You are on page 1of 8

C L I N I C A L A N D E X P E R I M E N T A L

OPTOMETRY
RESEARCH PAPER

Optical quality comparison among different Boston contact lens materials

Clin Exp Optom 2016; 99: 39–46 DOI:10.1111/cxo.12323

Alberto Domínguez-Vicent MSc Background: The aim was to assess the optical quality of four Boston contact lens materials
Jose Juan Esteve-Taboada PhD with an optical device based on Schlieren interferometry.
Teresa Ferrer-Blasco PhD Methods: The NIMO TR1504 (Lambda-X, Nivelles, Belgium) was used to measure higher-
Santiago García-Lázaro PhD order aberrations and their corresponding root mean square values of four different rigid
Robert Montés-Micó PhD gas permeable contact lenses made from four different Boston materials: EO, ES, XO and
Optometry Research Group (GIO), Department of XO2. For each lens, 30 measurements were performed with two optical apertures: 3.0 mm
Optics, University of Valencia, Spain and 6.0 mm. The modulation transfer function, point spread function, Strehl ratio and a sim-
E-mail: alberto.vicent@uv.es
ulation of the image provided by the lens were computed from the Zernike coefficients mea-
sured up to the fourth order.
Results: The root mean square error of higher-order aberrations varied significantly with ma-
terial type for both optical apertures (p < 0.01). The largest difference was obtained between
the Boston EO and the Boston ES materials (for a 6.0 mm aperture), the mean difference be-
ing (8.3 ± 0.2) × 10-2 μm. The modulation transfer functions, point spread functions and Strehl
ratio values were similar among all Boston materials at the smaller optical aperture; however,
differences between each material were more apparent for the 6.0 mm aperture, with the Bos-
ton ES material exhibiting the best optical quality.
Submitted: 11 August 2014 Conclusions: In terms of all metrics analysed, all Boston materials examined showed compa-
Revised: 15 April 2015 rable optical quality for a 3.0 mm aperture but the Boston ES material displayed the best opti-
Accepted for publication: 19 April 2015 cal quality for a 6.0 mm optical aperture.

Key words: Boston materials, Boston optical quality, contact lens aberrations, rigid contact lenses

In vivo measurements have demonstrated the with the AERCOR chemical architecture. International Organization for Standardiza-
optical benefits of wearing rigid gas-perme- This permits the maintenance and increase tion (ISO) method for the measurement of
able (RGP) contact lenses1–4 in comparison of oxygen delivery while reducing silicone.7 monofocal contact lenses.10 The NIMO
to soft contact lenses and spectacles.2,4 Boston XO is a second generation fluoro sili- TR1405 has been used to measure the distri-
Roberts and colleagues5 reported that soft cone acrylate, which offers superpermeability8 bution of refractive power within the optical
contact lenses induced a significant increase and is as dimensionally stable as gas perme- zone of different soft contact lenses11 and to
in higher-order aberrations, when compared able lenses of much lower Dk. The newest assess the power profile of multifocal contact
with the naked eye and several studies have Boston material, Boston XO2, provides excel- lenses.12 No previous studies have used this
reported that RGP contact lenses improve lent oxygen permeability without compromis- device to assess the optical quality of RGP
the optical quality of the eye, especially those ing wettability, stability or comfort. Table 1 monofocal contact lenses.
dominated by corneal aberrations (for exam- summarises the main physical properties of The aim of this study was to assess the opti-
ple, keratoconus);1 however, several studies these materials; however, it has not yet been cal quality of four Boston contact lens mate-
have reported that the effect of RGP contact studied whether differences in the physical rials with the NIMO TR1405 to determine
lenses on the eye’s wavefront aberrations de- characteristics, chemical structure and com- whether differences in physical properties,
pends upon the habitual ocular aberra- position of these Boston RGP materials influ- polymer composition and chemical structure
tions.2,3,6 In this sense, if the habitual ence the optical quality of the lens. affect optical quality in vitro.
aberrations of the eye were high, RGP contact The NIMO TR1405 (Lambda-X, Nivelles,
lens correction would reduce them, whereas Belgium) is an optical device based on the METHOD
the opposite trend would be expected, if the Schlieren interferometric principle,9 which
habitual aberrations were low. can be used to perform in vitro measurements
Boston materials are used to manufacture to obtain the refractive power and optical ab- Contact lenses used
RGP contact lenses. The Boston EO and Bos- errations from spherical, toric and refractive The contact lenses studied were RGP
ton ES materials are manufactured with a multifocal contact or intraocular lenses. This monofocal BIAS-S (Conoptica, Barcelona,
fluoro silicone acrylate technology, combined device is more precise than any current Spain). Four lenses were included with the

© 2016 Optometry Australia Clinical and Experimental Optometry 99.1 January 2016
39
Optical quality among Boston materials Domínguez-Vicent, Esteve-Taboada, Ferrer-Blasco, García-Lázaro and Montés-Micó

coefficients (higher-order aberrations) and


BOSTON MATERIAL
the contact lens refractive power profile.
ES EO XO XO2 The measuring protocol was as follows.

Refractive index 1.443 1.429 1.415 1.424 1. The contact lens was removed from its blis-
Oxygen permeability (Dk) 18 58 100 141 ter and cleaned carefully with distilled water.
Oxygen transmissibility (Dk/t) 15 48 67 94 2. A small fan was used to remove all traces of
Silicone content 5-7 % 5-6 % 8-9 % 12-13 % distilled water without touching the lens.
o o o 3. The contact lens was transferred to the
Wetting angle 52 49 49 38o
NIMO’s dry cell cuvette with its back
Dynamic contact angle (advanced / receding) 52o / 50o 62o / 60o 59o / 58o 50o / 40o surface oriented downward (Figure 1).
This step was performed delicately using
Table 1. Material characteristics of each Boston material used in the study a pair of tweezers and special attention
was taken not to touch the optical zone
of the contact lens with the tweezers.
same optical design made from four different directions are used to characterize the con- 4. The contact lens was aligned with the
materials: Boston ES, Boston EO, Boston XO tact lens power within the optic zone.9–12 NIMO’s optical axis, and then one mea-
and Boston XO2 (Table 1). According to the With a single measurement, the NIMO is surement was taken.
data provided by the manufacturer, all lenses able to obtain the spherical and cylinder 5. Both the dry cell cuvette and the contact
had the same nominal characteristics: a back powers with their axes, Zernike lens were removed from the plate of the
vertex power of -3.00 D, a base curve of 7.90
mm, a total diameter of 9.60 mm and a nom-
inal centre thickness of 0.19 mm. The optical
quality of each material was assessed compar-
ing the optical outcomes of lenses fabricated
with the same design and nominal character-
istics. Only one lens made from each material
was examined to minimise the introduction
of possible bias due to the fabrication process
and this was not the purpose of the current
study. The fabrication process was assumed
to be within tolerance limits stipulated in
the ISO 18369-2:2006.13

Measurements
The NIMO TR1504 (Figure 1) uses a quan-
titative deflectometric technique and
combines the interferometric Schlieren
principle with a phase-shifting method9 to
measure the optical properties of contact
and intraocular lenses. A measuring light
with a maximum radiance peak of 546 nm
illuminates a liquid crystal display (LCD).
In addition, the image of the LCD, which
passes through the lens being measured,
is formed on a charged-coupled device
(CCD) camera, with a resolution of 1396
× 1040 pixels. The contact lens is placed
on the dry cuvette and the coupling lens-
cuvette is placed in the object plane of
the instrument (Figure 1). A sinusoidal pat-
tern is projected on the LCD, which is illu-
minated uniformly, when no lens is placed
in the object plane. Schlieren fringes are
projected on the CCD due to light devia-
tions or deflections once the lens is placed
in the object plane of the instrument. The
light beam deviation along both ‘x’ and ‘y’ Figure 1. Schematic layout of the NIMO TR1504

Clinical and Experimental Optometry 99.1 January 2016 © 2016 Optometry Australia
40
Optical quality among Boston materials Domínguez-Vicent, Esteve-Taboada, Ferrer-Blasco, García-Lázaro and Montés-Micó

measurement, and the contact lens was contact lens made of each material. The whereas, Boston EO displayed the largest
also removed from the dry cell cuvette. point spread functions were presented using value (2.38 ± 0.06) × 10-2 μm. Table 2 also
6. The contact lens was placed again in the a non-linear scale to enhance the side lobes includes the mean HORMS differences for
dry cell cuvette and the coupling lens- of the PSF functions. Finally, the Strehl ratio all between-material comparisons and their
cuvette was replaced on NIMO’s mea- was calculated in the frequency domain, as significance levels. The greatest difference
suring plate. the ratio between the area under the lens was observed between the Boston EO and
7. Steps four, five and six were repeated 30 MTF curve divided the area under the dif- Boston ES (0.92 × 10-2 μm, p < 0.01) and
times to obtain a total of 30 measurements fraction-limited MTF curve. the smallest difference was obtained
for each lens material. Prior to each mea- between Boston EO and Boston XO2 (0.16 ×
surement, the contact lens was realigned Statistical analysis 10-2 μm, p < 0.01).
with the NIMO’s optical axis. This proto- All data were analysed using SPSS v.17.0 Differences between Boston materials were
col was completed twice, for 3.0 mm and (IBM, New York, New York, USA). The more apparent when the optical aperture was
6.0 mm optical apertures. normality of the data was verified using set to 6.0 mm (Table 2). Statistically signifi-
the Shapiro-Wilk test and the variance cant differences (p < 0.01) were obtained
The root mean square (RMS), modula- equality with the Levene’s tests. A two- for all between-material comparisons. The
tion transfer function (MTF), point spread way analysis of variance (ANOVA) of four Boston ES displayed the smallest HORMS
function (PSF), and Strehl ratio were com- (materials) times two (optical apertures) value (10.14 ± 0.05) × 10-2 μm and the Boston
puted to assess the optical quality of each factorial analysis was performed. The de- EO material displayed the largest value
lens material with a custom-made MATLAB pendent variables were the RMS of each (18.40 ± 0.10) × 10-2 μm. Table 2 also includes
program (Mathworks, Natick, Massachu- Zernike coefficient, the HORMS error the mean HORMS differences for all be-
setts, USA). In addition, the convolution be- and the Strehl Ratio. The Holm-Sidak tween-material comparisons and their signifi-
tween an optotype eye chart and the point multiple comparison test was performed cance levels. The maximum difference was
spread function was also computed. The when the ANOVA revealed statistically sig- obtained between the Boston EO and Boston
average RMS value of each Zernike coeffi- nificant differences. Differences were con- ES materials (8.26 × 10-2 μm, p < 0.01),
cient was also calculated for the 30 measure- sidered statistically significant when the p whereas, the minimum difference was ob-
ments taken. In addition, the higher-order value was smaller than 0.01. tained between the Boston XO and Boston
root mean square error (HORMS) from XO2 (6.32 × 10-3 μm, p < 0.01).
third to fourth order, which includes trefoil Tables 3 and 4 show the RMS of each third-
RESULTS
(Z3-3; Z33), coma (Z3-1; Z31), tetrafoil (Z4-4; and fourth-order Zernike coefficient for each
Z44), secondary astigmatism (Z4-2; Z42) and Table 2 shows the HORMS errors for all material over the 3.0 mm and 6.0 mm optical
spherical (Z40) aberration, was also calcu- Boston contact lens materials for the 3.0 mm apertures. When the aperture was set to 3.0
lated. The MTFs were computed up to 60 and 6.0 mm optical apertures. Similar mm (Table 3), each Zernike coefficient term
cycles per degree and the retinal contrast HORMS values among each Boston material was similar among all four materials, as the
threshold values, which were measured at a were obtained for the 3.0 mm optical aperture mean difference for any given coefficient
retinal illuminance of 500 td (troland),14 with the standard deviation of each value was less than 0.003 μm, except for the spheri-
were also included in the modulation smaller than 2 × 10-3 μm. The multiple cal aberration term, where the mean differ-
transfer function curves. The intersection comparison test revealed statistically sig- ence was less than 0.03 μm. This table also
between each Boston MTF and the retinal nificant differences among all Boston ma- includes all between-materials comparisons
curve gives information about the neural terials (p < 0.01): Boston ES displayed the for each Zernike coefficient with statistically
cut-off frequency given by an eye wearing a smallest HORMS (1.46 ± 0.03) × 10-2 μm, significant differences (p < 0.01). In terms

Material 3.0 mm optical aperture 6.0 mm optical aperture

Mean ± SD Mean difference (significance) Mean ± SD Mean difference (significance)


-2 -2 -2
Boston EO (2.38 ± 0.06)·10 Boston EO-Boston ES: 0.92·10 (18.40 ± 0.10)·10 Boston EO-Boston ES: 8.26·10-2
-2 -2 -2
Boston ES (1.46 ± 0.03)·10 Boston EO-Boston XO: 0.56·10 (10.14 ± 0.05)·10 Boston EO-Boston XO: 5.47·10-2
Boston EO-Boston XO2: 0.16·10-2 Boston EO-Boston XO2: 6.11·10-2
-2 -2 -2
Boston XO (1.82 ± 0.04)·10 Boston ES-Boston XO: 0.36·10 (12.92 ± 0.08)·10 Boston ES-Boston XO: 2.79·10-2
-2
Boston ES-Boston XO2: 0.76·10 Boston ES-Boston XO2: 2.16·10-2
-2 -2 -2
Boston XO2 (2.22 ± 0.16)·10 Boston XO-Boston XO2: 0.40·10 (12.29 ± 0.07)·10 Boston XO-Boston XO2: 6.32·10-3

Table 2. Mean high-order root mean square value from third- and fourth-order for each Boston material at the 3.0 mm and 6.0 mm optical
apertures. The mean difference with its significance level was also included for all material pairs comparisons with statistical significant
differences. Significant differences were considered when p < 0.01. SD: Standard deviation. Both mean and mean differences were
expressed in μm.

© 2016 Optometry Australia Clinical and Experimental Optometry 99.1 January 2016
41
Optical quality among Boston materials Domínguez-Vicent, Esteve-Taboada, Ferrer-Blasco, García-Lázaro and Montés-Micó

of spherical aberration, all between-material


Zernike coefficient Zernike value, μm (Mean ± SD) Mean difference (significant level)
comparisons showed statistically significant
Z (3 , -3) Boston EO: (3.00 ± 2.00)·10-4 Boston EO – Boston ES: -0.78·10-3 differences: the Boston EO material demon-
-3
Boston ES: (1.08 ± 0.20)·10 Boston EO – Boston XO: -2.06·10-4 strated the largest spherical aberration value
-4 (2.36 ± 0.06) × 10-2 μm, whereas the Boston
Boston XO: (5.06 ± 2.00)·10 Boston EO – Boston XO2: 4.70·10-4
ES showed the smallest value (1.41 ± 0.02) ×
Boston XO2: (7.70 ± 3.00)·10-4 Boston ES – Boston XO: 0.58·10-3 10-2 μm. No statistically significant differences
Boston ES – Boston XO2: -0.31·10-3 (p > 0.01) were obtained in terms of vertical
-3
Z (3 , -1) Boston EO: (2.03 ± 1.60)·10 None coma among all Boston materials. The
Boston ES: (2.70 ± 0.60)·10-3 Boston EO showed the smallest value (2.0 ±
Boston XO: (2.38 ± 1.20)·10-3 1.6) × 10-3 μm; the Boston ES material demon-
strated the highest coma value (2.70 ± 0.60) ×
Boston XO2: (2.41 ± 0.80)·10-3
10-3 μm. For horizontal coma, statistically
Z (3 , 1) Boston EO: (2.21 ± 1.50)·10-3 Boston EO – Boston ES: 1.46·10-3 significant differences were obtained for all
-4
Boston ES: (7.51 ± 7.00)·10 Boston EO – Boston XO: 0.90·10-3 between-material comparisons (p < 0.01),
Boston XO: (1.31 ± 0.50)·10-3 Boston ES – Boston XO2: -2.23·10-3 except for the Boston EO – Boston XO2 and
-3
Boston XO2: (2.98 ± 0.80)·10 Boston XO – Boston XO2: -1.68·10-3 BostonES–BostonXOcomparisons(p> 0.05).
-4 On the other hand, when the optical aper-
Z (3 , 3) Boston EO: (2.55 ± 1.80)·10 Boston EO – Boston ES: -0.81·10-3
-3
ture was set to 6.0 mm (Table 4), the RMS er-
Boston ES: (1.07 ± 0.20)·10 Boston EO – Boston XO2: -0.91·10-3 rors corresponding to vertical and horizontal
Boston XO: (2.09 ± 1.03)·10-4 Boston ES – Boston XO: 0.87·10-3 coma were smaller than 2.5 × 10-2 μm; how-
-3
Boston XO2: (1.16 ± 0.40)·10 Boston ES – Boston XO2: -0.09·10-3 ever, the spherical aberration was larger than
Boston XO – Boston XO2: -0.95·10-3 0.1 μm and the other coefficients were smaller
Z (4 , -4) Boston EO: (6.85 ± 2.00)·10 -4
Boston EO – Boston XO2: 2.97·10-4 than 0.003 μm. Table 4 also displays all
between-Boston material comparisons with
Boston ES: (3.71 ± 2.00)·10-4 Boston ES – Boston XO2: -0.17·10-4
-4
statistically significant differences for each
Boston XO: (4.79± 3.00)·10 Boston XO – Boston XO2: 0.91·10-4 Zernike coefficient. The magnitude of spheri-
-4
Boston XO2: (3.87 ± 2.00)·10 cal aberration varied significantly between all
Z (4 , -2) Boston EO: (9.03 ± 2.15)·10-4 Boston EO – Boston ES: -0.28·10-4 Boston materials (p < 0.01): the Boston ES
Boston ES: (1.18 ± 0.20)·10-3 Boston EO – Boston XO: -0.69·10-3 displayed the smallest value (10.08 ± 0.03) ×
-3 10-2 μm, whereas the Boston EO showed the
Boston XO: (1.59 ± 0.60)·10 Boston EO – Boston XO2: 4.14·10-4
-4
largest value (18.25 ± 0.05) × 10-2 μm. On the
Boston XO2: (4.89 ± 1.60)·10 Boston ES – Boston XO: -0.41·10-3 other hand, vertical coma was statistically sig-
Boston ES – Boston XO2: 0.69·10-3 nificant for all between-material comparisons,
Boston XO – Boston XO2: 1.10·10-3 except for the Boston ES – Boston XO compar-
-2
Z (4 , 0) Boston EO: (2.36 ± 0.06)·10 Boston EO – Boston ES: 0.95·10-2 ison and comparable horizontal coma values
Boston ES: (1.41 ± 0.02)·10 -2
Boston EO – Boston XO: 0.57·10-2 (p > 0.01) were obtained between Boston ES
-2 – Boston XO2 and Boston EO – Boston XO.
Boston XO: (1.79 ± 0.06)·10 Boston EO – Boston XO2: 0.17·10-2
Figures 2 and 3 show the MTFs for each
Boston XO2: (2.18 ± 0.16)·10-2 Boston ES – Boston XO: -0.37·10-2 material at the 3.0 mm and 6.0 mm optical
Boston ES – Boston XO2: -0.77·10-2 apertures. Comparable curves were ob-
Boston XO – Boston XO2: -0.39·10-2 tained for the smallest optical aperture
-4
Z (4 , 2) Boston EO: (8.60 ± 3.0)·10 Boston EO – Boston ES: -0.94·10-3 (Figure 2), where the MTFs of all four Bos-
ton lens materials were close to the diffrac-
Boston ES: (1.80 ± 0.40)·10-3 Boston EO – Boston XO: 4.43·10-4
-4
tion-limited curve. Additionally, the Boston
Boston XO: (4.17 ± 3.00)·10 Boston ES – Boston XO: 1.38·10-3 ES material provided the closest match to
-4
Boston XO2: (8.06 ± 3.00)·10 Boston ES – Boston XO2: 0.99·10-3 the diffraction-limited MTF. For the 6.0
Boston XO – Boston XO2: -3.89·10-4 mm optical aperture (Figure 3), all curves
Z (4 , 4) Boston EO: (9.32 ± 7.00)·10-4 Boston EO – Boston ES: 1.37·10-4 moved away from the diffraction-limited
Boston ES: (7.95 ± 2.00)·10 -4
Boston EO – Boston XO: 2.60·10-4 MTF, with the MTF of Boston ES closest to
-4 the diffraction-limited MTF. The cut-off
Boston XO: (6.72 ± 4.00)·10 Boston ES – Boston XO2: 5.49·10-4
-4
frequencies were similar among all Boston
Boston XO2: (2.47 ± 1.07)·10 materials for the smallest optical aperture,
being about 50 cycles per degree. For the
6.0 mm optical aperture, the Boston ES
Table 3. Mean and standard deviation (SD) of each Zernike coefficient term from third- to
material displayed the largest cut-off fre-
fourth-order for all Boston materials at the 3.0 mm optical aperture. The third column
quency (about 40 cycles per degree) and
represents the mean difference and the level of significance for all between-material the Boston EO demonstrated the smallest
comparisons with statistically significant differences, which were considered when p < 0.01. frequency (about 25 cycles per degree).

Clinical and Experimental Optometry 99.1 January 2016 © 2016 Optometry Australia
42
Optical quality among Boston materials Domínguez-Vicent, Esteve-Taboada, Ferrer-Blasco, García-Lázaro and Montés-Micó

Similar Strehl ratio values were obtained


Zernike coefficient Mean ± SD (μm) Mean difference (significant level)
among Boston materials for the 3.0 mm op-
Z (3 , -3) Boston EO: (3.68 ± 2.00)·10-4 Boston EO – Boston ES: 0.19·10-4 tical aperture; however, statistically signifi-
-4
Boston ES: (3.49 ± 1.60)·10 Boston EO – Boston XO: -1.59·10-3 cant differences were obtained for all
-3 between-materials comparisons (p < 0.01).
Boston XO: (1.96 ± 0.70)·10 Boston EO – Boston XO2: -2.09·10-3
In general, the Strehl ratio values were
Boston XO2: (2.43 ± 0.70)·10-3 Boston ES – Boston XO: -1.61·10-3 close to one at the smallest aperture: the
Boston ES – Boston XO2: -2.09·10-3 Boston ES lens demonstrated the largest
Boston XO – Boston XO2: -0.47·10-3 value (0.96), whereas the Boston EO showed
-2
Z (3 , -1) Boston EO: (1.60 ± 0.20)·10 Boston EO – Boston ES: 6.96·10-3 the smallest value (0.89). When the optical
Boston ES: (9.00 ± 2.00)·10-3 Boston EO – Boston XO: 6.70·10-3 aperture was set to 6.0 mm, the Strehl ratio
-3 values also revealed statistically significant
Boston XO: (9.26 ± 4.00)·10 Boston EO – Boston XO2: -0.24·10-2
-2
differences for all between-material compar-
Boston XO2: (1.83 ± 0.20)·10 Boston ES – Boston XO2: -9.32·10-3 isons (p < 0.01): the Boston ES material
Boston XO – Boston XO2: -9.07·10-3 displayed the largest value (0.25) and Boston
Z (3 , 1) Boston EO: (2.01 ± 1.08)·10-2 None EO the smallest value (0.19).
Boston ES: (7.85 ± 5.00)·10-3 Figure 4 shows the point spread function
images of each Boston material for both opti-
Boston XO: (1.95 ± 0.70)·10-2
cal apertures. Compact (that is, similar to a
Boston XO2: (1.19 ± 0.30)·10-2 single point source), with high peak intensity
Z (3 , 3) Boston EO: (1.82 ± 1.60)·10-3 Boston EO – Boston ES: 1.47·10-3 PSFs were obtained for all Boston materials
-4
Boston ES: (3.57 ± 1.50)·10 Boston EO – Boston XO: -0.17·10-3 for the 3.0 mm optical aperture. Generally
-3
Boston XO: (1.99 ± 0.70)·10 Boston EO – Boston XO2: -0.24·10-3 speaking, this metric was similar among all
Boston XO2: (2.07 ± 0.80)·10 -3
Boston ES – Boston XO: -1.63·10-3 Boston materials. Broader point spread func-
tions with lower peak intensity were obtained
Boston ES – Boston XO2: -1.71·10-3
-3
when the optical aperture was set to 6.0 mm.
Z (4 , -4) Boston EO: (1.97 ± 0.50)·10 Boston EO – Boston XO: 1.43·10-3 The Boston ES material displayed the most
-3
Boston ES: (2.08 ± 0.30)·10 Boston ES – Boston XO: 1.55·10-3 compact PSF with the highest peak intensity,
-4
Boston XO: (5.31 ± 1.50)·10 Boston XO – Boston XO2: -1.18·10-3 while the Boston EO showed the broadest
Boston XO2: (1.71 ± 0.20)·10-3 PSF with the lowest peak intensity. The
Boston XO and XO2 materials displayed sim-
Z (4 , -2) Boston EO: (2.17 ± 0.80)·10-3 Boston EO – Boston XO: 1.11·10-3
-3
ilar point spread functions.
Boston ES: (1.78 ± 0.40)·10 Boston EO – Boston XO2: 1.49·10-3 Finally, Figure 5 shows the convolution of
-3
Boston XO: (1.06 ± 0.70)·10 Boston ES – Boston XO: 0.72·10-3 each PSF with an optotype chart. As ex-
Boston XO2: (6.85 ± 3.00)·10-4 Boston ES – Boston XO2: 1.09·10-3 pected, similar images were observed among
-2
Z (4 , 0) Boston EO: (18.25 ± 0.05)·10 Boston EO – Boston ES: 8.17·10-2 all Boston materials for the 3.0 mm optical ap-
Boston ES: (10.08 ± 0.03)·10 -2
Boston EO – Boston XO: 5.54·10-2 erture. When the optical aperture was set to
-2 6.0 mm, the Boston ES material displayed
Boston XO: (12.71 ± 0.02)·10 Boston EO – Boston XO2: 6.16·10-2
the best optotype chart and the Boston EO
Boston XO2: (12.09 ± 0.05)·10-2 Boston ES – Boston XO: -2.63·10-2 the worst. In addition, both Boston XO and
Boston ES – Boston XO2: -2.01·10-2 XO2 materials displayed similar images.
Boston XO – Boston XO2: 0.63·10-2
-3
Z (4 , 2) Boston EO: (1.14 ± 0.62)·10 Boston EO – Boston XO: -1.11·10-3
Boston ES: (1.48 ± 0.18)·10-3 Boston EO – Boston XO2: 0.36·10-3 DISCUSSION
-3 -3
Boston XO: (2.25 ± 0.50)·10 Boston ES – Boston XO: -0.77·10 The Zernike coefficients (Table 3), HORMS
Boston XO2: (7.81 ± 2.00)·10-4 Boston ES – Boston XO2: 0.70·10-3 errors (Table 2) and Strehl ratios were small
Boston XO – Boston XO2: 1.47·10-3 in absolute value but varied significantly be-
Z (4 , 4) Boston EO: (1.89 ± 1.09)·10-3 Boston EO – Boston ES: 0.16·10-3 tween all Boston materials for the 3.0 mm ap-
-3 erture. Nevertheless, these differences were
BostonES: (1.74 ± 0.40)·10 Boston EO – Boston XO2: 1.53·10-3
-3
not noticeable in the modulation transfer
Boston XO: (1.08 ± 0.30)·10 Boston ES – Boston XO: 0.66·10-3 functions (Figure 2), point spread functions
-4
Boston XO2: (3.67 ± 3.00)·10 Boston ES – Boston XO2: 1.37·10-3 (Figure 4) or the simulations of the images
provided by the different materials (Figure 5).
These statistically significant differences were
Table 4. Mean and standard deviation (SD) of each Zernike coefficient term from third- to most likely a result of the extremely low stan-
fourth-order for all Boston materials at the 6.0 mm optical aperture. The third column dard deviations obtained due to NIMO’s high
represents the mean difference and the level of significance for all between-material level of repeatability.10 Therefore, optical dif-
comparisons with statistically significant differences, which were considered when p < 0.01. ferences among these materials are most

© 2016 Optometry Australia Clinical and Experimental Optometry 99.1 January 2016
43
Optical quality among Boston materials Domínguez-Vicent, Esteve-Taboada, Ferrer-Blasco, García-Lázaro and Montés-Micó

likely not clinically significant over a 3.0 mm


optical aperture.
For the 6.0 mm aperture, the Boston ES
displayed the best optical quality among the
other Boston materials in terms of Zernike
coefficients, HORMS errors and Strehl ratio
values. The simulated images in Figure 5
(bottom row) suggest that these differences
may be clinically significant. This figure also
highlights that differences between the Bos-
ton XO and XO2 materials are most likely
not clinically significant. Since all lenses ex-
amined had the same design and nominal pa-
rameters, any observed optical differences
between these lens materials might be related
to the polymer structure of each material;
however, small manufacturing differences
may account for some of the observed optical
quality differences between materials, as the
ISO 18369-2: 200613 allows a variance of 0.05
mm in the back optic zone radius. Finally, it
should be taken into consideration that these
results were obtained under in vitro condi-
tions and neither the optical quality of the
Figure 2. Modulation transfer function versus spatial frequency expressed in cycles per de- eye nor the performance of the lens on eye
gree for a 3.0 mm optical aperture for Boston EO, ES, XO and XO2 materials. The diffrac- were taken into account during computa-
tion-limited curve and retinal contrast threshold curve obtained by Sekiguchi, Williams and tions. For this reason, further in vivo studies
Brainward14 at a retinal illuminance of 500 td are included. should be aimed to assess whether these
observed optical differences are clinically
significant to RGP wearers.
Applegate and colleagues15 studied the
effect of the RMS wavefront error on the
visual performance using a Hartman-Shack
wavefront sensor. They reported the number
of letters lost as a function of the RMS of the
whole eye. According to their results, RMS
wavefront errors of 0.10 μm, 0.15 μm and
0.20 μm correspond to mean letter losses
of two, three and four letters, respectively.
Based on Applegate and colleagues’ results,
differences in visual function between
Boston materials would not be expected for
the 3.0 mm optical aperture as the differ-
ences in the RMS among all materials tested
were lower than 0.01 μm. In addition, differ-
ences in the wearer’s visual function might
not be expected due to the HORMS differ-
ence between the lens materials, which was
less than 0.10 μm, for the 6.0 mm aperture.
As was stated previously, a further in vivo
study is needed to assess whether subtle
optical differences between materials affect
the subject’s visual performance.
Spherical aberration was the predominant
higher-order aberration for both optical aper-
Figure 3. Modulation transfer function versus spatial frequency expressed in cycles per de- tures and was negative in sign for all materials
gree for a 6.0 mm optical aperture for Boston EO, ES, XO and XO2 materials. The diffrac- (Tables 3 and 4). As the spherical aberration
tion-limited curve and retinal contrast threshold curve obtained by Sekiguchi, Williams and of the whole eye is in most cases positive,16
Brainward14 at a retinal illuminance of 500 td are included. these particular lenses would minimise the

Clinical and Experimental Optometry 99.1 January 2016 © 2016 Optometry Australia
44
Optical quality among Boston materials Domínguez-Vicent, Esteve-Taboada, Ferrer-Blasco, García-Lázaro and Montés-Micó

total spherical aberration of the combined


eye-contact lens system, as was reported by
Hong, Himebaugh and Thibos;4 however,
the spherical aberration of contact lenses in-
creases with increasing (absolute) contact lens
power.17
There are other technologies to measure
in vitro optical properties of RGP contact
lenses. Kollbaum and colleagues18 evaluated
the accuracy and repeatability of a commer-
cial Hartman–Shack aberrometer to measure
the aberrations of dry and wet contact lenses.
The accuracy of third-order aberrations was
verified by comparing the magnitude of in-
duced coma after decentring a lens with a
known amount of spherical aberration, while
the accuracy of fourth order was evaluated by
comparing the measured longitudinal spher-
ical aberration to that expected based on ray-
tracing through the lens design. This ap- Figure 4. Point spread functions computed from the wavefront aberrations for each
proach yielded accurate results with an error Boston contact lens material examined for both 3.0 mm and 6.0 mm optical apertures
lower than one per cent. The authors used
two different procedures to assess repeatabil-
ity. First, repeated measurements were taken
with the wet cell remaining in place, and sec-
ond, repeated measurements were taken af-
ter removing and reinserting the lens into
the wet cell. The variance using the first pro-
cedure was 3.9 × 10-7 μm2, when measuring
RGP contact lenses and 4.6 × 10-6 μm2 for
soft contact lenses. With the second proce-
dure, the variance was 2.9 × 10-6 μm2 for
RGP lenses and 2.0 × 10-5 μm2 for soft con-
tact lenses. Previous studies have assessed
the NIMO’s repeatability for the measure-
ment of monofocal contact lenses10 and the
power profiles of multifocal soft contact
lenses;19 however, no previous studies have
assessed the repeatability and accuracy of this
device for the measurement of Zernike coef-
ficient terms. Thus, further studies should as-
sess both repeatability and reproducibility of
the NIMO for measure aberrations of both
RGP and soft contact lenses.
In vitro measurements were conducted in Figure 5. Simulation of the images provided by the different lens materials for 3.0 mm and
this study; however, differences in on-eye 6.0 mm optical apertures
lens performance related to surface wettabil-
ity, oxygen permeability and the wearing
modality may impact on the quality of optical aperture; however, in vivo studies scholarship (Universidad de Valencia)
vision. Further studies should be aimed to are required to clarify these conclusions awarded to Alberto Domínguez Vicent.
assess whether differences among materials drawn from in vitro testing.
affect visual performance and to assess the REFERENCES
visual quality of these materials over time. 1. Dorronsoro C, Barbero S, Llorente L, Marcos M. On-
eye measurement of optical performance of rigid gas
In summary, it can be concluded that ACKNOWLEDGEMENTS
permable contact lenses based on ocular and corneal
no visual differences might be expected This research was supported in part by the aberrometry. Optom Vis Sci 2003; 80: 115–125.
among Boston materials for a 3.0 mm ap- Starting Grant funded by the European Re- 2. Lu F, Mao X, Qu J, Xu D, He J. Monochromatic
erture due to similar results in the optical search Council (ERC-2012-StG-309416- wavefront aberrations in the human eye with contact
lenses. Optom Vis Sci 2003; 80: 135–141.
metrics assessed. Conversely, visual differ- SACCO) to Professor Robert Montés-Micó 3. Choi J, Wee WR, Lee JH, Kim MKK. Changes of oc-
ences might be expected at the 6.0 mm and by an ‘Atracció de talent’ research ular higher order aberration in on- and off-eye of

© 2016 Optometry Australia Clinical and Experimental Optometry 99.1 January 2016
45
Optical quality among Boston materials Domínguez-Vicent, Esteve-Taboada, Ferrer-Blasco, García-Lázaro and Montés-Micó

rigid gas permable contact lenses. Optom Vis Sci 2007; 10. Joannes L, Hough T, Hutsebaut X, Dubois X, Ligot 15. Applegate RA, Ballentine C, Gross H, Sarver DJ,
84: 42–51. R, Saoul B et al. The reproducibility of a new power Sarver CA. Visual acuity as a function of Zernike
4. Hong X, Himebaugh N, Thibos L. On-eye evalua- mapping instrument based on the phase shifting mode and level of root mean square error. Optom
tion of optical performance of rigid and soft Schlieren method for the measurement of spherical Vis Sci 2003; 80: 97–105.
contact lenses. Optom Vis Sci 2001; 78: 872–880.
and toric contact lenses. Cont Lens Anterior Eye 2010; 16. Chakraborty R, Read SA, Collins MJ. Diurnal varia-
5. Roberts B, Athappilly G, Tinio B, Naikoo H, Asbell P.
33: 3–8. tions in ocular aberrations of human eyes. Curr Eye
Higher order aberrations induced by soft contact
11. Belda-Salmerón L, Madrid-Costa D, Ferrer-Blasco T, Res 2014; 39: 271–281.
lens in normal eyes with myopia. Eye Contact Lens
2006; 32: 138–142. Garcia-Lazaro S, Montés-Micó R. In vitro power pro- 17. Dietze H, Cox M. On- and off-eye spherical aberration
files of daily disposable contact lenses. Cont Lens of soft contact lenses and consequent changes of ef-
6. Shen J, Thibos L. Peripheral aberrations and image
quality for contact lens correction. Optom Vis Sci Anterior Eye 2013; 36: 247–252. fective lens power. Optom Vis Sci 2003; 80: 126–134.
2011; 88: 1196–1205. 12. Montés-Micó R, Madrid-Costa D, Domínguez-Vicent 18. Kollbaum P, Jansen M, Thibos L, Bradley A. Valida-
7. Bennett E, Weissman B. Clinical Contact Lens Practice. A, Belda-Salmerón L, Ferrer-Blasco T. In vitro power tion of an off-eye contact lens Shack-Hartmann
Philadelphia: Lippincott Williams & Wilkins, 2005. profiles of multifocal simultaneous vision contact wavefront aberrometer. Optom Vis Sci 2008; 85:
8. Benjamin WJ. EOP and Dk/L: The quest for hyper lenses. Cont Lens Anterior Eye 2014; 37: 162–167. E817–E828.
transmissibility. J Am Optom Assoc 1993; 64: 196–200. 13. ISO:18369-2. Contact lens. Part 2: Tolerances. Oph- 19. Domínguez-Vicent A, Marín-Franch I, Esteve-
9. Joannes L, Dubois F, Legros JC. Phase-shifting schlie- thalmic optics - contact lenses; 2006. Taboada J, Madrid-Costa D, Montés-Micó R. Repeat-
ren: high-resolution quantitative Schlieren that uses 14. Sekiguchi N, Williams DR, Brainnard DH. Efficiency ability of in vitro power profile measurements for
the phase-shifting technique principle. Appl Optics in detection of isoluminant and isochromatic inter- multifocal contact lenses. Cont Lens Anterior Eye 2015;
ference fringes. J Opt Soc Am A 1993; 10: 2118–2133. 38: 168–172.
2003; 42: 5046–5053.

Clinical and Experimental Optometry 99.1 January 2016 © 2016 Optometry Australia
46

You might also like