Photometer 5010 ve
DEPKES RI AKL 10102011427
Semi - automated photometric system
Flexsible cuvette concept
Effective temperature regulation system
Double-secured liquid control
Open system : high capacity for programmable methods
JRJDE IIE
Eratalogue
peaceType
Light
Wavelength
Wavelength
Selection
Photometric Range
Cuvette System
‘Temperature Control
Semi automatic, single-beam, double-beam filter photometer
Source Halogen lamp — 12V,20W
340 nm - 800 nm
‘Automatic via 9-position filter wheel :
6 standard interference fiters: 340 nm, 405 nm, 492 nm,546 nm,578 nm, and 623 nm;3 positions
for optional fiters of choice
0-250
Micro flow cell: 32_ 1,10 mm light path interchangeable with normal standard cuvettes ( macro of
semi-micro,disposable or special glass)
Internal Peltier element,temperature variable, pre-adjusted to 25 C,30 C, and 37 C Equilibration
eo for aspirated reaction mixture to reach 37 C from ambient temperature: 15 sec
Aspiration System
Sipping Volume
In peristaltic pump driven by stepper motor, programmable aspiration volume controlled by
infrared light barier (buble detector)
250 110 2000 |, programmable for sample as well a for washing in 50_ I steps
Presentation
Integrated Printer
Languages
Memory
Data Logging
Signal Port
Measurement Procedures
Measuring Time
Delay Time
Dimensions
Weight
Marking
‘Touchscreen, graphie display,240 X 128 dots, blue, backlight
Dot-matrix printer, 24 characters per line
English and Indonesian
General operating soffware contained in FLASH MEMORY can be updated by PC (data file downloaded
from Internet or CD ROM ) Capacity for 250 pre-programmed tests
Up to 230 tests can be stored in RAM. Import of data by keyboard or PC
Up to 50 non-linear calibration curves with max 20 sets of points can be store
Up to 2970 results of max 99 samples can managed in the memory
Serial port for connection to an external printer and/or PC
+ Absorbance
End point with factorstandard or multiple standard, with or without reagent blank and/or
sample blank
Kinetics with standard, factor or multiple standards, with or without reagent blank
Fixed time with standard, factor or multiple standards, with or without reagent blank
Turbidimetry
Single, Double and triple determinations
Curve fitting for non-linear standard curves
Available QC program
Kinetic: variable from 3 ~ 25 deltas, time per delta 4 ~ 255 sec in 1 sec steps
Fixed time: variable from 0 ~ 65535 sec in 1 see steps
Programmable from 0 ~ 65535 sec in 1 sec steps
Length 33 cm X width 34 em X height 18 em
kg
ce
fe 6/1105 motions
DISTRIBUTOR :
Gedung RNI, JI. Denpasar Raya Kav. Dill Kuningan
Raja Wali Nusindo :ssare'12380 indonesia
RNI Group
Phone : (021) 252 3820 (Hunting) Fax. (21) 52914239
www nusindo.co id