Use a mild detergent to clean the instruments. Using a ETO Concentration: 600 ± 25mg/L
softbristle brush, gently clean the devices while immersed Gas Exposure Time: 4 hours
in the cleaning solution. Thoroughly brush all passages of
the instruments. 12 hours aeration time @ 55°C
Stryker Endoscopy
5900 Optical Court 6. Pay extra attention to the exposed underside of the shafts 100% ETO
San Jose, CA 95138 USA and tips when cleaning with the brush. Wrapped in an instrument tray or fully perforated sterilization
1-800-624-4422
U.S. Patents: www.stryker.com/patents 7. Flush instruments thoroughly under running water. box.
8. Dry instruments with a lint free towel. Preconditioning Parameters
Temperature: 43° ± 2°C
Steam Sterilization Relative Humidity: 65% + 10/-20%
Precondition Time: 1 hr
Conquest Series Manual Instruments Note: Instruments to be thoroughly cleaned prior to Sterilization Parameters
Maintenance Manual sterilization.
Temperature: 55° ± 2°C
“Flash” (unwrapped method) @270°F - 272°F Relative Humidity: 70% ± 5%
Manual instruments are intended for arthroscopic surgical use (132°C - 134°C) ETO Concentration: 600 ± 25mg/L
to cut, resect and grab meniscus and tissue. Minimum Exposure Time: 30 minutes Gas Exposure Time: 4 hours
Gravity (wrapped method) @250°F - 254°F
Warning 12 hours aeration time @ 55°C
(121°C - 123°C)
1. Federal Law (USA) restricts these devices to use by, or on Minimum Exposure time: 100 minutes
the order of, a physician. Minimum Dry Time: 8 minutes Inspection
2. Avoid contacting instruments with metal objects, causing Gravity (wrapped method) @270°F - 272°F • Inspect instruments on a continual basis. If a problem is
the instruments to break and leave pieces in the wound, (132°C - 134°C) observed or suspected, it should be immediately set aside
which may be difficult to remove. Minimum Exposure Time: 60 minutes for evaluation and/or repairs.
3. Bending or prying with the instruments may cause the Minimum Dry Time: 8 minutes • Inspect all components for cleanliness. Disassembly for
instrument to break and leave pieces in the wound, which cleaning is recommended.
may be difficult to remove. Pre Vacuum (wrapped) @270°F - 272°F
(132°C - 134°C)
Warranty
Minimum Exposure Time: 4 minutes
Cleaning Minimum Dry Time: 8 minutes
Stryker Endoscopy warrants against defects in both materials
1. The Stryker Conquest Series Manual Instruments are and workmanship to the registered owner at the time of
supplied non-sterile and must be cleaned and sterilized Ethylene Oxide (ETO) Sterilization purchase. All components are covered by the warranty for a
prior to initial uses. period of one year from the date of purchase. This warranty
ETO/Oxyfume 2002 extends to all purchasers and is limited to the repair or
2. Lay units in a clean pan. Prepare an enzymatic detergent
Wrapped in an instrument tray or fully perforated sterilization replacement of the product without charge when returned to:
solution as recommended by the manufacture. Add a
box. Stryker Endoscopy, 5900 Optical Court, San Jose, CA 95138,
sufficient amount of the prepared solution to the pan to
USA.
completely cover the instruments. Preconditioning Parameters
This warranty does not extend to any unit which has been
3. Let the instruments soak for the recommended amount of Temperature: 55° ± 2°C subject to misuse, abuse, neglect, improper installation or
time. This will reduce blood, protein and mucus from the Relative Humidity: 70% ± 5% operation or that has been altered, adjusted or tampered with
instruments. Precondition Time: 1 hr by any person other than Stryker Endoscopy authorized service
4. After soaking, remove the instruments and rinse them personnel.
thoroughly under running water. Sterilization Parameters
Ethylene Oxide Carrier: Oxyfume 2002
Temperature: 55° ± 2°C
Relative Humidity: 70% ± 5%
Pressure (PSIG Start): 25.4 ± 1 PSIG