Follow the below mentioned steps to set the parameters, run the machine.
Press PROG key to enter the program mode and ensure ‘Prog’ on display for 2
seconds.
Ensure ‘A 0.0’ on display. A to E is the identification for the five programs, and also it
indicates that you are in the program mode.
After the last (E) program, press PROG key to get out from the program schedule
mode.
During SET mode, after releasing the SET key only it will get out from the Program
schedule mode.
Enter the Program schedule mode. Go to the program, from which the pump should
start to run.
Run
Press Stop key and ensure the RUN LED flashes.
After the target volume/ target time reached, pump will run in KVO mode. i.e. AL 2 and
“0.1vo” will alternatively display and pump will run in the set KVO flow rate.
Set only the last programs 4 {B, C, D and E} (or) 3 {C, D and E} (or) 2 {D & E}
(Or) 1 {Only E}, skip the initial programs by simply pressing the PROG key.
2. Press PCA key, ‘Prs’ and the ‘Number of pressing of PCA switch’ will be d
displayed alternatively.
3. Again press PCA key, ‘Act’ and the ‘Number of Activation of PCA’ will be
displayed alternatively.
4. Again press PCA key, the time duration between two pressings of PCA switch
will be displayed and ensure Hr.mt LED glows. The flashing of display indicates the
activation of PCA, constant display indicates the pressing of PCA and no activation was
there. For the next 20 pressing, the duration of the last 20 pressing will be displayed.
5. To get out from this view mode, before viewing 20 intervals, Press any other key.
Press BUZZER OFF key and then press SET Key simultaneously for 2 seconds.
Ensure display shows the volume level from OFF-9 Followed by letter “b”.
After pressing the Decimal increment key one time, Ensure display shows the
Occlusion level form OFF-9 followed by letter “O”.
After pressing the Decimal increment key two times, Ensure display shows the
Default data mode ON/OFF, followed by letter “d”.
Note:
Default ON – When we switch off and on the pump, pump will on with default
(Previous running) data.
After pressing the Decimal increment key three times, Ensure display shows the
Bolus mode A/P, followed by letter “bo”.
After pressing the Decimal increment key four times, Ensure display shows the
Kvo flow rate from 0.0 – 9.9 (mL/Hr), followed by letter “vo”.