Package Insert - 06267 - LH - en - 30406 PDF
Package Insert - 06267 - LH - en - 30406 PDF
VIDAS® LH (LH)
VIDAS LH is an automated quantitative test for use on the VIDAS family instruments for the determination of human
luteinizing hormone in human serum or plasma (lithium heparin), using the ELFA technique (Enzyme Linked Fluorescent
Assay).
bioMérieux SA English - 1
®
VIDAS LH (LH) 06267 O – en - 2015/01
Hazard statement
H318 : Causes serious eye damage.
H373 : May cause damage to organs through prolonged or repeated exposure.
H315 : Causes skin irritation.
H302 : Harmful if swallowed.
Precautionary statement
P280 :Wear protective gloves/protective clothing/eye protection/face protection.
P305 + P351 + P338: IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present
and easy to do. Continue rinsing.
P309 + P311 : IF exposed or if you feel unwell: Call a POISON CENTER or doctor/physician.
** Signal Word: DANGER
Hazard statement
H318 : Causes serious eye damage.
Precautionary statement
P280 :Wear protective gloves/protective clothing/eye protection/face protection.
P305 + P351 + P338: IF IN EYES: Rinse cautiously with water for several minutes. Remove contact lenses, if present
and easy to do. Continue rinsing.
For further information, refer to the Material Safety Data Sheet.
MATERIALS AND DISPOSABLES REQUIRED BUT WARNINGS AND PRECAUTIONS
NOT PROVIDED For in vitro diagnostic use only.
- Pipette with disposable tip to dispense 3ml, 2ml and For professional use only.
200 µl.
- Powderless, disposable gloves.
- For other specific materials and disposables, please
refer to the Instrument User’s Manual.
- Instrument of the VIDAS family.
bioMérieux SA English - 2
®
VIDAS LH (LH) 06267 O – en - 2015/01
This kit contains products of human origin. No - bilirubinemia (after spiking samples with bilirubin:
known analysis method can totally guarantee the 0 to 513 µmol/l).
absence of transmissible pathogenic agents. It is However, it is recommended not to use samples that are
therefore recommended that these products be clearly hemolyzed, lipemic or icteric and, if possible, to
treated as potentially infectious and handled collect a new sample.
observing the usual safety precautions (see
Laboratory Biosafety Manual - WHO - Geneva - Specimen stability
latest Edition). Samples can be stored at 2-8°C in stoppered tubes for up
This kit contains products of animal origin. Certified to 48 hours; if longer storage is required, freeze the sera
knowledge of the origin and/or sanitary state of the or plasma at -25 ° 6 C.
animals does not totally guarantee the absence of Avoid successive freezing and thawing.
transmissible pathogenic agents. It is therefore
recommended that these products be treated as INSTRUCTIONS FOR USE
potentially infectious and handled observing the usual For complete instructions, see the User’s Manual.
safety precautions (do not ingest or inhale).
Reading Master lot data
Do not use the SPRs if the pouch is pierced.
Do not use visibly deteriorated STRs (damaged foil or Before each new lot of reagents is used, enter the
plastic). specifications (or factory master data) into the instrument
Do not use reagents after the expiration date indicated using the master lot entry (MLE) data.
on the label. If this operation is not performed before initiating the
Do not mix reagents (or disposables) from different lots. tests, the instrument will not be able to print results.
Use powderless gloves, as powder has been reported
to cause false results for certain enzyme immunoassay Note: the master lot data need only be entered once
tests. for each lot.
Kit reagents contain sodium azide which can react with It is possible to enter MLE data manually or
lead or copper plumbing to form explosive metal azides. automatically depending on the instrument (refer to the
If any liquid containing sodium azide is disposed of in User’s Manual).
the plumbing system, drains should be flushed with
water to avoid build-up. Calibration
The wash buffer (well 9) contains a harmful agent Calibration, using the calibrator provided in the kit, must
(11.5% diethanolamine). Refer to the hazard statements be performed each time a new lot of reagents is opened,
"H" and the precautionary statements "P" above. after the master lot data have been entered. Calibration
The substrate (well 10) contains an irritant agent should then be performed every 14 days. This operation
(6.6% diethanolamine). Refer to the hazard statements provides instrument-specific calibration curves and
"H" and the precautionary statements "P" above. compensates for possible minor variations in assay signal
Spills should be wiped up thoroughly after treatment throughout the shelf-life of the kit.
with liquid detergent or a solution of household bleach The calibrator, identified by S1, must be tested in
containing at least 0.5% sodium hypochlorite. See the duplicate (see User’s Manual). The calibrator value must
User’s Manual for cleaning spills on or in the instrument. be within the set RFV "Relative Fluorescence Value"
Do not autoclave solutions containing bleach. range. If this is not the case, recalibrate.
The instrument should be regularly cleaned and
Procedure
decontaminated (see the User’s Manual).
STORAGE CONDITIONS 1. Only remove the required reagents from the
refrigerator and allow them to come to room
Store the LH VIDAS kit at 2-8°C.
temperature for at least 30 minutes.
Do not freeze SPRs, reagent strips or dilution buffer.
Store all unused reagents at 2-8°C. 2. Use one "LH" strip and one "LH" SPR for each
After opening the kit, check that the SPR pouch is sample, control or calibrator to be tested. Make sure
correctly sealed and undamaged. If not, do not use the that the storage pouch has been carefully resealed
SPRs. after the required SPRs have been removed.
Carefully reseal the pouch with the desiccant inside 3. The test is identified by the "LH" code on the
after use to maintain stability of the SPRs and return instrument. The calibrator must be identified by "S1",
the complete kit to 2-8°C. and tested in duplicate. If the control needs to be
If stored according to the recommended conditions, all tested, it should be identified by "C1".
components are stable until the expiration date indicated 4. If necessary, clarify samples by centrifugation.
on the label. Refer to the kit composition table for
special storage conditions. 5. Mix the calibrator, control and samples using a vortex-
type mixer (for serum or plasma separated from the
SPECIMENS pellet).
Specimen type and collection 6. For this test, the calibrator, control, and sample
Serum or plasma (lithium heparin). Do not use plasma test portion is 200 µl.
collected on EDTA.
7. Insert the "LH" SPRs and "LH" strips into the
None of the following factors have been found to instrument. Check to make sure the color labels with
significantly influence this assay. the assay code on the SPRs and the Reagent strips
- hemolysis (after spiking samples with hemoglobin: match.
0 to 300 µmol/l (monomer)),
8. Initiate the assay as directed in the User’s Manual. All
- lipemia (after spiking samples with lipids: 0 to 2 g/l
the assay steps are performed automatically by the
equivalent in triglycerides),
instrument.
bioMérieux SA English - 3
®
VIDAS LH (LH) 06267 O – en - 2015/01
9. Reclose the vials and return them to the required RANGE OF EXPECTED VALUES
temperature after pipetting. These figures are given as a guide; it is recommended
10. The assay will be completed within approximately that each laboratory establishes its own reference values
40 minutes. After the assay is completed, remove the from a rigorously selected population.
SPRs and strips from the instrument. Men: 1.1 - 7.0 mIU/ml
11. Dispose of used SPRs and reagent strips in an Women:
appropriate recipient.
- Ovulation peak (D0) 9.6 - 80.0 mIU/ml
RESULTATS AND INTERPRETATION - Follicular phase:
Once the assay is completed, results are analyzed First half (D-15 to D-9) 1.5 - 8.0 mIU/ml
automatically by the computer. Fluorescence is measured Second half (D-8 to D-2) 2.0 - 8.0 mIU/ml
twice in the Reagent Strip’s reading cuvette for each - Luteal phase (D+3 to D+15) 0.2 - 6.5 mIU/ml
sample tested. The first reading is a background reading Menopause: 8.0 - 33.0 mIU/ml
of the substrate cuvette before the SPR is introduced into
the substrate. The second reading is taken after PERFORMANCE
incubating the substrate with the enzyme remaining on Studies performed using VIDAS LH gave the following
the interior of the SPR. The RFV (Relative Fluorescence results:
Value) is calculated by subtracting the background
reading from the final result. This calculation appears on Measurement range
the result sheet. The measurement range of the VIDAS LH reagent
The results are automatically calculated by the instrument extends from 0.1-100 mIU/ml.
using calibration curves which are stored by the
instrument (4-parameter logistics model). The LH Detection limit
concentrations are expressed in mIU/ml (2nd IS 80/552). Defined as the smallest concentration of LH which is
Samples with a LH concentration greater than 100 mIU/ml significantly different from the zero concentration with a
should be retested after dilution to 1/2 or 1/4 in LH dilution probability of 95%: < 0.1 mIU/ml.
buffer (R1). Multiply the result by the dilution factor to
obtain the sample concentration. Hook effect
Interpretation of test results should be made taking into No hook effect was found up to LH concentrations of
consideration the patient’s history, and any other tests 10,000 mIU/ml.
performed.
Precision
QUALITY CONTROL
Within-run reproducibility:
A control is included in each LH VIDAS kit.
This control must be performed immediately after opening 5 samples were tested 30 times in a same run.
a new kit to ensure that reagent performance has not Sample 1 2 3 4 5
been altered. Each calibration must also be checked
using this control. The instrument will only be able to Mean concentration
1.6 7.1 25.9 37.8 66.1
check the control value if it is identified by C1. (mIU/ml)
Results cannot be validated if the control value deviates CV% 4.5 4.8 3.8 4.3 3.1
from the expected values.
Between-run reproducibility:
Note
5 samples were tested singly in 24 different runs on the
It is the responsibility of the user to perform Quality
same VIDAS over a period of 8 weeks.
Control in accordance with any local applicable regulation.
Sample 1 2 3 4 5
LIMITATIONS OF THE METHOD Mean concentration
1.7 7.2 26.3 33.3 64.7
Interference may be encountered with certain sera (mIU/ml)
containing antibodies directed against reagent
CV % 3.8 6.6 4.0 4.5 3.7
components. For this reason, assay results should be
interpreted taking into consideration the patient history,
and the results of any other tests performed. Specificity
Tested compound Cross-reactions (%)
LH 100
FSH < 0.1
TSH < 0.1
hCG < 0.01
hCG subunit < 0.01
bioMérieux SA English - 4
®
VIDAS LH (LH) 06267 O – en - 2015/01
Accuracy
Dilution test
Three samples were diluted in the LH dilution buffer (R1) and tested singly in 3 series. The ratio of the mean
concentration measured over the expected concentration is expressed as a mean recovery percentage.
Expected Mean measured Mean recovery
Sample no. Dilution factor concentration concentration percentage
(mIU/ml) (mIU/ml) (%)
1/1 30.3 30.33 100
1/2 15.2 15.19 100
1 1/4 7.6 7.27 96
1/8 3.8 3.51 92
1/16 1.9 1.76 93
1/32 0.9 0.84 93
1/1 39.9 39.91 100
1/2 20.0 20.92 105
2 1/4 10.0 9.76 98
1/8 5.0 5.14 103
1/16 2.5 2.53 101
1/32 1.2 1.30 108
1/1 43.9 43.95 100
1/2 22.0 24.50 111
3 1/4 11.0 11.57 105
1/8 5.5 5.84 106
1/16 2.7 3.06 113
1/32 1.4 1.44 103
bioMérieux SA English - 5
®
VIDAS LH (LH) 06267 O – en - 2015/01
Temperature limit
Use by date
Batch code
Date of manufacture
REVISION HISTORY
Change type categories :
N/A Not applicable (First publication)
Correction Correction of documentation anomalies
Technical change Addition, revision and/or removal of information related to the product
Administrative Implementation of non-technical changes noticeable to the user
Note: Minor typographical, grammar, and formatting changes are not included in the
revision history.
Release
Part Number Change Type Change Summary
date
INDEX OF SYMBOLS
Administrative REVISION HISTORY
2015/01 06267O CONTENT OF THE KIT (60 TESTS) – RECONSTITUTION
Technical OF REAGENTS
WARNINGS AND PRECAUTIONS
INSTRUCTIONS FOR USE
BIOMERIEUX, the blue logo, SPR and VIDAS are used, pending and/or registered trademarks belonging to bioMérieux, or one of its
subsidiaries, or one of its companies.
Any other name or trademark is the property of its respective owner.