EQUIPMENT LEGEND
DESCRIPTION
was
ur 7 iors
err
HetcHr
[SIOTTO CLASS TAS)
CD) [acquismion worxsrarion
Tw
S00 be
Beam iam
etd 26"
lam
iy
FD) [Stato cass (car)
(2) [verncat ear unit
kg
728 bs
Tar LOR AR | TELAT
sorsoa | ‘so.2-sea"
TEE na
sr2-39"
nove:
CEILING AND 8EL0M FLOOR.
NOTE 4: ‘GIOTTO CLASS’ VERTICAL X-RAY UNIT POW
RECEPTACLE FOR GANTRY POWER,
{UIPMENT INSTALLERS WILL USE EXISTING WIREWAYS WHEREVER POSSIBLE. HOSPITAL ENGINEERING DEPART
PROVIDE IMS SITE PANNING OFFICE WITH AS BUILT DRAWINGS, SHOWING WIREWAY LOCATIONS AND ROUTING ABOVE
NOTE 1: ALL RADIATION PROTECTION TO BE DESIGNED BY, OR APPROVED BY CUSTOMER'S LICESED RADIATION PHYSICIST.
NOTE 3: IMS STRONGLY RECOMMENDS TO ALLOW REMOTE ACCESS CONNECTION TO THE ACQUISITION WORKSTATION FOR
REMOTE DIAGNOSTIC SUPPORT. CONSULT WITH IMS CONNECTIVITY SPECIALIST REGARDING REQUI
RECOMMENDATION: SWITCH GIGABIT ETHERNET (100 BASE-T.
MENTS.
1S TO BE PERMANENTLY WIRED. IMS PROHIBITS USE OF PLUG AND
TIME OF INSTALLATION,
[NOTE 5: IN ACOOROANCE WITH IEC 6060-1, MINIMUM CLEARANCE OF 500 MM (20") MUST BE MAINTAINED BETWEEN THE TUBE
HEAD AT ANY POSITION OF ROTATION, AND ANY STATIONARY OBJECT (EXCLUDING FLOOR). F THIS REQUIREMENT 15,
NOT NET, ALTO C-ARM ROTATION MUST BE DISABLED AT T
SITE CONDITIONS FOR THE SYSTEM
II REQUIRED ROOM SIZE
2500 m= 3000 nen (S84 = BI")
INOW CEILING HEIGHT
2500 mm (38.4)
RECOMMENDED CEILING HEIGHT
2700 mm (106.3
OPERATING TEMPERATURE RANGE
(AINTAIN TEMP. RANGE 2sHRS/DAY, 7/¥K)
eC = OPC (SPre BF]
258C (77°F |= ROOM TEMP. FOR OPTIMAL PERFORMAN
NAx ALLOWABLE TENPERATURE CHANGE RATE: SC (9°F PER HOUR.
[OPERATING HUMIDITY RANGE
FR = 7585, WON CONDENSING
STORAGE TEMPERATURE (SYSTEM POW
JOR IN STORAGE)
AED DOWN
C= V0 [ PF = HAD) -REQUISTION WORKSTATION
208C = 4500C (°F + 41229F}- VERTICAL XRAY UNIT, EXCLUDING
[10°C + 240°C ( +50¢F + H1060F)- DETECTOR ASSEMBLY
VERTICAL Y-RAY UNIT SHIPPING CRATE
F600 ex 900 mx 1TBD ry (BROT x 35.87 70
> 800
ACQUISITION WORKSTATION SHIPPING CRATE
000 en x BOO mT nc126O (9A STST SO.
INH ENTRANCE
B00 mm x 2130 mm 31.5" 86.0"
RECOMMENDED ENTRANCE
B50 mmx 2130 mm 36.0" x 86.0"
CHANGES BASED ON CUSTOMER'S WISHES,
NOTE: TH EQUIPMENT LAYOUT, ROOM DIMENSIONS, ELECTRICAL AND MECHANICAL DETAILS, ARE BASED ON THE BES
INFORMATION AVAILABLE FROM THE STE, IN ADDITION TO THE CUSTOMER'S «NOW
OR ELECTRICAL CHANGES, INCLUDING RELOCATION OF EQUIPHENT ILLUSTRATED IN DTHIS DRAWING, ARE ALLOWED
ONLY WITH NOTIFICATION IN WRETING, AND REVIEW BY (hS SERVICE DEPT. MS RESERVES THE RIGHT TO MAKE
RE
CONSTRUCTION PROBLEMS, ETC
15, ARCHITECTURAL
‘GIOTTO CLASS'
EQUIPMENT PLAN
RECOMMENDED ROOM SIZE
2800
ri, $00
(OPERATOR AREA
rin, 500
‘SERVICE AREA
300
2.