GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
ii
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
Direction 2298380-100
Revision 9
iii
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
If you have any comments, suggestions or corrections to the information in this document,
please write them down, include the document title and document number, and send them to:
GENERAL ELECTRIC COMPANY
MEDICAL SYSTEMS
MANAGER - FIELD SERVICE DEVELOPMENT
ENGINEERING, X-RAY W-702
P.O. BOX 414
MILWAUKEE, WI 53201-0414
DAMAGE IN TRANSPORTATION
All packages should be closely upon discovery, or in any event, 8*285-3468 immediately after
examined at time of delivery. If within 14 days after receipt, and the damage is found. At this time be
damage is apparent, have notation contents and containers held for ready to supply name of carrier,
damage in shipment" written on inspection by the carrier. A transĆ delivery date, consignee name,
all copies of the freight or express portation company will not pay a freight or express bill number, item
bill before delivery is accepted or claim for damage if an inspection is damaged and extent of damage.
signed for" by a General Electric not requested within this 14 day
representative or a hospital receivĆ period. Complete instructions regarding
ing agent. Whether noted or claim procedure are found in
concealed, damage MUST be Call Traffic and Transportation, Section S" of the Policy &
reported to the carrier immediately Milwaukee, WI (262) 827-3468Ă/ Procedure Bulletins.
6/17/94
iv
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
TABLE OF CONTENTS
REVISION HISTORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
v
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
REVISION HISTORY
Title Page 9
i thru vi 9
Back page
vi
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
SECTION 1
INTRODUCTION
The purpose of this addendum is to provide updated information to the Prestige VH/
II AK/AD CDROM 2249768-200 revision 5. This addendum supplements the inĆ
formation contained on the CDROM. When the CDROM is revised, the information
contained in this addendum will be added to the CDROM.
SECTION 2
DUAL FOOTSWITCH FRU PART NUMBER
The part number listed in the CDROM for the Prestige dual footswitch (197799P005)
is incorrect. The correct part number is 2135299 (dual footswitch EMC version with
cover).
SECTION 3
ULTRANET SRI COLLIMATOR FRU PART NUMBER
The part number listed in the CDROM for the Prestige Ultranet SRI Collimator
(861333P015) is incorrect. The correct part number is 45296713.
1-1
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
SECTION 4
MODULE 2 CIRCUIT BOARDS FRU PART NUMBERS
Some of the part numbers listed in the CDROM for the Prestige Positioner Cabinet
Module 2 circuit boards are incorrect.
Positioner Cabinet Module 2 Incorrect FRU Part Number Correct FRU Part Number
Circuit Board
2A10 831271G015
2A38 Low Power Servo Board for 831827G035 831146G025
Collimator
2A46 On/Off Board 832907G015 2320694
831180G025 (U-Mask and Film
Marker Version)
2A80 CPU Board 831508G035 2320690
1-2
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
SECTION 5
FILM CASSETTE DETECTION BAR ADJUSTMENT
The RG0012 procedure listed in the CDROM for the Prestige Film Cassette DetecĆ
tion Bar Adjustment" is incorrect. Following is the correct procedure.
3. Refer to Illustration 1-3 and 1-2. The microswitch on/off engagement should
be set so that, when the front cassette detection arm center piece is raised to a
distance 17.5 mm (+/- 0.5 mm) from the top surface of the cassette tray, the
switch just switches between on/off. Loosen the two screws holding the microĆ
switch to the SFD frame and retighten when the correct on/off position is
achieved.
1-3
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
ILLUSTRATION 1–1
15.0 mm
2 mm
15.0 mm
1-4
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
ILLUSTRATION 1–2
CASSETTE
CASSETTE 1.0 mm
jaw
jaw
1.0 mm
1-5
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
SET SCREW B
2 SCREWS HOLDING
MICROSWITCH
ACTUATOR ARM
OF MICROSWITCH
SET SCREW A
SECTION 6
FLEXIBLE HOSE FRU PART NUMBER
There is no part number listed in the Prestige II CDROM for the "white flexible hose
on top of the tube arm." The correct part number is 853002P015. It is a FRU code 2.
SECTION 7
PREINSTALLATION DOCUMENT REFERENCE NUMBER
SECTION 8
SW133 AND SW139 SETTINGS ON 2A80CPU BOARD FOR CONNECTING LAPTOP
Add the following note to step 3 of Job Card IN0009, Tube Arm Installation:
Note: When you connect the laptop to the Prestige table using connector DB9 on
the 2A80 CPU board , you must check the settings of switches SW133 and
SW139. When using GPX cable (part number 2128821), set switches SW133
and SW139 on the 2A80 CPU board in position 2-3. When using MPG cable
(part number 2103512), set switches SW133 and SW139 on the 2A80 CPU
board in position 1-2.
SECTION 9
PALPATOR SPOON FRU PART NUMBER
1-6
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
SECTION 10
STATOR CABLE FRU PART NUMBER
The 18 meter long stator cable 2145653-2 part number has been replaced by
2145653-4, FRU 2.
The 24 meter long stator cable 2145653-3 part number has been replaced by
2145653-5, FRU 2.
SECTION 11
FLAT PANEL DISPLAY CONNECTOR BRACKET
Save this
bracket
When you replace the flat panel display (part number 2148762) on the Prestige II conĆ
sole, the cable connector bracket shown is not included. Remove the bracket and
install it onto the replacement flat panel display. The bracket is held on with rivets.
Remove the rivets carefully. Reinstall the bracket using nuts and bolts which must be
provided locally.
SECTION 12
TABLE TOP FRU CHANGE
The carbon fiber table top (model number 830948P022 manufactured by TOPKEY) is
no longer available as a FRU replacement. Instead order the phenolic table top model
number 832871P012. The reason is that the TOPKEY carbon fiber table tops may
show artifacts. FMI 10757 will replace all TOPKEY carbon fiber table tops with the
phenolic version.
1-7
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
SECTION 13
JOB CARD VF0002
Supplies: None
Tools:
Safety Precautions:
Take care not to catch your fingers.
Prerequisites:
Review all screws that fasten the Foot Table. Every single screw must be torqued by
hand pressure.
Procedure:
1-8
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
ILLUSTRATION 1–4
2. Illustration 1–5 shows a left view from Illustration. 1–4. Check the 4 square
head (6 mm) & 3 square head (5 mm) screws.
ILLUSTRATION 1–5
1-9
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
3. Next, Illustration 1-6 shows an upper view from Illustration 1-4. Check the two
indicated 5 mm hexagonal internal head screws.
ILLUSTRATION 1–6
4. Illustration 1-7 shows the joint for Jack 2. Check the four 5 mm square head
screws.
ILLUSTRATION 1–8
1-10
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
5. Illustration 1-9 shows the chain pulley for Jack 2. Check the 2.5 mm square interĆ
nal head screw, taking advantage of the hole that is in front of this pulley. IllustraĆ
tion 1-10 shows how this practice must be done.
ILLUSTRATION 1–9
Refer to Illustration 1-10 for checking the screw at the Chain Pulley for Jack 2.
ILLUSTRATION 1–10
1-11
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
6. Illustration 1-11 shows the other end of Jack 2. Ceck the two 5 mm square interĆ
nal head screws as shown.
ILLUSTRATION 1–11
7. Illustration 1-12 shows the joint between Jack 1 and its basement. Check the
three 8 mm hexagon head screws.
ILLUSTRATION 1–12
1-12
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
8. Illustration 1-13 shows the other end for Jack 1. Check the four 5 mm square
internal head screws.
ILLUSTRATION 1–13
9. Also, Illustration 1-14 shows three 3 mm square internal head screws that need
to be checked.
ILLUSTRATION 1–14
1-13
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
10. Illustration 1-15 shows four 5 mm square internal head screws to check at the
basement.
ILLUSTRATION 1–15
11. Illustration 1-16 shows six 8 mm hexagonal internal heads screws to check.
ILLUSTRATION 1–16
1-14
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
12. For the following Illustrations (1-17 to 1-19), use a Phillips Head Screwdriver
to check the screws. Screws to check are identified in each photo.
1-15
GE MEDICAL SYSTEMS PRESTIGE CDROM ADDENDUM
REV 9 DIRECTION 2298380–100
13. Illustration 1-20 shows four screws (left arrows) & 2 keys (right arrows) that hold
the 2 pulleys for Jack 1. Each of these need to be checked.
1-16
GE Medical Systems