You are on page 1of 7

Original article

Performance evaluation of the Sysmex haematology

J Clin Pathol: first published as 10.1136/jclinpath-2012-200930 on 31 July 2012. Downloaded from http://jcp.bmj.com/ on 9 April 2019 by guest. Protected by copyright.
XN modular system
Carol Briggs,1 Ian Longair,1 Punamar Kumar,1 Deepak Singh,2 Samuel J Machin1
1
Department of Haematology, ABSTRACT flow for red cell count, haematocrit and impedance
University College London Background The Sysmex XN haematology instrument platelet count. Fluorescence flow cytometry is
Hospital, London, UK
2
Department of Haematology,
performs automatic reflex testing, depending on sample used for leucocyte differential, NRBC, reticulocytes
The Doctors Laboratory, results. A nucleated red blood cell (NRBC) count is and fluorescence platelet count. Information on
London, UK provided on all samples. The instrument has a smaller immature cells is derived from the immature
footprint (34%) than previous Sysmex XE analysers. myeloid information (IMI) channel.
Correspondence to Methods An evaluation comparing all results to the Over recent years software upgrades have been
Carol Briggs, Department of
Haematology, University College Sysmex XE-2100 and manual microscopic differential and introduced; the first allowed reliable automated
London Hospital, Haematology morphology (n=390) was performed followed by a counting of immature granulocytes ( promyelo-
Evaluations, 5th Floor, 60 workflow study of 1000 samples to compare speed of cytes, myelocytes and metamyelocytes) in the dif-
Whitfield Street, London W1T operation and number of blood films reviews required ferential channel.2
4EU, UK;
carolbriggs@hotmail.com
from both systems. Later a new automated method to quantitate
Results The new features on the instrument are: (1) reticulated platelets, expressed as the immature
Accepted 18 June 2012 white cell and NRBC channel, all samples include the platelet fraction (IPF), was introduced.4 5 The IPF is
Published Online First NRBC count; (2) white cell precursor channel: false identified by flow cytometry with the use of a
31 July 2012 positive flags for blasts, abnormal lymphocytes and nucleic acid specific dye in the reticulocyte/optical
atypical lymphocytes are reduced significantly without a platelet channel. The clinical utility of this param-
statistical increase of false negatives; (3) low white cell eter has been established in the laboratory diagnosis
count mode: suggested setting of <0.5×109/l. An of thrombocytopenia due to increased peripheral
extended count is more precise and provides an accurate platelet destruction, particularly autoimmune
differential. Fluorescent platelet count is performed in a thrombocytopenic purpura and thrombotic throm-
dedicated channel. If the red cell or platelet size bocytopenic purpura, and as a predictor of platelet
histograms are abnormal or if the platelet count is low, recovery following haematopoietic progenitor
then a fluorescent platelet count is automatically cell transplantation.6 The latest software to be
performed. Good correlation with the XE-2100 and introduced was for the measurement of Ret-He,
manual differential was found and the improved results reticulocyte haemoglobin concentration. Ret-He is
compared to the reference flow cytometric analysis for a measure of the forward scatter of stained reticulo-
platelet counts, especially below 30×109/l (XE-2100, cytes and provides an indirect measure of the iron
R2=0.500; XN, R2=0.875). available for new red blood cell production over the
Conclusion The XN showed reduced sample previous 3–4 days. It also provides an early measure
turnaround time of 10% and reduced number of blood of the response to iron therapy, increasing within
films for examination, 49% less than the XE-2100 2–4 days of the initiation of intravenous iron
without loss of sensitivity with more precise and therapy.7
accurate results on low cell counts. The Sysmex XE-5000 was launched in 2007.This
analyser performs with blood cells and leucocytes
in the same way as the XE-2100 but with added
INTRODUCTION parameters. The instrument measures the haemo-
There is an increasing demand for haematology globin content of individual red cells, calculates the
tests with reduced turnaround times, along with percentage of hypochromic red cells (%Hypo He)
cuts in budgets for pathology laboratory. There and the percentage of hyperchromic red cells
have been advances in haematology instrumenta- (%Hyper He) and quantifies the proportion of
tion which have resulted in faster throughput. microcytic and macrocytic erythrocytes (%Micro
Abnormal cells that were previously only indicated R) and (%Macro R). The availability of extended
by an abnormal cell flags, such as immature granu- red cell parameters should allow earlier diagnosis of
locytes and nucleated red blood cells (NRBC), can abnormal iron metabolism and the response to iron
now be reliably enumerated on several top of the or folate supplementation.8
range analysers,1–3 leading to a reduction in the The XN Modular analyser was introduced in
number of microscopic counts. All haematology 2011. The instrument used in this evaluation was
analysers generate suspect flags in the presence of a prototype and there has been a software upgrade
abnormal cells, but false positive rates are high, for the instrument after this evaluation before
leading to unnecessary blood film reviews. The release to the market. The red cell parameters and
aim of any manufacturer is to increase specificity impedance platelets are measured in the same way
without losing sensitivity. as previous XE series instruments, but several new
The Sysmex XE-2100 (SYSMEX, Kobe, Japan) channels have been introduced: white cell
was introduced in 1999.1 It is an automated nucleated channel (WNR); white cell differential
haematology analyser using direct current sheath channel (WDF); white cell precursor channel

1024 J Clin Pathol 2012;65:1024–1030. doi:10.1136/jclinpath-2012-200930


Original article

(WPC); and fluorescent platelet channel (PLT-F). The reticulo- frequently there was insufficient sample for further testing for

J Clin Pathol: first published as 10.1136/jclinpath-2012-200930 on 31 July 2012. Downloaded from http://jcp.bmj.com/ on 9 April 2019 by guest. Protected by copyright.
cyte (RET) channel with related immature RET parameters this evaluation.
remains the same as on the XE series of instruments, including For the workflow study samples, 1000 were selected at differ-
an optical platelet count if desired. ent times of the day and on different days of the week to try
Using the WNR, a NRBC count is performed on all samples; to mimic 1 day’s workload. UCLH receives a high proportion
there is no longer a need to repeat the count in the NRBC of abnormal samples as a tertiary reference centre.
channel as before. Samples were analysed on the XE-2100 and the XN modular
The WDF includes the basophil count; there is no longer a system in Complete Blood Count and DIFF mode, with auto-
separate basophil channel and the separation of lymphocytes matic reflex testing on the XN as required.
and monocytes has been optimised. All NRBC are measured by
light scattering and fluorescence; impedance is no longer used Principle of leucocyte differential on the XE-2100
for white blood cells (WBC). The optical system on the XE-2100 employs a stable red diode
The WPC channel is only used for reflex testing when a laser producing a light beam of 633 nm wavelength and a poly-
sample shows a blast/abnormal lymphocyte flag. The flags will methine based fluorescent dye. Three signals are produced:
be separated into the specific flags of blast or abnormal forward scattered (FSC) light, providing information on cell size;
lymphocyte or not shown at all if found to be false positive side scattered light (SSc), providing information on internal cell
from the WDF channel; this reflex is automatic. structure; and side fluorescence (SFL), providing information on
The PLT-F channel can be selected for testing on any sample DNA/RNA content. The basophil count is derived in the WBC/
or only used as a reflex test if there are abnormal red cell or BASO channel, where basophils are resistant to the reagent
platelet size histograms or if the platelet count is below a system and retain their size and shape, forming a cluster of larger
pre-set limit, determined by the user, when an extended plate- cells distinct from other non-basophil nucleated cells whose
let count is performed to increase accuracy and precision. The membranes are perforated and cytoplasm is lost.
PLT-F is in a new dedicated channel using a RNA dye; it also In the IMI channel a combination of radio frequency and
quantitates the IPF, which is no longer performed in the RET impedance, direct current resistance methods are used in conjunc-
channel as on the XE-2100. tion with special reagents. Immature WBC contain less lipid
There is also a low WBC mode where samples with a WBC then mature cells. The effect of the lysing reagent differs with
of 0.5×109/l or less are reflexed for an extended count and a each type of immature cell, allowing quantitative differentiation.
leucocyte differential is reported; a differential is not reported NRBC are measured in a dedicated channel. This is per-
on WBC <0.5×109/l on the XE-2100. formed by flow cytometry and a polymethine dye, which
The aim of this study was to evaluate the performance char- stains the intracytoplasmic organelles and the nucleus of WBC
acteristics of the new methods, carryover, linearity and sample quite strongly while the staining of NRBCs is weak. This
stability as well as compare the results for all parameters to allows clear discrimination between the counts.
the XE-2100 and abnormal cell flagging with a manual diff-
erential and morphological examination and comparison of
platelet counts to the International Society for Heamatology/ Principle of measurements on the XN
International Council for Standardization in Haematology Fluorescent labelling of cells takes place after perforation of the
(ISLH/ICSH) reference flow cytometric method 9 with particular cell membrane caused by specific lysing reagents. After this a
regard to counts <20×109/l. polymethine dye enters the cell and binds to nucleic acid and
A further study was performed to evaluate the time for ana- bioreactive proteins in the cytoplasmic organelles.
lysis, number of repeats on the XE-2100 and reflex tests on the Platelet and IPF are stained by a fluorescent dye based on
XN. The number of abnormal cell flags on both instruments oxadine: SFL light which gives information on DNA/RNA
were also compared which would result in an examination of a content of the cell; side scattered light which gives information
blood film in the routine laboratory. One thousand samples, on intracellular structure of the cell; and FSC light which gives
selected to mimic the samples seen in our laboratory on a daily information on cell size. All three methods are used in the four
basis, were analysed. new channels which are the WNR, WDF, WPC and PLT-F
The ideal result would be faster throughput and an increase (figure 1) as well as the RET channel which is the same as on
in specificity of the flags without loss of sensitivity, and more the XE series of instruments.
accurate and precise results on low WBC and platelet counts.
Abnormal cell flags
Abnormal cells have differing characteristics, such as cell size,
PATIENTS AND METHODS nuclear size and granule content, from normal cells. In the
Patients presence of abnormal cells most instruments will generate an
For the comparative study, 390 residual K2EDTA anticoagulated abnormal or suspect flag. The instrument detects the cloud for
(Beckton Dickinson, San Jose, California, USA) samples were a particular cell population as having an abnormal size or shape
selected from the routine haematology laboratory at University by cluster analysis. The instrument will also detect events
College London Hospital (UCLH) Haematology after all routine outside the normal cell population areas. The flags are gener-
testing had been completed. Samples included 131 normal ated by pattern abnormalities from the DIFF channel and the
samples; the rest comprised various haematological diseases, IMI channel on the XE-2100, and from four different channels
such as leukaemias, lymphoma, myeloma, haemoglobinopa- on the XN. WNR flags for platelet clumps, abnormal NRBC
thies, idopathic thrombocytopenic purpura, thrombotic throm- and abnormal WBC.
bocytopenic purpura and interfering substances such as lipids The WDF channel flags for atypical lymphocytes and blast/
and bilirubin. Due to small sample volumes no paediatric abnormal lymphocytes.
samples were included in the study; although the XN only PLT-F flags the presence of platelet clumps and abnormal
requires 150 ml after testing in the routine laboratory, platelets.

J Clin Pathol 2012;65:1024–1030. doi:10.1136/jclinpath-2012-200930 1025


Original article

J Clin Pathol: first published as 10.1136/jclinpath-2012-200930 on 31 July 2012. Downloaded from http://jcp.bmj.com/ on 9 April 2019 by guest. Protected by copyright.
Figure 1 (A) The white cell nucleated channel (WNR) scatterplot showing the position of white blood cells (WBC), basophils (BASO), nucleated
red blood cells (NRBC), platelet clumps and ghosts. Side fluorescence (SFL) on the x-axis and forward scatter (FSC) on the y-axis. (B) The white cell
differential channel (WDF) scatterplot showing the position of neutrophils with basophils (NEUT+BASO), lymphocytes (LYMPH), monocytes (MONO),
eosinophils (EO), BASO, immature granulocytes (IG) and NRBC. The positions of atypical lymphocytes, blasts or abnormal lymphocytes are also
shown. Side scatter (SSC) on the x-axis and SFL on the y-axis. (C) The WPC scatterplot showing the position of abnormal lymphocytes, atypical
lymphocytes, NRBC, granulocytes, lymphocytes monocytes and blast cells. SSC on the x-axis and SFL on the y-axis. (D) The white cell precursor
channel (WPC) scatterplot showing the position of abnormal lymphocytes, atypical lymphocytes, granulocytes, lymphocytes, monocytes and blast or
basophils cells. SSC on the x-axis and FSC on the y-axis. (E) The fluorescent platelet channel (PLT-F) scatterplot showing the position of platelets
(PLT-F) IPF, WBC and RBC. SFL (SSC) on the x-axis and FSC on the y-axis. This figure is only reproduced in colour in the online version.

The WPC channel is only used when a reflex test is triggered WBC. Instrument generated abnormal cell flags were compared
in the presence of the abnormal lymphocytes/blasts. These with the morphological findings of the blood films.
two abnormal cells are separated into separate flags or no Platelet counts were compared to the ISLH/ICSH9 reference
longer shown if it was a false positive flag from the WDF flow cytometic method on samples, range 1–1728×109/l
channel. (n=185); 67 samples were at, or below, the platelet transfusion
threshold used at our hospital of 20×109/l (selected by the
Calibration and quality control impedance count on the XE-2100).
The XE-2100 and XN analysers were calibrated by their respect-
ive engineers before the study started. Manufacturer’s quality Low WBC mode
control material was run on both instruments on a daily basis. Thirty samples were selected and analysed on the XN and by
For the XE 2100 this was normal level of e-CHECK and for the manual differential. The samples were then diluted to a WBC
XN normal level of XN check. value less than 0.5×109/l. The samples were then re-run on the
XN; the low WBC mode was automatically activated. The dif-
Comparative study ferentials from these results were compared to the manual dif-
All results were compared to the XE-2100. ferentials and the automated counts on the whole blood and
Blood films were examined for every sample in accordance diluted blood.
with the recommended manual 400-cell differential from the
Clinical and Laboratory Standards Institute (CLSI) protocol Workflow study
H20-A2, 2007,10 including NRBC, and results compared with This was undertaken to evaluate the time saving possibilities
those from the XE-2100 and the XN. The absolute differential of using the XN in the routine laboratory. For the purpose of
counts for each class of leucocyte were obtained by multiplying this study any sample with an abnormal cell flag, or incom-
their relative microscopic percentages by the total XE-2100 plete differential results, was considered to need a manual

1026 J Clin Pathol 2012;65:1024–1030. doi:10.1136/jclinpath-2012-200930


Original article

blood film review. Quantitative flags alone, or numerical results Table 1 Correlation statistics of complete blood count results,

J Clin Pathol: first published as 10.1136/jclinpath-2012-200930 on 31 July 2012. Downloaded from http://jcp.bmj.com/ on 9 April 2019 by guest. Protected by copyright.
outside the normal range, were not considered to need a including nucleated red blood cell (NRBC), reticulocytes, immature
manual review. This allowed the assessment of the changes in granulocytes and immature platelet fraction (IPF%); XN compared to
flagging on the total number of blood films made in the labora- XE-2100
tory on a typical working day. The number of repeat tests Parameter R2 value Slope Intercept
needed on the XE-2100 and the automatic reflex tests on the WBC 0.99 1.07 −0.76
XN were recorded as well as the time taken for analysis on RBC 0.99 1.05 −0.24
each instrument. Hb 0.99 0.99 −0.05
HcT 0.99 1.04 −1.90
Carryover MCV 0.99 0.99 +0.08
Carryover was estimated following the method of Broughton MPV 0.87 0.99 +0.34
et al11 as recommended by ICSH. This was performed for PLT-I vs PLT-F 0.98 1.06 −13.70
PLT-O vs PLT-F 0.99 1.06 −1.50
WBC, NRBC, PLT-F and IPF, which are the methods changed
Neutrophils 0.99 1.00 −0.02
on the XN, by measuring a patient sample with high values in
Lymphocytes 0.99 1.14 −0.3
triplicate followed by a sample with low values in triplicate.
Monocytes 0.91 0.90 +0.09
Eosinophils 0.99 0.98 0.00
Precision Basophils 0.76 1.15 +0.01
Precision was performed on WBC, NRBC, PLT-F and IPF, the NRBC 0.99 1.04 0.00
changed methods on the XN. Low and normal values were Immature granulocytes 0.94 1.14 +0.02
used where possible for WBC, PLT-F and IPF and varying levels Reticulocytes 0.98 1.00 +1.32
for NRBC. Samples were tested 10 times according to the Ret-He 0.94 0.99 +0.71
ICSH guidelines.12 IPF 0.60 0.58 +1.0.
IPF, immature platelet fraction; PLT-F, fluorescent platelet channel; Ret-He, reticulocyte
Linearity haemoglobin; WBC, white blood cells.
This was performed on WBC, NRBC, PLT-F and IPF; both high
counts and low counts were examined. Linearity in the low
For platelet counting results for all samples, results were
range is very important for all parameters. Samples were
similar between instruments for the impedance method com-
diluted and analysed twice; the mean of these results were
pared to the reference method (XE-2100, R2=0.986; XN,
plotted against the calculated expected value for the dilution.
R2=0.972). The PLT-F method was superior to the XE-2100
optical method when compared to the flow cytometric refer-
Stability ence method on samples with a platelet count of ≤30×109/l
Stability was performed on all new measured parameters, as (XE-2100, R2=0.500; XN, R2=0.875); results are shown in
above. Samples were stored at room temperature and 4°C. figure 3.
Samples were tested immediately and 24, 48 and 72 h. All
samples were aliquoted and those stored at 4°C were warmed
for 15 min before testing. Workflow study
One thousand samples analysed on both analysers resulted in
a reduction of 49% of blood films for review when using the
RESULTS
results from the XN. This is due to a reduction in the blast,
Comparative study
abnormal lymphocyte flag and atypical lymphocyte flag.
Results for parameters compared to the XE-2100 are shown in
Results are shown in table 4. More re-runs/reflex tests were
table 1. All results were excellent apart from the IPF%. This
needed on the XN (3.7% more than the XE-2100) but this is
was caused by eight outlying samples. These had very high
preferable because of the 49% reduction in the number of blood
results on the XE-2100 but less than 2% on the XN.
films that would be reviewed on a daily basis compared to
Six samples were from patients on chemotherapy and the
results from the XE-2100. Turnaround time was shorter on the
other two had aplastic anaemia.
XN by 10% and if the time is included for the making of blood
Although there is no reference method for IPF it could be
films on the Sysmex SP 1000 (automated blood film and
assumed, using the clinical details from the patients, that the
stainer), this is reduced further.
XN results from the PLT-F channel are more correct. Apoptopic
WBC are sometimes stained by the RET stain on the XE-2100,
giving falsely high results for IPF%; an example is given in Low WBC mode
figure 2. This was from a patient with acute promyelocytic leu- Differentials on the diluted samples with a WBC count of
kaemia on chemotherapy with a WBC of 1.6×109/l. The <0.5×109/l showed no difference compared to the undiluted
optical platelet count from the RETIC channel on the XE-2100, samples when compared to the manual 400 cell differential;
where the IPF is also measured, is also falsely high compared to results are shown in table 5.
the flow cytometric reference counting method.
Results compared to the reference differential method are
Carryover
shown in table 2.
Results for all parameters tested were in the range 0.00–0.07%.
The flagging performance compared to microscopic morph-
No carryover was detected for any parameters tested.
ology comments, of the XN was greatly improved. For atypical
lymphocytes this was seen even when no reflex test was per-
formed to the instrument with the WPC channel. Results are Precision
shown in table 3.There is a 20% reduction in the blast/abnor- All results for all parameters within the normal values were
mal lymphocyte flag and 20% for the atypical lymphocyte flag. below manufacturer’s specifications.

J Clin Pathol 2012;65:1024–1030. doi:10.1136/jclinpath-2012-200930 1027


Original article

J Clin Pathol: first published as 10.1136/jclinpath-2012-200930 on 31 July 2012. Downloaded from http://jcp.bmj.com/ on 9 April 2019 by guest. Protected by copyright.
Figure 2 (A) Fluorescent platelet channel (PLT-F) scattergram from the XN of a patient with acute promyelocytic leukaemia. The impedance platelet
count and PLT-F agree with the reference flow cytometric method and the immature platelet fraction (IPF%) is 1.1. (B) The optical platelet
scattergram from the same patient. The impedance platelet count agrees with the reference flow cytometric method, but the optical count is falsely
high due to apoptotic/fragmented white blood cells (WBC). This has also caused a falsely high IPF%. This figure is only reproduced in colour in the
online version.

For low WBC counts, n=9 (range 0.66–1.21×109/l), the coef- as necessary, WPC and PLT-F and reticulocytes can be included,
ficient of variation (CV%) was on average 8.6%. or excluded, on any analyser as necessary, thereby saving
For low platelet counts, PLT-F, n=16 (range 9–39×109/l), the reagents when not required.
CV% average was 4.0%. With needs for faster turnaround times it is necessary to
For IPF normal counts, n=7 (range 1.25–5.88%), CV% was reduce the number of blood film reviews performed in the
6.8%, for high IPF% values; n=4 (range 12.6–34.29%), the CV laboratory, the challenge is to reduce them without missing
% was 4.5%. important diagnostic information. Using current haematology
For NRBC counts samples were tested on low (range 1–5/ analysers, 80% of blood film reviews are triggered by abnormal
100 WBC), medium (range 5–20/100 WBC) and high counts cell flags from the haematology analyser13 and 62% are due to
(greater than 20/100 WBC). false positive flags.14
For the low range the average CV% was 32%, medium range Comparative results against the Sysmex XE-2100 were all
7.4% and above 20/100 WBC, 8.7%. good apart from the IPF%. There is no current reference
method for the IPF, but due to the diagnosis of the eight
Linearity patients showing discrepant results compared to the XE-2100
Results for linearity for both high and low counts for WBC (low on the XN and high on the XE-2100) it may be that the
and PLT-F, IPF% and NRBC can be seen in table 6. results were more correct on the XN. Six patients were under-
going chemotherapy and two had aplastic anaemia. It has been
previously documented that apoptotic WBC or WBC fragments
Stability
interfere with optical platelet counts.15 Figure 2 clearly demon-
Results for the parameters tested, WBC, leucocyte differential,
strates a falsely high optical platelet count and calculated IPF%
NRBC, PLT-F and IPF% were all within precision limits over
on the XE-2100 compared to the XN and reference flow platelet
72 h at both room temperature and 4°C.
flow cytometric method. This patient was on chemotherapy

DISCUSSION
The XN is a user defined platform from a single instrument to
Table 3 Efficiency of abnormal white blood cells flags on the XE-2100
a line up of up to nine modules, so laboratories can adapt the
and XN in 390 patient samples
system to their needs. Analytical requirements can be adopted
Blast True False True False
cell flag positive positive negative negative Sensitivity Specificity
XE-2100 20 68 302 0 100% 82%
Table 2 Correlation statistics of differential results, including nucleated XN 19 20* 350 1 95% 99.7%
red blood cells (NRBC) and immature granulocytes; XN compared to Abnormal lymphocytes
the reference 400-cell differential XE-2100 4 30 352 4 50% 99%
Parameter R2 value Slope Intercept XN 3 8* 374 5 37.5% 98.7%
Neutrophils 0.98 0.97 −0.06 Atypical lymphocytes
Lymphocytes 0.99 1.10 −0.24 XE-2100 8 82 285 15 34.8% 95%
Monocytes 0.66 1.29 +0.05 XN 5 9* 358 18 78.3% 95.2%
Eosinophils 0.94 1.06 +0.02 XN 5 9* 358 18 78.3% 95.2%
Basophils 0.70 1.54 +0.03 without
WPC
Immature granulocytes 0.69 1.75 +0.08
NRBC 0.97 0.97 −0.13 *p<0.001.
WPC, white cell precursor channel.

1028 J Clin Pathol 2012;65:1024–1030. doi:10.1136/jclinpath-2012-200930


Original article

J Clin Pathol: first published as 10.1136/jclinpath-2012-200930 on 31 July 2012. Downloaded from http://jcp.bmj.com/ on 9 April 2019 by guest. Protected by copyright.
Figure 3 Results for the optical and
fluorescent platelet channel (PLT-F) on
samples selected with a platelet count
of 20×109/l using the impedance
count on the XE-2100. (A) Correlation
graph of the XE-2100 optical platelet
count compared to the reference flow
cytometric method. (B) Correlation
graph of the XN PLT-F count compared
to the reference flow cytometric
method.

for acute promyelocytic leukaemia. A reference method for reti- flag from 82 on the XE-2100 to nine on the XN without a stat-
culated platelets/IPF is clearly needed. istical increase in false negatives.
Results for the differential compared to the reference method For the workflow study abnormal cell flags for all cells, WBC,
were good. Monocytes and immature granulocytes show RBC and platelets, were used and the number of blood films
slightly less good correlation (R2=0.66 and 0.69, respectively; that would need to be examined using the rules previously
table 2); this was due to eight samples with thalassaemia described on 1000 samples were compared. Our findings show
major or intermedia with high numbers of NRBCs, 621.13 per that the number of blood films would be reduced from 199 on
100/WBC–69.2 NRBC per 100 WBC. All outliers were rejected the XE-2100 to 101 using the XN the WPC channel (49% fewer)
differentials on the XE-2100; no results reported. If the outliers but with an increase in reflex testing of 37 samples on the XN
are removed, correlation for monocytes is 0.74, and for imma- compared to repeat tests needed on the XE-2100 (table 4) in the
ture granulocytes is 0.89. On this prototype analyser the sup- 1000-sample set analysed. In the laboratory this automatic
pressed differential was not activated and on future versions reflex testing is preferable to the number of blood films exam-
these types of samples will not have the differential reported; ined microscopically. Nearly all the reduction in blood films is
however this would mean a manual differential will have to be due to the increased specificity of the blast and abnormal
performed. These are highly unusual samples not often lymphocyte/lymphoblast flag achieved using the WPC channel
encountered in most laboratories. without any loss of sensitivity.
The XN NRBC counts are reported on every sample form Impedance platelet counting results are similar on both ana-
the WNR channel with corrected WBC and lymphocyte lysers. This is to be expected as it is the same technology;
counts. The reporting of the NRBC saves repeat testing in the however flags suggesting an optical platelet count is needed on
presence of the flag. The NRBC count has been shown to be the XE-2100 were seen on 27% of samples but only 9% of
both accurate and precise. samples on the XN (figure 3) requiring a PLT-F. This is a reduc-
The flagging efficiency of the XE-2100 has been published tion of repeat sampling.
previously.16 17 The PLT-F method was superior to the XE-2100 optical
The flagging performance for the two analysers was assessed method when compared to the flow cytometric reference
by comparing microscopic morphological appearance of cells. method on samples with a platelet count of 30×109/l or less
Only the blast cell flag, and abnormal lymphocyte/lymphoblast (XE-2100, R2=0.500; XN, R2=0.875). These results are very
and atypical lymphocyte flags have been optimised on the XN important as this is the most common threshold for platelet
and these were the ones evaluated in the comparative study. transfusion and accuracy is needed to ensure patients receive, or
There were too few samples with abnormal lymphocytes do not receive, transfusions where indicated.
present to give meaningful information but all three flags The precision of platelet counts on the XN is also impressive
showed fewer false positives on the XN without a loss of sensi- due to the extended platelet counts. PLT-F precision showed a
tivity; the specificity was approximately the same on both ana- CV% of only 4% on platelet counts of between 9×109/l and
lysers. Even without reflex testing to the WPC channel there 39×109/l. This is far superior precision than previously docu-
was a reduction of false positives for the atypical lymphocyte mented on the impedance platelet counts from the Sysmex
XE-2100 with an average CV of 13.6% on counts in this range
of counts.18
Table 4 The total number of blood films from 1000 samples that
would have been made using flagging results from the XE-2100 and XN
n=100 XE-2100 XN with WPC-ch Table 5 Results for neutrophils
Blood films 199 101 Neutrophils R2 Lymphocytes R2 Monocytes R2
Reflex analysis 75 112 Undiluted sample 0.914 0.917 0.808
TAT (min) analysis 611 553 Diluted sample 0.968 0.948 0.878
TAT (min) analysis+SP 710 604
Lymphocytes and monocytes were compared to the reference 400-cell differential.
Table shows number of samples needing to be re-run and the time taken for analysis. Samples were then diluted to <0.5×109/l and re-run on the XN using the low WBC mode
TAT, turnaround time; SP, SP1000 automated film maker and stainer; WPC, white cell (extended count) and differential results compared to the original undiluted manual
precursor channel. differential.

J Clin Pathol 2012;65:1024–1030. doi:10.1136/jclinpath-2012-200930 1029


Original article

Table 6 Linearity results; both high and low levels were examined Contributors All authors contributed to either the practical work or writing of the
paper.

J Clin Pathol: first published as 10.1136/jclinpath-2012-200930 on 31 July 2012. Downloaded from http://jcp.bmj.com/ on 9 April 2019 by guest. Protected by copyright.
WBC×109/l High (range 385.65–38.5) Low (range 1.03–0.12)
R2=0.998 R2=0.998 Competing interests None.
PLT-F×109/l High (range 2561–256) Low (range 60–6) Provenance and peer review Not commissioned; externally peer reviewed.
R2=0.998 R2=0.998
IPF×109/l High (range 65–6.5) Low (range 6–0)
R2=0.988 R2=0.972 REFERENCES
NRBC Range 48.94–4.89 1. Briggs C, Harrison P, Grant , et al. New quantities parameters on a recently
R2=0.999 introduced automated blood cell counter—the XE-2100. Clin Lab Haem
2000;22:345–50.
IPF, immature platelet fraction; NRBC, nucleated red blood cell; PLT-F, fluorescent platelet 2. Fernandez B, Hamachi Y. Automated enumeration of immature granulocytes. Am J
channel; WBC, white blood cells.
Clan Pathol 2007;128:454–63.
3. Jean A, Boutet C, Lenormand B, et al. The new haematology analyzer DxH 800: an
The low WBC mode allows more precise counts below evaluation of the analytical performances and leucocyte flags, comparison with the
LH 755. Int J Lab Hematol 2011;33:138–45.
0.5×109/l with proven accurate differentials. The XE-2100 4. Briggs C, Hart , Kunka S, et al. Assessment of an immature platelet fraction (IPF)
automatically suppresses the differential at counts below in peripheral thrombocytopenia. Br J Haematol 2004;126:93–9.
0.5×109/l. The XN differential information may be useful for 5. Pons I, Monteagudo M, Lucchetti G, et al. Correlation between immature platelet
patients recovering from chemotherapy or stem cell/bone fraction and reticulated platelets. Usefulness in the etiology diagnosis of
marrow transplants and will be certainly more accurate and thrombocytopenia. Eur J Haematol 2010;2:158–63.
6. Zucker ML, Murphy CA, Rachel JM, et al. Immature platelet fraction as a predictor
less time consuming than performing a manual differential. of platelet recovery following hematopoietic progenitor cell transplantation. Lab Hem
Results for all performance characteristics, such as carryover, 2006;12:125–30.
precision and linearity were excellent. 7. Miwa N, Akiba T, Kimata N, et al. Usefulness of measuring reticulocyte hemoglobin
Stability was performed on the new parameters, WBC, differ- equivalent in the management of haemodialysis patients with iron deficiency. Int J
Lab Hematol 2010;32:248–55.
ential, NRBC, PLT-F and IPF; they were found to be stable up to 8. Urrechaga E, Borque L, Escanero JF. Potential utility of the new Sysmex XE 5000
72 h both at room temperature and at 4°C. Stability for other red blood cell extended parameters in the study of disorders of iron metabolism.
parameters will be the same for the XE-2100, as technology has Clin Chem Lab Med 2009;47:411–16.
not been changed, and these have been published previously.19 9. International Council for Standardization in Haematology Expert Panel on
With continuing pressure on laboratory resources and the Cytometry; International Society of Laboratory Hematology Task force on
platelet counting. Platelet counting by the RBC/platelet ratio method. A reference
need for faster turnaround times, the challenge is to reduce the method. Am J Clin Pathol 2001;115:460–4.
number of blood films examined without missing important 10. Clinical and Laboratory Standards Institute. Reference leukocyte differential
diagnostic information. Any analyser that decreases the count ( proportional) and evaluation of instrument methods. 2007: Approved
number of false positive flags without losing sensitivity will standard. CLSI document H20-A2. CLSI, Wayne, PA.
11. Broughton PM, Gowenlock AN, McCormack JJ, et al. Revised scheme for the
increase laboratory efficiency and this has been clearly demon-
evaluation of instruments for use in clinical chemistry. Ann Clin Biochem
strated for the blast cell, blast/abnormal lymphocyte and atyp- 1974;11:207–18.
ical lymphocyte flag. With an NRBC count reported on every 12. International Council for Standarization in Haematology Expert Panel on
sample this has further reduced the number of manual differen- Cytometry. Guidelines for the evaluation of blood cell analysers including those
tials or repeat tests required. More accurate and precise results used for differential leucocyte and reticulocyte counting and cell marker applications.
Clin Lab Haemat 1994;16:157–74.
are available for both platelets and low WBC counts from the 13. Novis DA, Walsh M, Wilkinson D, et al. Laboratory productivity and the rate of
PLT-F and WPC channels, and in addition a leucocyte differen- manual peripheral blood smear review: a College of American patholgists Q-Probes
tial is reported on WBC counts of <0.5×109/l. study of 95, 141 complete blood count determinations performed in 263
The use of the XN system improves workflow efficiency and institutions. Arch Pathol Lab Med 2006;130:596–601.
14. Barnes PW, McFadden SL, Machin SJ, et al. The International Consensus Group for
confidence of results in abnormal samples in the routine Hematology Review suggested criteria for action following automated CBC and WBC
haematology laboratory. differential analysis. Lab Heamatol 2005;11:83–92.
15. Zandecki M, Genevieve F, Gerard J, et al. Spurious counts and spurious results on
haematology analysers: a review. Part I: platelets. Int J Lab Hematol 2007;29:4–20.
16. Ruzicka K, Veitl M, Thalhammer-Scherrer R, et al. The new hematology analyser
Take-home messag Sysmex XE-2100: performance evaluation of a novel white blood cell differential
technology. Arch Pathol Lab Med 2001;125:391–6.
17. Stamminger G, Auch D, Diem H, et al. performance of the XE-2100 leucocyte
The Sysmex haematology XN modular system improves differential. Clin lab Haematol 2002;24:271–80.
workflow in a routine haematology laboratory. 18. De la Salle BJ, McTaggart PN, Briggs C, et al. The accuracy of platelet
counting in thrombocytopenic blood samples distributed by the UK National
External Quality Assessment Scheme for general haematology. Am J Clin Pathol
2012;137:65–74.
19. Imeri F, Herklotz R, Risch L, et al. Stability of hematological analytes depends on
Acknowledgements The assistance of Dr Jo Linssen, Sysmex Europe, is the hematology analyser used: a stability study with Bayer Advia 120, Beckman
acknowledged. Coulter LH 750 and Sysmex XE-2100. Clin Chem Acta 2008;397:68–71.

1030 J Clin Pathol 2012;65:1024–1030. doi:10.1136/jclinpath-2012-200930

You might also like