GALILEO Original and Upgrade 1 Test Report Name: Date (YYYY/MIM/OD} GALILEO serial number: Service Manual version 610207/__ Instructions: 1. Enter information on dotted lines. 2. Check (tick) check boxes if function correct. General Maintenance 3 Vitealstime clock battery replaced ‘Oxygen cell replaced Fan filter leaned or replaced Gas inlet filters replaced Nebulizer overhauled (using kit) LCD display backlight replaced Backup batteries in column replaced TSW 1: GMP Selftest wr ver: cP ver: gen ver: GIP ver: BIOs Ver: Date 08 ver: TSW 2: GPT-GMP Communication GPT Testhyte: Displays Done oa gen Rg: Displays Done oxo ubuuuLuL uououog Code Checksum: operating Hours: Displays 4 Increments Displays ox OKO. oka oKa LED on:ofé LED flashing LED Intensity Energency Relais on:off oxa oxa oa ox oxa TSW 3: Display All colors function oxa Max. 4 grouped bad pixels OK. Backlight switches on/off OKO. Vca-contz, ‘VeA-BI08: TSW 4: User Interface ‘Alarm Bigh oka Alarm Medium = OKO. Knob OKO. c= knob OKO Alazm Low oka alare OKO 100s 02 OK = MANUALOKO = NeBULIZER OKO TSW 5: GCP-GMP Communication DPRAM Test Displays ox OKO. GPC Comm Test Displays OK OKO GHP comm Test — Displays ox OKO. Busy Test. Displays OK OKO. Testhyte GMP TSW 6: A/D Conversion Selftest status Displyso OKO collisions Valueinrange OKO Collisions Valueinrange OKO Increments. OKO aoc 12 Displays 450105250 OK apc 14 Displays-990to-1010 OK ADC 15 Displays 990 to 1020 oxa ADC 16 Displays -10 to 10 oa TSW 7: D/A Conversion ave 9 All values in range OKO apc 10 Allvalues in range OKO. TSW 8: Zero and Fullscale Ppat Paw a Sezvo dp mixer ap Flow Sensor Zero calibration OK A Zero calibration OK Zero calibration OK Zero calibration OK Zero calibration OK Zera calibration OK OL Full-scale calibration OK Cl Fulkscale calibration OK Full-scale calibration OK C1 Fulkscale calibration OK Fulkscale check 0K Q. TSW 9: 02 Measurement 02 cal Calibration oka 02 tank Allfielésinange OKO Me Allfieldsin range OKO. oxygen Alifielés in range OKO TSW 10: Mixer Peank 300 to 260 mbar leak >8s OKO. Overpressure check <800 OKO. Ptank max Disolays 330 t0 350 oxa Flow_90ns Disolays 0 0 to 300 oxo Flow 100n= Disolays 0 to 200 oxa oxygen Displays as Oxygen “knob” 45% | OKO TSW 11: Auto Zero Valves a Flow Sensor Silver (ventilator side) connector closed, displays ~€0 to 40 and 5040 to $200 OK. Blue (patient side) connector closed, displays ~40 to 40 and 5040 to 5200 OKA. Rinse flow check (with glass of water) oxa TSW 12: Safety Valve Block Inspiratory valve airtightness check. Pressure <10 mbar (PN 155161) or < 4 mbar (PN 155491) oxa Safety-valve-block aitightness check, Pressure decrease from 90 to 43 mbar >10 seconds, oka TSW 13: Ambient Valve Ambient valve open: Ppat displays 0to-20 OKO. Youcan inhale = OK You cannot exhale OK “Ambient valve close You cannot oka TSW 14: Servo and Flow Sensor 2oml/s Pressure on gauge = pressure marked on capillary tube =3 mbar oka 500m1/s Corrected pressure on gauge = pressure marked on capillay tube #2 mbar OK Flow Sensor calibrates correctly oka cal status Displays 0 oka Flow pat Values in range during inspration-valve check oka Ppat Values in range during inspiration valve check oka TSW 15: Expiration Valve calibrate exp. valve Correctly calibrates vaNe OKO) cal status Displays 0 oKa Paw Displays values in range for PEEPICPAP settings OK Ppat Displays values in range for PEEP/CPAP settings OK Gd TSW 16: Nebulizer Paux Displays < 35.0 after 105 oKa Nebulizer pressure at gauge > 800 mbar oKa Nebulizer valve and compressor switch on and off correctly = OK Ptank Displays 150 to 350, oKa Preoperational check oka 3-month check oxa GALILEO Upgrade 2 Test Report Name: Date (YYYYMIMVDD}: GALILEO serial number: Service Manual version 610207/_ Instructions: 1, Enter information on dosted lines 2. Check (tick) check boxes if function correct. General Maintenance 3 V real-time clock battery replaced ‘oxygen call replaced Fan filter cleaned or replaced Gas inlet fiers replaced Nebulizer overhauled (using kit) LCD aisplay backlight replaced Backup batteries in column replaced ubuuvuLU oceuoucog Test 1, Microprocessor checks uP ver: Software checksum: Displays ok OKO scp ver: Controlboard rev. ger ver: GIP ver: BIOS ver: Date operating hours: 08 ver: Int, temp. Inrange OKO. Test 2, GPT-GMP communication & LED checks Selftest: Displays ¢ OKO. Increnented val Increments OKO. GPT-GuP increm. test: Displays Done oka LED on/ose oxa GPT-GHP intere test: Displays Done oka LED Flash/ofe = OKO. LED brightn oxa Bnergency alarm OKO Relay on/off = OKO. Test 3, LCD display checks VGA board ver: VGh-BI08 ver: All colors function oxa Max. 4 grouped bad pixels OKO. Backlg Test 4, User interface & alarmtone checks knob OKO. © ~ knob Alara OKO 1008 02 OKO switches onioff OK. High alam oka Medium alarm = OKO. oxa Low alare oka mowon. OKO wepuLazeR OKO Test 5, GCP-GMP communication checks DPRAM test: Displays OK OKO. Selftest Displayso OKO. Gup-ccr test: Displays OK OKO. Gcp-cP test: Displays ox OKO. DPRAM r/w test: Displays oK OKO DPRAK collisions Value inrange OK DDPRAM collisions Value inrange OKO Increnented val. Increments OK. Test 6, A/D converter check aoc 12 Displays 450105250 OK ADC 14 Displays -990to-1010 OK ADC 15 Displays 990 to 1010 oxa Ac 16 Displays -10 to 10 oxa Test 7, D/A converter check ave 9 Allvaluesin range OKO. apc 10 Allvalues in range OK Test 8, Zero and full-scale calibration Ppat Paux a Sezvo dp mixer ap Flow Sensor Zero calibration OK A Zero calibration OK A Zero calibration OK Zero calibration OK Zero calibration OK Zero calibration OK A Full-scale calibration OK Ol Fulkscale calibration OK Full-scale calibration OK Fulkscale calibration OK C1 Fulkscale check 0K Q. Test 9, 02 cell calibration & check Calibrate 02 celiCalibration OKO tank gas Allfielésinrange OKO Inlet aie Allfieldsinrange OKO. Inlet 02 Allfieldsinvange OKO. Test 10, Mixer calibration & checks Peank 300 to 280 mbar leak >8 5 oka Overpressure check <500 OKO Ptank max Disolays 330 10 350 oka 90ms flow Disolays 0 to 300 oka looms flow Disolays 0 to 200 oka oxygen Display as Oxygen “knob” 25% OKO Test 11, Flow Sensor circuit checks a Flow Sensor Siler (ventilator side) connector closed, displays ~€0 to 40 and 5040 to $200 OKO. Blue (patient side) connector closed, displays ~40 to 40 and 5040 to 5200 OKA Rinse flow check (with glass of water) oxa Test 12, Patient overpressure valve checks Inspiratory valve airtightness check. Pressure <10 mbar (PN 155161) or < 4 mbar (PN 155491) oxa Safety-valve-block aitightness check, Pressure decrease from 90 to 43 mbar >10 seconds, oka Test 13, Ambient valve check Ambient valve open Ppat displays 0to-20 OKO. Youcan inhale = OK You cannot exhale OK “Ambient valve close You cannot inhale OKO Test 14, Inspiration valve check 2oml/s Pressure on gaug 500m1/s Corrected pressure on gauge = pressure marked on capillary tube +2 mbar Flow Sensor calibrates correctly cal. statue Displays 0 Flow pat Values in range during inspration-valve check Ppat Values in range during inspiration valve check ressure marked on capillary tube =3 mbar oka oka oka oka oka oka Test 15, Expiration valve calibrat. & check calibrate exp. valve Correctly callorates vale oka cal. status Displays 0 oka Paw Displays values in range for PEEPICPAP settings oxa Ppat Displays values in range for PEEP/CPAP settings oka Test 16, Nebulizer compressor check Paux Displays < 38.0 after 105 oka Nebulizer pressure at gauge > 800 mbar oka Nebulizer valve and compressor switeh on and off correctly oka Ptank Displays 150 to 350, oKa Preoperational check oKa 3-month check oKa

You might also like