You are on page 1of 1

Programming sheet of Beckman coulter AU680

Test Name **LIPASE Type; Serum Operation Yes

Sample Volume; 4 uL Dilution; 0uL `ODLimit

Pre Dilution rate; 1 Min OD -0.100 Max OD 2.5000

Reagent Volume R1(R1-1) ; 200uL Dilution 0uL Reagent OD Limit

First Low -0.1000 High 2.5000

R2(R2-1) 50uL Dilution 0uL Last Low -0.1000 High 2.5000

Common reagent type None Name ___ Dynamic Range Low # High #

Wave length. Pri. 570nm Sec 700nm Correlation Factor A 1 B0

Method FIXED Factor b Maker A 1 B0

Reaction slope + On board stability Period 30 days 0 hrs

Measuring point 1 First 14 Last 19

Measuring point 2 First - Last -

Linearity Limit 15%

Lag Time Check YES