Syringe Pump
TE 331
• Syringe yang dapat digunakan:
10, 20, 30, 50 mL ( Terumo ) dan Syringe Khusus
• Flow Rate Range:
0,1 – 300 mL/h untuk Syringe 10, 20, 30 mL
0,1 – 1200 mL/h untuk Syringe 50 mL
• Display range:
0,1 – 999,9 mL
• Occlusion Detection Pressure:
III : 800 mmHg 200 mmHg
II : 500 mmHg 100 mmHg
I : 300 mmHg 100 mmHg
• Alarm:
Occlusion, Nearly – Empty, Low Battery, Plunger/Clutch Disengaged,
Syringe Disengaged, AC/DC disconnection
Code Error pada TE 331
Standard Calibration Tools
Jigs untuk Kalibrasi TE 331
Mode Setting untuk User
Mode Setting untuk BME
Blok Diagram TE 331
Blok Diagram 2
User Setting Mode
( DIP Switch Normal Mode )
• Setting the Max Flow Rate
• Enabling or Disabling Delivery Limit Function
• Setting time nearly empty alarm
• Changing Date and Time
• Changing Syringe Brand
• History
Setting Max Flow Rate
1. Mesin dalam keadaan mati, tekan tombol ON/OFF & STOP
simultan display “8888””USEr””0”. Pd saat tampil “0”
putar Dial dan enter pswd “331”
2. Tekan tombol STOP Display “rAL1”, tekan kembali
tombol STOP Display “1200”. Putar Dial 1 s/d 1200 mL/h
3. Jika ingin membatalkan, tekan dan tahan tombol cmL
Display “rAL1”
4. Menyimpan nilai setting, tekan dan tahan tombol START
selama 1,5 second Display “good”.
5. Matikan unit
Enabling or Disabling D Limit Function
1. Mesin dalam keadaan mati, tekan tombol ON/OFF & STOP
simultan display “8888””USEr””0”. Pd saat tampil “0”
putar Dial dan enter pswd “331”
2. Tekan tombol STOP Display “rAL1”, tekan tombol
DISPLAY SELECT Display “SPC1”
3. Tekan tombol STOP Display ” dL on “ atau “ dL off “
4. Tekan tombol STOP untuk memilih nilai setting
5. Tekan dan tahan tombol START selama 1,5 detik, untuk
menyimpan nilai setting, Display “ good “
6. Display kembali menampilkan “ SPC 1 “
7. Matikan unit
Setting time nearly empty alarm
1. Unit dalam keadaan mati, tekan tombol ON/OFF & STOP simultan
display “8888””USEr””0”. Pd saat tampil “0” putar Dial dan enter
pswd “331”
2. Tekan tombol STOP Display “rAL1”
3. Tekan tombol DISPLAY SELECT, Display “rAL1”
4. Tekan kembali tombol DISPLAY SELECT, Display “nEAr”
5. Tekan tombol STOP, Display “ - - - -”, putar dial ( 3 s/d 30 menit )
6. Jika ingin membatalkan, tekan dan tahan tombol cmL Display “nEAr”
7. Menyimpan nilai setting, tekan dan tahan tombol START selama 1,5
second Display “good”.
8. Display kembali menampilkan “ nEAr “
9. Matikan unit
Setting Date and Time
1. Unit dlm keadaan mati, tekan
Setting Tombol ON Blinking
tombol ON/OFF + STOP +
CmL, Display “8888” Tahun RATE XXXX
2. Lepas tombol ON/OFF, sementara Bln D Limit XXXX
tombol STOP + DISPLAY
Tgl D Limit XXXX
SELECT masih di tahan, Display
”8888” “bAUd”, lepas semua Jam mL XXXX
tombol Menit mL XXXX
3. Tekan tombol DISPLAY SELECT, Tekan DISPLAY Tekan PURGE
Display “tISt” SELECT untuk untuk merubah
merubah tombol
4. Tekan tombol STOP, Display
“RATE” berkedip
5. Putar Dial untuk setting ( Tabel )
6. Jika ingin membatalkan, tekan dan
tahan tombol cmL Display
“nEAr”
7. Tekan START selama 1,5 detik
8. Matikan unit
Mengganti Merk Syringe
1. Unit dalam keadaan mati Tekan tombol STOP + DISPLAY
SELECT simultan, tekan Power ON/OFF
2. Lepas tombol ON/OFF, tahan tombol STOP + DISPLAY
SELECT, Display “8888” “ syr”, lepas semua tombol
3. Masukan code syringe yang tertera pd unit
4. Putar Dial sebagai selektor
5. Tekan tombol START selama 1,5 detik untuk menyimpan ke
memory, Display “good”
6. Tekan ON/OFF untuk mematikan unit
Merk S yringe Ukuran S yringe
Terumo Japan 10, 20, 30, 50 ml
Terumo Europe 10, 20, 30, 50 ml
Terumo USA 10, 20, 30, 50, ( 60 ) ml
BD Precise 20, 50
BD Plastipak/USA 10, 20, 30, 50, ( 60 ) ml
Boin SAFTI 10, 20, 30, 50 ml
Nipro 10, 20 ml ( Thailand )
20, 50 ml ( Japan )
Vaccine 10, 20, 30, 50 ml
Monoject 10, 20, 30, 50, ( 60 ) ml
B Braun Omnifix 10, 20, 30, 50, ( 60 ) ml
JMS 10, 20, 30, 50 ml ( Japan )
TOP 10, 20, 30, 50 ml
Dispo Van 10, 20, 50 ml
Zeneca Diprivan PFS 50 ml
Penta Ferte 10, 20, 30, 50 ml
Chirana 50 ml
B Braun Perfusor 20, 50 ml
Codan Perfusion 50 ml
Melihat History
1. Unit dalam keadaan hidup, tekan tombol STOP + START
secara simultan, sampai terdengar bunyi buzzer, Display
“Hnnn”
“nnn” menunjukan nomor history dalam setiap pemakaian (
001 s/d 500 )
2. Putar Dial untuk memilih nomor history yang akan di lihat
3. Setiap kali anda menekan tombol DISPLAY SELECT maka
akan tertera:
a. Tanggal di operasikan
b. Jam berapa di operasikan
c. Detail pengoperasian
d. Berada pada nomor berapa History tsb di simpan
4. Tekan tombol STOP untuk keluar dari mode History
Self Diagnostic
1. Set DIP Switch 1 = ON dan 2 = OFF
2. Tekan tombol ON/OFF + STOP simultan, Display
”888.8” “FACt”
3. Lepas semua tombol, Display ”diAg”
4. Gunakan tombol PURGE sebagai tombol selektor
items ( lihat Tabel )
5. Tekan tombol STOP untuk mendiagnosa items yang
dipilih ( lihat Tabel )
6. Tombol START untuk eksekusi setiap items
Tabel Self Diagnostic
Diagnosis items Description
1 Software and EEPROM Version
2 LCD elements
3 Switch
4 Buzzer
5 Setting Dial
6 Plunger/Clutch
7 Motor Rotation and Motor Rotation Sensor
8 Nearly Empty Detection Sensor
9 Syringe Detection Sensor
10 Occlusion Detection Sensor
11 AC/DC Detection
12 Nurse Call Function
13 Total Operation Time
14 Number of Motor Rotation
15 Load on the Motor
16 Total Number Alarm & Error
Initializing E2PROM
(Menghapus Data dari Memory & Kembali ke Factory Setting)
1. DIP Switch 1 = ON & 2 = OFF
2. Tekan tombol ON/OFF + STOP, Display “8888”
”FACt”. Lepas semua tombol Display “dIAg”
3. Tekan tombol DISPLAY SELECT, 3 x, Display ”CLEA”.
4. Tekan tombol STOP, Display E atau J (Eropa/Jepang)
5. Tekan dan tahan tombol PURGE, 1,5 detik u/ memilih E.
MulaiDisplay: 0 0 0 0 0 0 0 0 0 0
6. Untuk memilih versi Jepang, tekan tombol START, Display
JP
7. Finish, Display “good”
8. Matikan unit
Bagaimana ???
Prosedur Kalibrasi
Step 3
Step 4
Step 5
B. Kalibrasi Deteksi Fungsi Nearly Empty
Step 1 dan 2
Step 3
C. Kalibrasi Deteksi Fungsi Oklusi
( Step 1 dan 2 )
Step 3 dan 4
Step 5
Cek Fungsi setelah melakukan Kalibrasi
( Kembalikan DIP Switch ke Mode Normal )
A. Cek fungsi Kalibrasi Ukuran Syringe
( Step 1 )
Step 2
B. Cek fungsi Kalibrasi Deteksi Nearly Empty
Step 1
Step 2
Step 3
C. Cek fungsi Kalibrasi Deteksi Oklusi
Step 1 dan 2
Step 3
Step 4
Step 5
Periodical Check
( DIP Switch, Set pada Normal Mode)
Part to be Inspected Periodically