You are on page 1of 5

ARTICLE IN PRESS

Applied Radiation and Isotopes 64 (2006) 485–489


www.elsevier.com/locate/apradiso

18
Calibration of the Capintec CRC-712M dose calibrator for F
L. Moa,b,, M.I. Reinharda, J.B. Daviesa,b, D. Alexieva, C. Baldockb
a
Australian Nuclear Science and Technology Organisation (ANSTO), New Illawarra Road, Lucas Heights, NSW 2234, Australia
b
Institute of Medical Physics, School of Physics, University of Sydney, NSW 2006, Australia
Received 6 January 2005; received in revised form 26 July 2005; accepted 28 September 2005

Abstract

Primary standardisation was performed on a solution of 18F using the 4pb–g coincidence counting efficiency-tracing extrapolation
method with 60Co used as a tracer nuclide. The result was used to calibrate the ANSTO secondary standard ionisation chamber which is
used to disseminate Australian activity standards for gamma emitters. Using the secondary activity standard for 18F, the Capintec CRC-
712M dose calibrator at the Australian National Medical Cyclotron (NMC) Positron Emission Tomography (PET) Quality Control
(QC) Section was calibrated. The dial setting number recommended by the manufacturer for the measurement of the activity of 18F is
439. In this work, the dial setting numbers for the activity measurement of the solution of 18F in Wheaton vials were experimentally
determined to be 443712, 446712, 459711, 473715 for 0.1, 1, 4.5 and 9 ml solution volumes, respectively. The uncertainties given
above are expanded uncertainties ðk ¼ 2Þ giving an estimated level of confidence of 95%. The activities determined using the
manufacturer recommended setting number 439 are 0.8%, 1.4%, 4.0% and 6.5% higher than the standardised activities, respectively. It
is recommended that a single dial setting number of 459 determined for 4.5 ml is used for 0.1–9 ml solution in Wheaton vials in order to
simplify the operation procedure. With this setting the expended uncertainty ðk ¼ 2Þ in the activity readout from the Capintec dose
calibrator would be less than 6.2%.
Crown Copyright r 2005 Published by Elsevier Ltd. All rights reserved.
18
Keywords: F; Solution; Activity measurement; Dose calibrator; Ionisation chamber; Dial setting number

1. Introduction activity values indicated by the instrument depend on the


induced ionisation current and a conversion factor in the
Cyclotron produced 18F in the form of 2-[18F]-fluoro-2- form of a dial setting number that converts pA to MBq.
deoxyglucose [(18F)FDG)] is mainly used in positron The instrument manufacturer has determined the dial
emission tomography (PET) imaging for assessment of setting numbers from calibration curves obtained by
abnormal glucose metabolism to assist in the diagnosis of measuring the response of the chamber to 57Co and 60Co
cancer. Accurate administration of radiopharmaceuticals (Capintec, Inc., 1990). The dial setting number that is
for diagnosis in nuclear medicine requires the instrument recommended by Capintec for 18F is 439. This is only valid
for activity measurement, the dose calibrator, to be for the geometry of 5 ml of solution in a 5 ml ampoule.
calibrated and traceable to national standards. Investigation of dial settings of Capintec dose calibrators
The Australian National Medical Cyclotron (NMC) has been performed on various geometries on 90Y (Coursey
routinely produces (18F)FDG. The ionisation chamber et al., 1993) and 117mSn, 62Cu, 188Re, 125I, 99mTc, 133Xe,
153
used by the NMC PET Quality Control (QC) Section is the Gd, 153Sm and 18F (Zimmerman et al., 1998, 1999, 2001;
commercial re-entrant dose calibrator, model CRC-712M, Zimmerman and Cessna, 1999, 2000). It was found that
manufactured by Capintec Inc (Pittsburgh, PA, USA). The most of the dial settings recommended by the manufacturer
result in an overestimation of the activity by up to 50%. In
Corresponding author. Australian Nuclear Science and Technology the case of 18F, up to 6% of overestimation of activity was
found depending on the sample geometries.
Organisation (ANSTO), New Illawarra Road, Lucas Heights, NSW 2234, 18
Australia. Tel.: +61 2 9717 9078; fax: +61 2 9717 9265. F decays by positron emission with an intensity of
E-mail address: lmx@ansto.gov.au (L. Mo). 96.86% and electron capture with an intensity of 3.14% to

0969-8043/$ - see front matter Crown Copyright r 2005 Published by Elsevier Ltd. All rights reserved.
doi:10.1016/j.apradiso.2005.09.006
ARTICLE IN PRESS
486 L. Mo et al. / Applied Radiation and Isotopes 64 (2006) 485–489

the ground state of stable nuclide 18O. The half-life of 18F is subtraction of the tracer activity from the extrapolated
1.8288 h (Be et al., 2004). The positron has a maximum absolute total b count rate of the mixture of 18F and 60Co
energy of 633.5 keV and an average energy of 249.3 keV. radionuclides. The activity concentration of the master
The 511 keV annihilation radiation following the emission solution was then obtained and used to calibrate the
of a positron has an intensity of 193.72% (Be et al., 2004). ANSTO Secondary Standard ionisation chamber. Radium
Liquid scintillation counting with the CIEMAT/NIST and check sources were used in all secondary standard
TDCR methods and 4pb–g coincidence counting technique ionisation chamber measurements to correct for the
have been used to standardise this radionuclide (Woods instability of the ionisation chamber and the inaccuracy
and Baker, 2004; Ratel et al., 2005). of the electrometer. The details of the primary standardisa-
In this work, the primary standardisation of 18F was tion procedures are described by Reinhard et al. (2002).
performed using the 4pb–g coincidence counting efficiency- A second batch of (18F)FDG stock solution was
tracing extrapolation method with 60Co as the tracer. The obtained from the NMC for the calibration of CRC-
standardised solution was then used to calibrate the 712M dose calibrator at the NMC at a later time. The
ANSTO secondary standard ionisation chamber. The stock solution was diluted with 0.1 M HCl to produce a
ANSTO Secondary Standard ionisation chamber is a re- 15 ml master solution with an activity of approximately
entrant, high-pressure TPA ionisation chamber (Sharpe 20 GBq. The master solution was then dispensed into a
and Wade, 1951). It is used as a working chamber for the series of 0.1, 1, 4.5 and 9 ml solutions in Wheaton vials.
routine dissemination of the activity standards for photon The samples were measured in the dose calibrator in the
emitting radionuclides. The CRC-712M dose calibrator same position with the readout of the Capintec dose
located at the NMC PET QC section was calibrated for the calibrator displayed in units of activity. Measurements
geometry of 0.1, 1, 4.5 and 9 ml solutions in 10 ml Wheaton were made as a function of the dial setting from 420, in
vials and the standard activity of the solution was increments of 10, to 500. A radium check source was used
determined using the ANSTO secondary standard ionisa- for stability measurements.
tion chamber. The samples were then dispensed into 5 ml BIPM
standard glass ampoules with nominally 3.6 ml of
2. Experimental (18F)FDG solution and the activity determined using the
secondary standard ionisation chamber.
A solution of (18F)FDG in an isotonic solution of Detection of photon emitting impurities was carried out
sodium chloride was obtained from the NMC. The using a HPGe crystal detector (Model no.: GMX-30190-P).
solution was diluted by the addition of 0.1 M HCl to The crystal dimension was 54.8 mm in diameter and
produce the F-18 master solution. The master solution had 76.8 mm in length. The source was positioned at 90 mm
an activity concentration of approximately 980 kBq/g in away from the detector end cap (93 mm from crystal head).
12 ml. Nominally 0.8 g of the 18F solution was taken from The signals from the detector were processed by the
the master solution and mixed with the nominally 0.2 g of ORTEC 92X Spectrum MASTERTM. The software used
the 60Co solution in a vial. From the mixture, a set of six 4p was ORTEC MAESTROTM for Windows.
counting sources was prepared on Pd/Au plated VYNS The linearity of both secondary standard ionisation
source mounts for the primary standardisation to deter- chamber and CRC-712M dose calibrator was checked by
mine the activity concentration of the master solution. The measuring the response to a 99mTc source over eight half-
exact mass dispensed was determined by weighing the lives.
syringe pycnometer before and after dispensing. Another The ‘‘calibration curve method’’ (Zimmerman and
set of three samples was prepared with nominally 3.6 g Cessna, 2000) was employed for the experimental determi-
taken from the master solution and dispensed into the 5 ml nation of the dial setting number. It is described below.
BIPM standard glass ampoules for the calibration of the For each sample, the dose calibrator activity readout
ANSTO secondary standard ionisation chamber. The mass Acap was divided by the standard activity Astd estimated by
dispensed from the master solution into the ampoules was the secondary standard ionisation chamber, giving the
determined by weighing the ampoules together with the cap ratio R ¼ Acap =Astd . The values of R vs dial setting number
before and after dispensing. The ampoules were then flame- were then fitted to the equation,
sealed.
R1 ¼ a  ðDial settingÞ þ b. (1)
The primary standardisation was performed using a
conventional 4pb(PC)–g coincidence counting system. A 4p Solving the fitting equation for R ¼ 1 enables the
gas-flow proportional counter was used for the b detection determination of the experimental dial setting number for
and a Nal(Tl) scintillator used for the g detection. The 4pb- each sample.
g coincidence counting efficiency-tracing extrapolation The uncertainties were evaluated in accordance with the
method (Baerg, 1973) was applied and 60Co was used as ISO GUM (International Organisation for Standardisa-
the tracer. Al foils were used to attenuate the b+ from the tion, 1993) and are quoted here as standard uncertainties
18
F and b from the 60Co in order to vary the b detection with a coverage factor of k ¼ 1 providing a confidence level
efficiency. The activity of 18F was determined by the of 68% unless otherwise specified.
ARTICLE IN PRESS
L. Mo et al. / Applied Radiation and Isotopes 64 (2006) 485–489 487

3. Results and discussion Table 2


Uncertainty components for the calibration factor for (18F)FDG for the
ANSTO secondary standard ionisation chamber
It is noted that the impurity of 48V was found in the
18
( F)FDG produced at the PET Center of the National Source of uncertainty Standard uncertainty
Institutes of Health in the USA (Zimmerman et al., 2001). components (%)
However, neither this work, nor the NMC during their 12
Type A Type B
years of production of (18F)FDG detected any g-emitting
impurities above the detection limit of the g-ray spectro- Activity concentration of primary solution 1.4
scopy system. ESDMa of sample current 0.05
The activity concentration of the (18F)FDG primary ESDM of background current 0.02
ESDM of Ra check source current at 0.03
solution determined using the 4pb–g coincidence counting
measurement time tm
efficiency-tracing extrapolation method was found to be Ra check source current at calibration time tc 0.30
1144.8 kBq/g71.4% at the reference time. The uncertainty Non-linearity of the chamber 0.35
components for the activity concentration of the primary Half life of 18F 0.02
source of (18F)FDG are given in Table 1. It can be seen Quadratic sum 0.36 1.4
Combined standard uncertainty 1.4
that the dominant contributors were the uncertainty in the
efficiency extrapolation (0.90%) and the standard devia- a
Experimental standard deviation of the mean.
tion of the samples (0.92%). The accuracy of the primary
standardisation was confirmed through an international
comparison of the activity of 18F coordinated by the Table 3
National Physical Laboratory in the UK (Woods and Determined dial setting number and its uncertainty for different
Baker, 2004). ANSTO’s result differed from the key geometries
comparison reference value by 0.60% (Ratel et al., 2005). Volume (ml) Dial setting Uncertaintya
A calibration factor of 29.774 pA/MBq71.4% was
determined for the ANSTO Secondary Standard ionisation 0.1 443 12
1 446 12
chamber for the geometry of 3.6 ml solution in a 5 ml
4.5 459 11
ampoule. The uncertainty components for the calibration 9 473 15
factor are given in Table 2. The dominant uncertainty
a
contributor was the activity concentration of the primary The uncertainty is an expanded uncertainty with coverage factor k ¼ 2
giving 95% confidence level.
solution (1.4%). The other contributions listed in the
Table 2 were insignificant in comparison with the large
uncertainty in the activity concentration of the primary
solution. reference time. The ratio of the standard activity Astd to
The standard activity concentration of the second batch the activity readout from the Capintec CRC-712 M dose
of (18F)FDG determined by the secondary standard calibrator Acap, R1 ¼ Astd =Acap , was plotted against the
ionisation chamber was 1334.82 kBq/g71.5% at the dial setting number and fitted to Eq. (1) using Microsoft
ExcelTM. The corrected dial setting for each volume was
determined by taking an average of two samples. See Table
Table 1
3 for the determined dial setting number and its uncertainty
Uncertainty components for the activity concentration of the primary for each sample volume.
source of (18F)FDG standardised using the 4pb–g coincidence counting The values for R ¼ Acap =Astd have a standard uncer-
efficiency-tracing extrapolation method tainty of 1.6%. The uncertainty components contributed to
Source of uncertainty Standard uncertainty
the combined uncertainty in R are listed in Table 4. The
components (%) calibration factor for the secondary standard ionisation
chamber contributed 1.4% of the combined uncertainty in
Type A Type B R-value. The calibrated current of radium check source for
Counting statistics 0.17 the secondary standard ionisation chamber contributed
Dead time 0.01 0.30% and the activity of the radium check source for the
Activity concentration of 60Co tracer 0.38 dose calibrator contributed 0.40%. The non-linearity for
Mass of 4p counting source 0.01 the Secondary Standard ionisation chamber and dose
Mass of 18F and 60Co mixture 0.03
calibrator contributed 0.35% and 0.50% respectively. The
Mass of 18F master solution 0.04
Half life of 18F 0.04 variations in wall thickness of the Wheaton vial contrib-
18
F b+ emission probability 0.17 uted 0.30%. The other uncertainty components have
Intercept from fitted curve 0.90 minimal influence. The standard uncertainty in the dial
Standard deviation of samples 0.92 setting was determined by solving the fitting Eq. (1) for the
Quadratic sum 1.4 0.18
dial setting values for R ¼ 1  U R , using the approach of
Combined standard uncertainty 1.4
Zimmerman et al. (2001).
ARTICLE IN PRESS
488 L. Mo et al. / Applied Radiation and Isotopes 64 (2006) 485–489

Table 4
Uncertainty components for the R ¼ Acap =Astd value determined for the Capintec CRC-712M dose calibrator for (18F)FDG

Source of uncertainty Standard uncertainty components (%)

Type A Type B

Calibration factor for the secondary standard chamber 1.4


ESDM of sample current in secondary standard chamber 0.04
ESDM of background current 0.02
ESDM of Ra check source current at measurement time tm 0.03
Ra check source (for secondary standard chamber) current at Ra calibration time tc 0.30
Non-linearity of the secondary standard chamber 0.35
Non-linearity of the dose calibrator 0.50
Activity of Ra check source (for dose calibrator) 0.40
ESDM of activity readout from dose calibrator 0.07
Sample mass 0.05
Variation in wall thickness of Wheaton vial 0.30
Half life of 18F 0.08
Quadratic sum 0.62 1.4
Combined standard uncertainty 1.6

Although the international comparison of the activity of 480


18 475
F confirmed the accuracy of ANSTO’s primary standar-
disation of the nuclide, 60Co might not be an ideal tracer to 470
Dial setting

465
be used for the absolute measurement of activity of 18F.
460
The uncertainty in the activity concentration determined
455
by the tracing method using 60Co as a tracer was as large as 450
1.4%, mainly due to the large uncertainty in the efficiency 445
extrapolation and standard deviation of different samples. 440
With 1.4% standard uncertainty being propagated to the 0 2 4 6 8 10
calibration factor and then the standard activity value Solution volume (ml)
determined using the secondary standard ionisation Fig. 1. Plot of dial setting number as function of solution volume. The
chamber, a large uncertainty was consequently obtained solid line is a linear fit.
for the dial setting. This is the reason why the expanded
uncertainties in the dial setting in this work were larger by a single dial setting is required for any volume between 0.1
factor of two compared with the work of Zimmerman et al. and 9 ml solution in Wheaton vials, a dial setting of 459
(2001). determined for the 4.5 ml would be recommended.
The dial setting recommended by the manufacturer, and When the dial setting was plotted against solution
currently used by the NMC, is 439 which is lower than the volume, it was found that the dial setting increased linearly
dial setting determined in this work resulting in over- with sample volume (Fig. 1).
estimation of sample activities by 0.8%–6.5% for the For photon-emitting nuclides one would expect that the
sample volume of 0.1–9 ml. Use of a larger volume results dial setting [or calibration factor (pA/MBq)] would
in a greater discrepancy. increase as the sample volume decreases. The opposite
The experimental results showed that if the dial setting of trend, observed in Fig. 1, indicates that the dial settings
459 for 4.5 ml was used for the 0.1, 1, and 9 ml solutions, were affected by the variations in geometry within the
the R-value would be 1.030, 1.025, 1.025. This would give chamber well. The relatively high-energy (511 keV) annihi-
rise to an uncertainty of p3% for all volumes of solution lation photons may not significantly affect the dial setting
investigated. However, even with consideration of other with changing solution volume.
uncertainty sources such as the linearity of the dose
calibrator (0.50%), stability correction for the dose 4. Conclusion
calibrator using the radium check source (0.40%) and the
variation of wall thickness of the Wheaton vials (0.30%), Primary standardisation was carried out on a solution of
the combined standard uncertainty would be 3.1% (or the (18F)FDG using the 4pb–g coincidence counting efficiency-
expanded uncertainty of 6.2% with k ¼ 2 for an estimated tracing extrapolation method. The result agreed with the
level of confidence of 95%) for the activity readout from key comparison reference value within 0.60% as verified in
the dose calibrator. In the nuclear medicine environment, an international comparison. The result was used to
as the ‘‘rule of thumb’’, the acceptable tolerance of the calibrate the ANSTO secondary standard ionisation
error for activity determination is 10%. Therefore, if a chamber.
ARTICLE IN PRESS
L. Mo et al. / Applied Radiation and Isotopes 64 (2006) 485–489 489

Using the secondary activity standard for the (18F)FDG, Capintec, Inc., Pittsburgh, PA, 1990. CRC-712 Radioisotope Calibrator
the Capintec CRC-712M dose calibrator located at the Owner’s Manual.
Coursey, B.M., Calhoun, J.M., Cessna, J.T., 1993. Radioassays of
NMC PET QC section was calibrated. New dial setting
yttrium-90 used in nuclear medicine. Nucl. Med. Biol. 20 (5), 693–699.
numbers, 443712, 446712, 459711, 473715, for the International Organisation for Standardisation, 1993. Guide to the
Capintec dose calibrator have been experimentally deter- Expression of Uncertainty in Measurement.
mined for (18F)FDG for the 0.1, 1, 4.5 and 9 ml solutions in Ratel, G., Michotte, C., Woods, M.J., 2005. Comparisons CCRI(II)-
a Wheaton vial, respectively. To simplify the operation K3.F-18 and APMP.RI(II)-K3.F-18 of activity measurements of the
procedure, a dial setting of 459 for 4.5 ml is recommended radionuclide 18F and links to the key comparison reference value of the
BIPM.RI(II)-K1.F-18 comparison. Metrologia 42, Tech. Suppl.,
for any volume between 0.1 and 9 ml solution in Wheaton 06007.
vials. The uncertainty in the activity readout from the Reinhard, M.I., Mo, L., Alexiev, D., 2002. Primary Standardisation of 18F
Capintec dose calibrator by using a single dial setting is by the Efficiency Tracing Technique using 60Co, ANSTO AP/TN-273.
well below 10%, an acceptable error tolerance of activity Sharpe, J., Wade, F., 1951. TPA Mk II ionisation chamber, AERE EL/
measurement in the nuclear medicine environment. R806.
Woods, M.J., Baker, M., 2004. Establishing equivalence for activity
standards of short-lived radionuclides using the NPL secondary
Acknowledgments standard radionuclide calibrator. Appl. Radiat. Isot. 60, 499–504.
Zimmerman, B.E., Cessna, J.T., 1999. The standardisation of 62Cu and
experimental determinations of dose calibrator settings for generator-
The author would like to thank Daniel Adam Hatton produced 62CuPTSM. Appl. Radiat. Isot. 51, 515–526.
and Amy Le of the Australian National Medical Cyclotron Zimmerman, B.E., Cessna, J.T., 2000. Experimental determinations of
for the sample preparation and experimental schedule commercial ‘dose calibrator’ settings for nuclides used in nuclear
arrangement. medicine. Appl. Radiat. Isot. 52, 615–619.
Zimmerman, B.E., Cessna, J.T., Schima, F.J., 1998. The standardisation
of the potential bone palliation radiopharmaceutical
References 117m
Sn(+4)DTPA. Appl. Radiat. Isot. 49, 317–328.
Zimmerman, B.E., Cessna, J.T., Unterweger, M.P., Li, A.N., Whiting,
Baerg, A.P., 1973. The efficiency extrapolation method in coincidence J.S., Knapp Jr., F.F., 1999. A new experimental determination of the
counting. Nucl. Instr. Methods 112, 143–150. dose calibrator setting for 188Re. J. Nucl. Med. 40 (9), 1508–1516.
Be, M.-M., Chiste, V., Dulieu, C., Browne, E., Chechev, V., Kuzmenko, Zimmerman, B.E., Kubicek, G.J., Cessna, J.T., Plascjak, P.S., Eckelman,
N., Helmer, R., Nichols, A., Schonfeld, E., Dersch, R., 2004. Table of W.C., 2001. Radioassays and experimental evaluation of dose
radionuclide, Monographie BIPM-5. calibrator settings for 18F. Appl. Radiat. Isot. 54, 113–122.

You might also like