IRIA CHECKINGS

1. Check the same tube in all the positions, do you find the L.E. error message in all the
positions?
2. Open the instrument cover with the 4 screws located on the backside of the
instrument.
3. Remove the mechanical part by the two screws showed on the pictures.

4. Then check the reading board, the photodiode couples have to be orientated one in
front of the other.
5. If all the photodiodes are fine, please, check the cable that connects the reading board
to the main board.
6. If the cable is fine and the photodiodes also maybe there are some problem on the
reading board, so if you have some other unit of IRIA please, try to exchange the board
to check if the problem is solved.