You are on page 1of 9

Analytica Chimica Acta 1000 (2018) 163e171

Contents lists available at ScienceDirect

Analytica Chimica Acta


journal homepage: www.elsevier.com/locate/aca

The differences in matrix effect between supercritical fluid


chromatography and reversed phase liquid chromatography coupled
to ESI/MS
Alfred Svan a, *, Mikael Hedeland a, b, Torbjo
€ rn Arvidsson a, c, Curt E. Pettersson a
a
Division of Analytical Pharmaceutical Chemistry, Uppsala University, BMC, Box 574, SE-751 23 Uppsala, Sweden
b
National Veterinary Institute (SVA), Dept. of Chemistry, Environment and Feed Hygiene, SE-751 89 Uppsala, Sweden
c
Medical Products Agency, Box 26, SE-751 03 Uppsala, Sweden

h i g h l i g h t s g r a p h i c a l a b s t r a c t

 Matrix effects were compared using


screening methods with SFC/ESI-MS
and RPLC/ESI-MS.
 Blood plasma, horse urine and
influent/effluent wastewater were
investigated.
 Through the use of post-column in-
fusions, matrix effect profiles were
generated.
 Quantitative and qualitative infor-
mation was compared, interferences
tentatively identified.
 Ion suppressions were generally
more common for SFC, and en-
hancements for LC.

a r t i c l e i n f o a b s t r a c t

Article history: For many sample matrices, matrix effects are a troublesome phenomenon using the electrospray ioni-
Received 8 August 2017 zation source. The increasing use of supercritical fluid chromatography with CO2 in combination with the
Received in revised form electrospray ionization source for MS detection is therefore raising questions: is the matrix effect
3 October 2017
behaving differently using SFC in comparison with reversed phase LC? This was investigated using urine,
Accepted 18 October 2017
Available online 23 October 2017
plasma, influent- and effluent-wastewater as sample matrices. The matrix effect was evaluated using the
post-extraction addition method and through post-column infusions. Matrix effect profiles generated
from the post-column infusions in combination with time of flight-MS detection provided the most
Keywords:
Matrix effects
valuable information for the study. The combination of both qualitative and semi-quantitative infor-
Supercritical fluid chromatography mation with the ability to use HRMS-data for identifying interfering compounds from the same exper-
Electrospray ionization iment was very useful, and has to the authors' knowledge not been used this way before. The results
Liquid chromatography showed that both LC and SFC are affected by matrix effects, however differently depending on sample
Ion enhancement matrix. Generally, both suppressions and enhancements were seen, with a higher amount of enhance-
Ion suppression ments for LC, where 65% of all compounds and all sample matrices were enhanced, compared to only 7%
for SFC. Several interferences were tentatively identified, with phospholipids, creatinine, and metal ion
clusters as examples of important interferences, with different impact depending on chromatographic

Abbreviations: E-WW, effluent wastewater; I-WW, influent wastewater; PCI,


post-column infusion; PCIMP, post-column infusion matrix profiles; PEA, post-
extraction addition.
* Corresponding author.
E-mail address: alfred.svan@farmkemi.uu.se (A. Svan).

https://doi.org/10.1016/j.aca.2017.10.014
0003-2670/© 2017 Elsevier B.V. All rights reserved.
164 A. Svan et al. / Analytica Chimica Acta 1000 (2018) 163e171

technique. SFC needs a different strategy for limiting matrix interferences, owing to its almost reverse
retention order compared to RPLC.
© 2017 Elsevier B.V. All rights reserved.

1. Introduction both studies ME was quantified using SRM with a tandem quad-
rupole mass spectrometer, finding generally lower MEs for SFC-MS
Matrix effects (ME) constitute a well-known phenomenon, than for LC-MS.
mainly when using electrospray ionization (ESI) connecting liquid The aim of this study was to investigate how the matrix effects
chromatography (LC) to mass spectrometry (MS). ME has been in ESI/MS differ between SFC and reversed phase LC, using drug
called the Achilles heel of LC/ESI-MS, and is generally described as substances, four complex sample matrices and commonly used
an alteration of the ionization efficiency by co-eluting molecules screening conditions for both techniques.
[1]. The term matrix effects includes both enhancement and sup-
pression of the signal and although the underlying mechanisms are 2. Material and methods
not fully understood, several theories exist [2]. These signal alter-
nations can lead to incorrect quantifications and dramatically Since the aim was not to develop and validate methods for
higher LODs, since e.g. a severe signal suppression might leave only screening and/or compound quantification, the conditions used
a few percent of the signal in comparison with that of a neat were gathered from the literature. For LC, a short UHPLC C18-
standard [3]. The most common method to handle ME is through column was chosen, often used for gradient elution in modern
normalization with isotopically labelled internal standards or screening methods, with eluents of water and methanol containing
reduction of the matrix interferences through sample preparation, 0.1% (v/v) formic acid as primary mobile phase. For SFC, a 2-picolyl-
however with an increase in cost and time consumption. Despite amine column was chosen, using CO2 and an 8e45% gradient of the
the use of matrix matched calibrators a quantification might be modifier containing methanol and ammonium formate.
problematic without a suitable internal standard, due to the relative The set of drug compounds chosen for this study was selected to
matrix effect: i.e., the signal difference between samples containing include different chemical properties and to ensure a spread in
the same matrix but of different origin, e.g. blood plasma from retention times over the chromatographic time scale. Some sub-
different individuals [4,5]. An alternation of the chromatographic stances were also included owing to previously observed severe
method to example avoid co-elution of interferences with the an- matrix effects in one of the study matrices, e.g. gemfibrozil [20],
alyte could also be an alternative. In multiresidue methods com- acetazolamide and miconazole [18].
mon in e.g. environmental analysis, this might however be
problematic due to the high number of compounds, which also may 2.1. Chemicals and reagents
limit the amount of internal standards.
Several sample materials analyzed in e.g. bioanalytical and Ammonia solution (4.0 M in methanol), ammonium acetate
environmental analysis are known to contain components which (Bio-extra 98%), ammonium formate (Fluka, LC-MS, 99%), formic
often give rise to ME, such as blood plasma, urine, wastewater acid (Fluka, LC-MS grade), methanol (Fluka Chromasolv, LC-MS
(WW) and extracts from food, soil and tissues [2,6]. Even if the ME grade), sodium chloride (99.5), acetazolamide (>99%), amiloride
may be decreased with sample preparation, such as extraction or hydrochloride-hydrate (98%), atenolol (98%), carbamazepine
sometimes simply by dilution, its impact on the analytical results (>99%), diclofenac sodium (>99), enalapril maleate (98%), fluox-
has to be tested. This is generally done by an estimation of the etine hydrochloride (99.9%, Riedel-de Hae €n), gemfibrozil
matrix effect influence, as recommended when validating analyt- (98.5%), hydrochlorthiazide (99%), mefenamic acid (98%),
ical methods according to e.g. EMA or FDA guidelines [7,8]. metoprolol tartrate (99%), miconazole (99%), propranolol hy-
Generally two different methods are used for ME determination, drochloride (99%), were purchased from Sigma-Aldrich (St. Louis,
where quantification of ME through post-extraction addition (PEA) MO, USA). Acetonitrile (LC-MS grade) was purchased from Fischer
is the most common, providing the ME% for a specific compound at scientific (Pittsburgh, PA, USA). Water was obtained from a Milli-Q
its retention time [4]. The second method, with post-column Q-POD-system from Millipore (Billerica, MA, USA). Carbon dioxide
infusion (PCI) of the analyte to determine the effect on signal in- (purity 99.99%) was obtained from Air Liquide (Paris, France).
tensity during the whole chromatographic run, is mainly qualita-
tive [9]. A combination of these two methods, forming the so-called 2.2. Sample preparation
matrix effect profile, has also been described using PCI experiments
with standard and matrix injections to determine the ME% for each 2.2.1. Plasma samples
data point in the chromatogram [10]. Pooled human plasma with Na-EDTA (3H Biomedical, Uppsala,
The new generation of SFC instruments with wider application Sweden) kept at 80  C during storage, was thawed in room
possibilities through the use of CO2 and polar organic solvents and temperature. The proteins were precipitated by mixing 500 mL
additives has created a new interest for this technique [11]. The plasma with 1000 mL ice cold acetonitrile for 15 s, and centrifuga-
electrospray ion source is generally used to combine SFC with MS tion for 5 min (12 100 g). A portion of the supernatant (1200 mL)
[12], which also is the interface where ME has its highest impact. was removed and evaporated at 40  C under a N2 gas stream. The
Few studies have however been published discussing the matrix samples used for SFC analysis were reconstituted in 500 mL aceto-
effects using SFC/ESI-MS. Most applications have only included ME nitrile:water 75:25 (v/v) and the LC samples in 500 mL water with
as a validation parameter, evaluated through an estimation of the 0.1% formic acid (v/v) and transferred to vials.
ME using PEA [13e15] or PCI [16,17]. To best of the authors'
knowledge, two studies comparing ME in SFC and LC-MS have been
2.2.2. Urine samples
presented [18,19], developed for doping control of urine samples. In
Surplus blank horse urine was obtained from National
A. Svan et al. / Analytica Chimica Acta 1000 (2018) 163e171 165

Veterinary Institute (SVA, Uppsala, Sweden) and stored in a freezer picolyl-amine) 1.7 mm, 100  3.0 mm column (Waters) at 50  C. A
(26  C) prior to analysis. After thawing in water bath, 100 mL were 515 HPLC-pump (Waters) added a post column flow of methanol at
transferred to Eppendorf tubes and diluted 1:9 (v/v) with the in- a flowrate of 0.45 mL min1 through a T-connection, prior a second
jection medium 900 mL acetonitrile:water 75:25 (v/v) for SFC and T-connection splitting the flow between the back-pressure regu-
900 mL water þ 0.1% formic acid (v/v) for LC. The tubes were mixed lator and the MS (for dimensions and schedule, see Fig. S-1). The
using vortex-mixer and centrifuged 10 min (12 100 g). The super- mobile phase consisted of CO2 and a modifier gradient: methanol
natants (300 mL) were transferred to vials for analysis. with 20 mM ammonium formate and 2% (v/v) water, made from a
stock solution of 1.0 M ammonium formate in water. The alterna-
2.2.3. Influent and effluent wastewater samples tive modifier contained methanol and 20 mM NH3. The 10 min
The influent (raw, I-WW) and effluent (treated, E-WW) waste- gradient started with 0e6 min from 8% modifier to 45% linearly,
water samples were collected as grab-samples in amber glass hold at 45% between 6 and 8.5 min, before equilibration at 8%
bottles at Kungsa€ngsverket wastewater treatment plant (Uppsala, modifier between 8.5 and 10 min. The flow rate was 1 mL min1,
Sweden). The water was filtered (0.7 mm glass fiber filter, Millipore) the back-pressure 150 bar, and the injection volume 4 mL. As weak
and pH-adjusted to pH 4 using formic acid. The SPE was performed wash, 1200 mL 2-propanol was used, with 800 mL methanol as
using Oasis® HLB columns (6 cm3 200 mg, Waters Corporation, strong wash.
Milford, MA, USA) and a general protocol, using 300 mL of effluent The MS detection was performed using the same detectors and
or 100 mL influent WW. After sample application and 5 min drying settings for both SFC and LC.
with vacuum, 5.0 mL water þ5% (v/v) methanol was used as wash, The Q-ToF/ESI MS system was used in both positive and negative
prior to a second drying (10 min) and the elution using 8.0 mL ionization mode, scanning from m/z 50 to 1200 using MSE. Using
methanol. The methanol was evaporated at 40  C and under a this scan-mode, the scans were altered between the use of low
stream of N2 gas. When dry, the samples were reconstituted as the collision energy and every second scan with high collision energy
plasma samples (see above). (20 eV in the transfer-cell together with a ramp of 20e40 eV in the
All samples were prepared in triplicate and analytical blanks trap-cell) for the Q-ToF. Scan time was set to 0.2 s, data were
were made for each matrix and chromatographic method, con- collected in resolution mode, and stored as centroid data. In ESIþ,
sisting of a corresponding volume of pure water treated as the the capillary voltage was set to 0.8 kV, the cone voltage to 40 V, the
samples. When performing the matrix effect semi-quantification source temperature to 120  C, and the nitrogen desolvation gas
using the PEA-method, a compound mixture containing 500 nM flow to 800 L h1 with a temperature of 650  C. In ESI-, a capillary
of the study compounds were prepared in water þ0.1% formic acid voltage of 2.0 kV was used, with remaining settings as for ESIþ.
for LC and acetonitrile:water 75:25 (v/v) for SFC. The mixture was The QqQ/ESI MS system was run in the selected reaction
used as reference and as reconstitution solvent for the WW and monitoring (SRM) mode using one precursor and one product ion
plasma samples and as dilution solvent for the urine samples, when for each compound (Tables Se1). In ESIþ, the capillary voltage was
semi-quantifying ME. To prepare the urine sample reference, 100 mL 3.0 kV, source temperature 120  C, desolvation gas flow 900 L h1
purified water was added instead of urine. with a temperature of 450  C. In ESI-, a capillary voltage of 2.5 kV
was used, with remaining settings as for ESIþ. When post-column
2.3. Instrumentation and analysis infusions were performed using QqQ/ESI MS, one single compound
was monitored per analysis, using a dwell time of 0.5 s. Both MS-
The LC/ESI-Q-ToF MS system consisted of an I-class Acquity instruments were equipped with ion sources using Z-spray
UPLC system connected to a Synapt G2-S Q-ToF mass spectrometer, geometry.
from Waters. A Kinetex C18 1.3 mm, 50  2.1 mm column (Phe- Instrument control and data evaluation for all instruments were
nomenex, Torrance, CA, USA) with a guard column with the same performed using the Mass-Lynx software (version 4.1, from
stationary phase was used and kept at 50  C. A 10 min gradient with Waters).
A: 0.1% (v/v) formic acid in water and B: 0.1% (v/v) formic acid in
methanol was used, at a flow rate of 0.3 mL min1. The gradient 2.4. Matrix effect determination
consisted of following steps: 4% B in 1 min, 1e3 min to 32% B lin-
early, holding 32% B in 1 min, 4e8 min to 85% B with curvature 8, The matrix effects (ME) were investigated using three methods.
8e9 min holding 85% B, and equilibrating 9e10 min with 4% B, the With the QqQ-system, the post extraction addition (PEA) method
starting condition. The injection volume was 7.5 mL. As sample was applied, semi-quantifying ME by comparing the response from
manager wash, 90:10 methanol:water was used for 10 s pre- and injected compound mixtures dissolved in pure solvent with the
12 s post-injection. Water:methanol 90:10 was used as purge- same mix dissolved in sample matrix, added after sample pre-
solvent. treatment. ME% was calculated in accordance with Matuszewski
Two mobile phases were used for both techniques, a primary for et al. [4]: ME ¼ B/A 100, where B corresponds to the peak area of a
all measurements and one alternative using the same gradient, for compound dissolved in a sample matrix, and A the corresponding
validating that the observed effect was not dependent on the mo- area from the compound in neat solvent.
bile phase. The alternative mobile phase used for LC consisted of A: The tandem quadrupole and ToF MS were both used for post-
5.0 mM ammonium acetate adjusted with ammonia to pH 8.5, and column infusion (PCI) experiments as described by Bonfiglio et al.
B: methanol. [9]: for LC, a 600 nM in H2O þ 0.1% formic acid mixture containing
The LC/ESI-QqQ MS system consisted of an Acquity UPLC con- three of the study compounds was infused using a syringe pump at
nected to a Quattro Micro tandem quadrupole mass spectrometer 5 mL min1, into the LC flow through a T-coupling, prior entrance to
(Waters). The same settings and conditions were used as for the LC the ESI probe. The chosen concentration was tested using calibra-
connected to the Q-ToF system, with the exception of the sample tion curves, to ensure linear response (Fig. S-2). For SFC, the com-
manager: the original Acquity UPLC system used a sample loop pound mixture was instead added to the post column make-up
instead of a “flow through needle”-system, as the I-class system. solution (6.7 nM). For both techniques, the signal intensity of the
The same SFC instrument (Acquity UPC2, Waters) was used for infused compounds were then monitored during injections of
both mass spectrometers. An external column block heater (Jones sample matrix and neat solvent. In addition to the generated
Chromatography, Wales, UK) was used to hold the Torus 2-PIC (2- infusion profiles, the signals from the PCI experiments were
166 A. Svan et al. / Analytica Chimica Acta 1000 (2018) 163e171

processed as first described by Stahnke et al. [10] to generate the positive ionization provided a higher overall signal intensity and
post column infusion matrix profiles (PCIMP): Initially the signal was more sensitive to ME [2]. The following results and discussion
was smoothed by using the mean intensity of five scans for every are mainly based on ESIþ, as the signal intensity achieved for ESI-
scan (SI(M5)). Thereafter, the means of each scan were calculated when affected by the sample matrices were too low to enable ac-
from the triplicates of the blank matrix injections and divided by curate measurements.
the mean of the corresponding scans from the analytical blank (see
Eq. (1)). The analytical blank was analyzed immediately prior to and 3.1. Plasma
after the samples in the sample list. By using the PCIMP, the
approximated ME% can be illustrated for every scan in the chro- The semi-quantification of ME deriving from PEA-experiments
matogram, and is therefore a hybrid of the traditional PCI and PEA using plasma LC/ESIþ QqQ MS/MS showed a strong enhancement
methods to measure ME [10]. (>50%) for 2 compounds, slight enhancement (15e50%) for 4
  compounds, whereas the rest were unaffected by ME (Table 1). The
SIM5 ðsample extractÞ signal of atenolol was enhanced by 30% (Fig. S-6), in accordance
MESI ð%Þ ¼  100  100 (1)
SIM5 ðanalytical blankÞ with the PCIMP using ToF MS, where an enhancement was
observed at the retention time for atenolol (Fig. 1). Fig. 1 also shows
When using ToF and PCI data, the signal of a compound was
that the different infused compounds are affected similarly by the
measured as the exact m/z with a window of 0.05 Da. In a few cases,
interferences, although compound related differences exist, one
the signal was obviously interfered by isobaric compounds, and the
example being the aforementioned enhancement for atenolol be-
m/z window was then narrowed to 0.02 Da.
tween 1 and 2.5 min which is not seen for the other compounds.
To identify background ions suspected to be involved or
This demonstrates that surrogate internal standards in multi-
responsible for ME, their retention time were compared with the
residue methods have to be experimentally evaluated, otherwise
retention time window affected of ME. If matching, their accurate
poor standardization and large errors might be the result.
m/z and product ions from the high energy function were used and
The semi-quantification of ME from plasma using SFC gave
compared with literature and freely available databases (MassBank,
generally small effects, with 4 compounds showing slight sup-
Human Metabolome Database, and METLIN). The identification was
pression (-15 - -50%) and one compound (amiloride) which was
therefore only tentative, since their structures were generally not
highly suppressed (<-50%). The overall semi-quantification results
confirmed using standards.
for plasma were very much in line with the summary of all matrices
(Table 1), with generally more suppression for SFC and enhance-
3. Results and discussion ment for LC.
The PCIMP generated from SFC/ESIþ ToF MS (Fig. 1) confirmed a
The sample preparation and chromatographic methods were suppression at the retention time of atenolol, and also showed
chosen to be representative for general screening methods of drugs several other areas with severe ion suppression. The strongly
in complex matrices often causing matrix effects. For LC, the suppressed amiloride with only 3% signal in the PEA-experiment
method chosen was containing column and mobile phases (Fig. S-6), has generally a similar ME% profile as the other com-
commonly used in screening methods [21e23], with a gradient pounds in the post-infusion experiments. Its retention time
intended to spread out the compounds over the chromatographic (6.04 min) is however in an area where all measured compounds
time scale (Fig. 1). The injection solvent was chosen to be less have close to 100% suppression (5.75e6.10 min).
eluting than the mobile phase, and similar to the starting compo- Fig. 2 summarizes the proposed major signal suppression
sition of the LC gradient. sources from plasma samples, tentatively identified using online
For SFC, the situation is more complicated, since the number of database and literature searches of accurate m/z values for pro-
published screening methods using MS-detection is so far only a tonated compounds and product ions. Although ToF MS has pre-
fraction of the number of LC-methods available. There is yet no such viously been used to create matrix effect profiles [30], this
thing as a standard column, and method development for SFC fo- integrated way of studying ME and identifying interferences has, to
cuses to a higher degree on column selection compared to LC [24]. author's knowledge, not been described before and provides a
The chemistry of the chosen 2-picolyl-amine (2-PIC) column is very useful tool for ME studies.
similar to that of the established 2-ethylpyridine-column. However, For LC (Fig. 2) high levels of polyethylene glycol (PEG) is
the 2-PIC provides better peak shapes [25], and it has been evalu- responsible for most of the suppression, due to the ability of the LC
ated using pharmaceutical-like compounds for screening [26,27]. method to efficiently spread out these polymers over the chro-
The injection solvent (75:25 acetonitrile:water) was chosen to in- matographic time scale. PEG is most likely a contamination origi-
crease the solubility of polar compounds, but still maintain an nating from the plastic plasma collection tubes [31]. The
acceptable peak shape [18]. A single column was chosen for SFC, anticoagulant used in the plasma (Na-EDTA), together with endo-
although the common practice in SFC method development is to genic polar compounds created a major suppression zone at t0,
scan several columns. The reason to this was to keep focus on the however it did not affect the first eluting compound acetazolamide
composition reaching the ESI source, and limit the scope to this at 1.60 min. The hydrophobic phospholipids eluted together at the
aspect, as column selectivity in SFC previously been investigated highest % of methanol in the gradient, and were therefore affecting
[24]. The choice of modifier and additive (20 mM ammonium gemfibrozil and mefenamic acid, although all hydrophobic com-
formate with 2% water in methanol (v/v)) was based on recently pounds eluting late would suffer from suppression according to the
published screening methods using SFC-MS [26,28,29]. The SFC ME% profiles (Fig. 1). Even though the drug compounds and sample
method spread out the compounds in almost reverse order matrices analyzed are the same for LC and SFC, the presumed cause
compared to LC (Fig. 1). The alternative mobile phases used for both of ME was rather different (Fig. 2). The PEGs, causing a scattered ion
LC and SFC were generally showing the same results as the primary, suppression zone in LC, were not well separated using the SFC
with differences in retention time. Therefore, results from the method and therefore only caused ion suppression in a single area
alternative mobile phases are only mentioned when significant of the chromatogram. The corresponding but reversed effect was
differences in ME between the mobile phases were observed. observed for the phospholipids in SFC, eluting together in LC, but
Both ESIþ and ESI- were used in the experiments, although scattered in SFC (Fig. 2). The highest level of suppression, co-eluting
A. Svan et al. / Analytica Chimica Acta 1000 (2018) 163e171 167

Fig. 1. The PCIMP from injections of pre-treated blank plasma samples, using ToF/ESI-MS with LC and SFC. The infusion profiles show ME in % after the injection of pre-treated blank
plasma, with ME ¼ 0 indicating no effect. The dotted lines illustrate the gradients used (B%). The chromatographic peaks are added from injections of standard mixtures, arbitrarily
set to 100% intensity.

Table 1 in comparison with SFC, where suppression was more common.


Summary of the PEA semi-quantification of ME for all four matrices, using QqQ MS. This was in consistency with our study, as seen in the summary of
Enhancement No ME Suppression all matrices (Table 1 and Fig. S-5).
Strong Slight Slight Strong
In LC, the PCIMP shows high suppression in conjunction with
the elution of the void, as might be expected from urine: polar
LC: Plasma 2 4 5 0 0
compounds as e.g. uric acid and carnitine eluted here (Fig. 3). A
LC: Urine 8 1 0 2 0
LC: I-WW 6 0 1 2 2 strong enhancement is seen for the signal of enalapril between 1
LC: E-WW 8 0 1 2 0 and 2 min in the PCIMP using the acidic mobile phase which is an
LC: Summary 24 5 7 6 2
exception among the study compounds, confirming that ME in
several cases is compound specific. This enhancement is not seen
SFC: Plasma 0 0 6 4 1
using the alternative alkali mobile phase, which generally created
SFC: Urine 1 1 3 2 4
SFC: I-WW 0 0 1 4 6 more suppression than the acidic mobile phase for urine samples
SFC: E-WW 1 0 1 6 3 using LC (Fig. 4).
SFC: Summary 2 1 11 16 14
For SFC, the majority of the interferences from urine eluted later
in comparison with those from the other matrices, reflecting the
n ¼ number of compounds, of totally 11. Strong enhancement: >50%, slight
hydrophilic character of the urinary constituents (Fig. 3). Fewer
enhancement: 15e50%, no ME: ±15%, slight suppression: 15- -50% and strong
suppression: <-50%. interferences could be tentatively identified in urine than in
plasma, but the alkali-metal clusters detected in plasma were
creating severe suppression also in the urine analysis, now joined
with e.g. amiloride was however caused by two clusters of different by an Mg2þ cluster. The metal ion content in urine is naturally
adducts involving sodium and potassium ions, naturally abundant fluctuating depending on homeostasis, and will therefore
in blood plasma. These highly suppressing clusters were unique for contribute to the sample-to-sample variation of this matrix.
the SFC experiments, but not for plasma. Urine samples gave rise to considerably more carry-over con-
In summary, both techniques suffered from ME, SFC mainly from taminants (comparing total ion chromatogram from PCI experi-
ion suppression and LC from both enhancement and suppression. ments) in SFC in comparison with other matrices and LC. This is
The interfering compounds causing these effects were however most likely also a reflection of the polar content of urine,
different and related to the chromatographic selectivity of the demanding a more intense autosampler washing procedure for the
techniques. normal phase-like SFC than for reversed phase LC.
Overall, the polar nature of urine created late-eluting in-
3.2. Urine terferences for SFC in comparison with LC. The suppression
observed in SFC was often severe, especially at the elution times of
Urine as a sample matrix is varying depending on source, when the different metal ion clusters (Fig. 3). SFC-MS analyses of urine
collected and what the subject had been eating and drinking. Due would therefore strongly benefit from a sample extraction prior
to the polar content of urine, LC-MS is the current technique of analysis, to remove salts and other very polar components.
choice for e.g. doping screening of urine samples. SFC has histori-
cally been used for more hydrophobic compounds, but was recently
evaluated and compared with LC for the analysis of doping com- 3.3. Effluent and influent wastewater
pounds in human urine [18]. In the study, enhancement was
observed to a higher degree for the urine samples analyzed with LC The PEA method for ME investigations is biased with
168 A. Svan et al. / Analytica Chimica Acta 1000 (2018) 163e171

Fig. 2. The blue solid line is the PCIMP for atenolol (extracted m/z 267.1703) in plasma, from experiments using LC/ESIþ ToF MS, and the green line shows corresponding data for
SFC/ESIþ ToF MS. ME% ¼ 0 means no effect on signal. Through investigation of background-ions, the origins of some of the major suppression zones are propose. (For interpretation
of the references to colour in this figure legend, the reader is referred to the web version of this article.)

Fig. 3. The solid lines are the PCIMP for atenolol, amiloride and enalapril during an injection of diluted urine, using ToF/ESI-MS with LC and SFC. ME% ¼ 0 means no effect on signal.
Through investigation of background-ions, the cause of some areas with major suppression is proposed.

enhancements in both effluent and influent WW, due to the total ion chromatogram (TIC) of the PCI using ToF MS shows that
occurrence of several compounds used in the society served by the the untreated wastewater (I-WW) contains more hydrophobic
WW-treatment plant (Figs. S3eS4). The occurrence of analytes in compounds (early eluting in SFC and late in LC) than the E-WW. The
the WW does however not substantially affect the PCI- high hydrophobic content is reflected in the PCIMP for SFC, where I-
experiments, more than through the peaks observed at the ex- WW gives more ion suppression during the first minutes of the
pected retention time of the monitored compound. chromatogram in comparison with E-WW (Fig. 5). The same
Although both the influent and effluent WW samples are pattern is seen for LC (Fig. 5), with an overall suppressed signal for
extracted using SPE, the high sample volumes (100 and 300 mL for I-WW, and slightly less for E-WW, with the largest difference in the
I-WW and E-WW respectively) and the complexity of the matrices end of the chromatogram. Some compounds responsible for ion
forms sample extracts with severe matrix effects. This gives a suppression in urine and plasma (e.g. carnitine) were detected at
generally higher signal intensity with more abundant peaks low intensities in I-WW, however not correlated to a matching loss
compared to the other matrices, despite the SPE (data not shown). in sensitivity. The overall high suppression affecting the signal in
Some differences could be noted between the matrices, e.g. the the I-WW samples was instead caused by other, unidentified,
A. Svan et al. / Analytica Chimica Acta 1000 (2018) 163e171 169

Fig. 4. PCIMP for enalapril (extracted m/z 377.2071) in urine samples, using LC/
ESIþ ToF. Comparison between the two different mobile phases, the original (blue line)
containing A: H2Oþ0.1% FA B: MeOHþ0.1% FA and the alternative phase (purple): A:
5.0 mM AcNH4 pH 8.5 B: MeOH. This is an exception: both the enhancement seen for
enalapril in the original mobile phase between 1 and 2.5 min, and the difference be-
tween the mobile phases, that normally showed similar ME, with variations in
retention. (For interpretation of the references to colour in this figure legend, the
reader is referred to the web version of this article.)

interferences.
The chromatographic orthogonality of reversed phase LC and
SFC techniques is also creating a similar but reversely directed ef- Fig. 5. A comparison of PCIMP generated from influent and effluent WW, using ami-
fect in comparison with the polar urine: the hydrophobic I-WW loride (extracted m/z 230.0557) and LC/ESIþ ToF MS (dark and light blue) and SFC/
shows generally less suppression using SFC compared to LC. This is ESIþ ToF MS (dark and light green). ME% ¼ 0 means no effect on signal. The retention
probably due to the strong elution of hydrophobic compounds early time of amiloride from an injection of a standard is marked by the red dashed line. (For
interpretation of the references to colour in this figure legend, the reader is referred to
in the gradient, washing these interferences away, while these the web version of this article.)
hydrophobic contaminants are more difficult to remove using LC.
This was confirmed by comparison of the TIC received from the
solvent blanks analyzed prior to and immediately after the I-WW would also explain why the enhancement mainly is observed early
samples. In SFC no carry-over was seen for I-WW, while high in- in the LC chromatogram: the methanol content is lower here and
tensity carry-over peaks were observed in the end of the LC- organic compounds from the sample matrix will therefore have a
gradient during several blank injections (data not shown). higher impact here. Finally, it would also explain why the
To investigate potential instrumental differences between the enhancement is seen more often with the QqQ MS equipped with
ToF MS and the QqQ MS/MS used in this study, a PCI experiment the older generation ESI source than for the newer ESI source on the
was performed using QqQ with SRM detection for the I-WW ToF mass spectrometer (Fig. S-7). This theory, if valid, might also
samples. The SFC analysis gave a high degree of similarity of the add important information on the mechanism behind
PCIMP between the QqQ and the ToF instruments, however the LC enhancement.
runs did not (Fig. S-7). Instead of the overall suppression received In summary, both techniques suffered from ion suppression
analyzing I-WW with ToF MS (Fig. 5), an enhancement was noted in when WW-samples were analyzed. The hydrophobic content in
parts of the chromatogram for QqQ MS. The SRM matrix effect especially I-WW could however provide some benefits for SFC over
profile for I-WW had similarities to the profiles received from the LC, due to the ability of SFC to remove these interferences more
urine and plasma-experiments with LC-ToF MS (Figs. 2 and 3) efficiently, providing less suppression in the end of chromatogram
showing areas of enhancement around 1e2.5 and 5e7 min. An and less memory-effects. The difference noted between the QqQ-
additional experiment using the quadrupole integrated in the ToF- and ToF-MS when performing PCI-experiments is most likely
MS instrument to monitor a single nominal precursor m/z at a time related to the different generations of the ESI sources.
was performed, ensuring that the difference seen did not originate
from the different detection (SRM for QqQ MS or MS-scan using ToF
MS) of the ions, but from the ionization source. Older generation 4. Conclusions
ESI-sources like the source on the QqQ MS has previously proven to
benefit more from a higher organic solvent content in the mobile When comparing the matrix effects from SFC and LC in combi-
phase than newer generation ESI sources, as the one on the ToF MS nation with ESI-MS, the PCIMP using ToF-MS and full-scan acqui-
system [32]. MS instruments with an older generation ESI has also sition was found very useful. Through this technique, both
showed a generally higher gain in S/N-ratio using the highly vola- qualitative and semi-quantitative data could be extracted, with the
tile CO2 in SFC than the water based reversed phase-LC, a difference ability to identify the interfering compounds from the same data.
not seen using a newer ESI source [18,28]. Except from direct This made it possible to not only see differences between the
interference with the ion formation in the liquid phase, it is chromatographic techniques, but also to help understanding what
believed that interfering compounds might affect the droplet for- caused them.
mation in ESI [33], possibly through modifications of boiling point, In our study, we could see that a sample matrix containing
viscosity and surface tension. The high content of organic com- mainly polar interferences, such as urine, will mainly affect the
pounds in the I-WW could hypothetically have a beneficial effect on eluting compounds in the first part of the chromatogram in
the ionization process, by e.g. lowering the surface tension. This reversed phase LC, while the first minutes were relatively unaf-
would explain why the enhancement is only seen for LC, as the fected for SFC. The chromatographic selectivity for interferences
mobile phase in SFC is already highly volatile due to the CO2. It will have a major impact on the matrix effect. For example, in
samples containing PEG contamination (e.g. plasma) the LC method
170 A. Svan et al. / Analytica Chimica Acta 1000 (2018) 163e171

was able to separate different PEGs, causing suppression over a chromatography with tandem mass spectrometry for the analysis of amino
acids and related compounds: comparing electrospray ionization and atmo-
wider time scale in comparison with SFC, where all PEGs eluted as
spheric pressure chemical ionization, Anal. Chim. Acta 981 (2017) 106e115,
one peak. The use of a different SFC column will however alter the https://doi.org/10.1016/j.aca.2017.05.005.
selectivity, and thereby also the retention time for the areas [13] R.A. Coe, J.O. Rathe, J.W. Lee, Supercritical fluid chromatography-tandem mass
affected by ME. The overall tendency of enhancement in LC and spectrometry for fast bioanalysis of R/S-warfarin in human plasma, J. Pharm.
Biomed. Anal. 42 (2006) 573e580, https://doi.org/10.1016/j.jpba.2006.05.025.
suppression in SFC, also observed in a previous study [18], could [14] Y. Tao, F. Dong, J. Xu, X. Liu, Y. Cheng, N. Liu, Z. Chen, Y. Zheng, Green and
also be confirmed. Hypothetically, this enhancement in LC/ESI-MS sensitive supercritical fluid chromatographic-tandem mass spectrometric
can be connected to an improved droplet formation, as this is method for the separation and determination of flutriafol enantiomers in
vegetables, fruits, and soil, J. Agric. Food Chem. 62 (2014) 11457e11464,
probably not improved by organic compounds using SFC where no https://doi.org/10.1021/jf504324t.
enhancement was seen. Although, this needs a more targeted [15] T. Berg, L. Kaur, A. Risnes, S.M. Havig, R. Karinen, Determination of a selection
investigation to be confirmed. of synthetic cannabinoids and metabolites in urine by UHPSFC-MS/MS and by
UHPLC-MS/MS, Drug Test. Anal. (2015), https://doi.org/10.1002/dta.1844 n/a-
Using an unselective sample preparation, there is probably no n/a.
quick-fix for matrix effects when analyzing compounds with [16] A. Svan, M. Hedeland, T. Arvidsson, J.T. Jasper, D.L. Sedlak, C.E. Pettersson,
varying chemical properties using the ESI source. For compounds Rapid chiral separation of atenolol, metoprolol, propranolol and the zwitter-
ionic metoprolol acid using supercritical fluid chromatographyetandem mass
with more similar properties, there might sometimes be some spectrometry e application to wetland microcosms, J. Chromatogr. A 1409
tricks. For example, when analyzing relatively hydrophobic com- (2015) 251e258, https://doi.org/10.1016/j.chroma.2015.07.075.
pounds found in a polar matrix as urine, SFC would probably pro- [17] Y. Hsieh, F. Li, C.J.G. Duncan, Supercritical fluid chromatography and high-
performance liquid chromatography/tandem mass spectrometric methods
vide both a short runtime and small impact from matrix. This is as
for the determination of cytarabine in mouse plasma, Anal. Chem. 79 (2007)
long as the compounds are eluting prior to the interferences, and 3856e3861, https://doi.org/10.1021/ac062441s.
the interferences are successfully washed away between analyses. [18] L. Novakova , M. Rentsch, A. Grand-Guillaume Perrenoud, R. Nicoli, M. Saugy,
SFC in combination with MS is still something relatively new in J. Veuthey, D. Guillarme, Ultra high performance supercritical fluid chroma-
tography coupled with tandem mass spectrometry for screening of doping
comparison to LC-MS, and future applications of SFC-MS will agents. II: analysis of biological samples, Anal. Chim. Acta 853 (2015)
contain both advantages compared to existing techniques, and 647e659, https://doi.org/10.1016/j.aca.2014.10.007.
[19] V. Desfontaine, L. Nova kova
, F. Ponzetto, R. Nicoli, M. Saugy, J.L. Veuthey,
difficulties, sometimes unexpected. The application to more polar
D. Guillarme, Liquid chromatography and supercritical fluid chromatography
analytes from water-based samples might for example bring to as alternative techniques to gas chromatography for the rapid screening of
some unexpected effects, as the metal cluster formations observed anabolic agents in urine, J. Chromatogr. A 1451 (2016) 145e155, https://
in this study when analyzing plasma and urine samples, created doi.org/10.1016/j.chroma.2016.05.004.
[20] M.J.M. Bueno, A. Agüera, M.J. Go mez, M.D. Hernando, J.F. García-Reyes,
heavily suppressed areas in the chromatogram. A.R. Fernandez-Alba, Application of liquid chromatography/quadrupole-linear
Ion trap mass spectrometry and time-of-flight mass spectrometry to the
Appendix A. Supplementary data determination of pharmaceuticals and related contaminants in wastewater,
Anal. Chem. 79 (2007) 9372e9384, https://doi.org/10.1021/ac0715672.
[21] H.H. Maurer, Current role of liquid chromatographyemass spectrometry in
Supplementary data related to this article can be found at clinical and forensic toxicology, Anal. Bioanal. Chem. 388 (2007) 1315e1325,
https://doi.org/10.1016/j.aca.2017.10.014. https://doi.org/10.1007/s00216-007-1248-5.
[22] R. Díaz, M. Ib ~ ez, J.V. Sancho, F. Herna
an ndez, Building an empirical mass
spectra library for screening of organic pollutants by ultra-high-pressure
References liquid chromatography/hybrid quadrupole time-of-flight mass spectrometry,
Rapid Commun. Mass Spectrom. 25 (2011) 355e369, https://doi.org/10.1002/
[1] P.J. Taylor, Matrix effects: the Achilles heel of quantitative high-performance rcm.4860.
liquid chromatographyeelectrosprayetandem mass spectrometry, Clin. Bio- [23] C. Bueschl, B. Kluger, M. Lemmens, G. Adam, G. Wiesenberger, V. Maschietto,
chem. 38 (2005) 328e334, https://doi.org/10.1016/j.clinbiochem.2004.11.007. A. Marocco, J. Strauss, S. Bo € di, G.G. Thallinger, R. Krska, R. Schuhmacher,
[2] A. Furey, M. Moriarty, V. Bane, B. Kinsella, M. Lehane, Ion suppression; a A novel stable isotope labelling assisted workflow for improved untargeted
critical review on causes, evaluation, prevention and applications, Talanta 115 LC-HRMS based metabolomics research, Metabolomics 10 (2014) 754e769,
(2013) 104e122, https://doi.org/10.1016/j.talanta.2013.03.048. https://doi.org/10.1007/s11306-013-0611-0.
[3] J.-P. Antignac, K. de Wasch, F. Monteau, H. De Brabander, F. Andre, B. Le Bizec, [24] E. Lesellier, C. West, The many faces of packed column supercritical fluid
The ion suppression phenomenon in liquid chromatographyemass spec- chromatography e a critical review, J. Chromatogr. A 1382 (2015) 2e46,
trometry and its consequences in the field of residue analysis, Anal. Chim. Acta https://doi.org/10.1016/j.chroma.2014.12.083.
529 (2005) 129e136, https://doi.org/10.1016/j.aca.2004.08.055. [25] V. Desfontaine, J.-L. Veuthey, D. Guillarme, Evaluation of innovative stationary
[4] B.K. Matuszewski, M.L. Constanzer, C.M. Chavez-Eng, Strategies for the phase ligand chemistries and analytical conditions for the analysis of basic
assessment of matrix effect in quantitative bioanalytical methods based on drugs by supercritical fluid chromatography, J. Chromatogr. A 1438 (2016)
HPLCMS/MS, Anal. Chem. 75 (2003) 3019e3030, https://doi.org/10.1021/ 244e253, https://doi.org/10.1016/j.chroma.2016.02.029.
ac020361s. [26] A. Dispas, P. Lebrun, P.-Y. Sacre , P. Hubert, Screening study of SFC critical
[5] A. Van Eeckhaut, K. Lanckmans, S. Sarre, I. Smolders, Y. Michotte, Validation of method parameters for the determination of pharmaceutical compounds,
bioanalytical LC-MS/MS assays: evaluation of matrix effects, J. Chromatogr. B J. Pharm. Biomed. Anal. 125 (2016) 339e354, https://doi.org/10.1016/
Anal. Technol. Biomed. Life Sci. 877 (2009) 2198e2207, https://doi.org/ j.jpba.2016.04.005.
10.1016/j.jchromb.2009.01.003. [27] E. Lemasson, S. Bertin, P. Hennig, H. Boiteux, E. Lesellier, C. West, Development
[6] H. Trufelli, P. Palma, G. Famiglini, A. Cappiello, An overview of matrix effects in of an achiral supercritical fluid chromatography method with ultraviolet
liquid chromatography-mass spectrometry, Mass Spectrom. Rev. 30 (2011) absorbance and mass spectrometric detection for impurity profiling of drug
491e509, https://doi.org/10.1002/mas.20298. candidates. Part II. Selection of an orthogonal set of stationary phases,
[7] EMA, Guideline on bioanalytical method validation, EMEA, Comm. Med. Prod. J. Chromatogr. A 1408 (2015) 227e235, https://doi.org/10.1016/
Hum. Use 44 (2012) 1e23 doi:EMEA/CHMP/EWP/192217/2009. j.chroma.2015.07.035.
[8] Guidance for industry: bioanalytical method validation, FDA, 2001. http:// [28] L. Novakov a, A. Grand-Guillaume Perrenoud, R. Nicoli, M. Saugy, J. Veuthey,
www.fda.gov/downloads/Drugs/GuidanceComplianceRegulatoryInformation/ D. Guillarme, Ultra high performance supercritical fluid chromatography
Guidances/UCM070107.pdf. Accessed 4 April 2017. coupled with tandem mass spectrometry for screening of doping agents. I:
[9] R. Bonfiglio, R. King, T. Olah, K. Merkle, The effects of sample preparation investigation of mobile phase and MS conditions, Anal. Chim. Acta 853 (2015)
methods on the variability of the electrospray ionization response for model 637e646, https://doi.org/10.1016/j.aca.2014.10.004.
drug compounds., Rapid Commun, Mass Spectrom. 13 (1999) 1175e1185, [29] M. Ishibashi, Y. Izumi, M. Sakai, T. Ando, E. Fukusaki, T. Bamba, High-
https://doi.org/10.1002/(SICI)1097-0231(19990630)13:12<1175::AID- throughput simultaneous analysis of pesticides by supercritical fluid chro-
RCM639>3.0.CO;2e0. matography coupled with high-resolution mass spectrometry, J. Agric. Food
[10] H. Stahnke, T. Reemtsma, L. Alder, Compensation of matrix effects by post- Chem. 63 (2015) 4457e4463, https://doi.org/10.1021/jf5056248.
column infusion of a monitor substance in multiresidue analysis with LCMS/ [30] O. Gonza lez, M. van Vliet, C.W.N. Damen, F.M. van der Kloet, R.J. Vreeken,
MS, Anal. Chem. 81 (2009) 2185e2192, https://doi.org/10.1021/ac802362s. T. Hankemeier, Matrix effect compensation in small-molecule profiling for an
[11] A. Tarafder, Metamorphosis of supercritical fluid chromatography to SFC: an LCeTOF platform using multicomponent postcolumn infusion, Anal. Chem. 87
Overview, TrAC - Trends Anal. Chem. 81 (2016) 3e10, https://doi.org/10.1016/ (2015) 5921e5929, https://doi.org/10.1021/ac504268y.
j.trac.2016.01.002. [31] R. Weaver, R.J. Riley, Identification and reduction of ion suppression effects on
[12] D. Wolrab, P. Frühauf, C. Gerner, Direct coupling of supercritical fluid pharmacokinetic parameters by polyethylene glycol 400, Rapid Commun.
A. Svan et al. / Analytica Chimica Acta 1000 (2018) 163e171 171

Mass Spectrom. 20 (2006) 2559e2564, https://doi.org/10.1002/rcm.2629. https://doi.org/10.1016/j.chroma.2014.06.066.


[32] A. Periat, I. Kohler, A. Bugey, S. Bieri, F. Versace, C. Staub, D. Guillarme, Hy- [33] R. King, R. Bonfiglio, C. Fernandez-Metzler, C. Miller-Stein, T. Olah, Mecha-
drophilic interaction chromatography versus reversed phase liquid chroma- nistic investigation of ionization supression in electrospray ionization, J. Am.
tography coupled to mass spectrometry: effect of electrospray ionization Soc. Mass Spectrom. 11 (2000) 942e950, https://doi.org/10.1016/S1044-
source geometry on sensitivity, J. Chromatogr. A 1356 (2014) 211e220, 0305(00)00163-X.

You might also like