Professional Documents
Culture Documents
Digesdahl® Digestion
Apparatus
Models 23130-20, -21
Instrument Manual
© Hach Company, 1989-91, 1995-97, 1999. All rights reserved. Printed in the U.S.A. hm/ct 2/97 8 ed
aa/dk Rev. 2, 9/99
TRADEMARKS OF HACH COMPANY
2
CERTIFICATION
Product Safety
Per 73/23/EEC LVD: certified compliant by Hach Company to EN 61010-
1 (IEC1010-1), supporting test records by ETL. Listed by ETL to UL
3101-1 (Listing # H0492805390). Certified by CSA to CSA C22.2 No.
1010.1 (Certification # H0492805390).
Immunity
EN 50082-1 (European Generic Immunity Standard) per 89/336/EEC
EMC: Supporting test records by Hach Company, certified compliance by
Hach Company.
Standards include:
EN 61000-4-2 “1995” (IEC 801-2) Electro-Static Discharge
EN 61000-4-4 “1995” (IEC 801-4) Electrical Fast Transients/Burst
EN 61000-4-5 “1995” (IEC 1000-4-5) Surge
EN 61000-4-11 “1994” (IEC 1000-4-11) Voltage Dips,
Interruptions and Variations
ENV 50140 “1993” (IEC 801-3) Radiated RF Electro-Magnetic Fields
ENV 50141 “1993” Conducted Disturbances Induced by RF Fields
ENV 50204 “1995” Radiated Electro-Magnetic Field from Digital
Telephones.
Emissions
EN 50081-1 (Emissions) per 89/336/EEC EMC: Supporting test records
by TÜV Product Services and Hach Company, certified compliance by
Hach Company.
Standards include:
EN 55014 (CISPR 14) Emissions, Class B Limits
EN 60555-2 Harmonic Disturbances Caused by Electrical Equipment
EN 60555-3 Voltage Fluctuation (Flicker) Disturbances Caused by
Electrical Equipment
iii
CERTIFICATION, continued
Canadian Interference-Causing Equipment Regulation, IECS-003,
Class A:
Supporting test records by TÜV Product Services, certified compliance
by Hach Company.
This Class A digital apparatus meets all requirements of the Canadian
Interference- Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du
Règlement sur le matériel brouilleur du Canada.
FCC Part 15, Class "A" Limits: Supporting test records by TÜV Product
Services, certified compliance by Hach Company.
(1) this device complies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions:
(2) this device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may
cause undesired operation.
Changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to Part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is
likely to cause harmful interference, in which case the user will be
required to correct the interference at his own expense.
The following techniques of reducing the interference problems are
applied easily:
1. Disconnect power from the Digesdahl to verify that it is the source of
the interference.
2. If the Digesdahl is plugged into the same outlet as the device with
which it is interfering, try another outlet.
3. Move the Digesdahl away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving
the interference.
5. Try combinations of the above.
iv
TABLE OF CONTENTS
CERTIFICATION .................................................................................................................... iii
SPECIFICATIONS.................................................................................................................. vii
SAFETY PRECAUTIONS .................................................................................................... viii
OPERATION
v
TABLE OF CONTENTS, continued
DIGESTION PROCEDURES
GENERAL DIGESDAHL DIGESTION ............................................................................. 29
SAMPLE TYPE AND SIZE................................................................................................... 35
MAINTENANCE
SECTION 5 MAINTENANCE............................................................................................ 81
5.1 Fuse Replacement ............................................................................................................. 81
5.2 Kit Replacement Parts ...................................................................................................... 81
GENERAL INFORMATION
REPLACEMENT PARTS....................................................................................................... 85
HOW TO ORDER .................................................................................................................. 87
REPAIR SERVICE ................................................................................................................. 88
WARRANTY.......................................................................................................................... 89
vi
SPECIFICATIONS
Temperature Range:
Variable from 100 to 480 ° C (212 to 896 ° F)
Temperature Control:
Within ± 15 ° C of set point*
Power Requirements:
115 and 230 Vac models, 50/60 Hz, 250 W** (use only single phase
for 230 V)
Aspirator Capacity:
11.5 L/min (3.0 gal/min) at water flow rate of 6.5 L/min (1.7 gal/min).
Minimum pressure 51.7 kPa (7.5 psi). Drain required.
Dimensions:
14 cm x 16.5 cm x 33.6 cm (5.5” x 6.5” x 13.25”).
Total height: approximately 61 cm (24”) when assembled for use.
Weight:
Net: 3.85 kg (8.5 lbs)
Shipping: 4.8 kg (10.6 lbs)
Operating Conditions:
15-35 ° C (59-95 ° F); 0-85% relative humidity
* The Digesdahl apparatus may be influenced by electric field radiation of 3 volts per meter or greater at
frequencies of 100 ±15 MHz and 180 ±15 MHz. The temperature specification of ±15 ° C was not exceeded by
more than 10 ° C at these radio frequencies.
** Use only single phase power for 230 V models. The over-temperature protection device may not interrupt power
when using poly-phase power.
vii
SAFETY PRECAUTIONS
Notice
Before attempting to unpack, set up or operate this instrument, please read
this entire manual.
The digestion procedures in this manual involve the use of strong acid
and oxidizer at high temperatures. To avoid personal injury, observe
all warning messages.
DANGER
Indicates either a potentially or imminently hazardous situation which,
if not avoided, could result in death or serious injury.
CAUTION
Indicates either a potentially or imminently hazardous situation which,
if not avoided, could result in minor or moderate injury.
NOTE
Information that requires special emphasis.
Precautionary Labels
Please pay particular attention to labels and tags attached to the instrument.
4.2 Digestion
viii
OPERATION
DANGER
Handling chemical samples, standards, and reagents can be dangerous. Review the
necessary Material Safety Data Sheets and become familiar with all safety procedures
before handling any chemicals.
DANGER
La manipulation des échantillons chimiques, étalons et réactifs peut être dangereuse.
Lire les fiches de données de sécurité des produits nécessaires et se familiariser avec
toutes les procédures de sécurité avant de manipuler tout produit chimique.
1
2
SECTION 1 GENERAL DESCRIPTION
1.1 Introduction
The Hach Digesdahl Digestion Apparatus shown in Figure 1 is designed
to digest a wide variety of sample types for subsequent determination of
total Kjeldahl nitrogen, several minerals, and nutrients. Sample types
include food products, feeds, grains, wastewater sludges, plating baths,
plant tissues, fertilizers, beverages, and oils.
DANGER
Ce produit n’a pas de protection contre la surchauffe lorqu’il est utilisé
sur une alimentation électrique en triphasé. Utiliser cet appareil
seulement sur alimentation 230 V monophasé.
3
Figure 1 Digesdahl Digestion Apparatus
CAPILLARY
FUNNEL
VENT TRAP
CAP
CAPILLARY
FUNNEL
ADAPTOR
VENT TRAP
BODY
COLLUMN
BAFFLE
FRACTIONATING
COLUMN
COLUMN
RECEPTACLE
SAFETY
VERTICAL SHIELD
SUPPORT
FLASK
HEAT WEIGHT
SHIELD
DIGESTION
FLASK
TEMPERATURE
CONTROL
TEMPERATURE
INDICATOR
POWER
SWITCH
MODE
SWITCH
HEATER
ASSEMBLY
4
SECTION 2 PREPARATION FOR USE
2.1 Unpacking
Remove the Digesdahl Digestion Apparatus and accessories from the
shipping container and inspect each item for any damage that may have
occurred during shipping.
If any items are missing or damaged, please contact the Customer Service
Department, Hach Company, Loveland, Colorado (the toll-free number is
800-227-4224). For customers outside the U.S.A., contact the Hach office
or authorized distributor serving you. Please do not return the instrument
without prior authorization from Hach.
5
2.2 Assembly
2.2.1 Heater Assembly
Assemble the heater assembly as shown in Figure 2.
Figure 2
Heater Assembly
6
2.2.2 Fractionating Head System
The fractionating head system is shown in Figure 3 and includes the
fractionating column, column baffle, the capillary funnel, and the exhaust
system components. Assemble as follows:
8
2.3 Selecting a Temperature Setting
Note: During a The particular temperature applicable for the sample to be digested is
digestion, the given in the appropriate analysis procedure manual. The steps below
temperature explain how to set the heater temperature.
reading will vary
slightly above and
1. Connect the power cord to the line voltage receptacle and set the
below the
selected POWER switch to ON. The TEMPERATURE indicator will light.
temperature. This
2. Set the Mode switch to SET. The TEMPERATURE indicator will
will not affect the
digestion or the display the current temperature set point in degrees Celsius.
accuracy of the
3. If a different temperature is desired, adjust the Temperature ADJ
final analysis.
When idling (no control for the desired temperature in the digital display.
digestion flask in
4. Set the Mode switch to READ and allow time (approximately
place), the
temperature may 10 minutes) for the heater to reach the selected temperature. With the
fluctuate ± 15 ° C Mode switch in the READ position, the temperature display shows the
from the set point. current operating temperature.
9
10
SECTION 3 SAFETY & ENVIRONMENTAL
CONSIDERATIONS
Each person performing laboratory tests is responsible for safety. The
analyst should develop and practice good safety habits to minimize
chances for accidents by practicing good laboratory techniques.
• Be sure to keep the heat shield and the Digesdahl’s safety shield in
place during use.
• Always wear safety glasses or goggles- be sure they have side shields.
• Wear protective gloves for during digestion procedures. Use tongs or
finger cots to transfer hot apparatus.
• During digestion, use the heat setting and digestion time specified in
the instructions. Do not leave the Digesdahl unattended during use.
11
• When digesting a new substance for the first time, begin with a smaller
size and work up to the optimum quantity for digestion. Do not permit
the flask to boil to dryness.
• Allow the Digesdahl to cool naturally (in ambient air). Cold water may
cause hot glassware to shatter.
12
Proper handling and storage procedures involving hydrogen peroxide
should always address two major characteristics of the product:
13
• Do not use unapproved materials (brass, copper, carbon steel, rubber,
etc.) for transfer or storage systems.
Caps on the reagent bottles are made with a special porous liner that
allows venting of gas. The venting cap always must be used on the bottle
of hydrogen peroxide. As a precaution, the reagent bottles are shipped in a
plastic bag. If there is evidence of leakage during shipment, wear gloves
when removing the bottle from the bag and rinse the bottle with water
when removed from the bag. Rinse the bag before disposal.
While cleaning a spill or leak, please follow the safety measures below:
1. DO NOT attempt to clean the apparatus if it is hot; this could cause the
glass to shatter. Let the apparatus cool before cleaning it.
14
2. Unplug the Digesdahl before cleaning.
3. Wear gloves and goggles when handling the glassware. Carefully rinse
the glassware several times with water to decontaminate it.
4. Wipe the exterior surfaces of the Digesdahl apparatus (heating mantle,
electrical controls, etc.) and other laboratory surfaces several times
with a damp or wet cloth or paper towel.
5. Do not rinse or spray the apparatus directly; this could damage
the equipment.
6. Discard any paper towels or cloths in an appropriate manner; they may
be contaminated from the sample or chemical residues.
This manual does not contain all the regulatory requirements. Additional
state and local laws may apply to waste that you generate. It is each
generator’s responsibility to know which regulations apply to them and to
adhere to these regulations. Check with your environmental compliance
staff for specific instructions.
15
CONSIDERATIONS DE SECURITE ET D’ENVIRONNEMENT
Chaque personne effectuant des analyses de laboratoire est responsable de
la sécurité. L’analyste doit développer et appliquer de bonnes habitudes
de sécurité pour minimiser les risques d’accidents en pratiquant de bonnes
techniques de laboratoire.
16
• NE PAS ajouter d’alcool, d’acétone ou autre solvant organique dans la
fiole de digestion avant ou après minéralisation.
17
L’eau oxygénée (30% ou 50%) est un oxydant puissant et ne doit jamais
être stocké près de matières inflammables. Comme l’acide sulfurique,
il peut provoquer des brûlures à la peau et aux yeux. En cas de contact
avec les yeux ou la peau, laver les yeux et/ou la peau à l’eau pendant
15 minutes. Retirer les vêtements contaminés. Appeler un médecin.
L’eau oxygénée est extrêmement corrosive et doit être lavée à l’eau si elle
est répandue sur des appareils ou sur la paillasse. Lire et observer tous les
avertissements sur les étiquettes des réactifs et les fiches de données de
sécurité des produits.
• Stocker dans les récipients d’origine avec les bouchons fournis et les
maintenir fermés lorsqu’ils ne sont pas utilisés. (S’assurer que les
récipients sont ventilés. Les flacons d’eau oxygénée Hach sont livrés
avec un joint de bouchon perméable spécial).
18
• Laver la peau contaminée rapidement à grande eau. Retirer rapidement
les vêtements contaminés et les laver soigneusement avant de les
réutiliser. Les laver régulièrement.
Les bouchons des flacons de réactifs sont prévus avec un joint poreux
spécial qui permet l’échappement de gaz. Le bouchon d’origine doit
toujours être utilisé sur le flacon d’eau oxygénée. A titre de précaution,
les flacons de réactifs sont expédiés dans un sac plastique. S’il existe des
traces de fuites pendant le transport, porter des gants pour retirer le flacon
du sac et rincer le flacon à l’eau après l’avoir sorti du sac. Rincer le sac
avant élimination.
19
ou la peau à l’eau pendant 15 minutes. Retirer les vêtements
contaminés. Appeler un médecin.
6. Eliminer les papiers et tissus de façon appropriée, ils peuvent avoir été
contaminés par l’échantillon ou les résidus chimiques.
20
SECTION 4 OPERATION
4.1 Apparatus Preparation
See SECTION 3 for more safety information. Prepare to perform the
digestion either behind a laboratory safety shield or inside a fume hood. If
a fume hood is used, check the fume exhaust system to verify it is
working property.
Safety glasses and protective clothing are mandatory.
Turn on the water to the aspirator to maximum flow. Remove the
capillary funnel and place a finger over the opening in the vent trap cap. A
distinct suction should be felt.
4.2 Digestion
Procedures for using the Digesdahl Digestion Apparatus vary with sample
type. Most procedures use a two-phase digestion process involving
concentrated sulfuric acid and 50% hydrogen peroxide. Sulfuric acid
dehydrates and chars the sample. Hydrogen peroxide is added via the
capillary flow funnel to complete sample decomposition. The capillary
funnel feeds hydrogen peroxide into the digestion flask at a rate of 3 mL
per minute. This allows the analyst to control the amount of time sample
is exposed to the hydrogen peroxide (digestion time) by varying the
volume of hydrogen peroxide used.
DANGER
Wear protective eye glasses and clothing. A strong acid (concentrated sulfuric acid) and
a strong oxidant (50% hydrogen peroxide) are used in the digestion reaction. These
chemicals can cause burns if splashed on the skin or permanent eye damage if allowed
to contact the eyes. If the chemicals are hot, effects are considerably more severe.
Immediately rinse any affected area thoroughly with water and contact a physician.
DANGER
Porter des lunettes et vêtements de protection. Un acide fort (acide sulfurique concentré)
et un oxydant puissant (peroxyde d'hydrogène à 50%) sont utilisés dans la réaction de
minéralisation. Ces produits chimiques peuvent causer des brûlures s'ils sont projetés sur
la peau ou des blessures irréversibles aux yeux s'ils sont mis au contact des yeux. Si ces
produits sont chauds, les effets sont considérablement plus graves. Immédiatement rincer
toute partie atteinte abondamment à l'eau et consulter un médecin.
21
A general procedure, incorporating application specific requirements, is
provided in this section of the manual. To ensure complete sample
digestion, consider the variables described in the following paragraphs.
22
Table 1 Digestion Guidelines of Specific Sample Types
Preheat
Sample Vol. of Vol. of
Sample Weight Time Special Instructions
Type Acid Peroxide
(Step 5)
Plant tissue 0.25 to 0.5 g 4 mL 4 min. 10 mL Use Nitrogen-free paper to
weigh samples.
Meat & 0.5 g or 4 mL or as in 4 min. 10 mL —
Poultry predigestion predigest
Fluid 0.1 to 0.25 g 4 mL 4 min. 10 mL Add 0.4 g Kjeldahl Reduction
Fertilizers Powder to flask before adding
sulfuric acid. Place the flask in
an 80 °C oven 15 minutes before
digestion. Use N-free paper to
weigh samples.
Feed & 0.25 g 4 mL 4 min. 10 mL —
Forage
Dairy 0.25 to 2.0 g 4 mL 4 min. 10 mL —
Cereal 0.25 to 0.5 g 4 mL 4 min. 10 mL Use Nitrogen-free paper to
weigh samples.
Beverage about 5 g 4 mL 1 min. 10 mL Preheat acid for 1 minute then
(pipet into funnel add sample through funnel. Heat
flask for 30 seconds after sample
is in the flask.
Sludge <2.5 g wet sludge 4 mL 3-5 min. 10 mL of Heat the diluted digest for 15
<0.5 g dried sludge increase in 5 minutes and filter.
mL increments
Water & not more than 0.5 g 3 mL until acid 10 mL or Water must evaporate before
Wastewater solid (mL = 40/C; is refluxing increase in 5 acid will reflux. Boiling chips
C= % solids) mL increments required.
Bath 0.3 to 10 mL 4 mL 4 min. 10 mL Water must evaporate before
Solutions acid will reflux. Boiling chips
required.
Edible Oils 0.25 to 0.5 g 4-6 mL 4 min. 5 mL Weigh samples into flake and
immediately record exact weight.
and 5 mL later
Ion equivalent of 0.25 g 10-15 mL 12 min. 20 mL Digest will be clear with particles
Exchange dry resin on bottom if metal oxides are not
Resins soluble in H2SO4. Add aqua
reagia or suitable solvent to
dissolve particles. If particles are
floating, start again using 15 mL
H2SO4 and longer char time.
23
Sulfuric acid (H2SO4) consumption depends on the anhydrous mass of
material and the chemical composition of the substance. Use of 4 mL
H2SO4 is suitable for many materials, but not all. Therefore, the analyst
must pay attention to the amount of residual H2SO4 for the type of sample
digested, and adjust the amount of acid or sample accordingly. Never use
less than 3 mL of concentrated sulfuric acid. Larger volumes of H2SO4
may be used, but avoid a large excess since sample pH adjustment is
required in most subsequent determinations.
4.2.4 Carbonization Period
A carbonization period prior to the addition of hydrogen peroxide
provides a reducing environment which helps convert organic nitrogen to
ammonia. In the presence of oxidizable carbon compounds, sulfuric acid
reacts to produce sulfur dioxide, which is the active reducing agent.
The reaction is:
H2SO4 → H2O + SO2 + ½ O2
24
If 50% hydrogen peroxide is not available, 30% peroxide may be used as
a last resort with caution. Because of the lesser strength, at least
1.67 times more volume must be used (i.e., 16.7 mL of 30% vs. 10 mL of
50% peroxide). Always run a digestion standard, either glycine
p-toluenesulfonate or nicotinic acid p-toluenesulfonate, when using
30 percent peroxide to check completion of the digestion. All
recommended safety precautions apply to both strengths of the hydrogen
peroxide. See Section 3, Safety Considerations.
25
Solid Samples
1. Weigh 0.25 grams of a Primary Standard for Kjeldahl Nitrogen.
Digest the standard following the general digestion procedure.
2. Any of the three standards may be used. Ammonium p-toluenesulfonate
(185.5 mg/L TKN) is the least difficult to digest. Glycine
p-toluenesulfonate (141.63 mg/L TKN) is moderately difficult to digest,
and Nicotinic Acid p-toluenesulfonate (118.58 mg/L TKN) is the most
difficult to digest.
Liquid Samples
1. Weigh 10.000 grams of a Primary Standard for Kjeldahl Nitrogen.
2. Transfer to a 1-liter volumetric flask and dilute to the mark.
3. Using a volumetric pipet, add 15 mL of the prepared solution to the
digestion flask and digest the standard following the general
digestion procedure.
4. Any of the three standards may be used. Ammonium
p-toluenesulfonate (111.03 mg/L TKN) is the least difficult to digest.
Glycine p-toluenesulfonate (84.98 mg/L TKN) is moderately difficult
to digest, and Nicotinic Acid p-toluenesulfonate (71.15 mg/L TKN) is
the most difficult to digest.
A primary standard set, containing one bottle of each of the standards,
may be purchased from Hach Company. Table 2 lists some of the
properties of the three primary standards.
Table 2 Some Properties of Kjeldahl Nitrogen Standards (#22778-00)
Ammonia PTSA Glycine PTSA Nicotinic Acid PTSA
Formula C7H11O3SN C9H13O5SN C13H13O5SN
Structure
26
DIGESTION PROCEDURES
27
Danger
Review Section 3 for important safety information before
performing this digestion procedure.
Danger
Lire au chapitre 3 les informations importantes pour la
sécurité avant d’effectuer cette technique de digestion.
28
GENERAL DIGESDAHL DIGESTION
29
GENERAL DIGESDAHL DIGESTION, continued
3-5 minutes
5. Heat 3-5 minutes. 6. Do not proceed if 7. After addition of 8. Take the hot flask off
Do not boil sample sulfuric acid is not hydrogen peroxide is the heater and allow the
to dryness. If sulfuric visible in the flask. Add complete, boil off excess flask to cool. Remove
acid is not present after 10 mL of 50% Hydrogen hydrogen peroxide by the fractionating column
heating, do not proceed Peroxide to the charred heating for one more from the digestion flask.
with Step 6. Discard the sample via the funnel on minute. Do not heat Note: Use finger cots to
sample if it evaporates to the fractionating head. to dryness. remove the digestion flask.
dryness. Start over and Note: Visually confirm the Note: If the sample goes Place it on a cooling pad
use a larger volume of presence of sulfuric acid in to dryness, turn off the for at least one minute.
sulfuric acid in Step 2 of the flask before adding Digesdahl and air cool to Then remove the column.
this procedure. Or chose hydrogen peroxide. room temperature. Add
a smaller sample amount Note: If the digest does not water to flask before
for digestion. turn colorless, add handling. Repeat the
5 mL increments of digestion from the Step 1
Note: Some organic
peroxide until the digest using a new sample.
samples may need more
than five minutes for becomes clear.
complete digestion. See Note: Use long-neck
Table 1 on page 22. glassware or a pipet
to transfer the
hydrogen peroxide.
30
GENERAL DIGESDAHL DIGESTION, continued
Use metals or
TKN procedure
9. Dilute the digest with 10. Turn the 11. If the sample has 12. Continue the
approximately 70 mL of temperature dial to a visible turbidity, filter analysis using one of the
deionized water. If not heat setting of 204 ° C or wait until the following procedures.
analyzing for aluminum, (400 ° F). Add 150 mL of turbidity settles, and the
nickel or iron, dilute to water to a 400-mL upper portion of the
the beaker. Place the beaker sample is clear.
100-mL mark with on the heater. Place the Filter as follows:
deionized water; skip flask in the beaker and
Step 10 and proceed to boil for 15 minutes. a) Place a filter paper into
the filter holder with
Step 11. If analyzing for Air cool to room
wrinkled surface upward.
nickel, aluminum or temperature and dilute to
iron, go to Step 10. the 100-mL mark with b) Place the filter holder
deionized water. Invert assembly in the filtering
Note: Add deionized
flask and wet the filter with
water slowly at first. Cool several times to mix.
deionized water to ensure
the flask if necessary
adhesion to the holder.
for handling.
c) While applying a
vacuum to the filtering
flask, transfer the sample to
the filtering apparatus.
d) Slowly release the
vacuum from the filtering
flask and transfer to
another container.
31
GENERAL DIGESDAHL DIGESTION, continued
Metals Procedure
Note: If analyzing aliquots smaller than 0.5 mL, pH adjustment is unnecessary.
32
GENERAL DIGESDAHL DIGESTION, continued
TKN, Colorimetric Methods
Consult the spectrophotometer or colorimeter procedure to complete the
TKN analysis. The following is only a guide to use if a procedure is
not available.
5. Continue to add 1 N KOH in this manner until the first permanent blue
color appears.
6. Add deionized water to the volume indicated in the colorimetric
procedure.
7. Continue with the colorimetric procedure.
33
34
SAMPLE TYPE AND SIZE
Although the Digesdahl can digest many types of samples, a specific
digestion is required for specific sample types. To classify the sample as a
liquid, oil or solid, use the following charts. Follow the appropriate
procedure indicated for each sample type.
Sample size varies, depending on the sample type and the parameter being
measured. Once the correct sample type is decided, use the tables
following each of the digestion procedures to determine the sample size.
Chart 1
For Aqueous Liquids
35
SAMPLE TYPE AND SIZE, continued
Chart 2
For Oils
36
SAMPLE TYPE AND SIZE, continued
Chart 3
For Solids*
37
Danger
Review Section 3 for important safety information before
performing this digestion procedure.
Danger
Lire au chapitre 3 les informations importantes pour la
sécurité avant d’effectuer cette technique de digestion.
38
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS
39
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
4:00
5. Boil 4 more minutes. 6. Do not proceed if 7. After addition of 8. Take the hot flask off
Do not boil the sample sulfuric acid is not hydrogen peroxide is the heater and allow the
to dryness. If sulfuric visible in the flask! Add complete, boil off excess flask to cool. Remove
acid is not present in the 10 mL of 50% Hydrogen hydrogen peroxide by the fractionating column
flask after the 4 minute Peroxide to the charred heating for one more from the digestion flask.
heating, do not proceed sample via the funnel on minute. Do not heat Note: Use finger cots to
with Step 6! Discard the fractionating column. to dryness. remove the digestion flask.
sample if it evaporates to Note: Visually confirm the Note: If the sample goes Place it on a cooling pad
dryness. Start over and presence of sulfuric acid in to dryness, turn off the for at least one minute.
use a larger volume of the flask before adding Digesdahl and air cool to Then remove the column.
sulfuric acid in Step 2 of hydrogen peroxide. room temperature. Add Do not add water to the
flask until it has cooled.
this procedure. Or Note: If the digest does not water to flask before
choose a smaller sample turn colorless, add 5 mL handling. Repeat the
amount for digestion. increments of hydrogen digestion from the
peroxide until the digest beginning using a new
becomes clear or does not sample.
change color.
Note: If sample foams
during peroxide addition,
remove the Digesdahl
digestion flask and
fractionating column (use
finger cots). Starting with
Step 1, repeat the digestion
adding only 2 mL of
hydrogen peroxide. Then
follow with 8 mL of
hydrogen peroxide.
Note: Do not heat
to dryness.
40
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
9. Dilute the digest to 10. If analyzing for 11. Turn the 12. If the sample has
approximately 70 mL aluminum, nickel or temperature dial to a visible turbidity, filter or
with deionized water. iron, continue to Step 11. heat setting of 204 ° C wait until the turbidity
Note: Add deionized water If analyzing for other (400 ° F). Add 150 mL of settles, and the upper
slowly at first. Cool the substances, dilute to the water to a 400-mL portion of the sample
flask if necessary 100-mL mark with beaker. Place the beaker is clear.
for handling. deionized water; skip on the heater. Place the Filter as follows:
Step 11 and go to flask in the beaker and
Step 12. boil for 15 minutes. Air a) Place a filter paper into
the filter holder with
cool to room
wrinkled surface upward.
temperature and dilute to
the 100-mL mark with b) Place the filter holder
deionized water. Invert assembly in the filtering
flask and wet the filter with
several times to mix.
deionized water to ensure
adhesion to the holder.
c) While applying a
vacuum to the filtering
flask, transfer the sample to
the filtering apparatus.
d) Slowly release the
vacuum from the filtering
flask and transfer to
another container.
Continue with the
analysis using the
appropriate procedure
below.
Metals Method
Note: If analyzing aliquots smaller than 0.5 mL, pH adjustment is not necessary.
1. Pipet the appropriate analysis volume into the appropriate mixing
graduate cylinder. See Sample and Analysis Volume Tables for
Aqueous Liquids following this procedure to determine the
analysis volume.
Note: Some methods require pipetting into a volumetric flask or a regular
graduate cylinder.
41
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
2. Dilute to about 20 mL with deionized water.
3. Add one drop of 2,4 Dinitrophenol Indicator Solution.
4. Add one drop of 8 N Potassium Hydroxide (KOH) Standard Solution,
swirling between each addition, until the first flash of yellow appears
(pH 3). If analyzing for potassium, use 5 N sodium hydroxide instead.
Do not use a pH meter if analyzing for potassium or silver.
5. Add one drop of 1 N KOH. Stopper the cylinder and invert several
times to mix. If analyzing for potassium, use 1 N sodium hydroxide
instead.
Note: Use pH paper to insure the pH is 3. If it is higher than 4, do not readjust with
acid; start over with a fresh aliquot.
6. Continue to add 1 N KOH in this manner until the first permanent
yellow color appears (pH 3.5-4.0).
Note: High iron content will cause precipitation (brown cloud) which will co-
precipitate other metals. Repeat this procedure with a smaller aliquot volume.
7. Add deionized water to the volume indicated in the colorimetric
procedure for the parameter you are analyzing. Fill a second graduated
mixing cylinder to the same volume with deionized water.
8. Continue with the colorimetric procedure for the parameter you
are analyzing.
42
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
6. Add deionized water to the volume indicated in the
colorimetric procedure.
7. Continue with the colorimetric procedure.
A × 5000
------------------------ = mg/L Total Al
B× C
A = mg/L reading from instrument
A × 5000
------------------------ = mg/L Total Al
B× C
A = mg/L reading from instrument
43
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
Cadmium, Dithizone (Method 8017)
A × 25 = µg/L Total Cd
-----------------
B× C
A = µg/L reading from instrument
A × 2500
------------------------ = mg/L Total Cr
B× C
A = mg/L reading from instrument
44
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
Cobalt (Method 8078)
A × 2500
------------------------ = mg/L Total Cu
B× C
A = mg/L reading from instrument
45
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
Iron, 1,10 Phenanthroline (Method 8008)
A × 2500
------------------------ = mg/L Total Fe
B× C
A = mg/L reading from instrument
46
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
Lead, Dithizone (Method 8033)
A × 25 = µg/L Total Pb
-----------------
B× C
A = µg/L reading from instrument
A × 2500
------------------------ = mg/L Total Mn
B× C
A = mg/L reading from instrument
47
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
Nickel, PAN (Method 8150)
Expected
Sample Analysis
Nitrogen Dilute to
amount (mL) Volume (mL)
conc. (mg/L)
0.5-28 40.0 10.0* 25.0 mL*
2-112 20.0 5.00* 25.0 mL*
11-560 10.0 2.00* 25.0 mL*
45-2250 5.00 1.00* 25.0 mL*
425-22500 1.00 0.50* 25.0 mL*
A × 75
----------------- = mg/L TKN
B× C
A = mg/L reading from instrument
48
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
Phosphorus, Ascorbic Acid (Method 8048)
A × 2500
------------------------ = mg/L Total PO4
B× C
A = mg/L reading from instrument
49
DIGESTION PROCEDURE FOR AQUEOUS LIQUIDS, continued
Silver (Method 8120)
A × 5000
------------------------ = mg/L Total Zn
B× C
A = mg/L reading from instrument
50
51
Danger
Review Section 3 for important safety information before
performing this digestion procedure.
Danger
Lire au chapitre 3 les informations importantes pour la
sécurité avant d’effectuer cette technique de digestion.
52
DIGESTION PROCEDURE FOR OILS
1. Transfer 0.25 g or 2. Add 4 mL 3. Turn the temperature 4. Place the flask weight
less of sample into a concentrated sulfuric dial to a heat setting of followed by the
100-mL Digesdahl acid (spec. gravity 1.84) 440 ° C (825 ° F). When fractionating column
digestion flask; see to the digestion flask. the proper temperature is with funnel on the flask.
Sample and Analysis Note: Use only Hach reached, turn on the Place the flask on the
Volume Tables for Oils Digesdahl digestion flasks. water to the aspirator heater and boil 4
following this Volumetric flasks with and make sure there is minutes. Do not boil to
procedure. concave bottom should not suction to the dryness! If sulfuric acid
be used. fractionating column. is not present in the
Note: Take care to
ensure you have a Note: Safety glasses and Note: Wait for the proper flask after the boiling
homogenous sample. a safety shield placed temperature to be reached period, do not proceed
between the operator and before sample is placed on to Step 5! Discard the
the Digesdahl are required. the heater. sample and use more
sulfuric acid for the
digestion procedure in
Step 2. Or choose a
smaller sample size for
digestion from the
Sample and Analysis
Volume Tables for Oils,
following this procedure.
Note: If sample foams up
into the neck of the flask,
lower temperature to
335 ° C (635 ° F). Continue
heating at lower
temperature until all
water is evaporated. Then
return to original
digestion temperature.
53
DIGESTION PROCEDURE FOR OILS, continued
5. Do not proceed if 6. After addition of 7. Take the hot flask off 8. Dilute the digest
sulfuric acid is not visible hydrogen peroxide is the heater and allow the to approximately 70 mL
in the flask. Add 10 mL complete, boil off excess flask to air cool. Remove with deionized water.
of 50% hydrogen hydrogen peroxide by the fractionating column Note: Add demineralized
peroxide to the charred heating for one more from the digestion flask. water slowly at first. Cool
sample via the funnel on minute. Do not heat Note: Use finger cots to the flask if necessary
the fractionating head. to dryness. remove the digestion flask. for handling.
Note: Visually confirm the Note: If the sample goes to Place it on a cooling pad
presence of sulfuric acid in dryness, turn off the for at least one minute.
the flask before adding Digesdahl and air cool Then remove the column.
hydrogen peroxide. completely. Add water Do not add water to the
to flask before handling. flask until it has cooled.
Note: If the digest does not
turn colorless, add 5 mL Repeat digestion from the
increments of peroxide until beginning using a new
the digest becomes clear or sample.
does not change color.
Note: If sample foams
during hydrogen peroxide
addition, stop the peroxide
flow and remove the
digestion flask and
fractionating column (use
finger cots). Cool for
30 seconds and return
apparatus to the heating
block. Start peroxide
addition with 2 mL,
then follow with the
remaining peroxide.
54
DIGESTION PROCEDURE FOR OILS, continued
9. If analyzing for 10. Turn the 11. If the sample is 12. Continue the
aluminum, nickel or temperature dial to a visibly turbidity, filter or analysis using the
iron, continue to Step heat setting of 204 ° C wait until the turbidity appropriate procedure
10. If analyzing for other (400 ° F). Add 150 mL of settles, and the upper below.
substances, dilute to the water to a 400-mL portion of the sample is
100-mL mark with beaker. Place the beaker clear.
deionized water; Skip on the heater. Place the Filter as follows:
Step 10 and go to flask in the beaker and
Step 11. boil for15 minutes. Cool a) Place a filter paper into
the filter holder with
to room temperature and
wrinkled surface upward.
dilute
to the mark with b) Place the filter holder
demineralized water. assembly in the filtering
flask and wet the filter
Invert several times
with demineralized water
to mix.
to ensure adhesion to
Note: When using a the holder.
Digesdahl Digestion
Apparatus system without c) While applying a
temperature control dials, vacuum to the filtering
reset to a lower setting that flask, transfer the sample to
gently boils the water. the filtering apparatus.
d) Slowly release the
vacuum from the filtering
flask and transfer to
another container.
Metals Method
Note: If analyzing aliquots smaller than 0.5 mL, pH adjustment is not necessary.
55
DIGESTION PROCEDURE FOR OILS, continued
3. Add one drop of 2,4 Dinitrophenol Indicator Solution.
4. Add one drop of 8 N Potassium Hydroxide (KOH) Standard Solution,
swirling between each addition, until the first flash of yellow appears
(pH 3). If analyzing for potassium, use 5 N sodium hydroxide instead.
Do not use a pH meter if analyzing for potassium or silver.
5. Add one drop of 1 N KOH. Stopper the cylinder and invert several
times to mix. If analyzing for potassium, use 1 N sodium hydroxide
instead.
Note: Use pH paper to insure the pH is 3. If it is higher than 4, do not readjust with
acid; start over with a fresh aliquot.
5. Continue to add 1 N KOH in this manner until the first permanent blue
color appears.
56
DIGESTION PROCEDURE FOR OILS, continued
6. Add deionized water to the volume indicated in the
colorimetric procedure.
7. Continue with the colorimetric procedure.
A × 5000
------------------------ = mg/kg Total Al
B× C
A = mg/L reading from instrument
A × 5000
------------------------ = mg/kg Total Al
B× C
A = mg/L reading from instrument
57
DIGESTION PROCEDURE FOR OILS, continued
Cadmium, Dithizone (Method 8017)
A × 25
----------------- = mg/kg Total Cd
B× C
A = µg/l reading from instrument
58
DIGESTION PROCEDURE FOR OILS, continued
Cobalt (Method 8078)
A × 2500
------------------------ = mg/kg Total Co
B× C
A = mg/L reading from instrument
59
DIGESTION PROCEDURE FOR OILS, continued
Iron, 1,10 Phenanthroline (Method 8008)
A × 2500
------------------------ = mg/kg Total Fe
B× C
A = mg/L reading from instrument
60
DIGESTION PROCEDURE FOR OILS, continued
Lead, Dithizone (Method 8033)
A × 25
----------------- = mg/kg Total Pb
B× C
A = µg/l reading from instrument
61
DIGESTION PROCEDURE FOR OILS, continued
Nickel, PAN (Method 8150)
A × 2500
------------------------ = mg/kg Total Ni
B× C
A = mg/L reading from instrument
A × 75
----------------- = mg/kg TKN
B× C
A = mg/L reading from instrument
62
DIGESTION PROCEDURE FOR OILS, continued
Phosphorus, Ascorbic Acid (Method 8048)
A × 2500
------------------------ = mg/kg Total PO4
B× C
A = mg/L reading from instrument
63
DIGESTION PROCEDURE FOR OILS, continued
Silver (Method 8120)
A × 5000
------------------------ = mg/kg Total Ag
B× C
A = mg/L reading from instrument
64
65
Danger
Review Section 3 for important safety information before
performing this digestion procedure.
Danger
Lire au chapitre 3 les informations importantes pour la
sécurité avant d’effectuer cette technique de digestion.
66
DIGESTION PROCEDURE FOR SOLIDS
1. Transfer 0.50 g or 2. Add 4 mL 3. Turn the temperature 4. Place the flask weight
less of sample into a concentrated sulfuric dial to a heat setting of followed by the
100-mL Digesdahl acid (spec. gravity 1.84) 440 ° C (825 ° F). When fractionating column
digestion flask; see to the digestion flask. the proper temperature is with funnel on the flask.
Sample and Analysis Note: Use only Hach reached, turn on the Place the flask on the
Volume Tables for Digesdahl digestion flasks. water to the aspirator heater and boil 4
Solids following Volumetric flasks with and make sure there is minutes. Do not boil to
this procedure. concave bottom should not suction to the dryness! If sulfuric acid
be used. fractionating column. is not present in the
Note: Be sure you have a
homogenous sample. Solid Note: Safety glasses and a Note: Wait for the proper flask after the boiling
sample should be finely safety shield placed temperature to be reached period, do not proceed
ground or chopped and between the operator and before sample is placed on to Step 5! Discard the
mixed well. the Digesdahl are required. the heater. sample and use more
sulfuric acid for the
digestion procedure in
Step 2. Or choose a
smaller amount from the
Sample and Analysis
Volume Tables for
Solids.
Note: If sample foams up
into the neck of the flask,
lower temperature to
335 ° C (635 ° F). Continue
heating at lower
temperature until all
water is evaporated. Then
return to original
digestion temperature.
Note: White acid vapors
accompanied with a reflux
line indicate that the
sulfuric acid is boiling
Note: Some organic
samples may need more
than 5 minutes for
complete digestion.
67
DIGESTION PROCEDURE FOR SOLIDS, continued
5. Do not proceed if 6. After addition of 7. Take the hot flask off 8. Dilute the digest to
sulfuric acid is not hydrogen peroxide is the heater and allow the approximately 70 mL
visible in the flask. Add complete, boil off excess flask to air cool. with deionized water.
10 mL of 50% hydrogen hydrogen peroxide by Remove the Note: Add deionized
peroxide to the charred heating for one more fractionating column water slowly at first. Cool
sample via the funnel on minute . Do not heat to from the digestion flask. the flask if necessary
the fractionating dryness. Note: Use finger cots to for handling.
column. Note: If the sample goes to remove the digestion flask.
Note: Do not heat to dryness. dryness, turn off the Place it on a cooling pad
for at least one minute.
Note: Visually confirm the Digesdahl and air cool to
Then remove the column.
presence of sulfuric acid in room temperature. Add
water to flask before Do not add water to the
the flask before adding
handling. Repeat the flask until it has cooled.
hydrogen peroxide.
digestion from the
Note: If the digest does not beginning using a new
turn colorless, add 5 mL sample.
increments of peroxide until
the digest becomes clear or
does not change color.
Note: If sample foams
during hydrogen peroxide
addition, stop the peroxide
flow and remove the
digestion flask and
fractionating column
(use finger cots). Cool for
30 seconds and return
apparatus to the heating
block. Start peroxide
addition with 2 mL,
then follow with the
remaining peroxide.
68
DIGESTION PROCEDURE FOR SOLIDS, continued
9. If analyzing for 10. Turn the temperature 11. If the sample has 12. Continue with the
aluminum, nickel or dial to a heat setting of visible turbidity, filter or analysis using the
iron, continue to Step 204 ° C (400 ° F). Add wait until the turbidity appropriate procedure
10. If analyzing for other 150 mL of water to a settles, and the upper below.
substances, dilute to the 400-mL beaker. Place portion of the sample
100-mL mark with the beaker on the heater. is clear.
deionized water; skip Place the flask in the Filter as follows:
Step 10 and go to beaker and boil for 15
a) Place a filter paper into
Step 11. minutes. Air cool to
the filter holder with
room temperature and wrinkled surface upward.
dilute to the mark with
deionized water. Invert b) Place the filter holder
assembly in the filtering
several times to mix.
flask and wet the filter
Note: When using a with deionized water to
Digesdahl Digestion ensure adhesion to
Apparatus system without the holder.
temperature control dials,
reset to a lower setting that c) While applying a
gently boils the water. vacuum to the filtering
flask, transfer the sample to
the filtering apparatus.
d) Slowly release the
vacuum from the filtering
flask and transfer to
another container.
Note: If small aliquots
(≤1.0 mL) are analyzed,
filtration is not needed.
Metals Method
Note: If analyzing aliquots smaller than 0.5 mL, pH adjustment is not necessary.
69
DIGESTION PROCEDURE FOR SOLIDS, continued
2. Dilute to about 20 mL with deionized water.
3. Add one drop of 2,4 Dinitrophenol Indicator Solution.
4. Add one drop of 8 N Potassium Hydroxide (KOH) Standard Solution,
swirling between each addition, until the first flash of yellow appears
(pH 3). If analyzing for potassium, use 5 N sodium hydroxide instead.
Do not use a pH meter if analyzing for potassium or silver.
5. Add one drop of 1 N KOH. Stopper the cylinder and invert several
times to mix. If analyzing for potassium, use 1 N sodium hydroxide
instead.
Note: Use pH paper to insure the pH is 3. If it is higher than 4, do not readjust with
acid; start over with a fresh aliquot.
5. Continue to add 1 N KOH in this manner until the first permanent blue
color appears.
70
DIGESTION PROCEDURE FOR SOLIDS, continued
6. Add deionized water to the volume indicated in the colorimetric
procedure.
7. Continue with the colorimetric procedure.
A × 5000
------------------------ = mg/kg Total Al
B× C
A = mg/L reading from instrument
A × 5000
------------------------ = mg/kg Total Al
B× C
A = mg/L reading from instrument
71
DIGESTION PROCEDURE FOR SOLIDS, continued
Cadmium, Dithizone (Method 8017)
A × 25 = mg/kg Total Cd
-----------------
B× C
A = µg/l reading from instrument
A × 2500
------------------------ = mg/kg Total Cr
B× C
A = mg/L reading from instrument
72
DIGESTION PROCEDURE FOR SOLIDS, continued
Cobalt (Method 8078)
A × 2500
------------------------ = mg/kg Total Cu
B× C
A = mg/L reading from instrument
73
DIGESTION PROCEDURE FOR SOLIDS, continued
Iron, 1,10 Phenanthroline (Method 8008)
A × 2500
------------------------ = mg/kg Total Fe
B× C
A = mg/L reading from instrument
74
DIGESTION PROCEDURE FOR SOLIDS, continued
Lead, Dithizone (Method 8033)
A × 25 = mg/kg Total Pb
-----------------
B× C
A = µg/l reading from instrument
A × 2500
------------------------ = mg/kg total Mn
B× C
A = mg/L reading from instrument
75
DIGESTION PROCEDURE FOR SOLIDS, continued
Nickel, PAN (Method 8150)
A × 75
----------------- = mg/kg TKN
B× C
A = mg/L reading from instrument
76
DIGESTION PROCEDURE FOR SOLIDS, continued
Phosphorus, Ascorbic Acid (Method 8048)
A × 2500
------------------------ = mg/kg Total PO4
B× C
A = mg/L reading from instrument
A × 2500
------------------------ = mg/kg Total K
B× C
A = mg/L reading from instrument
77
DIGESTION PROCEDURE FOR SOLIDS, continued
Silver (Method 8120)
A × 5000
------------------------ – mg/kg Total Zn
B× C
A = mg/L reading from instrument
78
MAINTENANCE
79
80
SECTION 5 MAINTENANCE
5.1 Fuse Replacement
The fuse holder is located on the back panel within the power cord
receptacle. See Figure 2. To replace a fuse, turn off the instrument power
and disconnect the power cord from the back panel. Removal of the
vertical support is necessary. Remove the fuse holder by prying with a
small screwdriver or other pointed tool at the small opening at the top of
the fuse holder. Replace the fuse with the same type and rating used
originally. Refer to Replacement Parts.
81
82
GENERAL INFORMATION
83
84
REPLACEMENT PARTS
REQUIRED REAGENTS
Quantity Required
Description Per Digestion Unit Cat. No.
Hydrogen Peroxide, 50% ................................................... 10 mL .......... 500 mL .......21196-49
Sulfuric Acid, ACS,
(concentrated, spec. gravity 1.84).............................. ≥ 2.5 mL .............. 4 kg ...........979-09
Water, demineralized ...........................................................varies...................4 L ...........272-56
REQUIRED APPARATUS
Dispenser, pour-out, 10 mL.................................................... 1.....................each .......22200-38
Pipet, serological, 10 mL........................................................ 1.....................each ...........532-38
Pipet filler, safety bulb ........................................................... 1.....................each .......14651-00
Boiling chips, silicon carbide ..............................................varies............... 500 g .......20557-34
Safety glasses ......................................................................... 1.....................each .......18421-00
Safety shield, for Digesdahl ................................................... 1.....................each .......20974-00
OPTIONAL REAGENTS
Description Unit Cat. No.
Hydrogen Peroxide, 30% ......................................................................... 200 mL ...........144-45
Kjeldahl Reduction Reagent (for fluid fertilizers) ......................................... 40 g .......23653-04
2, 4-Dinitrophenol Indicator Solution ............................................ 100 mL MDB .........1348-32
Nitric Acid Solution, 1:1 .......................................................................... 500 mL .........2540-49
Potassium Hydroxide, 1 N .............................................................. 50 mL SCDB .......23144-26
Potassium Hydroxide Standard Sol’n, 8 N............................................... 500 mL ...........282-49
Sodium Hydroxide, 5 N ................................................................ 50 mL* SCDB .........2450-26
Sodium Hydroxide, 1 N ............................................................... 118 mL* MDB .........1045-37
TKN Indicator Solution................................................................... 50 mL SCDB .......22519-26
OPTIONAL APPARATUS
Description Unit Cat. No.
Balance, Precision, 115 V ..............................................................................each .......26104-00
Balance, Precision, 230 V ..............................................................................each .......26104-02
Beaker, 400 mL ..............................................................................................each ...........500-48
Beaker, Berzelius, 200 mL .........................................................................12/pkg .......22761-75
Bottle, Wash, 1 L............................................................................................each ...........620-16
Bulb, dropper, 2 mL ...................................................................................12/pkg .......21189-00
Cylinder, graduated, 50 mL............................................................................each ...........508-41
Dispenser, 1-5 mL (for H2SO4, meat)............................................................each .......23121-37
Dispenser, 10-50 mL (for H2SO4, meat)........................................................each .......23121-41
Filter discs, glass, 47 mm .........................................................................100/pkg .........2530-00
Filter holder, membrane .................................................................................each .........2340-00
85
REPLACEMENT PARTS, continued
OPTIONAL APPARATUS (continued)
Description Unit Cat. No.
Flask, filter, 500 mL....................................................................................... each........... 546-49
Flask, flat-bottom volumetric, 100 mL .......................................................... each....... 23125-42
Fume Scrubber Apparatus, 115 V.................................................................. each....... 23266-00
Fume Scrubber Apparatus, 230 V.................................................................. each....... 23266-02
Oven, laboratory, 120 V................................................................................. each....... 14289-00
Oven, laboratory, 230V.................................................................................. each....... 14289-02
Paper, weighing, 76 x 76 mm .................................................................. 500/pkg....... 14738-00
pH paper, pH 1-11..................................................................................5 roll/pkg........... 391-33
Pipet, Pasteur, disposable, 229 mm.......................................................... 200/pkg....... 21234-01
sension™1 Basic Portable pH Meter, with electrode ................................... each....... 51700-10
Spatula, stainless, 10 cm ................................................................................ each........... 561-64
Spoon, measuring, 0.05 g............................................................................... each........... 492-00
Stir bar, PTFE, 1”........................................................................................... each....... 20953-51
Stir plate, magnetic, 7X7, 115 Vac................................................................. each....... 23444-00
Stir plate, magnetic, 7X7, 230 Vac................................................................. each....... 23444-02
Stopper, hollow, size #5 ............................................................................... 6/pkg....... 14480-05
Syringe, 5 mL, plastic .............................................................................. 100/pkg....... 23433-33
Watch Glass, 65 mm .................................................................................. 12/pkg........... 578-97
86
HOW TO ORDER
87
REPAIR SERVICE
88
WARRANTY
89