Professional Documents
Culture Documents
^ _^ll. ll I
rrftrilffilfdl hll / i i fllhnm^n
i,
t
't
,D
G
Io
q
-
llll ldli|Jllttl]uI
i t,tiiuvu'
1]
Llld
|j /riiilfe[ ]n
"
-ruuvt1q|![Elf1/
t
:
U
-:
Prepared by
:
A. lecturer Nidhal Yousif Nasser
Seconcl year
3
University of Technology
Eleclromechanical. Eng. Dep.
Measuremcnt Lab.
2'"1 year
1.1. Objective
1.2. Theory
D, Arsonval Galvanometer
This type of galvanometers is used in the various niethods of resistance
fireasurelnent and in D.C. potentiometer u,ork. D, Arsonval galvanometer belongs to
the group of nroving coil instruments. So, this type of galvanometer consists of a coil
canying the current to be measured and sr,l,inging in the held of a permanent magnct.
Constrr,rctional f'eatures of the galvanontctcL are as follows:
I . Moving coil: Tlre coil of many tnrns of finc wire wound on an aluminum fornrcr
Itltry be eirhet' rcctangular or circular. "l he coil is suspenrled bctween the poles of
;, ptlllliillcrlt lrl.rqnct. Titere is u ilretl irr-rn core
insidc tltc .cctil, tlte eir g,rp.;
l'ci\\1ccl1 1lr.r irorr core iind the perrrrtrircnl rn:rgnct ale usLrlrll\ about 1.5 rnm. the
ilt'll core i. cylitrcirical irt casc ol'r'et'lLrncular c,.ril ancl is sphclical in case 1i'
ciloulirr coil.
S uspension
2. Damping: In case ol'a coil with a rnctar former, mainly alnrninum, eddy
current
dan.rping introducecr. 'rhe cr'rping torque wliich opposes the ,iotion is
is
produced due to the rcaction of eclcly current with the pcrmiinent
magnet-filccl.
Tlre edcly current is clLre to the e.nr.f. indr-rced i' the fornrcr by the fi,-.ld of thc
ironnanent-nr il.qlret as thc firrnrcl nrtntcrs irr it. ;\ltcr.natil e ly., claniping pray bc
.btaired b1'c.rncctirr,q a sl*ra1 rcsislurcc nc.oss tlrer ga1r.a'.ar,.:lcr.
3 lndication. In lnost cilscs. a btatr ol light Js plo.lec:lcd to a srrlill mirror
attirclrecl
t() the s,s.pc.rsion :trip, as shor.r'r i. liisure ( l), thr..ugh a glass
windou, in tlrc
();llei cas('01'1lle instrrttrcnt.'J-lre bcanr o1'light is rcflcctecl pr thc
fr.6 mi[()r n1 to a
scale.
4. Zc ro-:rd.iu s tnr cnt. ;\ tolsion rread, to * hitrh the crld of the sr:spension str.ip is
attachecl. is providcci lirr ad.juslinr thc coil
Ir)sitjon arrcl hencc, rhe zero settilrg.
5 ( lilil cllrnrlling .\ 5115;''.'r''rir1.1 p.irlvrnoircl.
i hls ,r rl,:licntc i.rrovilrg s...strr.n and
hence thcy are protected fi.om mechanical injury when it is moved.
D,C. Voltmeter
The voltmeter .ses a high series resistance with the permanent magnet moving
coil meter.
-l'he
range of a voltmeter can be increased by resistance in series with the
voltmeter as shown in Figure (2). If the instrument gives full scale deflection when
a
current I is passed through it then:
Where:
V: voltage to be measr-rred
R: meter lesistance
ll.: series resis',ance
\/: il{: r'oltagc drop acr.oss thc instruntent for fuil scalc deflection
__j
l:i,gtrtL t)1. 1).( . I ulttttt't,,t.
D.C. Amnteter
I ne permanent magnet moving coil instrurnent is basically used fbr the
l)rcasurellent of current. As the nrovin" coil is small and light it can carry very srnall
cun'ent. Therefore, it is necessary, lbr measurement of high current,
to employ a
device which allows a known fraction of the current to be measured
to pass through
the coil. This known fraction must be within the range of the instrument.
A shunt
resistance is used for this purpose. Shunts are precise low resistances.
i /'-\ R
^ \___
\l'-l
I
_ 7\,\/\\__ r I
o
R.
3' Change the source voltage to 10V and notice the deftection of the
galvanometer.
4' Yary the power supply voltage until full scale deflection is reached. Record
the vaiue ofvoltage.
5' Repeat step 1-4 to design a voltmeter that can measure the voltages (15V
and
20v).
6. Compute the absolute error and relative error.
(2)Ammeter
1. Connected the experiment circuit shown in Figure (3).
2. connected the ammeter to measure the current in the circuir.
3 ' Cornpute the resistance R,6 theoretically to getting full scale deflection
(F.S.D.)
7' Change the powe| suppiy voitage to 0.5mA ancl notice the dellection of the
galvanonreter.
8' Vary the souLcc voltagc until full scale <lcf'lection is reachecl. Itec:orcl
the value
o i cU n.cn t.
Calibration of Voltmeter
4.t. Objective
l. To leam the procedure of calibration of a voltmeter.
2. To determine the relative effor.
4.2."theory
'I'here are many methods in caribrating the measuring devices and these
methods should have a high degree ofaccuracy during calibration procedure.
I llt'
l I
t l
I
In, 1- Iu P|tr,rtr,rrrul,
()r I
r
I
I )tl- ll,ll
I
l
v l
I
4-1
A simple method o1 calibrating a dc voltmeter is shown in Figure(l), where the
voltage (E) across dropping resistor R2 is accurately measured with
a potentiometer
or digital voltmeter. The meter to be calibrated is connected across the same two
points as the potcntiometer and should therefore indicate the
same voltage. The exact
value of the voltage V on the terminals of the calibrated vortmeter
is
Y=(1-+p /p \FlO //l\
L '' ' "l ' "2t-digital " '\t )
,
By changing the variable resistor R so that several points on the voltmeter
scale can be calibrated. vortmeter tested with the method
of Figure (1) can be
calibrated with accuracy of+0.01 percent.
4.3. Procedure
4.4. l)iscussion
,1-2
University of Technology
lllectromechanical. Eng Dep.
Ivleasulement Lab.
z
^nd vear
Exp No (3)
Calibration of Ammeter
3.1. Objective
3.2. Theory
l)/C Supply
'l'o - F -------
Pr r Icn lion r.'ter crr I) ig itll
voltnletitr
I": E/Rs
Thus the results of this calcuiation is compared to the actual
reading of the
ammeter under calibration and inserted in the circuit.
The circuit consist of a
precision dc power supply and a resistor R" (which is
the intemal resistance of the
ammeter) and a variable resistor R is placed in the circuit
to control the currenr ro
any desired value, so that different points on the meter
scale can be calibrated.
Finally the difference of the comparison between the exact
readings with the
measured reading of the current dete'nines the enor
in the meter. correction
curwes c.uld be drown between the measured current
I and the relative error E. as
shown in Figure (2)
3.4. Discussion
2. Theory
Wien Bridge is one ofthe simplest and best-klown oscillators and is basically used extensively in
oircuits for audio applications. Figure (1) shows the wien bridge circuit configuration. on the
positive side, this circuit has only a few components and good frequency stability.
Because of its sirnplicity and stability, it is the most commonly used audio frequlncy oscillator. in
the figurel shown the Wien Bridge circuit is connected between the amplifier inputierminals and
the or{put lerminal.
The bridge has a series RC netrvork in one arm and a parallel RC network in the adjoiling arm. In
the remaining two arms of the bridge, resiskrr R3 and Rp are connected.
1'he phase angle criterion for oscillation is that the total phase shift around the circuit must be
0".
'I'his condition occurs only when
the bridgc is balanced, that is at resonancc. Thc frequency of
.scillation Fe is exactly the resonant l'requcncy o1'the bala'ced wien Bridge and is given by
F-
to
- '>* [n -., n
a/ l.1t JI1 f(2 L-- 1 L 2
If Rt- R2- R
uncl C1-, C.-.C
Iro - 1 /2rr1l(. - 0. I 5g/Rc
AssLrrrins tlltt ihc resistors at.i c(luitl in tlre value, ancl llrc capacitors are equlrl
in the virlLre irr the
t.rrt. LirL icg r,l'tlrr.: \\'icp llricirrr .\t r tis lcr;rrencl, lhe r:airr rcclrrircd
lirr sitsllrilcLl oscilirrtiorr rs
g.i\ell lr.\'
,1.,.-. 1/Lt -l
i.c.. I ' Il 1, i/|.; -l,L,rRr 2R,
Figure (l) Llien-Bridge Circuit Schematic
3 - Equipments Needed:
l. Oscilloscope
2. Patch Cords
4 - Circuit diagram:
l'ldn Bfl
':lgs
i:rrcntshrr lrtijnr,r
Observation Tqble:
Frequency
6 - Discussiotr
1. Whet is the icas,,r of llte fletlitetrcl' Iiittiiatiorl in \\'i:rl Ilridgc?
-1. ( oLrltl rto usc tlrc bridgo ftrr capacitancc, tttcrtsrttctnettt/ lfycs, crirllin tlrc ntaitt cotttlititr:r 1tr
acliievc thlrt.
University of Technology
Electromechanical. Eng. Dep.
Mcasurement Lab.
a vear
Exp. No. (5 )
10.1. Objective
10.2. Apparatus:
10.3. Theory
l't,, Vu,
Vn, I/^^ "'(1)
I,R, I,R,
t n, - 11 "' t'';
Where:
l. I lke the valuc o1'I11..-1t2. 1 00O. ancl thc: vrLr.iablc resistance box.
3. Sot D.C. porver supl.rly o1. (lr_t 0V).
.1. \,'rrry lil
r'alue to olr,.:rin [lre lorv value for.r,r,rltnrcte
r.
5 \\i iren balarrce of the briclgc is obtainecr,
carculate trre varr-re of R.* fronr equatio'
(3).
10,5. Discussions
Ery. Na (6)
AC Yoltmeters
1. Objective
To design, constuc! and calibrate a basic AC voltmeter circuit using both half wave and
frrll wave rectifi cation.
2. Theory
There are several types of meters available in measuring altemating currents and voltage.
So far the most widely used is the rectifier type of meters. This instrument can be constucted by
using the basic DC movement in conjunction with rectifiers. It is an average responding type
with scale calibrated in terms of the r.m.s value of a sin wave.
It is important to review some basic definitions related to a.c quantities.
'Average value of ac quantity is defiaed as that value which is obtained by averaging all the
instantaneous values over a period of a half cycle.
For the symmetrical a.c quantity, the average value over a complete cycle is zero as both positive
and negative half cycles are exactly identical. Hence average value is calculated over a half cycle. If
the a.c quantity is continuous then aveftrge value can be expressed mathematically using integration
a.s,
,/z
ooo:
2t oudt
i J0
Where the interval (TD) indicates the average over hslf a cycle.
rf
RMSrf stands for Root mean square, and, is away of expressing an A.C quantity of voltage <ir
current in terms functionally equivalent to DC. For example, 10 volts A.c RMS is the amount of
A.C voltage that would produce the sarre amount of heat dissipation a cross a resistor cif given
value as a 10 volt D,C power supply. Also known as the uequivalent" or D.C equivalent value of an
A.C voltage or current. Mathematically the r.m.s value of the continuous as voltage having time
period T is given by.
T
1r
11^2tt
Er.m-s:
IJ
o
A.C Vohmeter using half wave RectiJier.
The A.C Voltneter using half wave Rectifier is achieved by introducing a diode in a basic d.c
volfrneter.
This is shown in the fig (1). The diode D conducts only during positive halfcycle
And meter moyement is by passed for another cycle. Hence it responds to half the average value of
the a.c input.
E* = E. /tt
Also E^ = '!28-,.^.,
E^ = 0.45Er.^.,
or Ea" = 0.458,.^." , Ea" = 0.318 E.
Thus the value ofseries multiplier can be obtained for a.c voltrneter as.
R" = Rr- (RD + Rn)
Where
R7= Ea" / Ipsp
R2 = Forward resistance of the diode D.
.R,n = Movement intemal resistance,
The meter a.c sensitivity (O,/tQ equals 0.9 of d.c sensitivity for firll scale.
1. Calculate theoretically the value of R,, for the following r.m.s AC voltage ranges (3, 5' 7
Vr.m.s) using half-wave rectification.
2. Construct the half-wave AC voltmeter circuit shown in Figure (1). Using variable resistor as
multiplier resistors (Rs).
3. Apply the r.m.s AC voltages range (3, 5, 7 Vr.m.s) to your AC voltmeters and adjust the
resistors R., for firll-scale deflection on each range.
4. With a firll-scale reading on the 5V r.m.s range, connect the oscilloscope across the movement
4. Discussion
1. Could we use AC voltneter to measure DC voltage? How, explain.
2. What is the advantage of the rectifier?
3. Calculate the AC sensitivity of the half wave and full wave circuits ammeter movement.
4. Do you think the value of the Diode forward resister Re is fixed or variable with input
voltage?
lsoli.riion
lronsiorme
lo t{
Transfomrer ?s K
5n K
Exp. No.@
Study and Calibration of a Single - Phase kilo watt hour Meter (Energt
Meter)
1. Objecfive
l. To study the principal operation of single - phase kilo watt hour meter.
2. To draw the model graph of the percentage error versus a unity power factor load
crxrent,
2. Theory
Energy is the total power taken through a period of time:
t
y,y
= lvidt ----a)
0
Were:
W= work
v= Voltage in volt.
i: Cunent in Amp.
t= Time in second.
The unit of energy is Joule or watt - second. It is one watt during a period of time equal
to (1) second and, if the unit of time is hour then the energy was measured is watt hour. In high
energy loads, a kilo watt -hour units is used which is the value of(1000) watt in one hour.
The watt hour meter is not often found in a laboratory situation but it is widely used in
commercial measurement of electrical energy. In facl it is evident wherever a power company
supplies an industrial or domestic consumer with electrical energy.
Energy meter is an integrating instrument and takes into account both quantities i.e the
power, time and the product of them, which is the energy. An energy meter keeps a record
of the
total energy consumed in a circuit during a particular period of time but it does not give any ide4
about the variation in the rate of energy consumption during that period.
't
The inductive energy meter device consists of two electrical magnets. One of them is the
load current coil, and the other is the voltage coil connected across the line potential.
A light aluminum disk is suspended in the air gap between current coil and potential coil.
Both electromagnets coil produces a flux passes through the disc. The two fluxes induce e.m.f.s.,
in the disc which further produces the circulatory eddy currents. The reaction of the eddy
currents create a torque on the disc, causing it to rotate. The developed torque is proportional to
the field strength of the voltage coil and the field strength of the current coil.
Braking of the disk is provided by two small permanent magnets located opposite each
other at the rim ofthe disk.
Pxn;p=hn
Where n : speed of the disc in rev/ time.
ft = constant of proportionality.
Calibrating the energy meter means to find out the error in the measurement of energy by
energy meter .every energy meter has its own characteristics constant specified by the
manufacturer which relates the energy measured in KWh and the number of revolutions of the
disc. But practically the revolutions are very large and can not measure in the laboratory. Hence
using this constant energy recorded for certain less number ofrevolutions say 5, is calculated in
the lab. For the calibration purpose, this energy is denoted as Ea (Actual energy).
To have zero error the Actual energy consumed by the load for the time corresponding
to the 5 revolutions must be same as Ea. This energy is called the true energy denoted as E1. For
various loads, the time required to complete the 5 revolutions of disc is measured with the help
True energt measured in htatlsec = power (p) in watt x time (t) in seconds
Actaal energy mewured in watt sec = (3.6x106 t no. of revolutions in time (t)) / N
Where:-
t = time in seconds for n cycles of meter under test
3. Procedure
l- Observe and record the Energy meter constant (N) in revolution / KWh
2- Connect the circuit as shown in Figure (l). Select the suitable range for the voltnoeter,
arnmeter and watt meter.
3- Change the load current in steps of 0.5 A to maximum curent of the KWh meter (current
coil) and write down (A, v, w) of each step, also the time taken by the disc for 5
revolutions (use stop watch).
4- Calculate the relative enor for every value of load current.
Tabulation:
2. Theory
Thermistor is a contraction of a term thermal resistor. Although positive
temperature co-efficient
(P'T'c) of unit exhibit an increase in the value ofresistance with increase
in temperature are available,
most a negative temperature coefficient i.e. their resistance decrease
with increase in temperature. In
some materials the resistance of Thermistor at room temperature
may decrease as much as 6%o for l.C
rise in temperature. This high sensitivity to temperature change
make the Thermistor extremely well
suited to precision tempemtue measurement, contol & compensation.
Therefore, especially in lower
temperature range of-100'C too 300"c, or to detect very
small changes in temperature which cannot
be observed with an RTD or a therrnocouple Thermistor
are composed of a sintered mixture of
metallic oxides, such as Mn, Ni, Co, Cq Fe, & U. Their resistance
r.ange from 0.5O to 75MO and thev
are available in wide variety of shapes and sizes.
! eod
(3) Rd
(b) Red
Leoo5
(c) Disc
(d) Probc
Figure (1)
Charocteristic 4350K
Temp.
Constru.tion Dlcctric.lCircuit
(a) (b)
Figure (2)
The resistance of the NTC Therrnistor varies over a wide range for the temperature rang available
within the heated enclosure. If resistance reading is to be taken at rcgular interval of 1 minute, the
reading must be obtained very quickly. The method selected connects the Thermistor in series with a
calibratsd resistance to the +5 supply.
For each reading the variable Thermistor is adjusted until the voltage at the junction of the Thermistor
and resistance is half of the supply voltage. For this setting there will be the same voltage drop across
the Thermistor and the resistance and the since the same Current flows in each their resistance must be
equal. Hence, the value of the resistance read from the calibrated resistance scale is the same as the
resistance of the Thermistor.
3. Procedure
1- Connect the circuit as shown in figure (4).
a. The A output of NTC Thermistor to C socket of 10 tum potenticimeter.
b. Corurect a dieital multi-meter as voltmeter between socket B of 10 tum Dotentiometet and
ground.
c. Connect socket A of 10 turn potentiometer to Gnd.
2- Switch 'On' the power supply and note the temperature by connecting the voltmeter temporarily to
the IC temperature sensor output adjust the l0 tum potentiometer until the volt"ge indicated by--
tlre voltmeter is 2.5V and then note the dial reading.
Nofe.' Since there is a lK resistance in the output lead of the potentiometer the total
Resistance will be lOxDial reading +lK Ohms.
3- Connect the +12V supply to the heater element input socket and at I minute intervals note the
values the dial reading to ptoduce 2.5V across the resistance and also the temperature from the IC
temperature sensor. Record the values in above Table.
4- Record the values of dial reading & temperature in below Table.
5- Switch'Of the power supply and disconnect the Heater element supply (+l2V).
Time(minutes) 0 I ) 3 4 J 6 7 8 9 10
Temperuture
(from IC
temPeralure
sensor)"K
Tempercture
(from IC
temperature
sensor)oc
Dial reading
for 2.5V
Thermistor
resistonce
(10x dial
reading +lK
Ohm)
Tlici'mistor
Rcsistancc
K(o)
10 20 30 40 50
'I't'l|| pcriltll o(:
I {.
Figure (3)
4. Results and Discussion
1- Plot the graph the Thermistor against temperature. It should resemble the graph
below in Figure
(3). Make a comment on the results obtained.
2- Explain the difference between Thermistor and resistance thermometer
(RTD).
3- Discussion the various limitations of Thermistor.
Figarc
f)
University of Technology
Electromechanical. Eng. Dep.
Measurement Lab.
2nd yeu
1. Objective
To be able to analyze t}te working frequency response ofAC meters.
2, Theory
General propose AC voltrneters have a flat frequency response curve over a limited range
of frequencies, the meter response inaccurate above or below this range of fiequencies. This is due
to the frequency characteristics of components such as rectifier diodes; instrument-type rectifiers,
capacitors, and wire-wound resistors in voltmeter circuit. The major effect is generally attributed to
capacitance associated with the recti$'ing elements in AC voltrneters that use a d'Arsonval meter
movement.
The frequency response is presented in a form of a graph that shows output amplitude
plotted versus frequency. Tlpical plot of the voltage gain ofan amplifier versus frequency is shov,,n
in Figure (1). The gain is null at zero frequency, then rises as frequency increases, level off for
frther increases in frequency, and then begins at high frequencies. The fiequency response of an
amplifier can be divided into tbree frequency regions.
The frequency response begins with the lower tequency region designated between OHz
and lower cutoff fiequency, F6 the gfi is equal to 0.707 An;a is a constant mid-band gain obtained
from the mid-band frequency region. The upper frequency region covers frequency between upper
cutoff frequency and above. Similarly, at upper cutoff frequency, F11 the gain is equal to 0.707 Ania.
After the upper cutoff frequency, the gain decreases with frequency increases and dies off
eventuallv.
Midband region
3. Procedure
1. Connected the experimental circuit shown in Figure (2), use any one of the two meter
movement.
2. Set the sine wave generator output to 2V,-s at I KHz.
3. Decrease the generator frequency until the voltneter reading decreases to l.4vms.
Record this tequency until as F1.
4. Increase the generator frequency until the voltmeter reading again decreases to l.4v*.
record this frequency as F11.
5. Set the generator to each of frequency shown in the data table and record the voltmeter
reading.
6. Repeat steps I through 5 with the second voltrneter.
4. Discussion
Plot the frequency response curve for each meter movement. Discuss the curves.
Sine AC
Wave Voltmeter
Generator
-t a
Digital Voltmeter
Exp. No.@
6.1. Objective
Employing the oscilloscope to measure AC voltage, and description signal
waveform.
6.2. Theory
6-l
In order to use the oscilloscope to rnake AC voltage measurements, which
means the observation and measurement of time varying voltage signals. In Figure
Sinusoidal AC Voltage
0.6
0.4
10 15 20
'lime in Seconds
6-2
6.3. Procedure:
l. Connect the oscilloscope with function generator.
6.4. Discussions
6-3
University of 1'echnology
Electromechanical. Eng. Dep.
Mcasurement Lab.
2no year
1. Objective
To study the Characteristics of tungsten Filament Lamp
2. Theory
the light source used in ST230l Trainer is a imported Tungsten filament lamp with a mixture of
nitrogen and argon gas under low pressure. 1'hc lamp glows more brightly as the power feeding
1he lamp is increased. Two factors will be aff'ected as the lamp voltage is increased:
I- The temperattrre of the filament is proportional to the input power. Power varies with the
squalc of the voltage and is also affected by the resistance of the lamp, which increases as
f rlament temperature increases.
2 - Thc spectral response of the lamp varies with the filamcnt temperature. At lorv temperatures
lltc light is in the infrared region of the visitrlc spectrum and the light graclually increases ur
fiecluency (red-olange-yellow. . . ) as the tenrferature is raised. These lactors makc it difficult 1o
bo plccisc about the responsc of the sensor.s which will be investicated.
'l hc Limp is oonnected
as shou n below
Itrnrrl
Larnir
OV
Lamp Filament
3. Procedure
e. Connect the other input of Filament Lamp to +ve input of moving Coil Meter.
I Connect the -ve input of moving Coil Meter to 0V.
2- Switch'On'the power supply & set the 10 KQ wire wound potentiometer to minimum for zero
voltage (on Moving Coil Meter) from the power Amplifier.
3- Take readings of Filament Lamp current as indicated on the Digital Multi-meter as lamp voltage
is increased in 1V steps. Record the results in below table.
4- Switch 'Off thc power supply.
4. Ra,^ ults and Discussion
I - Calculate thc conesponding values of Lamp Filament pow-er (Vxl) and resistance (V=l) record
thc results in the given table.
I - Plot the grzrphs of lamp powcr and rcsistance against applied voltage. lt should resemble the one
givcn in graph I . Write clou'n your contment.
3 - Plot the I-V characte|istics and compare to an ideal resistor.
4- State the equation of temperature influence on metal resistance and write a brief comments on
filament resistance bel.ravior on temp rise.
Lamp
l{esistance
(f))
: YN
'/ \)
a 2 o
E
:
r-t n
I U- '6
a O
A
o,7\)
o
Ci/.\ :
<l? l*
?_?
lo
l- JfuWVJ
{tr
c-
F
il
P
t- o
q)
o U
-t
o
A
an
\-
l*
b lrl
G .'' EF
Y,r
I
c
tj
{
.l
il
o:
T?
t::
o I
iJi
:'
.1.;
1>
:l
ffi
ll I----'l i rol..rlr irr)a
a JoilJili.roJ
0g
University of TechlologY '
Electromechanical. Eng. DeP. .-
Measurement Lab.
2"d year
Exp No (11)
De Sauty's bridge
13.1. Objective
MeasurementoftheunknowncapacitancebyusingDeSauty'sbridge'
13.2. TheorY
Bridge is used for the accurate measurement of electrical quantities:
upon the excitations used, the bridges are classified as AC bridges & DC bridges.
De-sauty Bridge comes under category of AC bridges and it is used for
measurement of capacitance, General form of AC bridges consists of four arms of
impedances & AC excitation.
At balanced condition:
(l/jwCl) * q4: (l/jwC2)*R3"' (l)
SeParating real & imaginary Part,
c1: c2*(R4/R3) ... (2)
l3-l
Figure (I 3- 1): Experimental Circuit
13.3. Procedure
1. Connected the experimental circuit shown in Figure (13-1).
2. Set the null point of galvanometer by adjusting the variable resistor R3.
13.4. Discussion
l. What is the limitation of this bridge?
2. What is the disadvantage for this bridge?
I t-L
Exp.
'r /" (! lr-l()
dnol
)estTn ns lr,"Kti- n .P Ar',o\ague-
A"lLim"L.r
D.. A^m.Lt"
'}-c, Anmelrr fclhqe
o
o -S zt A
o^5 n/
Lr= s' 4A
Qrh ''
Z!-"x5o X(;< = 3ooJL
.-a
=
I -?
o 9X lo- - $oX\o"
l* = o -51 ntA
tr: ''f(-<o'$ cz -5 |
x/"/ t o =o-ol1
f) '<. V o ll *ot'r
D.. Vol(.ne_(d ro(riqe
U
l^ r,
/uv
Ro
o
,l
I
I
lq=
o
5"'u
77-"xso X lo(
l\< -{
L fo X /c)
q7^3Ktr-
V'ol - /o-l v
7
\ 'Qo = //a - | (t = u-l
2-
\)(
/o'l - /c X4 o" - "'" "11
=
(o'l
--f,x.l
f - ,Uo(2)
A/ /t'r'
=- : -Jt;,
->(
t_ ':> ''--' --
:. 1r 6; .i2 r:-i -- r\
Lu /: c,
1t \) CS
\. L
C.
(-'
Obsemafion Table:
R'1
R"t
., *)J-S =Z-LS
R1 E.^.., 'Z 7\
-J1
- 72)'
Lt,.,\ t v
I A \o
.zz 5,,
9-t --=-_- : \S ->
.->
E r^s
f-.tt t.., .-\,-\ e.
f,Ln-
r\s
\x lI\.
-J
-1 1) cl
/a1
l'1\* c> -) u
\\ \
K: x'tR'I \\
Llloo : J51-. \€Ra-t(,.,
R.-d r\z5.iL
=
Exp. No. (7)
Study and Calibration of a Single - Phsse kilo watt hour Meter (Energt
Meter)
.,
Time(minute 0 I 3 4 l 6 7 8 9 10
s)
Temperature
)-5 2 7oL:J 7lo 3t3 j tt"1 '3r1 .r 3t t.? 3L) 7 j tt.5 j26 1 t28
(from IC
temperature
sensor)oK
Temperature
(from IC 3z-"6 T9 9/.85 n.8t p.7S $b) 18-tt ," 5t 5235 ,3-+5 hst
temperature
sensor)oC
Dial reading
for 2.5V 66o 6qu ,q(, 5qo LA1. ,Lz 5tt # 9so qotLl 9.12-
Thermistor
resistance
4/
+'o +'11
Lqa b+ 6qL 6n (rb 6." L 6- 51'\ 51L
(10x dial
reading +1K
Ohm)
Exp. No.lQ)
Analog Voltmeter
vo/4- vrAeatqfe^"nt
)
(
Ly u>i") cs 5 c-\ LLo> C.,F e /
F= \kl+z
-9
T= ZY.o'5 Xt.
\ \ .-.\ oo.' ltZ'- \ l=ftZ
'rr I 2 Y:o'S X tr-l
V,1q X\Y-
?ta
1(v
\,r= Jon x * /- 7V
r qv .La v
V= b ?|Yt A X'
Lamp 0 1 2 3 4 5 6 7 I 9 10
Filament
voltage (V)
Lamp
Filament Jo.1e 1'1o dl 8, I or{ \i rrl l{ (fl.IX 881'z t"t5
o
current(mA)
Lamp
Filament pt6 2(J q) qtL"16 ,8t'|t 7a bt q&.\ \
5'1n lLo4'\
power (mw)
Lamp
Resistance o n6) 18.20 3:"l.lb 19 31 vtt \5 qt )o 5t 53 tJ
(o)
Exp. No. (l)
De Sauty's bridge