■ CMV-ACMV
To ensure the patient’s pro-
per adaptation once the
sedation effect has gone
and upon the patient’s
request, HORUS 4 adapts
the insufflated flow.
■ SIMV
This mode combines controlled and spontaneous cycles
in synchronization with the patient’s pulmonary activity.
■ PS-PEEP
The patient’s muscular fatigue is considerably reduced
thanks to a remarkable short response time of the
inspiratory and expiratory valves, and a very low expiratory
resistance.
■ MRV Insist on
Mandatory Rate Ventilation NIV key
Automatic adjustment of the pressure support
according to the patient’s respiratory rhythm, appeases
the latter’s ventilation. This is a useful mode for patient
weaning from long term ventilation treatment.
■ APRV-Bi level
Spontaneous airway pressure release ventilation
This pressure release mode enables better CO2 elimination
and a drop in the average intrathoracic pressure.
SPECIFICATIONS HORUS 4
Le Square
Applications Ventilation modes
Adult - Child - Infant Volume Controlled/Assisted Controlled....................................CMV - ACMV
Pressure Controlled/Assisted Controlled.............................PCMV - APCMV
Resuscitation – Intensive Care Units - Recovery - Synchronized Intermittend Mechanical Ventilation...............................SIMV
Post-operative Treatment – Emergency Units Spontaneous with Pressure Support and PEEP.............................PS - PEEP
Pressure Regulated Volume Control * ................................................ PRVC
Mandatory Rate Ventilation *..................................................................MRV
- Non contractual document: the manufacturer reserves the right to alter specifications -
Technical specifications Airway Pressure Realease Ventilation *................................................APRV
Electrical supply ..................................................... 100 to 240 V; 50 to 60 Hz Spontaneous with Guaranteed Minimum VT,*...........................PS - VT min
75 VA, Class I, Type B Non-Invasive Ventilation, NIV…………….........………..CMV – ACMV (NIV)
Medical gases..................................................................................2.8 - 6 bar ……………….………PCMV – APCMV (NIV)
Weight of ventilator................................................................................ 14 kg …………......................……………SIMV (NIV)
Weight of castor base ............................................................................24 kg ……………............………….PS – PEEP (NIV)
Dimensions, ventilator (H x W x D) ...............................200 x 430 x 440 mm
Dimensions, ventilator and castor base .....................1300 x 585 x 610 mm
Patient Monitoring
Performances Real-time curves
Tidal volume ............................................................................ 20 to 2000 mL Trends
Inspiratory flowrate ................................................................. 1 to 200 L/min Alarm history
Maximum inspiratory flowrate ................................................... > 200 L/min Ventilation parameters
Frequency ......................................................................... 5 to 80 cycles/min Resistance/Compliance
SIMV frequency................................................................. 1 to 60 cycles/min Auto PEEP
Inspiratory plateau................................................................... 0 to 60 % of Ti Plateau pressure
Inspiratory time ........................................................ 10 to 80 % of total time
PEEP .................................................................................... 0 to 25 cm H2O
Pressure support .................................................................. 0 to 35 cm H2O Alarms
Insufflation pressure.............................................................. 0 to 60 cm H2O
P max
Maximum inspiratory pressure ................................................... 90 cm H2O VT min/max
Pressure support slope ............................................. 50 to 150 cm H2O/sec FiO2 min/max
FiO2 ............................................................................................ 21 to 100 % Patient disconnection
Inspiratory flowrate trigger ...................................................... 0.1 to 5 L/min f min/max
Inspiratory pressure trigger ................................................. 0.5 to 5 cm H2O Apnea ventilation
Expiratory flow rate trigger ....................................................... 0 to 30 L/min V min/max
Inspiratory pause................................................................................ 0 to 5 s Power Supply failures
Expiratory pause............................................................................... 0 to 10 s Automatic alarm threshold setting
Sustained exhalation ...................................................................... 0 to 2 min
■ Straight-forward settings
• Two configuration screens
Both ventilation and monitoring dedicated screens are designed to
offer a more rational approach of settings. Furthermore, to provide
an acute analysis, HORUS 4 is easily equipped with its special
Ventilation Monitor ("MdV").
• Thanks to its direct access button design, you can immediately
select the patient, ventilation or patient interface type.
■ Back-up ventilation
• The option to set in PS-PEEP a guaranteed minimum breathing
rate and/or tidal volume provides the patient with optimum
protection against hypoventilation.
This is particularly useful in the ventilation-weaning phase or in
the post-operative recovery phase.
• A Pmaxi alarm and a VTmaxi alarm respectively avoid barotraumas
and volotraumas. Indeed, as soon as the previously set limits are
reached, this security alarm stops insufflation.
■ A User-friendly product
Compact and light, with both its suspended and fixed versions,
HORUS 4 perfectly fits with your environment.
The Ventil
Much
■ Curve analysis
Curve freezing function + cursors, with the option
to perform all of the necessary zooms.
ilation Monitor
h more than a screen
Monitoring display of
patient adaptation to
ventilator settings