You are on page 1of 4

CONFERENCE REPORT

Analytical Methods Validation: Bioavailability, Bioequivalence,


and Pharmacokinetic Studies

VINODP. SHAH', KAMALK. MIDHA*', SHRIKANT DIGHE', IAIN J. M C G I L V E R AJEROME


~, P. SKELLY*,
AVRAHAMYACOBI~,THOMAS LAYLOFF~, C. T. VISWANATHAN',C. EDGARCOOK1',R. D. MCDOWALL**,
KENNETHA. pllTMAN**, AND SIDNEY SPECTOR@
Received November 12, 1991, from the 'Center for Drug Evaluation and Research, Food and Drug Administration, Rockville, MD 20857, the
*Universi of Saskatchewan, Saskatoon, Saskatchewan, Canada S7N 0WO, the "Health Protection Branch, Bureau of Drug Research, Ottawa,
Ontario, zanada KIA OL2, the TAmericanCyanamid Company, Lederle Laboratories, Pearl River, NY 10965, the aDivision of Drug Analysis,
Center for Drug Evaluation and Research, Food and Drug Administration, St. Louis, MO 63101, the 11 Research Trian le Institute, Research
Triangle Park, NC 27709, the ** Wellcome Research Laboratones, Beckenham, Kent, BR3 3BS, England, the **Bris!ofM ers Squibb Company,
Pharmaceutical Research Institute, P.O. Box 4755, Syracuse, NY 13221-4755, and the 55 Vanderbilt University Medical Jenter, Nashville, TN
37232. Accepted for publication November 26, 1991.

Abstract 0 This is a summary report of the conference on "Analytical necessary to validate the analytical methodb) at each site and
Methods Validation: Bioavailability, Bioequivalence and Pharmacoki- provide appropriate validation information for different sites
netic Studies." The conference was held from December 3 to 5,1990,in to establish interlaboratory reliability. Unless a method is
the Washington, D.C., area and was sponsored by the American used on a regular basis, providing confidence in its continued
Association of Pharmaceutical Scientists, the US. Food and Drug validity, it is essential to document that the method is still
Administration, Federation International Pharmaceutique, Health Pro- valid before analysis of samples in the study. Adequate
tection Branch (Canada), and the Association of Official Analytical validation for methods not used on a regular basis often
Chemists. The report presents our assessment of the major agreements consists of running a standard curve with new quality-control
and issues discussed at the conference. The report is also intended to samples to show that the responses, relationship, and general
provide guiding principles for validation of analytical methods used in characteristics of the method are similar to previous valida-
bioavailability, bioequivalence, and pharmacokinetics studies in humans tion results.
and animals. The objectives of the conference were as follows: (1) to
reach a consensus on what should be required in analytical methods Validation of Analytical Methods
validation and the procedures to establish validation; (2) to determine
processes of application of the validation procedures in bioavailability, Method validation includes all of the procedures required to
bioequivalence, and pharmacokinetics studies; and (3) to develop a demonstrate that a particular method for the quantitative
report on analytical methods validation that may be referred to in determination of the concentration of an analyte (or series of
developingfuture formal guidelines. Acceptable standardsfor document- analytes) in a particular biological matrix is reliable for the
ing and validating analytical methods with regard to processes, param-
eters, or data treatments are discussed because of their importance in intended application. Some of the more commonly used
assessing pharmacokinetic, bioavailability, and bioequivalence studies. bioanalytical techniques include (1) chemical methods [e.g.,
Other topics that were considered essential in the conduct of pharma- chromatography (GC and HPLC) and various procedures with
cokinetic studies or in establishing bioequivalency criteria, including MS (direct MS, MSMS, and combination techniques such as
measurement of drug metabolites and stereoselective determinations, GC-MS and LC-MS)] and (2) biological methods [e.g., those
are also discussed. based on immunoassay procedures (RIA, enzyme-multiplied
- ___ -
. ~. ~ _ _
immunoassay technique [EMIT], enzyme-linked immunosor-
bent assay [ELISA]), and microbiological methods)]. Many of
the principles, procedures, and requirements are common to
Analytical methods that are used for the quantitative all types of analytical methods.
determination of drugs and their metabolites in biological The parameters essential to ensure the acceptability of the
samples play a significant role in evaluation and interpreta- performance of an analytical method are stability of the drug
tion of bioavailability, bioequivalence, and pharmacokinetic in the matrix under the storage conditions of the study,
data. It is essential to use well-characterized and fully accuracy, precision, sensitivity, specificity (selectivity), re-
validated analytical methods to yield reliable results that can sponse function, and reproducibility. Although there are
be satisfactorily interpreted. Analytical methods and tech- various stages in the development and validation of an
niques are constantly being changed and improved; in many analytical procedure, the validation of the analytical method
instances, these methods are at the cutting edge of the can be envisaged to consist of two distinct phases: (1) the
technology. It is also important to emphasize that each development phase, in which the assay is defined, and (2) the
analytical technique has its own characteristics, which will application phase, in which the method is applied to actual
vary from drug to drug. Moreover, the appropriateness of the analysis of samples from pharmacokinetic, bioavailability,
technique may be influenced by the ultimate objective of the and bioequivalence studies.
study. Specific validation criteria are needed for methods Development of Analytical Methods: Chemical Assays-
intended for analysis of each analyte (drug and/or metabo- The following principles of validation of analytical methods
lite). Although validation of each method will be independent provide steps for developing a new method or establishing an
of those of other methods, there may be situations in which existing method in a particular laboratory for the first time.
comparison of the methods will be necessary (e.g., when more Any modification of an analytical method would require
than one method has been used in a long-term study). When revalidation of the procedures. Validation of analytical meth-
sample analysis is conducted at more than one site, it is ods should be performed to support pharmacokinetic,

This article not subject to U.S. copyright. Journal of Pharmaceutical Sciences I 309
Published 1992, American Pharmaceutical Association Vol. 81, No. 3, March 1992
bioequivalence, and related studies in a new drug application mining the coefficient of variation andlor appropriate confi-
(NDA) or an abbreviated new drug application (ANDA).Full dence interval. The LOQ should serve as the lowest concen-
validation of methods may not be necessary in conducting tration on the standard curve and should not be confused with
exploratory pharmacokinetic studies. It is suggested that the limit of detection (LOD; see glossary).
validation include investigation of samples from dosed sub- Specific Recommendations for Method Validation-The
jects. following recommendations should be considered when vali-
Principles for Establishing a Valid Method-The follow- dating a method.
ing principles should be used to establish a valid method. (1)The stability of the analyte in the biological matrix a t
(1) A specific, detailed description and protocol of the the intended storage temperature(s) should be established. In
method should be written (standard operating procedure). addition, the influence of freezethaw cycles (a minimum of
(2) Each step in the method should be investigated to two cycles a t two concentrations in duplicate) should be
determine the extent to which environmental, matrix, mate- studied.
rial, or procedural variables, from the time of collection of the (2) The specificity of the assay method should be established
material up to the time of analysis and including the time of with six independent sources of the same matrix.
analysis, may affect the estimation of analyte in the matrix. (3)The accuracy and precision should be determined with
Variability of the matrix due to its physiological state may a minimum of five (excluding blank sample) determinations
need to be considered. per concentration. The mean value should be within 2 15%of
(3)A method should be validated for the intended use with the actual value, except a t LOQ, where it should not deviate
an acceptable protocol. All experiments used to make claims by more than t20%. The precision around the mean value
or draw conclusions about the validity of the method should should not exceed 15%coefficient of variation (CV),except for
be presented in a report (method validation report). LOQ, where it should not exceed 20% CV. Other methods of
(4) Whenever possible, the same biological matrix as that in determining accuracy and precision that meet these limits
the intended samples should be used for validation purposes. may be equally acceptable.
(For tissues of limited availability, such as bone marrow, (4) The standard curve should consist of five to eight
physiologically appropriate proxy matrices may suffice.) The standard points, excluding blank, with single or replicate
stability of the analyte (drug and/or metabolite) in the matrix samples. The standard curve should cover the entire range of
during the collection process and the sample storage period expected concentrations.
should be assessed, preferably before sample analysis. It is (5) The simplest relationship for response versus concen-
recommended that the stability of the analyte in the matrix tration (response function) should be determined, and the fit
from dosed subjects be confirmed. Accuracy, precision, repro- should be statistically tested. The function should be repre-
ducibility, response function, and the specificity of the method sented with an appropriate algorithm or graphical technique.
with respect to endogenous substances, metabolite(s), and Application to Routine Drug Analysis-Many of the
known degradation products should be established with principles for establishing a valid method are relevant to
reference to the biological matrix. In regard to specificity, prestudy validation. This section will emphasize the valida-
there should be evidence that the substance being quantitated tion parameters that should be performed during routine
is the intended analyte. application of a method to a particular study.
(5)The concentration range over which the analyte will be In general, with acceptable variability as defined by vali-
determined must be defined in the method, on the basis of dation data, analysis of biological samples can be done by
evaluation of actual standard samples over the range (includ- single determination without a need for duplicate or replicate
ing their statistical variation). This procedure defines the analysis. The need for duplicate analysis should be assessed
standard curve. on a case-by-case basis. For example, for a robust procedure
(6)It is necessary to use a sufficient number of standards to of low variability, with accuracy and precision routinely well
define adequately the relationship between concentration and within tolerances, single analysis would suffice. For a difficult
response. The relationship between response and concentra- procedure with a labile analyte, when the precision and
tion must be demonstrated to be continuous and reproducible. accuracy tolerances are difficult to achieve, duplicates may be
The number of standards to be used will be a function of the essential. A procedure should be developed that documents
dynamic range and the nature of the concentration-response the reasons for reanalysis.
relationship. In many cases, five to eight concentrations A standard curve should be generated for each analytical run
(excluding blank values) may define the standard curve. More for each analyte and should be used to calculate the concentra-
standard concentrations may be necessary for nonlinear tion of analyte in the unknown samples assayed with that run.
relationships than would be for linear relationships. It is important to use a standard curve that will cover the entire
(7)The accuracy and precision with which known concen- range of concentrations in the unknown samples. Estimation of
trations of analyte in the biological matrix can be determined unknowns by extrapolations of standard curves below the low
must be demonstrated. Within- and between-run accuracy standard or above the high standard is not recommended.
and precision should be calculated with commonly accepted Instead, it is suggested that the standard curve be redetermined
statistical procedures. Determination of accuracy and preci- or samples be reassayed a h r dilution. The quality-control(QC)
sion can be accomplished by analysis of replicate sets of samples should be used to accept or reject the run. These QC
analyte samples of known concentrations from a n equivalent samples are matrix spiked with analyte.
biological matrix. At least three concentrations representing In summary, (1) a standard curve should consist of five to
the entire range of the calibration curve should be studied: eight standard points, excluding blank (either single or
one near the lower limit of quantitation (LOQ), one near the replicate), covering the entire range; (2) the response function
center, and one near the upper boundary of the standard is determined by appropriate statistical tests based on the
curve. For a method to be considered valid, specific criteria actual standard points during each run in the validation; and
must be set for accuracy and precision over the range of the (3) the system suitability is based on the analyte and tech-
standard curve. nique (a specific procedure or sample can be identified to
( 8 ) The LOQ is the lowest concentration on the standard assure the optimum operation of the system employed).
curve that can be measured with acceptable accuracy, preci- Acceptance Criteria for the Run-Accuracy and Preci-
sion, and variability. The LOQ should be determined by using sion-The acceptance criteria are not more than 15%CV for
at least five samples independent of standards and by deter- precision and not more than 15%deviation from the nominal

310 I Journal of Pharmaceutical Sciences


Vol. 81, No. 3, March 7992
value for accuracy. However, a t the LOQ, 20% is acceptable replicate samples. When replicate samples need to be mea-
for both precision and accuracy. It is desirable that these sured during validation to improve accuracy, the same pro-
tolerances be provided both for intraday and interday or cedure must be followed for unknown samples.
interrun experiments. (6) If there are intermediate steps between the plasma (or
QC Samples-QC samples in duplicate at three concentra- other biological matrices) and the final assay (such as extrac-
tions (one near the LOQ, one in midrange, and one approach- tion of biological sample followed by immunoassay) and if
ing the high end of the range) should be incorporated into each parallel processed standards in the biological matrix are not
run. The results of the QC samples provide the basis of being used, it is necessary to establish recovery and use it in
accepting or rejecting the run. determining results. Possible approaches to assess efficiency
At least four of the six QC samples must be within 20% of and reproducibility of recovery are use of radiolabeled tracer
their respective nominal values; two of the six QC samples analyte (quantity too small to affect the assay), advance
(not both a t the same concentration) may be outside the 220% establishment of reproducible recovery, and use of an internal
respective nominal value. A confidence-interval approach standard that is not recognized by the antibody but can be
yielding comparable accuracy and precision is an acceptable measured by another technique.
alternative. (7) Correction for nonspecific matrix effects can be accom-
Repeat Analysis-The protocol for repeat analysis should be plished with separation techniques that remove the effect or
established a priori. Some aberrant values can be identified, the matrix. These techniques may be used in defining the
which can be attributed to processing errors, equipment standard curve for both controls and samples. The use of
failure, poor chromatography, or QC samples outside pre- standards in the matrix is recommended. This approach will
defined tolerance. Cautious use of “pharmacokinetic fit”, such obviate many of the earlier mentioned concerns.
as a double peak, may call for repeat analysis of some samples Other Issues-Commerciul Kits-Commercial kits are
in the study, but the reasoning should be clearly documented. available for both immuno- and microbiological assays, and
the analytical methods based on such kits should be validated.
Validation of Analytical Methods: Immuno- and The validation assures that the bioassay kit is applicable to
Microbiological Assays the study problem and that subsequent batches or lots of kits
have performance characteristics similar to the original
Many of the parameters for and principles of analytical validated kit or test. Any modifications and extensions of
validation for chemical methods are also applicable to immu- assays from one kit (or test) to another must be validated.
no- and microbiological methods, but there are some specific Measurement of Metubolite(s)-The complex area of deter-
differences. In immuno- and microbiological assays, the re- mination of drug metabolites in bioavailability studies to
sponse must be shown to be related to the concentration of the support drug submissions was discussed in the conference.
analyte in question. The questions differed somewhat according to the objective of
Selectivity Issue-As with chromatographic methods, it the application of the bioanalytical measurement (e.g.,
must be demonstrated that the bioassay is selective for the bioequivalence versus pharmacokinetic profiling). Some sit-
analyte. An alternative method, if rigorously established, uations exist in studies of bioavailability and bioequivalence
may be used to compare the results of the bioassay. For in which (1) the parent drug cannot be measured in biological
bioassay, an appropriate combination of other techniques samples and only the metabolite can be measured, (2) the
may be used to show selectivity, including the following: (1) parent drug and active and/or inactive major metabolite(s1
comparison of standards in biological fluids with standards in can be measured, (3)more than one metabolite is present, or
buffer to detect matrix effects; (2) parallelism of diluted (4) the accumulation of metabolite is augmented (e.g., in the
clinical samples with diluted standards to detect presence of case of renal impairment). Under such situations, should one
closely related compounds; (3) serial separation techniques measure the metabolite(s)? Can decision criteria be developed
(e.g., extraction) and chromatography, with the bioassay as for measuring the metabolite in such situations? From the
detector, to demonstrate that the response is due only to the discussions, the following suggestions were made:
analyte in question; and (4) metabolite (or endogenous com- (1) All methods applied for measuring drug and metabo-
pound) cross-reaction assessed initially by comparison of lite(s) should be validated for that particular study matrix,
displacement curves (in critical cases, metabolite cross- with the same general parameters mentioned earlier (accu-
reaction should also be assessed by addition of metabolite to racy, precision, specificity, recovery, and reproducibility).
analyte). Similar criteria will be applicable when the drug is (2) Pharmacokinetic, bioavailability, and bioequivalence
concomitantly administered with other drugs. studies should be based on the moieties that contribute
Quantitation Issues-These issues are summarized as significantly to the pharmacologic or therapeutic effect.
follows: Assay of Stereoisomers-The need for stereoselective
(1) Criteria for precision and accuracy of immuno- and determination in bioavailability and bioequivalence studies
microbiological assays should be based on the requirements of was another issue that was discussed. There are many drugs
the study and should match those of chromatographic meth- that are administered as racemic mixtures, and they may
ods. Any decision to run the sample analysis in single, undergo stereoselective metabolism and/or elimination. One
duplicate, or triplicate should be based on variability. isomer may be more active than the other. Under what
(2)Immunoassay standard curves are essentially nonlinear circumstances should one measure individual drug isomers
and, in general, require more concentration points to define and/or metabolite(s) isomers from the biological matrix? The
the fit over the range claimed. following suggestions were made: (1) all methods used for
(3)It should be established that an acceptable curve-fitting measurement of stereoisomer should be validated (with em-
model is being used by examining statistics for goodness of fit phasis on stereospecificity); (2) for bioequivalence studies of
and by back-calculating results of standards and control an existing racemic product, a stereospecific assay is not
samples. required if the rate and extent profiles are superimposable
(4) Both upper and lower LOQ values must be defined by (within usual statistical boundaries); and (3)for new chemical
acceptable accuracy, precision, or confidence-interval criteria entities, the pharmacokinetic profiles for the stereoisomer
based on the study requirements. should be characterized in normal subjects.
(5) For all assays, the key factor is the accuracy of the Pharmacodynamic Measurements-The final difficult issue
reported results. This accuracy may be improved by use of identified was the area of pharmacodynamic measurements.

Journal of Pharmaceutical Sciences I 311


Vol. 81, No. 3, March 7992
The suggestions w e r e as follows. (1) all pharmacodynamic Acknowledgments
procedures used for definitive bioequivalence o r related stud- To reach wide circulation and to have information available to
ies m u s t be fully validated under controlled conditions a n d scientists by prior arrangement, this document was submitted to the
should include a placebo a n d (2) the pharmacodynamic effect followingjournals for publication: Pharmaceutical Research; Journal
measured for bioequivalence studies should b e related to the of Pharmaceutical Sciences; International Journal of Pharmacy;
actual pharmacologic (therapeutic) end point of the activity of International Journal of P h a r m e u t i c s ; European Journal of Drug
the drug. Metabolism and Pharmacokinetics; and Journal of Association of
Ofticia1Analytical Chemists. The following individuals contributed to
this conference report: Kenneth S. Albert (Forest Laboratories, New
Glossary York, NY); Sanford Bolton (St. Johns University, Queens Campus,
Jamaica, NY); C. Ed ar Cook (Research Triangle Institute, Research
Accuracy: closeness of determined value to the t r u e value. Triangle Park, NC). hhrikant Dighe (Center for Dru Evaluation and
Generally, recovery of added analyte over an appropriate Research, Food and Drug Administration, Rockvilg, MD); Michael
range of concentrations is taken as an indication of accuracy. Dobrinska (Merck, Sha and Dohme, West Point, PA); William Doub
Whenever possible, the concentration range chosen should (Division of Dru Analfrsis, Center for Drug Evaluation and Re-
search, Food a n d Drug Administration, St. Louis, MO); Michael
bracket the concentration of interest. Eichelbaum (Dr. Mar arete Fischer-Bosch Institute, Auer-
Analyte: a specific, unique chemical moiety i n the f o r d s ) bachstrasae, Stuttgart, &*an ); John W. A. Findlay (Burroughs
it would be found i n a biologic matrix. Wellcome Company Research &angle Park, NC); Keith Gallicano
(Health Protection branch, Bureau of Drug Research, Ottawa, On-
Biological matrix: a unique material of biological origin tario, Canada); William Garland (Hoffmann-La Roche, Inc., Nutley,
that c a n be prepared i n a reproducible manner (e.g., blood, NJ); Dwight J. Hardy (Universit of Rochester Medical Center,
serum, plasma, urine, feces, saliva, sputum, a n d various Rochester, NY); James D. Hulse ( d r r i s Laboratories, Inc., Lincoln,
discrete tissues). NE); H. Thomas Karnes (Medical Colle e of Virginia, Richmond,
VA); R. D. McDowall (Wellcome Researchfaboratones, Beckenham,
Limit of detection (LOD): the lowest concentration of an Kent, U.K.); Ron Lange (Glaxo Inc., Research Triangle Park, NC);
analyte that the analytical process c a n reliably differentiate Thomas Layloff (Division of Drug Analysis, Center for Drug Evalu-
from background levels. ation and Research, Food and Drug Administration, St. Louis, MO);
Limit of quantitation (LOQ): the lowest concentration of an William D. Mason (University of Missouri, Kansas Cit ,MO); Gordon
McKa (Universit of Saskatchewan, Saskatoon, gaskatchewan,
analyte that c a n be measured w i t h a stated level of confi- Canat&); Iain J. d G i l v e r a y (Health Protection Branch, Bureau of
dence. Drug Research, Ottawa, Ontario, Canada); Kamal K. Midha (Uni-
Linear range: generally t a k e n as the range over which the versit of Saskatchewan, Saskatoon, Saskatchewan, Canada); Eric
procedure has been demonstrated to give a linear detector Ormsg (Health Protection Branch, Bureau of Drug Research, Ot-
response. A reproducible, nonlinear response curve, however, tawa, Bntario, Canada); James Overpeck (Center for Drug Evalua-
tion and Research, Food and Drug Administration, Rockville, MD);
c a n also be acceptable (nonlinearity is certainly the case with Kenneth A. Pittman (Bristol-Myers Squibb Company, Pharmaceu-
immunological procedures). tical Research Institute, Syracuse, NY); H. D. Plattenberg (L. A. B.
Method: a set of a l l of the procedures involved in the GmbH and Company, We enerster, Nen-Ulmr, Germany); Vinod P.
collection, processing, storage, and analysis of a biological Shah (Center for Drug #valuation and Research, Food and Dru
Administration, Rockville, MD); Gerald Shiu (Center for Drug Eva?
matrix for an analyte. uation and Research, Food and Drug Administration, Rockville,.MD);
Precision: the closeness of replicate determinations of an Daniel Sitar (Department of Pharmacolo University of Manitoba,
analyte by an assay. Precision c a n be further subdivided into Winnepeg, Manitoba, Canada); Jerome P’Skelly (Center for 9 r u g
within-day o r intra-assay precision a n d between-day or inter- Evaluation and Research, Food and Drug Administration, Rockville
MD); Fritz Sorgel (Institute of Biomedical and Pharmaceuticai
assay precision. Research, Nuremberg, German ); Sidney Spector (Vanderbilt Uni-
Specificity: the ability of method to measure only w h a t i t is versit Medical Center, Nashvise, TN); James T. Stewart (Univer-
intended t o measure. sity o r Georgia, Athens, GA); C. T. Viswanathan (Center for Drug
Standard curve: the relationship between the experimental Evaluation and Research, Food and Drug Administration, Rockville,
MD); Avraham Yacobi (American C anamid Company, Lederle
response value a n d the analytical concentration; also referred Laboratories, Pearl River, NY); and L. e u h (Warner LambeMarke-
to as a calibration curve. Davis Company, Ann Arbor, MI).

312 I Journal of Pharmaceutical Sciences


Vol. 81, No. 3, March 1992

You might also like