DIO 1000 PA
Ultrasonic Flaw Detector
This manual may not be copied in whole or in any portion by any means without the
express written consent of STARMANS electronics, Ltd. The material in this manual
is for informational purposes and is subject to change without notice.
In no event will STARMANS electronics, Ltd. be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect in this documentation,
even if advised of the possibility of such damages.
Drawings and pictures in this manual contain confidential information and are
supplied by STARMANS electronics, Ltd., the owner of such information. In
accepting a drawing, the user agrees that it is for the user’s sole use. It will not be
reproduced or distributed to others and that the drawings or the information
contained herein will not be used in any manner detrimental to STARMANS
electronics, Ltd.
To switch ON the Phased Array Mode press the button “?” and press the control key
Phased Array.
To switch ON the wedge press the button “PA WEDGE OFF” to the position ON.
1.3 Aperture
To find the menu Aperture press the button “?” and press the button “PA Mode Chan
ALL”. The instrument DIO 1000 PA has 16 parallel channels. Therefore when “PA
Mode Chan ALL” is set up the 16 element probe works as 16 transmitters and 16
receivers.
To change the Aperture press the black part of the button “PA Mode Chan ALL”. The
PA Mode is changed to “PA MODE Chan SEL TX-RX” as you can see on the
following picture. When is selected this mode here is possible to change the Aperture
of probe.
The instrument has function of Angle Corrected Gain. To enable the function press
the button “?” and then the button “PA ACG OFF” to the position ON.
When function ACG is ON (on the screen is showed PA ACG DEFAULT) in the S-
scan (sector scan) the gain is automatically corrected for each angle.
When the PA Mode is switched ON, on the basic screen of the instrument appears
the following screen. When the wedge is also switched ON the refracted angle in the
steel is showed on the display.
When the S-scan function is switched ON then is possible to change the following
functions: PA Steps; Start and Stop of the Sector scan; PA Angle and PA Focus. By
changing the PA Angle is moving the cursor with displayed Gate 1 which is used for
measuring detected defects. (Note: A-scan is displayed bellow the S-scan in the
same time.)
When is adjusted in the SETTING submenu material thickness then in the S-scan is
displayed this thickness as 1T and doubled material thickness as 2T.
Other functions of instruments are the same as for DIO 1000 SFE and can be found
in the Operational Manual for DIO 1000 SFE. For instance on the following pictures
are showed DAC and DGS curves in phased array mode.