HPP Inter 1

You might also like

You are on page 1of 7

TAGEDENS E M I N A R S I N P E R I N A T O L O G Y 43 (2019) 11 17

Available online at www.sciencedirect.com

Seminars in Perinatology
www.seminperinat.com

Postpartum hemorrhage: early identification


challenges
Maria Andrikopoulou*, and Mary E. D’Alton
Department of Obstetrics and Gynecology, College of Physicians and Surgeons, Columbia University, New York, NY

A R T I C L E I N F O AB STR ACT

Postpartum hemorrhage is the leading cause of maternal morbidity and mortality world-
wide. The majority of maternal deaths associated with hemorrhage could be preventable.
The accurate assessment of blood loss, identification of risk factors and timely recognition
of postpartum hemorrhage remain major challenges in obstetrics. It is important to review
available modalities for estimation and quantification of peripartum blood loss, the value
of risk assessment tools as well as the challenges in early recognition of clinical signs and
symptoms of postpartum hemorrhage.
Ó 2018 Elsevier Inc. All rights reserved.

treatments with associated emotional and economic costs.


Introduction Thus, early, timely and precise identification of PPH is vital,
although limited due to lack of accurate methods to measure
Postpartum hemorrhage (PPH) is the leading cause of maternal blood loss.
morbidity and mortality worldwide, and affects up to 10% of all
deliveries.1 Although historically many definitions have been
used, the American College of Obstetricians and Gynecologists’
reVITALize program, in an effort to standardize clinical defini- Estimated blood loss
tions in obstetrics, defines postpartum hemorrhage as blood
loss greater than or equal to 1000 mL or blood loss with signs or “Estimated blood loss” is perhaps the most common method
symptoms of hypovolemia within 24 h of delivery whether used to assess blood loss at the time of delivery. Also known
cesarean section or vaginal birth.2 as “eyeball” estimation, estimated blood loss (EBL) mainly
Postpartum hemorrhage can lead to maternal deaths, the relies on a physician’s opinion based on clinical experience
majority of which could be preventable.3 However, the accu- and use of visual aids such as sponges, kidney dishes and
rate assessment of blood loss still remains a major challenge sanitary pads. Despite the widespread utilization of estimated
in the care of women in labor and delivery. Depending on the blood loss, there have been inaccuracies and controversy in
clinical circumstance, clinicians are prone to either underesti- the literature described for many years on the utility and pre-
mate or overestimate maternal blood loss. Underestimation is cision of visual aids. Buckland and Homer reported that visual
particularly problematic as it results in delayed recognition of aids such as kidney dishes or smaller containers were associ-
postpartum hemorrhage, delayed initiation of therapy leading ated with more accurate estimation of blood loss than estima-
to increased maternal morbidity and mortality. On the other tion of blood when soaked into linen or pads.4 On the other
hand, overestimating blood loss leads to unnecessary hand, the prospective simulation study by Brooks et al.

* Corresponding author: Columbia University Irving Medical Center, 622 West 168th Street, New York 10032 USA.
E-mail address: ma3659@cumc.columbia.edu (M. Andrikopoulou).

https://doi.org/10.1053/j.semperi.2018.11.003
0146-0005/Ó 2018 Elsevier Inc. All rights reserved.
12 S E M I N A R S I N P E R I N A T O L O G Y 43 (2019) 11 17

showed that the addition of visual aid led to overestimation of anesthesia providers were randomized to estimate simulated
blood loss compared to just the use of a collector bag and baby blood loss in calibrated or noncalibrated vaginal delivery drapes,
scale.5 Larson and colleagues reported that standard estima- showed that the use of calibrated drapes improved the accuracy
tion of blood loss with visual aids, pads, swabs and diapers of EBL compared to noncalibrated drapes, with error of <15%.18
have been proven to be imprecise and inaccurate both for Visual estimation of blood loss was inaccurate compared to
vaginal delivery and cesarean section.6 Visual estimation is quantitative methods, irrespective of experience, level of train-
thought to be inaccurate amongst health care professionals ing or specialty. The discrepancy was higher with increasing
and does not correlate with years of training and experience.7 blood volume. A randomized controlled trial by Patel et al. has
Simulation, teaching sessions and clinical reconstructions also reported superiority of QBL.19 In this study, women were
have been widely used in the clinical setting in an effort to randomized to visual or drape estimation of blood loss at the
improve clinical blood loss estimation skills.8 11 However, the time of delivery. EBL was shown to underestimate postpartum
long-term benefit of such teaching tools has not been proven blood loss by 33% when compared to QBL. Additionally, the
yet. A study by Toledo et al. included 44 participants who authors performed spectrometry in 10 patients and found a
underwent web-based training of blood loss estimation. The high correlation between blood loss and QBL. Similarly, a pro-
post-training test, which was repeated 9 months later, spective study by Kadri et al., which included 150 women who
showed a decline and inaccuracy of blood loss estimation had a vaginal delivery compared visual assessment of blood
skills remote from the didactic training.12 Simulation exer- loss versus gravimetric measurement via measurement of
cises increase the vigilance and accuracy of health care pro- weight of soaked materials. Health care providers underesti-
fessionals initially, however education does not produce mated blood loss by 30%, when visually calculating blood loss
sustainable improvements in blood loss estimation. compared to gravimetric measurement.20
Different specialties and disciplines have also reached to However, other studies have questioned whether QBL can
similar conclusions on the challenges of accurately estimating actually reduce risk associated with postpartum hemorrhage. A
blood loss. A prospective blinded cross sectional study among cluster randomized trial in 13 European countries investigated
emergency medical services (EMS) personnel, showed that the use of plastic collector bag in the measurement of blood
EMS professionals were also inconsistent with the amount of loss.21 Maternity units were randomized to use a collector bag
blood loss.13 There was no correlation between accuracy and versus visual estimation of blood. The collector bag failed to
level of training. The conclusion was that time is better to be reduce postpartum hemorrhage. Similarly, Hackock et al.
spent attending to patient rather than visually estimating reviewed 36 studies on blood loss during delivery and provided
blood loss given poor estimates of bleeding. Similarly, a study no evidence that blood loss measurement led to improved PPH
on anesthesia providers, showed comparable results, with diagnosis, prevented progression to severe PPH or was corre-
accuracy of estimated blood loss not related to provider train- lated to improved maternal outcomes.22 Additionally, when
ing, years of education, years of experience, gender, or ethnic- QBL was used as a predictor of hemoglobin, it did not improve
ity.14 These results are reinforced by the report of Rothermel the prediction of hemoglobin 12 h postpartum in deliveries with
and Lipman who showed that visual estimation of operative average blood loss and it was proven to be time and resource
blood loss is unreliable and inaccurate amongst anesthesia intensive, requiring specific protocols and providers’ educa-
providers, surgeons, nurses and practitioners.15 tion.23 The poor sensitivity of weighted blood loss to detect PPH
compared to hemoglobin drop was also reported by Atukunda
et al., however, that study underlined that weighted blood loss
Quantitative blood loss could be valuable in areas with high PPH prevalence when labo-
ratory measurements cannot be promptly accessed.24
Given that estimation has been shown to be inaccurate in the Given the challenges of existing techniques to accurately cal-
evaluation of blood loss, an effort has been made for quantifica- culate the measurement of blood loss, recently, both in vitro
tion of blood loss during labor. The California Maternal Quality and clinical studies in obstetrics and also other surgical fields
Care Collaborative (CMQCC) developed an obstetric hemorrhage have investigated Triton, a novel mobile monitoring sys-
toolkit which emphasizes the need for cumulative quantitative tem.25 28 It measures hemoglobin loss absorbed by surgical
assessment of blood loss.16 Under-buttock drapes during vagi- sponges using a platform. Images of blood soaked sponges are
nal deliveries and gram scales to weigh blood soaked materials captured and transferred to a remote server. Feature Extraction
have been proposed to quantitate blood loss more precisely.16 It Technology is used to provide an accurate and precise measure-
is speculated that quantitative blood loss (QBL) has multiple ment of blood loss.25 26 It is speculated that it provides precise
advantages in comparison to EBL. It is believed that it leads to measurement of hemoglobin on surgical sponges compared to
earlier interventions and improves the culture of drawing atten- manual rinsing measurements and is more accurate than gravi-
tion to blood loss. Objective data can help the nurse mobilize metric evaluation.26 However, further studies, in a larger set of
the obstetrical team and additional resources earlier. Propo- patients, are needed to better evaluate the clinical use of this
nents highlight that even though resource needs for quantita- new technology and the role of Triton in labor and delivery.
tive blood loss measurement are higher than EBL, the use of
under buttock drapes or newborn scales to weight the soaked
material are overall economic and cost effective measures Risk assessment
which offer an advantage to labor and delivery units.17
Multiple studies have evaluated the role of QBL in labor and Both estimated and quantitative blood loss can be inaccurate
delivery. A study by Toledo et al., where 106 obstetrics and and cannot predict or reduce the risk of PPH. Thus, early
TAGEDENS E M I N A R S I N P E R I N A T O L O G Y 43 (2019) 11 17 13

identification of risk factors for postpartum hemorrhage can facilities in a timely fashion. Women with pregnancy com-
lead to both awareness and preparedness for high blood plications that are at risk for hemorrhage such as placenta-
loss.29 Risk assessments should be undertaken during prena- tion anomalies, should be transferred to centers that
tal visits, antepartum care, admission to labor and delivery, specialize in the management of such pregnancy complica-
during the labor and postpartum course, as risk factors can tions with the help of a multidisciplinary team.29 Even
change or evolve during labor course.29 Identifying the risk though there are multiple validated risk assessment tools
factors in each case provides the team the opportunity to available to help providers identify women at high risk,30 31
prepare, confirm that supplies and blood products are avail- (Table 1) these tools can only be used as a guidance, as they
able for use, get appropriate specialties involved in the care identify only a fraction of women who will develop postpar-
of patients and possibly transfer high risk patients to other tum hemorrhage.

Table 1 – Risk assessment tools to help providers identify women at high risk for hemorrhage.31
14 S E M I N A R S I N P E R I N A T O L O G Y 43 (2019) 11 17

Table 2 – Obstetric hemorrhage checklists: initial steps, medication use, blood transfusion and actions in dif-
ferent stages of hemorrhage.43

(continued on next page)


TAGEDENS E M I N A R S I N P E R I N A T O L O G Y 43 (2019) 11 17 15

Table 2 (continued.)

Vital signs and symptoms of hemorrhage rate (RR) 20 30 correlates with 15% of blood volume loss,
whereas decreased blood pressure, tachycardia over 120 bpm
In addition to meticulous estimation of blood loss in cases of and RR 30 40 correlates with 30 40% of blood volume loss.32
postpartum hemorrhage, careful observation of clinical signs Many studies have challenged the use of vital signs cut offs
is also vital. Low systolic blood pressure, tachycardia, and in the setting of hemorrhage. Brasel et al. in a study which
raised respiratory rate have been historically used as signs of included the records of more than 10,000 trauma patients,
hypovolemia. According to the Advance Trauma Life Support reported that increased heart rate over 100 bpm, did not pre-
(ATLS) calcification, tachycardia with heart rate over 100 beats dict the need for immediate intervention in trauma patients.33
per minute (bpm), decreased pulse pressure and respiratory Similarly, Victorino et al. also showed that tachycardia is not a
16 S E M I N A R S I N P E R I N A T O L O G Y 43 (2019) 11 17

reliable sign of hypovolemia in surgical trauma patients.34 vital but should not be the only focus for the management of
Even though it is independently associated with hypotension, postpartum hemorrhage. Other factors such as the rate of
absence of tachycardia is not indicative of absence of signifi- blood loss, clinical signs, patient symptoms, the shock index,
cant blood loss. Other studies have also questioned the validity and the physiological response to hemorrhage are crucial to
of vital signs in trauma patients, showing that an association early recognition of a high risk situation which may help pro-
exists between tachycardia, lower systolic blood pressure and mote optimal management.22
raised RR, but not to the degree reported in ATLS guidelines.35
A systematic review including 30 studies on the association of
blood loss with clinical signs and symptoms reported a signifi- Disclosure
cant variability between blood loss and clinical signs.36 Thus, it
can be very challenging to establish specific cut off points for Dr. Andrikopoulou has no conflict of interest to disclose. Dr.
vital signs which could alert providers and trigger an expedited D’Alton is acting as the Studywide Principal Investigator for a
team response and action. Prospective, Single Arm, Pivotal Clinical Trial Designed to Assess
Most studies in the literature include mainly surgical patients the Safety and Effectiveness of the InPress Device In Treating Pri-
and are not focused on the obstetric population. Changes in mary Postpartum Hemorrhage with InPress Technoligies, Inc.
maternal physiology such as increase in maternal blood volume
and cardiac output, could affect the role of vital signs as a surro- R E F E R E N C E S
gate for blood loss in pregnancy. Case reviews of maternal deaths
led to the proposed Maternal Early Warning Criteria, which indi-
cate vital sign parameters that should trigger prompt patient 1. Say Lale, Chou Doris, Gemmill Alison, et al. Global causes of
evaluation and management such as systolic and diastolic blood maternal death: a WHO systematic analysis. Lancet Glob
pressure, heart rate, respiratory rate, oxygen saturation, oliguria, Health. 2014;2(6):e323–e333.
2. Menard MK, Main EK, Currigan SM. Executive summary of the
maternal agitation, confusion or unresponsiveness.37 However,
reVITALize initiative: standardizing obstetric data definitions.
this list of parameters only provides guidelines to help alert pro-
Obstet Gynecol. 2014;124(1):150–153.
viders and prevent delay in recognition of severe hemorrhage or 3. Haeri S, Dildy GA III. Maternal mortality from hemorrhage.
other emergencies.37 Semin Perinatol. 2012, February;36(1):48–55: WB Saunders.
Since change of vital signs does not always correlate with 4. Buckland SS, Homer CS. Estimating blood loss after birth:
the amount of blood loss, other clinical signs that could have using simulated clinical examples. Women Birth. 2007;20
a predictive value in the setting of postpartum hemorrhage (2):85–88.
5. Brooks M, Legendre G, Brun S, et al. Use of a visual aid in addi-
have been studied. The shock index (SI) is calculated as the
tion to a collector bag to evaluate postpartum blood loss: a
heart rate divided by the systolic blood pressure and can prospective simulation study. Sci Rep. 2017;7:46333.
be an accurate predictor of cardiovascular changes secondary 6. Larsson C, Saltvedt S, Wiklund I, Pahlen S, Andolf E. Estima-
to blood loss even in patients who otherwise would be consid- tion of blood loss after cesarean section and vaginal delivery
ered normotensive.36 The SI represents a more reliable indi- has low validity with a tendency to exaggeration. Acta Obstet
cator of hemodynamic changes secondary to blood loss. Et Gynecol Scand. 2006;85(12):1448–1452.
7. Yoong W, Karavolos S, Damodaram M, et al. Observer accu-
Values over 0.9 have been associated with higher bleeding
racy and reproducibility of visual estimation of blood loss in
and mortality in trauma patients.38 39An elevated SI greater
obstetrics: how accurate and consistent are health-care pro-
than 0.9 has been shown to be associated with massive trans- fessionals? Arch Gynecol Obstet. 2010;281(2):207.
fusion,40 and is a strong predictor of ICU admissions41 and 8. Dildy GA, Paine AR, George NC, Velasco C. Estimating blood
adverse outcomes in women with PP hemorrhage.42 Further loss: can teaching significantly improve visual estimation?
prospective studies are needed to validate the role of SI in Obstet Gynecol. 2004;104(3):601–606.
obstetric population, but it is a promising index that could be 9. Bose P, Regan F, Paterson Brown S. Improving the accuracy of
estimated blood loss at obstetric haemorrhage using clinical
combined with the Maternal Early Warning Criteria and thus
reconstructions. BJOG Int J Obstet Gynaecol. 2006;113(8):919–924.
lead to earlier intervention and action. 10. Prasertcharoensuk W, Swadpanich U, Lumbiganon P. Accu-
Ultimately, even in the setting of normal vital signs, initiation racy of the blood loss estimation in the third stage of labor. Int
of treatment for suspected or evolving PPH should not be J Gynecol Obstet. 2000;71(1):69–70.
delayed. When a clinically significant hemorrhage is identified, 11. Maslovitz S, Barkai G, Lessing JB, Ziv A, Many A. Improved
even without change in vital signs, medical management with accuracy of postpartum blood loss estimation as assessed by
uterotonic medications should be initiated, adequate intrave- simulation. Acta Obstet Et Gynecol Scand. 2008;87(9):929–934.
12. Toledo P, Eosakul ST, Goetz K, Wong CA, Grobman WA. Decay
nous access should be established and transfusion should be
in blood loss estimation skills after web-based didactic train-
promptly started. With continued bleeding or abnormal vital ing. Simul Healthcare. 2012;7(1):18–21.
signs, aggressive escalation of therapy should be undertaken 13. Patton K, Funk DL, McErlean M, Bartfield JM. Accuracy of esti-
including massive transfusion protocol and transfer to the oper- mation of external blood loss by EMS personnel. J Trauma
ating room for surgical interventions (Table 2).43 Acute Care Surg. 2001;50(5):914–916.
14. Adkins AR, Lee D, Woody DJ, White WA. Accuracy of blood
loss estimations among anesthesia providers. AANA J.
2014;82(4):300–306.
Conclusion 15. Rothermel LD, Lipman JM. Estimation of blood loss is inaccu-
rate and unreliable. Surgery. 2016;160(4):946–953.
In conclusion, postpartum hemorrhage is a complicated, mul- 16. Lyndon A. Cumulative quantitative assessment of blood loss.
tifactorial process. Accurate quantification of blood loss is CMQCC Obstet Hemorrhage Toolkit Vers. 2015;2:80–85.
TAGEDENS E M I N A R S I N P E R I N A T O L O G Y 43 (2019) 11 17 17

17. Gabel KT, Weeber TA. Measuring and communicating blood 30. Bingham D, Melsop K, & Main E. CMQCC Obstetric Hemor-
loss during obstetric hemorrhage. J Obstet Gynecol Neonatal rhage Hospital Level Implementation Guide. The California
Nurs. 2012;41(4):551–558. Maternal Quality Care Collaborative (CMQCC) 2010.
18. Toledo P, McCarthy RJ, Hewlett BJ, Fitzgerald PC, Wong CA. 31. https://www.acog.org/-/media/Districts/District-II/Public/
The accuracy of blood loss estimation after simulated vaginal SMI/v2/HEMRiskAssessmentLDAdminIntrapartum.pdf?
delivery. Anesth Analg. 2007;105(6):1736–1740. dmc=1&ts=20180714T0246059507
19. Patel A, Goudar SS, Geller SE, et al. Drape estimation vs. visual 32. American College of Surgeons Trauma Committee. Advanced
assessment for estimating postpartum hemorrhage. Int J Gynecol Trauma Life Support For Doctors. 8th ed. Chicago, IL: Ameri-
Obstet. 2006;93(3):220–224. can College of Surgeons; 2008.
20. Al Kadri HM, Al Anazi BK, Tamim HM. Visual estimation versus 33. Brasel KJ, Guse C, Gentilello LM, Nirula R. Heart rate: is it truly
gravimetric measurement of postpartum blood loss: a prospec- a vital sign? J Trauma. 2007;62:812–817.
tive cohort study. Arch Gynecol Obstet. 2011;283(6):1207–1213. 34. Victorino GP, Battistella FD, Wisner DH. Does tachycardia cor-
21. Zhang WH, Deneux-Tharaux C, Brocklehurst P, Juszczak E, relate with hypotension after trauma? J Am Coll Surg.
Joslin M, Alexander S. Effect of a collector bag for measurement 2003;196:679–684.
of postpartum blood loss after vaginal delivery: cluster rando- 35. Guly HR, Bouamra O, Little R, et al. Testing the validity of the
mised trial in 13 European countries. Bmj. 2010;340:c293. ATLS classification of hypovolaemic shock. Resuscitation.
22. Hancock A, Weeks AD, Lavender DT. Is accurate and reliable 2010;81(9):1142–1147.
blood loss estimation the ’crucial step’ in early detection of 36. Pacagnella RC, Souza JP, Durocher J, et al. A systematic review
postpartum haemorrhage: an integrative review of the litera- of the relationship between blood loss and clinical signs. PLOS
ture. BMC Pregnancy Childbirth. 2015;15(1):230. one. 2013;8(3):e57594.
23. Hamm RF, Wang E, Romanos A, O’rourke K, Srinivas SK. 37. Mhyre JM, D’oria R, Hameed AB, et al. The maternal early warn-
Implementation of quantification of blood loss does not ing criteria: a proposal from the national partnership for mater-
improve prediction of hemoglobin drop in deliveries with nal safety. J Obstet Gynecol Neonatal Nurs. 2014;43(6):771–779.
average blood loss. Am J Perinatol. 2018;35(02):134–139. 38. Cannon CM, Braxton CC, Kling-Smith M, Mahnken JD, Carlton
24. Atukunda EC, Mugyenyi GR, Obua C, et al. Measuring post- E, Moncure M. Utility of the shock index in predicting mortal-
partum haemorrhage in low-resource settings: the diagnostic ity in traumatically injured patients. J Trauma Acute Care Surg.
validity of weighed blood loss versus quantitative changes in 2009;67(6):1426–1430.
hemoglobin. PLOS one. 2016;11(4):e0152408. 39. Olaussen A, Blackburn T, Mitra B, Fitzgerald M. Shock Index
25. Konig G, Holmes AA, Garcia R, et al. In vitro evaluation of a for prediction of critical bleeding post trauma: a systematic
novel system for monitoring surgical hemoglobin loss. Anesth review. Emerg Med Australas. 2014;26(3):223–228.
Analg. 2014;119(3):595. 40. Sohn CH, Kim WY, Kim SR, et al. An increase in initial shock
26. Holmes AA, Konig G, Ting V, et al. Clinical evaluation of a index is associated with the requirement for massive transfu-
novel system for monitoring surgical hemoglobin loss. Anesth sion in emergency department patients with primary post-
Analg. 2014;119(3):588. partum hemorrhage. Shock. 2013;40(2):101–105.
27. Sharareh B, Spencer Woolwine SS, Abraham P, Schwarzkopf 41. Nathan HL, El Ayadi A, Hezelgrave NL, et al. Shock index: an
R. Real time intraoperative monitoring of blood loss with a effective predictor of outcome in postpartum haemorrhage?
novel tablet application. Open Orthop J. 2015;9:422. BJOG Int J Obstet Gynaecol. 2015;122(2):268–275.
28. Doctorvaladan SV, Jelks AT, Hsieh EW, Thurer RL, Zakowski 42. El Ayadi AM, Nathan HL, Seed PT, et al. Vital sign prediction of
MI, Lagrew DC. Accuracy of blood loss measurement during adverse maternal outcomes in women with hypovolemic
cesarean delivery. AJP Rep. 2017;7(2):e93. shock: the role of shock index. PLOS One. 2016;11(2):e0148729.
29. Main EK, Goffman D, Scavone BM, et al. National partnership 43. https://www.acog.org/-/media/Districts/District-II/Public/SMI/v3/
for maternal safety: consensus bundle on obstetric hemor- 18sm07a170130StagesHemorrCheck3.pdf?dmc=1&ts=20180909T
rhage. J Obstet Gynecol Neonatal Nurs. 2015;44(4):462–470. 2026018285

You might also like