo-
se observaron mayores fuerzas con un número decreciente de facial Implants 10: 188–198.
implantes de soporte. Los momentos de torsión fueron más Brånemark, P.-I., Svensson, B. & van Steenberghe, D. (1995)
altos cuándo se usaron sólo tres implantes. Ten-year survival rates of fixed prostheses on four or six
implants ad modum Brånemark in full edentulism. Clinical
Oral Implants Research 6: 227–231.
Brånemark, P.-I., Engstrand, P., Öhrnell, L.-O., Gröndahl, K.,
Nilsson, P., Hagberg, K., Darle, C. & Lekholm, U. (1999)
Brånemark NovumA: a new treatment concept for rehabili-
tation of the edentulous mandible. Preliminary results from
a prospective clinical follow-up study. Clinical Implant Den-
tistry and Related Research 1: 2–16.
Duyck, J., Van Oosterwyck, H., Vander Sloten, J., Lievens, S.,
De Cooman, M., Puers, R. & Naert, I. (2000) Three-dimen-
sional force measurements on oral implants: a methodolog-
ical study. Journal of Oral Rehabilitation (in press).
Esposito, M., Hirsch, J.-M., Lekholm, U. & Thomson, P. (1998)
Biological factors contributing to failures of osseointegrated
oral implants. (II) Etiopathogenesis. European Journal of
Oral Sciences 106: 721–764.
Falk, H., Laurell, L. & Lundgren, D. (1989) Occlusal force pat-
tern in dentitions with mandibular implant-supported fixed
cantilever prostheses occluded with complete dentures. In-
ternational Journal of Oral and Maxillofacial Implants 4: 55–
62.
Falk, H., Laurell, L. & Lundgren, D. (1990) Occlusal inter-

474
Occlusal forces on oral implants supporting fixed prostheses
ference and cantilever joint stress in implant-supported pros- International Journal of Oral and Maxillofacial Implants 4:
theses occluding with complete dentures. International 277–283.
Journal of Oral and Maxillofacial Implants 5: 70–77. Lundqvist, S. & Haraldson, T. (1992) Oral function in patients
Fløystrand, F., Kleven, E. & Øilo, G. (1982) A novel miniature wearing fixed prosthesis on osseointegrated implants in the
bite force recorder and its clinical application. Acta Odontol- maxilla: 3-year follow-up study. Scandinavian Journal of
ogica Scandinavica 40: 209–214. Dental Research 100: 279–283.
Glantz, P.-O., Rangert, B., Svensson, A., Stafford, G.D., Arnvi- McAlarney, M.E. & Stavropoulos, D.N. (1996) Determination
darson, B., Randow, K., Lindén, U. & Hultén, J. (1993) On of cantilever length–anterior–posterior spread ratio as-
clinical loading of osseointegrated implants. Clinical Oral suming failure criteria to be the compromise of the pros-
Implants Research 4: 99–106. thesis retaining screw-prosthesis joint. International Journal
Haraldson, T. & Zarb, G. (1988) A 10-year follow-up study of of Oral and Maxillofacial Implants 11: 331–339.
the masticatory system after treatment with osseointegrated Mericske-Stern, R., Assal, P. & Buergin, W. (1996) Simul-
implant bridges. Scandinavian Journal of Dental Research taneous force measurements in 3 dimensions on oral endos-
96: 243–252. seous implants in vitro and in vivo. Clinical Oral Implants
Hobkirk, J.A. & Havthoulas, T.K. (1998) The influence of man- Research 7: 378–386.
dibular deformation, implant numbers, and loading posi- Naert, I., Quirynen, M., van Steenberghe, D. & Darius, P.
tion on detected forces in abutments supporting fixed im- (1992) A study of 589 consecutive implants supporting com-
plant superstructures. Journal of Prosthetic Dentistry 80: plete fixed prostheses. Part II: Prosthetic aspects. Journal of
169–174. Prosthetic Dentistry 68: 949–956.
Hoshaw, S.J., Brunski, J.B. & Cochran, G.V.B. (1994) Mechan- Patterson, E.A., Burguete, R.L., Hue Thoi, M. & Johns, R.B.
ical loading of Brånemark implants affects interfacial bone (1995) Distribution of load in an oral prosthesis system: an
modelling and remodelling. International Journal of Oral in vitro study. International Journal of Oral and Maxillofacial
and Maxillofacial Implants 9: 345–360. Implants 10: 552–560.
Isidor, F. (1996) Loss of osseointegration caused by occlusal Quirynen, M., Naert, I. & van Steenberghe, D. (1992) Fixture
load of oral implants. A clinical and radiographic study in design and overload influence marginal bone loss and fixture
monkeys. Clinical Oral Implants Research 7: 143–152. success in the BrånemarkA system. Clinical Oral Implants
Isidor, F. (1997) Histological evaluation of peri-implant bone Research 3: 104–111.
at implants subjected to occlusal overload or plaque ac- Sertgöz, A. & Güvener, S. (1996) Finite element analysis of the
cumulation. Clinical Oral Implants Research 8: 1–9. effect of cantilever and implant length on stress distribution
Lindhe, J., Berglundh, T., Ericsson, B., Liljenberg, B. & Mar- in an implant-supported fixed prosthesis. Journal of Pros-
inello, C. (1992) Experimental breakdown of peri-implant thetic Dentistry 76: 165–169.
and periodontal tissues. A study in the dog. Clinical Oral Skalak, R. (1983) Biomechanical considerations in osseointe-
Implants Research 3: 9–16. grated prostheses. Journal of Prosthetic Dentistry 49: 843–
Lindquist, L.W., Carlsson, G.E. & Glantz, P.-O. (1987) Re- 848.
habilitation of the edentulous mandible with a tissue-inte- Smedberg, J.-I., Nilner, K., Rangert, B., Svensson, S.A. &
grated fixed prosthesis: a six-year longitudinal study. Quin- Glantz, P.-O. (1996) On the influence of superstructure con-
tessence International 18: 89–96. nection on implant preload: a methodological study. Clinical
Lindquist, L.W., Rockler, B. & Carlsson G.E. (1988) Bone re- Oral Implants Research 7: 55–63.
sorption around fixtures in edentulous patients treated with Tashkandi, E.A., Lang, B.R. & Edge, M.J. (1996) Analysis of
mandibular fixed tissue-integrated prostheses. Journal of strain at selected bone sites of a cantilevered implant-sup-
Prosthetic Dentistry 59: 59–63. ported prosthesis. Journal of Prosthetic Dentistry 76: 158–
Lindquist, L.W., Carlsson, G.E. & Jemt, T. (1996) A prospective 164.
15-year follow-up study of mandibular fixed prostheses sup- van Zyl, P.P., Grundling, N.L., Jooste, C.H. & Terblanche, E.
ported by osseointegrated implants. Clinical results and (1995) Three-dimensional finite element model of a human
marginal bone loss. Clinical Oral Implants Research 7: 329– mandible incorporating six osseointegrated implants for
336. stress analysis of mandibular cantilever prostheses. Interna-
Lundgren, D., Falk, H. & Laurell, L. (1989) The influence of tional Journal of Oral and Maxillofacial Implants 10: 51–57.
number and distribution of occlusal cantilever contacts on Weinberg, L.A. (1993) The biomechanics of force distribution
closing and chewing forces in dentitions with implant-sup- in implant-supported prostheses. International Journal of
ported fixed prostheses occluding with complete dentures. Oral and Maxillofacial Implants 8: 19–31.

475

You might also like