You are on page 1of 1

Quantachrome QuadraWin - Data Acquisition and Reduction

for QuadraSorb SI
�2000-16, Quantachrome Instruments. All rights reserved.
version 7.1

Analysis Report
Operator: SSTTI Date:Thu Oct 03 08:50:35 2019Operator:admin
Date:2019/10/03
Sample ID: : A1X Filename: 021019E1.RAW
Sample Desc: Y.Budyk_1931040_MP3_9.3%Comment:
Outgas Time: 8.0 hrs Outgas Temp: 250.0 �C Sample
Weight: 0.0726 g
Analysis gas: Nitrogen Molec. Wt: 28.0134 g Non-ideality:
6.58e-05 1/Torr
Analysis Time: 1174.4 min Instrument: Autosorb Station 1 Bath temp.:
77.4 K
Data Reduction Parameters
Thermal Transpiration: offEff. mol. diameter (D): 3.54 �Eff. cell
stem diam. (d): 4.0000 mm

Adsorbate modelNitrogen Temperature 77.350K


Molec. Wt.: 28.013 Cross Section: 16.200 Ų Liquid
Density: 0.808 g/cc

Relative Volume @ STP 1 /


[ W((Po/P) - 1) ]
Pressure
P/Po cc/g 1/g

3.32938e-02 22.0767
1.2482e+00
3.82755e-02 22.7738
1.3982e+00
5.07467e-02 24.8682
1.7200e+00
9.97316e-02 30.2326
2.9318e+00

BET summary
Slope = 25.142 1/g
Intercept = 4.289e-01 1/g
Correlation coefficient, r = 0.999823
C constant= 59.622

Surface Area = 136.191 m�/g