ECP 20-2PV
Technical specifications and illustrations are without M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
obligation, subject to modifications. 05.97/06.06 info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111 6.2
Example application for ECP 20-2
1 Heated filter sample probe SP2000-H 10
2 Heated sample line 4M4/6
3 Cooler ECP20-2 Testgas
IN
4 3-way ball valves 3L/PV-1 4
5 Peristaltic pumps SR25.1 8
6 Diaphragm pump MP47
7 Fine filter FP-2T-D with liquid alarm LA1
8 Aerosol filter CLF-5/W optional according to application ECP20-2 3 9
9 Flow meter FM10, 50-500 Nl/hr Sample Gass IN +5°C
10 Analyser, e. g. PMA1000
7 6
1 2
5
Alternative:
Trap with float or
Condensate OUT
condensate vessel
Dimensions
60
ø7
M&C Jet-Stream
heat exchanger
Cooling block
+5°C
280 385
C
ON
C
GL 25
PG 11 PG 9
Power -in Alarm contact
Condensate - OUT
Dimensions in mm
20
60 °C -1 40 °C -2 ence
40 °C -1 60 °C -3
16 for the gas cooler ECP20-2 with heat exchangers out of:
60 °C -2 40 °C -3
1 Glass
12
2 PVDF
3 Stainless steel 316Ti
8
in dependence on the gas flow rate and
40 °C and 60 °C sample inlet dew point
4 at 25 °C ambient temperature.
0 50 100 150 200 250 300 350 400 450 500 550
Sample gas flow rate Nl/hr
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany Technical specifications and illustrations are without
6.2 info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111 obligation, subject to modifications. 05.97/06.06
Technical Data
Gas Cooler series ECP® ECP 20-2
Part No. Gas Cooler ECP 20-2 01K8100 01K8140 01K8120
Material of heat exchanger Duran glass® PVDF SS 316Ti
Gas pressure max.-bar abs. 2) with GL adapter 2 / 32) 3 10**
Sample gas connection for ø 6 mm o.d. tube G1/4"i
Condensate connection for ø 12 mm o.d. tube G3/8"i
∆P at 500 Nl/hr flow rate 5 mbar
Stagnant space per heat exchanger approximately 70 ml
Gas stream / Gas flow rate 1x max. 500 Nl/hr *
Ambient temperature +5 to +45 °C
Storage temperature -20 to +60°C
Sample outlet dew point range of adjustment: +2 °C ..... +15 °C, factory setting: +5 °C
Dew point stability at constant conditions < ± 0,1 °C
Sample inlet temperature max. 180 °C*
Sample inlet dew point max. 80 °C*
Total cooling capacity at +25 °C ambient 80 kJ/h
Main power connection / Power consumption 230V 50Hz or 115V 60Hz switchable, 250VA
Ready for working <20 min.
Electrical connection terminals 2,5 mm2, 2x PG11 cable gland
Status alarm 1 changeover contact, potential free contact rating 250V, 2A, 500VA, 50W alarm point: ∆ ±3 °C to T SOLL
GL-adapter and tube fittings for the connection of different tube diameters at the heat exchanger see chapter 11, data sheet 11.5 and 11.6.
Technical specifications and illustrations are without M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
obligation, subject to modifications. 05.97/06.06 info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111 6.2