RADIANT WARMER
MODEL NO. YD-RW-07SCA+ BABY CRIB
SERVICE/OPERATION MANUAL
I. INSTALLATION :
(*Install by following below order and page 19)
STEP 5.
TAKE THE SCREW AND PAD FROM THE
UP-SHAFT. CONNECT HEATER AND UP-SHAFT.
TIGHT THE SCREW BACK TO THE UP-SHAFT.
STEP 4.
FIX THE CONTROL BOX ON THE UP-SHAFT.
STEP 3.
CONNECT UP-SHAFT AND BOTTOM-SHAFT.
STEP 2.
FIX THE BABY CRIB FRAME ON THE
BOTTOM-SHAFT (OPTIONAL ACCESSORY).
STEP 1.
CONNECT FIRST THE BOTTOM-SHAFT AND THE
BASE, LEAVE THE SPRING INSIDE OF THE
BOTTOM-SHAFT.
2. CONDITIONS (NORMAL) :
A) INFANT CARE ROOM’S CONDITION: ROOM TEMPERATURE IS USUALLY KEPT AT
24~28℃. 28℃ IS IN SUMMER; 24℃ IS IN WINTER. THE RELATIVE HUMIDITY IS
MAINTAINED AT 65%RH. AND GOOD AIR CIRCULATION IS REQUIRED.
2
YON DON
3. PREPARING :
A) HEIGHT ADJUSTMENT : HOLD THE LAMP
HOUSING AND PULL OUT THE SW ITCH TO
ADJUST TO THE NEEDED DEGREE (FIVE
SECTIONS AVAILABLE). RELEASE THE
SHAFT TO FINISH THE PROCESS.
NOTE : 1. KEEP 40 ~ 60 CM OF DISTANCE
BETWEEN THE PATIENT AND THE RADIANT HEAT SOURCE. ADJUST IT BY
REFERR ING THE HEATING SETTING AND ENVIRONMENT TEMP. 2. P LACE THE
PATIENT AT A SAFE-BEDDING AND PREVENT POSSIBLE FALLING RISK FROM
THE BEDDING.
C) MOVING DEVICE BY US ING TWO HANDS, HOLD THE STAND AND WARMER
HOUSING. (MAKE SURE THE HOUSING IS COOL BEFORE TOUCHING TO
AVOID SCALD).
D) BEFORE INSERTING INTO THE ELECTRICAL OUTLET, MAKE SURE THE POWER
SWITCH IS IN OFF POSITION.
F) THERE IS A TAG WITH THE CORRECT POWER VOLTAGE ON THE PLUG. PLEASE DO
NOT DAMAGE THE TAG AND MAKE SURE THE VOLTAGE OF THE ELECTRICAL OUTLET
IS INDICATED ON THE TAG. IF THE VOLTAGE IS MISMATCHED, THE MACHINE WILL BE
DAMAGED.
3
YON DON
THERE ARE TWO CONTROL MODES OF THIS CONTROLLER : SKIN CONTROL AND
MANUAL CONTROL. PRESS SKIN / MANU KEY FOR 3 SECONDS, INTO CHOSE SKIN
CONTROL OR MANUAL CONTROL. WHEN SWITCH ON, MICRO PROCESSOR IN THE
CONTROL BOX WILL AUTO SET INTO THE LAST CONTROL MODE YOU SET.
INCREASE / DECREASE
PUSH 3 SECONDS (OUTPUT POWER OF HEATER 1 ~ 99%)
TO SET OUTPUT POWER
4
YON DON
(THE ADJUSTABLE RANGE IS FROM 1 TO 99%). PLEASE REFER TO THE CHART OF NEXT
PAGE OF THE MANUAL MODE OF ITS HEATER OUTPUT, WHICH CAN GIVE YOU A
ROUGH IDEA TO SET IT (GENERALLY SETTING WILL BE 60~80%). AFTER COMPLETE
SETTING, PRESS SET KEY AGAIN TO LEAVE THE SETTING STATUS. NOTE : UNDER THIS
MODE, TEMP. LEVEL CANNOT BE PRECISELY CONTROLLED BUT CAN BE MONITORED
BY USING SKIN TEMP. PROBE
SELECT UNDER SKIN TEMP. MODE : PLUG IN THE SKIN PROBE FOR DISPLAYING
PROMPT INFANT SKIN TEMPERATURE ON THE CONTROL PANEL (MORE DETAILS
ABOUT SKIN SENSOR PROBE, SEE STEP E.). PRESS SET KEY FOR 3 SECONDS TO GO
INTO SETTING TEMPERATURE. USE △ OR ▽ KEY TO ADJUST IT (AVAILABLE RANGE
℃~40℃
IS 25℃ ℃, 36℃
℃~36.5℃
℃ IS THE MOST COMMONLY SELECTED). AFTER COMPLETE
SETTING, PRESS SET KEY AGAIN TO LEAVE THE SETTING STATUS. NOTE : A) SKIN TEMP.
COMPARE WITH PRACTICAL TEMP. IS ALWAYS LESS THAN 0.5℃℃~0.8℃℃. B) SKIN TEMP.
℃~42℃
DISPLAY RANGE : 25℃ ℃,BELOW 25℃ ℃ APPEAR 25--,OVER 42℃ ℃ APPEAR 42U。 。C)
UNDER THIS MODE, FURTHERMORE, THE CONTROLLER WILL MAINTAIN THE BABY
SKIN TEMPERATURE IN SETTING RANGE AUTOMATICALLY.
6
YON DON
E-1. A) INSERT THE SKIN SENSOR PROBE INTO THE SOCKET (THE PLUG MUST BE FULLY
INSERTED INTO THE SOCKET TO ENSURE A GOOD ELECTRICAL CONTACT).
B) CLEAN THE SKIN AT THE INTENDED PROBE SITE (BETWEEN THE UMBILICUS AND
ENSIFORM) WITH ACETONE (CH3COCH3) TO REMOVE SMEGMA EMBRYONUM AND SOIL.
C) ATTACH THE PROBE TO THE INFANT (MAKE SURE THAT THE FLAT METAL SIDE OF
THE PROBE HEAD IS IN COMPLETE CONTACT WITH THE INFANT) PROCESS ALONG THE
VENTRIMESON (THE MIDDLE LINE ON THE VENTRAL SURFACE) WITH NON-IRRITATIVE
ADHESIVE TAPE.
D) TEAR OFF THE STICKER OF THE HEATER REFLECTION PATCH TO STICK ON THE TAPE
(ON THE BACK OF THE SKIN PROBE). (USING HEAT REFLECT SLICE TO REFLECT THE
HEAT COMING FROM THE HEATER, ENABLING THE SKIN PROBE TO DO ACCURATE
MEASUREMENT OF SKIN TEMP.).
7
YON DON
F. ALARM
F.1. M-TIMER S.H.F. 15M :
A). UNDER SKIN <SERVO> MODE, HEATER CONTINUES ITS FULL AMOUNT OUTPUT
OVER 15 MIN. LED OF M-TIMER S.H.F.15M WILL ON. REMOVE IT BY PRESS MUTE KEY
AS WELL AS INSPECTS THE REASON.
B). SELECT UNDER MANUAL CONTROL MODE, CONTROL SYSTEM WILL AUTO ALARM
ONCE EVERY 15 MINUTES, LED OF M-TIMER S.H.F.15M WILL ON. PRESS MUTE KEY TO
REMOVE TONE, LED OFF.
F-2. PROBE FAIL :
A). IF SYSTEM FAILURE, LED OF SYSTEM FAILURE WILL ON. TURN OFF THE UNIT, AND
THEN TURN ON THE UNIT, IF IT CANNOT BE NORMAL OPERATION, PLEASE INFORM
YON DON DISTRIBUTOR OR OUR OFFICE DIRECTLY.
B). WHEN HEATER FAIL LIGHT ON, DISCHARGE THE TERMINAL END, CHECK IF THE
HEATER HAS DISCONNECTED.
C). IF POWER FAILURE OR POWER MALFUNCTION, WILL ALARM BY SOUND.
G. ILLUMINATION
RED SWITCH
(OUT OF WORKING)
SYSTEM ILLUMINATION
G-1. ON WORKING, PRESS LIGHT TO TURN ON/OFF THE LAMPS FOR ILLUMINATION,
LED OF LIGHT ON WILL ON/OFF. IF YOU TURN ON THE LAMPS BUT THE LED OF LIGHT
ON IS GO OUT, PLEASE CHECK THE LAMP OR SWITCH IS FAILURE OR NOT, IF ALL THE
CHECKS FAIL TO DETECT ANY MALFUNCTION. PLEASE CONTACT WITH YON DON
DISTRIBUTOR OR OUR OFFICE DIRECTLY.
G-2. OUT OF WORKING, TURN ON/OFF RED SWITCH BESIDE THE CONTROLLER PANEL
FOR ILLUMINATION.
9
YON DON
RESET
TONE / TIME
ON / OFF
H-1. PRESS TIME ON/OFF TO START TIMER, SOUND AT THE FIRST ONE-MINUTE AND
THEN CONTINUE TO SOUND AT A MULTIPLE OF 5 (5, 10, 15, 20…); AT THE SAME TIME,
TWO PILOTS IN THE MIDDLE OF ELAPSED TIMER FLASHING.
H-2. PRESS RESET FOR RE-TIMING & REVERTING TO ZERO
PRESS TONE ON/OFF TO TURN ON/OFF THE ALARM SOUND. LED OF TONE ON WILL
ON/OFF.
I. HIDDEN SETTING
10
YON DON
I-2 .ALARM LASTING TIME SETTING OF FULL HEATING TIME AND MANUAL MODE,
CONTINUE J-1) PRESS SET KEY, SKIN TEMP. APPEAR In. 2 AND HEATER OUTPUT%
APPEAR 15 (ORIGINAL RATE, IF PRESS KEYS TO LONG, THE RATE MAY BE CHANGED),
ADJUST IT BY PRESS △ ▽ (SETTING RANGE IS 1~99 MINUTES). AFTER SETTING,
PRESS SET KEY TO LEAVE THE SETTING STATUS AND TO HEATER OUTPUT RATIO
SETTING, (J-3).
I-3 HEATER OUTPUT RATE SETTING.
CONTINUE J-2) PRESS SET KEY, SKIN TEMP. APPEAR In. 3 AND ELAPSED TIMER APPEAR
00 (ORIGINAL RATE), ADJUST IT BY PRESS △ ▽ (SETTING RANGE –45 ~ 45, EACH 0.01
℃ INCREASE OR DECREASE 2). AFTER SETTING, PRESS SET KEY TO FINISH ALL
SETTING. SITUATION : THE SKIN TEMP. AND SET TEMP. IS KEEPING CONSTANT BUT
THE SKIN TEMP IS HIGHER/LOWER THAN SET TEMP.
VI. MAINTENANCE
1. MAINTENANCE :
A) MAINTENANCE MUST BE PERFORMED ON A REGULAR BASIS. A REGULAR ANNUAL
MAINTENANCE IS REQUIRED AFTER TWO YEARS OF OPERATION.
B) SUGGESTED CLEANING AND DISINFECTING SOLUTIONS ARE SHOWN BELOW :
a. 0.2~0.5% BENZALKONIUM CHLORIDE SOLUTION.
b. 0.2~0.5% BENZETHONIUM CHLORIDE SOLUTION.
c. 0.02~0.5% CHLORHEXIDINE SOLUTION.
TOOTH PASTE CAN ALSO BE USED TO CLEAN THE ADHESIVE SUBSTANCE LEFT BY
ADHESIVE TAPE/PLASTER.
CAUTION : 1) THE USE OF ALCOHOL FOR CLEANING AND DISINFECTING IS STRICTLY
PROHIBITED. ALCOHOLIC SOLUTIONS CAN DAMAGE THE PLEXIGLASS COMPONENTS.
2) DO NOT USE UNDILUTED OR UNEVENLY DILUTED SOLUTIONS TO CLEAN THE UNIT.
WIPE OUT THE LEFT OVER SOLUTIONS WITH A WET, SOFT, CLEAN CLOTH TO MAINTAIN
THE QUALITY OF THE UNIT AND EXTEND ITS LIFE TIME. 3) STERILIZATION : IN
OCCASIONS, INTENSIVE CARE ROOM STERILIZATION IS NEEDED. UV LAMP CAN BE
USED TO ACHIEVE THIS TASK. CONSULT YOUR DEALERS FOR YD-P-616UV / 626UV/
626UVMP PRODUCT IF YOU ARE CONCERNED ABOUT STERILIZATION.
11
YON DON
IV. APPENDIX
1. CIRCUIT DESCRIPTION.
2. ASSEMBLY DIAGRAMS, STARTING AFTER CIRCUIT DESCRIPTION.
Contact Us
YON DON ENTERPRISE CO., LTD.
Website: http://www.yondon.com
2F, No. 34, Lane 80, Section 3, Nan-Kang Road, Nan-Kang,
Tel: 886-2-2785-9800.
Fax: 886-2-2786-5539.
E-Mail : info@yondon.com
Warranty Service. If the system has been working until recently, please contact your agent (the
company that sold the system to you) first. We will be more than glad to help should you
require further assistance after contacting your local agent.
Thank You
12
YON DON
13
YON DON
14
YON DON
15
YON DON
16
YON DON
17
YON DON
18
YON DON
19
YON DON
System Board
Display Board
20