How to solve the problem that LCD screen shows abnormally

Regarding the problem that LCD screen shows abnormally, please check the following
solutions step by step:

1. Please reconnect the cable which connects the LCD screen with the main board, and
make sure it is tightly connected.

2. Please restart the instrument & press “Del” button on the keyboard immediately. You
know, what you enter is the “BIOS” menu (although you can not see it). Please press
“←”button once & press “↓”button three times & press “F10” & press “Enter”. The above
procedure is used to let the instrument load optimal defaults to try to solve this problem.

3. Please follow the below suggestions to check the problem.

(1). Please remove the six screws which fix the front cover. Open the front cover gently

Front cover

Figure 1
(2). Connect one CRT display to the main board & turn on the instrument, if the CRT
display shows abnormally, it means that the main board is defective. May I know if have a
workable main board in your stock? If yes, please exchange a workable main board to try
to solve the problem.
Main board

CRT
connector

Figure 2
(3). If the CRT display shows normally, please restart the instrument press the “Del” on
the keyboard enters the display setting & select the “Chipset” and “Boot Display Type” &
press “enter” key & select “Panel+ CRT”. Save and exit as figure 3 and figure 4.

Figure 3
Figure 4
(4). Restart the instrument to check if LCD screen will show normally, if LCD screen still
shows abnormally, it means that LCD screen is defective.