Chapter 7.

Working in TOP View (Optional Feature)

7.1.2 Setting Up TOP View for TIMED, or TIMED CONTINUOUS Operation (cont.)
5. Press to activate and set the NEW DATA Function in the IMAGE Submenu (located in the DISPLAY Menu).
The selection sets the screen location (LEFT, RIGHT, TOP, or BOTTOM) at which newly acquired TOP VIEW
data is added to the display screen. This function controls the display screen scrolling direction.
6. Press to activate and set the AMP COLOR PALETTE in the IMAGE Submenu (located in the DISPLAY
Menu). Select one of the four standard palettes to represent with various color scales the amplitude of echoes
captured in A- or B-Gate. Alternatively, select CUSTOM to choose the user inputted, remotely generated,
customized color palette.
7. Press to activate and set the TOF COLOR PALETTE in the IMAGE Submenu (located in the DISPLAY Menu).
Select one of the four standard palettes to represent with various color scales the time-of-flight measurement to the
A- or B-Gate triggering echo. Alternatively, select CUSTOM to choose the user inputted, remotely generated,
customized color palette.
8. Press to activate and set the DATA SOURCE Function in the VIEW Submenu (located in the DISPLAY Menu).
Select from four types of gated data to represent in the TOP View including:
• AMP A – Amplitude of echoes captured in the A-Gate

• AMP B – Amplitude of echoes captured in the B-Gate

• TOF A – Time of flight duration to A-Gate triggering echoes

• TOF B – Time of flight duration to B-Gate triggering echoes

9. Press to activate TOP View by setting the VIEW MODE Function in the VIEW Submenu (located in the
DISPLAY Menu) to TOP. The instrument will immediately begin acquiring data and filling the TOP View display.

Figure 65: Make these settings for TIMED TOP View operation. Most TOP View specific settings are accessed
from the CLOCK, VIEW, and IMAGE Submenus.

148 Phasor XS User’s Manual