The removal and replacement instructions for L/N parts are available in the CELL-DYN Ruby
Operations Manual.
ISA 170-003U, March 2016 ©2012, 2016 Abbott Laboratories, Abbott Park, IL 60064 Page 1 of 6
For Internal Use Only
CELL-DYN Ruby is a trademark of Abbott Laboratories in various jurisdictions.
CELL-DYN Ruby PM #1
Reminders ShortNote Picture/info on right CD Ruby Service Manual
General Attention Activator CD Ruby Operator’s Manual
SafetyWarning
2 Decontaminate Instrument
2.1 Flush flow cell and waste lines to WC-1
through V56 using 10% bleach. Let sit
minimum 5 minutes. Replace WC and
tubing from flow cell to WC-1 if damaged
3
2.2 Run 3 tubes 10% bleach through SL to
clean closed aspiration pathway and
decontaminate the blood pathways.
3 Fans
3.1 Inspect 4 external fans for proper operation.
Replace if non-functional & date 3 2.1 Flushing Optical Flow Cell
3.1a Replace fans every 2 years & date CD Ruby Flushing the Optic Flow Cell and WC1.wmv
3.2 Perform Page 469 - Clean Fan Filter p
9-49
4 Solenoids 6
4.1 In Diagnostic, close all Solenoids, VP-09
Solenoid Operation Verifications
4.2 Press on individual solenoids check that
they are fully closed. (Solenoids no
addition movement)
4.3 Note any solenoids that do click closed for
further action later
NOTE: Do this before replacing tubing
4.4 Inspect each solenoid for salt buildup,
corrosion or rust
4.5 Note any solenoids that show size of rust or
ISA 170-003U, March 2016 ©2012, 2016 Abbott Laboratories, Abbott Park, IL 60064 Page 2 of 6
For Internal Use Only
CELL-DYN Ruby is a trademark of Abbott Laboratories in various jurisdictions.
CELL-DYN Ruby PM #1
salt build up. 4 8.2 Bleaching Waste Chambers
5 Front Panel CD ruby cleaning WC3.wmv
6 Mixing Chambers
6.1 Inspect RBC & WBC mixing chambers for
corrosion, leaks or excessive buildup
6.2 Confirm metal inlet ports are not loose
7 Reservoir
7.1 Inspect for leaks
7.2 Confirm metal posts are intact & no salt
buildup
7.3 Confirm inlet tubing for diluent is not
crimped or restricted
7.4 Correct as needed
8 Waste Chambers
8.1 Inspect all WC for cracks, leaks, plumbed
correctly
8.2 If WC-3 has a lot of buildup, remove &
clean with bleach 4 5 9.1 Waste Line Check
NOTE: This may take 30 minutes or more
8.3 Replace WC as needed
9 Syringe Drives
9.1 Inspect and Clean Syringe Drives as
needed
10 Peristaltic Pump
10.1 Clean shoe and rollers
11 Solenoids 6
11.1 Blow out each solenoid with canned air
11.2 Clean or Replace damaged solenoids
identified in section 4
ISA 170-003U, March 2016 ©2012, 2016 Abbott Laboratories, Abbott Park, IL 60064 Page 3 of 6
For Internal Use Only
CELL-DYN Ruby is a trademark of Abbott Laboratories in various jurisdictions.
CELL-DYN Ruby PM #1
13.3 Replace any damaged component 6 4, 12, & 13 Solenoids , Tubing & Check Valves
14 Pressure Sensor CD Ruby Tubing Map
Kit Bag PN 9360024 9360025
14.1 Replace pressure sensor. Tubing PN 93476-01 93476-01
14.2 Turn instrument on and prime. Length (Inch) 1.3 1.5
Quantity 50 15
15 Vacuum /Pressure Subassembly Valve # 1-1 2-7 4-4 6-3 1-5
15.1 Inspect Vacuum & Pressure accumulators 1-2 3-1 4-5 6-4 2-3
for cracks, excessive salt buildup 1-4 3-2 4-6 6-7 2-8
1-7 3-3 5-1 9-1 3-4
16.1a Replace pressure accumulator 1
1-8 3-6 5-2 9-3 6-8
every two (2) years.
2-1 3-7 5-4 9-4 9-2
15.2 If needed, vacuum accumulators can be 2-2 3-8 5-5 9-5 9-7
checked by confirming the resistance 2-4 4-1 5-7 9-6 V3
between the 2 leads is more than 20 2-5 4-2 6-1 9-8
MegaOhms. 2-6 4-3 6-2
15.3 Perform VP-44 Vacuum Accumulator 1
and 2 Rinsing Procedure if needed
Kit Bag PN 9360239 9360240 9360241 9360242
15.4 Perform VP-16 Vacuum & Pressure Level
Tubing PN 93476-01 93476-01 93476-01 93476-01
Verification/Adjustment Length (Inch) 2.0 3.0 3.5 4.0
Quantity 1 1 1 2
16 Laser Bench Valve # 5-6 1-3 6-5 V2
V5
16.1 Confirm background within spec 1
16.2 Run at least 1 sample to confirm proper
scatter before performing optical Kit Bag PN 9360243 9360244 9360245
checks/alignment Tubing PN 93476-01 93177-01 93177-01
Length (Inch) 5.5 3.0 8.0
16.3 Verify Optical WBC CVs Quantity 1 1 1
referencing VP-18 Optics Bench Valve # 6-6 V1 V4
Alignment Procedure beginning at step
“Perform Verification” using diluted 7.0
SRP 8 7 14 HGB Flow Cell
16.4 If CVs are out of spec: Ruby VP 39 Flushing HGB Flow Cell.wmv
16.4a Perform complete VP-18 Optics
Bench Alignment Procedure
NOTE: : If CVs are within specs& no Diff
issue reported by customer, there is no need to
clean or align bench
17 Autoloader
17.1 If excessive diluent on platen, verify
operation of detent pins
17.2 Clean detent pins 9
17.3 Replace tube grippers every 24 months
17.4 Inspect cable and Tube Sensor board for
corrosion replace if needed 10
17.4a Clean or adjust Tube Sensor as
needed VP-58 Tube Sensor Adjustment
17.5 Clean Forward Indexer & Pawls. Confirm
spring action 11
17.6 Clean Sweep Arm Bushings. Do not oil
or lube 12
17.7 Inspect Spring on Sweep Arms. Replace
every 2 years or as needed F1.07
Autoloader Sweep Arm Spring
ISA 170-003U, March 2016 ©2012, 2016 Abbott Laboratories, Abbott Park, IL 60064 Page 4 of 6
For Internal Use Only
CELL-DYN Ruby is a trademark of Abbott Laboratories in various jurisdictions.
CELL-DYN Ruby PM #1
Replacement Procedure
17.8 Clean Bar Code Reader. Adjust position
as needed VP-11 Bar Code Reader
8 17.3 WBC Mean Channel & CV Specifications
Verification/Alignment
17.9 Perform the following VPs:
17.9a VP-23 Tower Unit Stop Solenoid
Verification
17.9b VP-24 Bar Code Spin Assembly
Verification
17.9c VP-25 Tube Height Sensors (S1/S2)
Verification
17.9d VP-26 Mixer Up/Down Verification
17.9e VP-27 Mixer Head Rotation
Verification
17.9f VP-28 Mixer Bladders Verification
Offset Specifications
17.9g VP-29 Rack Advance & Tube Sensors
Verification Pull the flow cell out of the beam path. Adjust the front mirror Y
17.9h VP-30 Cross Transfer Arms & Rack screw for minimum reading on the DMM. The end result
Sensors Verification should be a reading of < 50 mV for the OFFSETS for both the
17.9i VP-31 Mixer Bladders Pressure 0 and 10 degree photodiodes. With Diluent in the flow cell the
Verification/Adjustment reading should be <100 mV.
Talsico® Process Picture Maps™ & all associated intellectual property are owned by Talsico International, ABN 20 419 167 619, & subject to licensing Agreement
End of document
ISA 170-003U, March 2016 ©2012, 2016 Abbott Laboratories, Abbott Park, IL 60064 Page 6 of 6
For Internal Use Only