You are on page 1of 2

Technical Specifications; Microwave synthesizer

Sl. No. Item descriptions: Specifications


1. Application Range Microwave assisted synthesis system; should be included
routine Organic, Organome tallic, Nano materials
synthesis, fluorination, caustic solutions, catalysts using
palladium, non‐polar solvents like toluene, hexane etc.

2. Power Output Microwave power of minimum 800 W or higher.

3. Microwave Power 5000 Watts/liter or more


field density
4. Maximum Temperature 300 ℃ or better

5. Heating Must reach the desired temperature in around 5 min.


rate/performance
6. Temperature IR measurements with accuracy of ±1 ℃ or better
measurements
7. Maximum Pressure 20 bar or better
8. Pressure sensor Integrated Pressure Sensor to measure pressure.
9. Magnetic Stirrer Should have inbuilt magnetic stirrer; variable speed
from 0 rpm up to at least 2000 rpm or more
10. Solvent range System must be able to effectively heat polar as well as
non-polar solvents to elevated temperature without
heating aids.
11. Glass or quartz vials Must includes 50 ml or better reaction vials
for non corrosive reac
tions

12. Teflon reaction vessel Must includes 50 ml or better reaction vials


for corrosive reactions:

13. Display Inbuilt Touch-screen display


14. Consumables: 1. Magnetic Stir Bars for 50ml
2. Glass Vessels (50ml or better)
3. Caps or stoppers for reaction vessel
15. Service Facility Supplier should clearly mention about their service set up
in India for prompt service support. Direct service support
from the company is desired.

16. Installation and training Free of cost at site


17. Warranty The Microwave Synthesizer including its all accessories
should be under comprehensive warranty for three (3)
years from the date of installation.

You might also like