90501 4:00:02 AM 3113 Change Hydraulic Fluid
From Buffer
To DI
ID of Batch on SSA 1: 2338
ID of Batch on SSA 2: 2337
ID of Batch on SSA 3: 2336
90502 4:02:52 AM 3032 Mutex event
Batch: -68
Mutex long wait 18
ID of Batch on SSA 1: 2338
ID of Batch on SSA 2: 2337
ID of Batch on SSA 3: 2336
22-Dec-2019
12/27/2019 9:02 AM 303/1013
From: 12/20/2019 9:01:00 AM
To: 12/27/2019 9:01:00 AM
91641 9:21:18 AM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Bond Dewax Solution (position 4 of 8)
91642 9:21:20 AM 3113 Change Hydraulic Fluid
From Dewax
To Alcohol
24-Dec-2019
91926 11:43:56 AM 3081 The heater relay has been forced off due to:
Description: Relay stuck off
91927 11:43:56 AM 3044 The PM has entered a System Error state
Error number 3
The slide heater relay is detected as
permanently off. Power to the slide heaters
has been removed for safety reasons.
91960 11:55:08 AM 3019 Computer disconnected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92009 1:11:24 PM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Alcohol (position 7 of 8)
92010 1:11:26 PM 3112 Prime bulk fluid
Bulk container *Bond Dewax Solution (position 4 of 8)
92011 1:11:26 PM 3113 Change Hydraulic Fluid
From Alcohol
To Dewax
92012 1:12:13 PM 3037 LLS retry event
92014 1:12:35 PM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Bond Dewax Solution (position 4 of 8)
92015 1:12:37 PM 3113 Change Hydraulic Fluid
From Dewax
To Alcohol
92016 1:13:02 PM 3112 Prime bulk fluid
Bulk container *Bond Wash Solution (position 6 of 8)
92017 1:13:02 PM 3113 Change Hydraulic Fluid
From Alcohol
To Buffer
92018 1:14:32 PM 3112 Prime bulk fluid
Bulk container *Bond ER Solution 1 (position 2 of 8)
92019 1:14:32 PM 3113 Change Hydraulic Fluid
From Buffer
To HIER1
92020 1:15:43 PM 3112 Prime bulk fluid
Bulk container *Bond ER Solution 2 (position 3 of 8)
92025 1:29:56 PM 3081 The heater relay has been forced off due to:
Description: Relay stuck off
92026 1:29:56 PM 3044 The PM has entered a System Error state
Error number 3
The slide heater relay is detected as
permanently off. Power to the slide heaters
has been removed for safety reasons.
92027 2:17:57 PM 3019 Computer disconnected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92028 2:17:57 PM 3009 Communication error?
Description: 5 Service Event messages from PM have been lost
92029 2:17:57 PM 3054 Processing Module started
Firmware version number 05.01.00
Config pack version number
Normal mode
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92030 2:17:57 PM 3018 Computer connected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92054 2:29:25 PM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Alcohol (position 7 of 8)
92055 2:29:27 PM 3112 Prime bulk fluid
Bulk container *Bond Dewax Solution (position 4 of 8)
92056 2:29:27 PM 3113 Change Hydraulic Fluid
From Alcohol
To Dewax
92057 2:30:14 PM 3037 LLS retry event
92059 2:30:35 PM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Bond Dewax Solution (position 4 of 8)
92060 2:30:38 PM 3113 Change Hydraulic Fluid
From Dewax
To Alcohol
92072 2:42:13 PM 3024 The separator bottle is full (LLS probes triggered)
Standard separator
Separator full
92073 2:44:33 PM 3081 The heater relay has been forced off due to:
Description: Relay stuck off
92087 3:15:37 PM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Alcohol (position 7 of 8)
92088 3:15:39 PM 3112 Prime bulk fluid
Bulk container *Bond Dewax Solution (position 4 of 8)
92089 3:15:39 PM 3113 Change Hydraulic Fluid
From Alcohol
To Dewax
92090 3:16:25 PM 3037 LLS retry event
92092 3:16:47 PM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Bond Dewax Solution (position 4 of 8)
92093 3:16:49 PM 3113 Change Hydraulic Fluid
From Dewax
To Alcohol
92094 3:17:16 PM 3112 Prime bulk fluid
Bulk container *Bond Wash Solution (position 6 of 8)
92095 3:17:16 PM 3113 Change Hydraulic Fluid
From Alcohol
To Buffer
92096 3:18:45 PM 3112 Prime bulk fluid
Bulk container *Bond ER Solution 1 (position 2 of 8)
92097 3:18:45 PM 3113 Change Hydraulic Fluid
From Buffer
To HIER1
92098 3:19:56 PM 3112 Prime bulk fluid
Bulk container *Bond ER Solution 2 (position 3 of 8)
92099 3:19:56 PM 3113 Change Hydraulic Fluid
From HIER1
To HIER2
92103 3:45:13 PM 3081 The heater relay has been forced off due to:
Description: Relay stuck off
92104 3:45:13 PM 3044 The PM has entered a System Error state
Error number 3
The slide heater relay is detected as
permanently off. Power to the slide heaters
has been removed for safety reasons.
92105 4:34:44 PM 3017 Bulk bottle state has changed
Bulk container *Bulk Waste (position 8 of 8)
Bulk container removed
92106 4:43:06 PM 3019 Computer disconnected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92107 4:43:06 PM 3009 Communication error?
Description: 5 Service Event messages from PM have been lost
92108 4:43:06 PM 3054 Processing Module started
Firmware version number 05.01.00
Config pack version number
Normal mode
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92109 4:43:06 PM 3018 Computer connected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92116 4:46:08 PM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Alcohol (position 7 of 8)
92117 4:46:10 PM 3112 Prime bulk fluid
Bulk container *Bond Dewax Solution (position 4 of 8)
92118 4:46:10 PM 3113 Change Hydraulic Fluid
From Alcohol
To Dewax
92119 4:46:56 PM 3037 LLS retry event
92121 4:47:18 PM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Bond Dewax Solution (position 4 of 8)
92122 4:47:21 PM 3113 Change Hydraulic Fluid
From Dewax
To Alcohol
92123 4:47:47 PM 3112 Prime bulk fluid
Bulk container *Bond Wash Solution (position 6 of 8)
92124 4:47:47 PM 3113 Change Hydraulic Fluid
From Alcohol
To Buffer
92125 4:49:16 PM 3112 Prime bulk fluid
Bulk container *Bond ER Solution 1 (position 2 of 8)
92156 5:36:14 PM 3081 The heater relay has been forced off due to:
Description: Relay stuck off
92157 5:36:14 PM 3044 The PM has entered a System Error state
Error number 3
The slide heater relay is detected as
permanently off. Power to the slide heaters
has been removed for safety reasons.
92170 5:54:52 PM 3019 Computer disconnected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92171 5:54:52 PM 3009 Communication error?
Description: 5 Service Event messages from PM have been lost
92172 5:54:52 PM 3054 Processing Module started
Firmware version number 05.01.00
Config pack version number
Normal mode
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92173 5:54:52 PM 3018 Computer connected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92187 5:59:14 PM 3019 Computer disconnected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92207 6:02:47 PM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Bond Dewax Solution (position 4 of 8)
92208 6:02:49 PM 3113 Change Hydraulic Fluid
From Dewax
To Alcohol
92209 6:03:15 PM 3112 Prime bulk fluid
Bulk container *Bond Wash Solution (position 6 of 8)
92210 6:03:15 PM 3113 Change Hydraulic Fluid
From Alcohol
To Buffer
92211 6:04:32 PM 3112 Prime bulk fluid
Bulk container *Bond ER Solution 1 (position 2 of 8)
92212 6:04:32 PM 3113 Change Hydraulic Fluid
From Buffer
To HIER1
92214 6:05:44 PM 3112 Prime bulk fluid
Bulk container *Bond ER Solution 2 (position 3 of 8)
92215 6:05:44 PM 3113 Change Hydraulic Fluid
From HIER1
To HIER2
92216 6:06:53 PM 3112 Prime bulk fluid
Bulk container *Deionized Water (position 5 of 8)
92217 6:06:53 PM 3113 Change Hydraulic Fluid
From HIER2
To DI
92218 6:11:56 PM 3055 Processing Module initialized
25-Dec-2019
92226 8:51:59 AM 3019 Computer disconnected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92227 8:51:59 AM 3009 Communication error?
Description: 5 Service Event messages from PM have been lost
92228 8:51:59 AM 3054 Processing Module started
Firmware version number 05.01.00
Config pack version number
Normal mode
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92229 8:51:59 AM 3018 Computer connected
Event times indicate when events were
logged on the controller. This may differ
from the actual event times on the processing
module.
92237 8:53:30 AM 3112 Prime bulk fluid
Bulk container *Alcohol (position 7 of 8)
92238 8:53:30 AM 3113 Change Hydraulic Fluid
From Unknown
To Alcohol
92247 8:54:42 AM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Alcohol (position 7 of 8)
92248 8:54:44 AM 3112 Prime bulk fluid
Bulk container *Bond Dewax Solution (position 4 of 8)
92249 8:54:44 AM 3113 Change Hydraulic Fluid
From Alcohol
To Dewax
92252 8:55:31 AM 3037 LLS retry event
92254 8:55:52 AM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Bond Dewax Solution (position 4 of 8)
92255 8:55:54 AM 3113 Change Hydraulic Fluid
From Dewax
To Alcohol
92258 8:56:21 AM 3112 Prime bulk fluid
Bulk container *Bond Wash Solution (position 6 of 8)
92259 8:56:21 AM 3113 Change Hydraulic Fluid
From Alcohol
To Buffer
92267 8:57:50 AM 3112 Prime bulk fluid
Bulk container *Bond ER Solution 1 (position 2 of 8)
92268 8:57:51 AM 3113 Change Hydraulic Fluid
From Buffer
To HIER1
92271 8:59:01 AM 3112 Prime bulk fluid
Bulk container *Bond ER Solution 2 (position 3 of 8)
92272 8:59:01 AM 3113 Change Hydraulic Fluid
From HIER1
To HIER2
26-Dec-2019
92345 4:00:01 AM 3113 Change Hydraulic Fluid
From Buffer
To DI
ID of Batch on SSA 1: 2350
ID of Batch on SSA 2: 2348
ID of Batch on SSA 3: 2349
92352 4:03:35 AM 3032 Mutex event
Batch: -26
Mutex long wait 20
ID of Batch on SSA 1: 2350
ID of Batch on SSA 2: 2348
ID of Batch on SSA 3: 2349
92358 4:06:13 AM 3032 Mutex event
Batch: -20
Mutex long wait 23
ID of Batch on SSA 1: 2350
ID of Batch on SSA 2: 2348
ID of Batch on SSA 3: 2349
92360 4:09:04 AM 3032 Mutex event
Batch: -21
Mutex long wait 39
ID of Batch on SSA 1: 2350
ID of Batch on SSA 2: 2348
ID of Batch on SSA 3: 2349
92362 4:09:46 AM 3032 Mutex event
Batch: -26
Mutex long wait 34
ID of Batch on SSA 1: 2350
ID of Batch on SSA 2: 2348
ID of Batch on SSA 3: 2349
93524 9:27:07 AM 3031 Priming of the bulk fluid line could not be confirmed. Check the
syringe, reagent tubing and bulk bottle for leaks. Verify LLS.
Bulk container *Bond Dewax Solution (position 4 of 8)
93525 9:27:10 AM 3113 Change Hydraulic Fluid
From Dewax
To Alcohol
93526 9:27:36 AM 3112 Prime bulk fluid
Bulk container *Bond Wash Solution (position 6 of 8)
93527 9:27:36 AM 3113 Change Hydraulic Fluid
From Alcohol
To Buffer
27-Dec-2019
93747 4:00:01 AM 3113 Change Hydraulic Fluid
From Buffer
To DI
ID of Batch on SSA 1: 2361
ID of Batch on SSA 2: 2360
ID of Batch on SSA 3: 2359
93749 4:04:20 AM 3032 Mutex event
Batch: -53
Mutex long wait 19
ID of Batch on SSA 1: 2361
ID of Batch on SSA 2: 2360
ID of Batch on SSA 3: 2359