*Delfi
. . . ..dctl(1ffledcail .com
Part# 9-2100-001
U.S. PATENTS 6,213,939; 5,931,853; 5,649,954; 4,469,099; PAT. PEND. and FOREIGN
PATENTS
.
Delfl medical
* innovations inc.
ATION OF REMEDIES . . . .
LIMIT Delfi Medical Innovations Inc. be liable for any special incidental or ~o~sequential d
In no case shall b h of warranty or other legal theory. Some states do not allow hm1tations on amages
whether bas_ed ofon brea~h in certain transactions. In such states, the limits in this paragraph and the warra~
or on remedies r rea Precet,,ng
paragraph do not apply.
WARRANTY CLAIMS . .
1 th ent of a warranty claim within the warranty period please take the following steps:
i" N::; customer service Department, Delfi Medical Innovations Inc. at 800-933-3022. Please provide
details about the nature of the problem and include the product serial number. Upon receipt of this informati
Delfi will provide a date for service or a return shipping authorization. on,
2. Upon receipt of the shipping authorization, forward the equipment, freight prepaid, to the location specified
in the shipping authorization.
Your compliance with these steps will help assure that you receive prompt warranty service for your product.
AC Powers -----
upply Serial Number
TABLE OF CONTENTS
SECTION 1.0: DELFI PORTABLE TOURNIQUET SYSTEM (P.T.S.) GENERAL INFORMATION
1. 1 INTENDED USE ........ ..... ..... .. .. .. ..... .... .. .... .. .. .. .. .. .. .. .... ... ..... .. .... .. .... .. .. .. .. .. .... .. .. .... .. .. 2
1.2 CONTRAINDICATIONS. ........ ... ..... .... ........................ ................ ............... ..... ........ .. 2
1.3 PRECAUTIONS IN USE .. ... .. .. .. .. .. .. .. .. ........ .. .. .. ...... .. ......... .. .. ... .... ........ ..... ... .... .. .... . 2
1.4 ADVERSE EFFECTS .. .... .... ... ............ ...... .... ....... ... ..... .. ... .. .. .... .. .. ... ..... .. ... .. ... .. .. .. ... . 3
SECTION 2.0: P.T.S. INSTALLATION AND OPERATING INSTRUCTIONS
2.1 SPECIFICATIONS
2.1.1 POWER REQUIREMENTS .. ... ................. .. .. ............... .......... .. .. .... .... .. .. ....... 4
2. 1.2 PERFORMANCE .... ... .... ... .. .... ...... ... ....... .. ... ... ........ .. .. .... ... ... ........ .. ... ........ 4
2.1.3 SIZE ... .. ...... ...... ... ......... ............. .. ..... ..... .... ....... .. ....... .. ..... .. ......... ............... 4
2.2 INITIAL INSPECTION .. .. .. .. ..... ....... .. ... ....... ... .. ........ .. .. ... ...... .. ......... .. .. .. ....... .... .. ..... 4
2.3 CONTROLS, INDICATORS, DISPLAYS, AND CONNECTORS ...... .. .. .. .. ... .. .. .. ... .. 5
2.4 INITIAL SETUP ..... .. .................. .... .. ....................... .... ....... ...... ... ... .... .... .... ......... ..... 9
2.5 FUNCTIONAL AND CALIBRATION CHECK ....... .. ............. ........ .... ....... ........ .. ..... . 9
2.5.1 AC POWER CONNECTION AND INDICATOR .. ... ....... .. ...... .. .. ....... .. .. ...... 9
2.5.2 POWER-UP SEQUENCE AND SELF-TESTS ....................... .................... 9
2.5.3 PRESSURE SET POINT ADJUSTMENT .............................. .. .. ................ 9
2.5.4 TIME LIMIT SET POINT ADJUSTMENT ......................... .................. ...... .. 10
2.5.5 CALIBRATION CHECK .. ... .......... ...... .... .. ... .... ........ .. ...... .. .. .. .. ........ ........... 10
2.5.6 "PRESSURE LOW' and "CUFF LEAK" ALARM CHECK .............. ............ 11
2.6 PRESSURE AND TIME DEFAULTS .. ................................................................. ... 11
2. 7 OPERATION ..... ....... ... ..... .... ... .. .... ... ... .. .. .. ............ .... .. ... ..... ..... .. ........ .. .... .. .... ...... ... 11
2.8 ALARM CONDITIONS ............................................................................................ 13
2.8.1 ALARM SILENCE FUNCTIONS ............................................. ................... 17
. 2.8.2 INTERNAL HARDWARE FAILURES ........................... .......... ................ ... 17
2.8.3 PRESSURE HIGH or LOW ALARMS .. .. .................... ...... .......................... 17
2.8.4 LEAK ALARMS ... ....................................................................................... 17
SECTION 3.0: P.T.S. MAINTENANCE AND SERVICE INSTRUCTIONS
3.1 GENERAL MAINTENANCE INFORMATION ..... ... .. ....... ........ ... .. ...... .... .... .... ... ...... ...... 18
3.2 PERIODIC MAINTENANCE ............... .................................................... ........ ............... 18
3.2.1 CLEANING ......... ....................................................................................... 18
3.2.2 INSPECTION .................................. .. ........... ...................................... ........ 18
3.2.3 FUNCTIONAL AND CALIBRATION CHECKS .......................................... 18
3.3 CALIBRATION .......................................................... ................. .. .................................. 18
3.4 LEAK TESTING ................................................................................. ............................ 19
3.5 BATTERY TESTING and REPLACEMENT ................................................................... 20
3.5.1 BATTERY TESTING ........ .. .. .... .... ... ..... .. ...... ......... ... .. ... ... .... ........ .... .. .. .... .. 20
3.5.2 BATTERY PACK REPLACEMENT ................................... .. ............. .......... 20
3.6 TROUBLESHOOTING GUIDE ...... ............... .................................................................. 21
3.7 INTERNAL HARDWARE SERVICING AND MAINTENANCE .... ....... ............................ 23
3.8 WARNINGS, CAUTIONS, LABELS, and SYMBOL DEFINITIONS ....... ............. .... ....... 23
FIGURES and TABLES:
Figure 2.1: P.T.S. Controls, Indicators, Displays, and Connectors ................................................... 7
Figure 2.2: P. T.S. Pressure and Time Display Detail ... .. .. .. ...... ... ... ...... .. .... ..... ..... ........... .... ...... .. .... .. 8
Figure 2.3: P.T.S. Message Display Detail ........................................................................................ 8
Figure 3.1: Labels ................................................ ..... ....... ....................................... .. ......... .... ... .... . ... . 23
Table 2.1: Alarm Conditions .............................................................................................................. 13
Table 3.1: Troubleshooting ................................................................................................................ 21
1
LE TOURNIQUET SYSTEM (P.T.S.)
Fl poRTAB
r 10N 1.0: DEL Ar 10N to the pollcle■ In your practice •ettlng, T
sEC ERAL INFORM y1t1rn according ntralndlcatlons, and adver9e effecta are hffe follow,
GEN hi• tourn tqut t •• pr1c1ut1on1, co o ''•d ,,"g
NOTE: U•• t 1nt1ndt d u• ' '
'"'°""'uon I~~ In thl• proc•••·
gu
ldt to•••. E
,
Is Intended to be used ~Y qualified medical prof . ,
INTENDED U~ uet system (P.T,S·l, extremity during surgical procedures on th ~ss1ona1s to
1.1 lfi portable rourn qd flow In a patlen sl sting less than 90 minutes. Tourniquets ha a extrerni.,,
n,e De de blOO rations a · , · h ve bee •1.
. parsrllY occlu . lly used tor ope Id In surgical procedures invo1ving t e extremities incl d' n founa
tern , ts are genera operation fie u 1ng:
T0 urn1que ploodtess
tul in producing 8 _ Kirschner wire removal
use , of certain fractures _Knee joint replac~men~s .
. Reductiond cyst excisions _Replacement of finger Joints
. rumor an of certain joints - Amputations
- Arthroscopy _Nerve injuries
9
- Bone taran:us tasciotomy _Total wrist joint replacement
- SubcU .
- Tendon repair
ontrol the distal flow of CO2 or any other gases u
nlquet cu ffs to C f ff sect a
ARNING: Do not use tour uffs have not been evalua!ed for -~a ety or e ectlveness in contro 'a
:stention media. Tournlqu,:ta~ site during arthroscopic msufflat1on procedures. Possible effe~lng
as flow beyond the s_urg . ner include serious subcutaneous emphysema proximal to th Of
g. tourniquet cuff m this man e CUff.
using a . .
NTRAINDICATIONS . . . . .
1.2 CO . .
t for possible contraindications to tourniquet use. A partial 11st 1s provided bel
Refer to ~he medical htehra ~re decision to use a tourniquet rests with the attending physician. ow,
however in every case t e ma I
- Post-traumatic lengthy hand reconstruction
_Open fractures of the leg h th · ·
. •inJ·ur·ies - Elbow surgery (w ere ere 1s concomitant excess swelling)
- Severe crus hin 9 . .
rtension - Diabetes melhtus
- Severe hype d'I d' t· · h d
_Skin grafts in which all bleeding points must_ be rea I y 1s 1_ ngu1s e
_compromised vascular circulation, e.g., ~enphe_ral artery disease
_Sickle cell disease or trait (relative contramd1cat1on, see PRECAUTIONS IN USE) .
• Secondary or delayed procedures after immobilization.
2
tourniquet time can also produce changes in the coagulability of the blood with increased
Prolonge d . . . t . t t·
. time Always min1m1ze ourn1que 1me.
clottin~ t paralysis may result from excessive pressure. Insufficient pressure may result in passive
• Tourniqt~en of the limb with possible irreversible functional loss. Always use the minimum effectiv~
conges 10 . . . .
tourniquet pressure, as described in the med1ca_lhteradture._ . It 'bl
fl f should be done rapidly to occlude arteries an veins as near s1mu aneous 1y as poss, e.
• In at~ and complete exsanguination reportedly prolongs pain-free tourniquet time and improves the
• Car~; of Intravenous Regional Anesthesia (Bier Block anesthesia). In the presence of infection and
qu~ ~ul fractures, after the patient has been in a cast, and in amputations due to malignant tumors,
~!:nguination before tourni~uet application may be done without the use of an elastic bandage by
levating the limb for 3 to 5 minutes.
• ~n case of failure, the tourniquet cuff must be fully deflated an~ the limb exsanguinated again before
inflation. Reinflation over blood-filled vasculature may lead to 1ntravascular thrombosis.
• ~urniquet users must . be familiar with the in~ation-deflation se~uence when using two tourniquet
cuffs and two P.T.S. units together for IVRA (Bier Block anesthesia) , so that the wrong tourniquet will
not be released accidentally.
• Test for hemoglobin type and level before using a tourniquet on patients with sickle-cell anemia. When
the tourniquet is used for these patients, the limb should be carefully exsanguinated and the P0 2 and
pH should be closely monitored.
• Select the proper cuff size to allow for the overlap recommended by the cuff manufacturer. Too much
or too little overlap may cause cuff rolling and telescoping, unexpected release of the cuff from the
limb, inability to maintain a bloodless field at normal pressures, and/or undesired pressure distribution
on the limb.
• The skin under the tourniquet cuff must be protected from mechanical injury by smooth , wrinkle-free
application of the cuff. If the tourniquet cuff is applied over any material that may shed loose fibers
(such as Webril) the fibers may become embedded in the contact closures and reduce their
effectiveness. Follow the cuff manufacturer's recommendations for limb protection material under the
cuff. In general, a limb protection sleeve designed specifically for the cuff provides the best protection.
• If skin preparations are used preoperatively, they should not be allowed to flow nor collect under the
cuff where they may cause chemical burns.
• Whenever the tourniquet cuff pressure is released, the wound should be protected from blood surging
back by applying pressure dressings and , if necessary, elevating the limb. Transient pain upon
tourniquet pressure release can be lessened by elevation of the limb. If full color does not return within
3 to 4 minutes after release, the limb should be placed in a position slightly below body level.
• The deflated cuff and any underlying limb protection material should be completely removed
as soon as tourniquet pressure is released. After the cuff has been fully deflated and removed
from the patient, the unit can be set to STANDBY. Even the slightest impedance of venous
return may lead to congestion and pooling of blood in the operative field.
• Whenever IVRA (Bier Block anesthesia) is used, it is recommended that the tourniquet remain inflated
for at least 20 minutes from the time of injection.
3
LATION AND OPERATING INSTRUCTIQ
. p T.S. 1NSTAL Ns
SECTION 2.0. .
5pECIFICA TIONS
2.1 REQUIREMENTS
2.1.1 poWER (AC powered Mode):
M In• Un• Voltage
~ 264 VAC . 47/63 Hz.
Lin• currenst: t 90 VAC input.
< o.sARM a
Input power:
20 watts maximum.
AC power cord: ssembly supplied by Delfi.
AC adapter/power cord a
AC power plug: ng straight blade, 15 amp.
Hospital grade, 3 pro .
Battery Type: d .d (NiMh) pack, 1500 milliamp hours (use only batteries supplied by O lfi
Sealed 4.SV nickel-metal-hy n e e ).
Battery Recharge Time:
a.ohours (typical).
2.1.2 PERFORMANCE
cuff Pressure Range:
50-4 75 mmHg, 5 mmHg increments.
pressure Accuracy:
+/-5 mmHg.
Pressure Regulation: . ..
+/-10 mmHg of set point (10 second average under non transient cond1t1ons).
nme Alarm Set Point Range:
0-240 minutes.
Timer Accuracy:
0.1% of elapsed time.
Inflation Rate:
34 inch cylindrical cuff applied to a 30 inch thigh inflates to 350 mmHg in less than 20 seconds.
Deflation Rate:
34 inch cuff applied to a 30 inch thigh deflates to less than 10 mmHg in less than 25 seconds.
2.1.3 SIZE
Height: 240 mm (9.45 inches)
Width: 116 mm (4.57 inches)
Depth: 50 mm (1.97 inches)
Weight: 500 g (17.5 oz)
4
CONTROLS, INDICATORS, DISPLAYS, AND CONNECTORS
2.3
Refer to Figure 2.1 for the locations of the unit's controls, indicators, displays and connectors described below:
1_'ON/STANDBY' Button
Turns the unit on or sets the_ unit to st~ndby. In standby m~de (powered off), the power to all instrument
functions (i.e. inflati~n, deflation, e_tc.) 1s off,. but power con~mues to supply the battery charging circuitry
whenever AC power 1s present. This button will n?t set the u_nit to sta~dby when there is pressure in the cuff.
Ensure the cuff is fully deflated, and the cuff and hmb protection material have been removed from the patient
prior to setting the unit to standby.
6. 'INFLATE' Button
Inflates the cuff to the pressure set point and starts the elapsed time monitor. Momentarily depressing the
'INFLATE' button immediately begins rapid inflation of the cuff.
7. 'DEFLATE' Button
Deflates the cuff and stops the elapsed time monitor. To prevent accidental deflation of the cuff, the
'DEFLATE' button has a delay and must be pressed and held for approximately 2 seconds before the unit will
deflate the cuff. A short tone is sounded after the 2 second delay to indicate that deflation has started and the
user may then release the 'DEFLATE' button . If the user momentary presses then releases the 'DEFLATE'
button, nothing happens. If the user releases the 'DEFLATE' button any time after deflation has begun, the
cuff continues to deflate to zero pressure.
5
9 Message Display .
· present the corresp
. ending words are. displayed and will remain d' 1
I conditions are , . ·1 bl VVh isp ayed
w,en a arrT_' , . orrected, as long as sufficient power 1s ava1a e. en operating on batte until the
alarm c~n~1t1on is c bol is shown (see below). ry Power the
battery indicator sym
4
2
----17
~
I- -
15 13
16-----
---
8
h
2. 4 INITIAL SETUP
t t 'ft
. h. ing and storage, the unit's battery could become weak. Prior to initial use, the unit must be
ounng AC power using the AC power supply and cord assembly until the battery is fully charged . This
P1~.9 h~~ge should take no more than 8 hours. The battery must be fully charged before any initial use,
:~~:~di~g any calibration checking procedures, initial checks, or tests performed by biomedical engineering at
your facility.
WARNING: use the AC power supply and cord assembly supplied with your P.T.S. only. Do not use
other AC power supply or cord. Use of an improper power supply may cause irregular operation
:::icould be hazardous to the patient and/or user, and may permanently damage the P.T.S. unit.
CAUTION: Avoid exposing the AC power supp_ly to li~ui~s. Do not immerse in fluid. Do not allow the
AC power supply to lie on the floor where pooling of liquids may occur. Clean by damp cloth (alcohol
or mild detergent wipe) only. The AC power supply is resistant to occasional splashing or dripping of
fluids, but is not fluid-tight. If immersed in or exposed to excessive amounts of liquids, the AC power
supply may fail and may pose an electrical shock hazard.
WARNING : Use battery packs supplied by Delfi only. Do not use any other batteries. Use of an
improper battery pack may cause irregular operation that could be hazardous to the patient and/or
user, and may permanently damage the P.T.S. unit.
9
, RESSURE + , button. The pressure display should read "250• ", and after
1
b) Press and .hold the P and then by 10 mmHg increments to the maximum pressure set point ~econd
begin Increasing In 5mmHg of 475• •
mmHg. , URE + , button. Within 2 seconds, the pressure display returns to "0".
c) Release the PRESS
'PRESSURE _ ' button and release (within 1 second). The pressure di
d) Momentarily press th e t point) for 2 seconds. The pressure display then returns to "0". splay reads
"475*" (the new pressure se
'PRESSURE_ , button. The pressure display should read "475* ", and aft 1
th 8
e) Press and ~old _ ~ Hg increments and then by 10 mmHg to the minimum pressure set po~rt :COnd
begin decreasing in mm in of so• •
mmHg. h d' I
, ESSURE _, button. Within 2 seconds, t e pressure Isp ay returns to "0".
nRelease the PR
TIME LIMIT SET POINT ADJUSTMENT:
2.5.4
. ss the 'TIME + ' button and release (within 1 second). The time display reads ,, 60• ,,
1
a) Mom~ntan. Y·tpreet point) for 2 seconds. Note that the trailing asterisk indicates that the time displ ~th e
default time 11m1 s b . fl t d Th t· d' ayed IS a
. t the current elapsed time that the cuff has een in a e . e 1me 1splay then returns to •0•
~~~00 .
d hold the 'TIME + ' button. The time display should read "60* ", and after 1 second beg·
~) Pre~s an,·n 5 minute and then by 1O minute increments to the maximum time limit set point of •240. in
increasing •
minutes.
c) Release the 'TIME+ ' button. Within 2 seconds, the time display returns to "0".
d) Momentarily press the 'TIME - ' button a~d rel~ase (within 1 second}- ,, The time display reads "240• " (the
new time limit set point) for 2 seconds. The time display then returns to O .
e) Press and hold the 'TIME - ' button_. Th~ time display sho~I~ read _"240_* ' ., and ~fter 1 second begin
decreasing in 5 minute and then by 10 minute increments to the minimum time hm1t set point of "5* " minutes.
nRelease the 'TIME-' button. Within 2 seconds, the time display returns to "0".
NOTE: Anytime a trailing asterisk (*) appears, the data being displayed is the set point. Set pressure
and time will revert to the default pressure and time when the unit is powered off and powered up
again.
~ ;~ess .8nd th
hold the 'D~FLATE' button for 2 seconds to deflate the unit. Disconnect the hose from e
· · ·unit.The pressure display should decrease too mmHg.
n If any stablized
displ d
. . . k the pressure
pressure reading 1s off by more than 5 mmHg during the calibration chec or
ay oes not return to zero, the unit must be calibrated. See Section 3.3 below.
10
-
2. 7 OPERATION
NOTE: The P.T.S. unit should be powered up at least once each day of use to ensure the self-test
routine is performed regularly. The P.T.S. should be powered off and left plugged in to AC power
when not in use.
a) Press the 'ON/STANDBY' button to turn the unit on. The unit will execute a self-check diagnostic test as
described in Section 2.5.2 of this manual. Observe that all the display segments and the red alarm indicator
light illuminate during the self-check as shown in Figs 2.2 and 2.3. Successful completion of the self-check
indicates that the unit is ready for use.
WARNING: If a connected cuff Is pressurized to 50 mmHg or more during power-up, the P.T.S. will
assume that a surgical procedure is in progress, adopt the pressure sensed in the cuff as the new set
point, and will automatically regulate the cuff at this pressu_re. To alert the operator of this condition,
the unit will sound the alarm tone, illuminate the red alarm Indicator light, flash the pressure display,
and show "PRESSURE" In the message display. The operator should immediately check the pressure
set point and readjust to the proper set point if necessary. The alarm will be cleared as soon as the
set point Is examined or adjusted ('PRESSURE + ' or 'PRESSURE - ' button pressed momentarily or
pressed and held).
b) Connect the hose assembly and a single port cuff to the unit at the hose connector.
11
Seleet the appropriate tourniquet pressure ~nd tourniquet time lim~ set ~ints for the 'Peciflc
~=g
c) lfied b the surgeon. The default set points for pressure and tlfTW: hmit are r~ed fr~ ~ e. as
rn1er up and are active until the u&er adjusts them (see Sections 2.5.3 and 2.S.4). ~~
values can be modified by the user (see SectJon 2.6). ~
WARNING: In 111 ca1t1 tt lt enentlal to check the prt11ure and time limit set PC>inta and
they 1 ,. the dt1lred valutt before Inflating the cuff. . C<>nftnn that
For each patient, tourniquet pressure required ~ occlude blood flow to operative site should be
.. "ective pressure The minimum effective pressure depends on many factors i"'"'Ud· . set to the
m,mmum e11, · d 1· of fit · ' '""' tng the in,.__ -
of the cuff (upper or lower limb), the type of cu~ an qu~ ity to the hmb, Whether the flrnb i s ~
'ed or obese the patient's preoperative systohc pressure, and the maXimum a ~ - .noona,
hypertrophI I , . th ed' I 1-+,., atu ' .'-.,OlecJ . . .
systolic pressure during the_procedure. Refer to e m 1ca .,LC:r re ,or current techniques for ~ - fl
the minimum effective tourniquet pressure for each case. ~
d) Prepare the patient in accordance with your established pr~ures an~ ~ manufacturers . .
The precautions of Section 1 an~ the followm_g are offered as ~ guide to assist m this process_I n ~-
a tourniquet cuff should be applied to the widest part of the lt~b to allow as much tissue as P0Ssibte cases
between the cuff and any nerves or vascular structures susceptible to damage. The optimum P0Sitions to ie
upper arm and the proximal third of the thigh. In certain cases of fore-foot surgery, the tourniquet CUff ~e the
applied around the calf or to the area proximal to the malleoli._For emergency surgery of the ~ be
sufficiently small tournique~ may b~ selected by the surgeo~ for fitting around the forearm. Apply a leak-k~
tourniquet cuff smoothly w1tho~~ wrinkles The hose conn~~~s should be ?laced so that the hose Wil not be
kinked when the limb is ~s1ti?ned for ~urgery. The ~1abllio/ of_the skin an~ deeper tissues should be
established prior to e~sangumat1on of ~e ~1rnb_ and tournI~uet mfl~on. Exsangu,nate f!1e limb by elevating it
for a minimum of 2 minutes and wrapping 1t, distal to proximal, using an Esmarch, Martin, or elastic bandage
The bandage should come_ up _ ?'
appr?ximately to 1 in. (2._5 cm) from th~ ~ge the tourniquet cuff. The -
bandage is removed foll_o~ing inflation of the cuff: If reg,on_al anesthesia 1s being used, the anesthetic agent or
nerve block is then administered. The surgeon will determine:
• VVhen the tourniquet is to be inflated;
• VVhat pressure is applied;
• How long the tourniquet is applied;
• VVhether to allow for intermittent aeration of tissue by deflating the cuff, and the duration of these
aeration periods;
• VVhen to deflate and remove the tourniquet.
Appropriate tourniquet time and the need for intermittent deflation of the cuff depend greatly on the patient's
anatomy, age, and absence of vascular disease. In many operating rooms, it is customary to prominently note
the time of inflation, and to warn the surgeon after a certain time has elapsed. This will allow the surgeon to
assess the need for further tourniquet time.
e) Press the 'INFLATE' button when the cuff must be inflated. The unit will pressurize the cuff to the pressure
set point and start the tourniquet time clock. If the unit cannot pressurize the cuff to within 15 mmHg of the set
point in less than 45 seconds, a leak alarm will be sounded (see Table 2.1 for information about possible
alarm conditions). The time display shows the tourniquet time, and the pressure display shows the current
pressure in the cuff.
~ When the cuff must be deflated, press and hold the 'DEFLATE' button. After holding the button for 2
seco nd s,. the al~rm tone will sound briefly, the pressure display will show the falling pressure of th~ cuff, aod
th e tourniquet time clock will stop and hold the display of the tourniquet time. Remove the toumiqu~t c~
and ~ny underlying limb protection material Immediately following final cuff deflation. The time
tourniquet cuff removal should be noted, and the circulation of the limb should be checked.
NOTE: The elapsed tourniquet time can be zeroed after the cuff is deflated by pressing and hold:g
th e 'TIME+ ' and 'TIME- ' buttons simultaneously. The alann tone will sound briefly to ind icate e
time hu been zeroed.
~~t~!er the cuff has ~n removed, disconnect the cuff from the P.T.S. During normal use, the P.T.S~::!
from th:et t? stand bY_ if pressure is present in the cuff. Once the cuff has been properly deflated, r
patient and disconnected from the P.T.S., the unit can be set to standby.
12
,..._. V
~ ;
::, C
Sc
c'., t'D~
2.8 ALARM CONDITIONS 3
Q)
There are a number of conditions for which the P.T.S. will produce visual and audible alarms. Those conditions, indications and appropriate actions are ~,.
shown in Table 2.1. The appropriate actions indicated are based on the most probable causes and should only be used as a guide. Other causes of alarm
conditions may indicate a need for other actions.
.
Table 2.1: Alarm Conditions
Message Action to silence
Alarm Pressure Time Action required to correct
Condition Description display alarm tone for 30
tone display display alarm condition
(steady) seconds
High The unit has detected PRESSURE Constant Current cuff Normal Momentarily Press Press and hold 'PRESSURE+ ' to
Pressure high pressure in the HIGH pressure, 'PRESSURE+ ', adjust set pressure to within
above set cuff. High pressure is flashing 'PRESSURE- ', 1SmmHg of sensed pressure
point defined as pressure TIME+ ', or
that is + 15mmHg above TIME- '. OR
the pressure set point
continuously for more Press and hold 'DE FLATE' for 2
than 1 second . seconds to deflate the cuff.
Current cuff Normal Momentarily Press Press and hold 'PRESSURE - ' to
Low The unit has detected PRESSURE Constant
pressure, 'PRESSURE+ ', adjust set pressure to within
Pressure low pressure in the cuff. LOW 1SmmHg of sensed pressure
flashing 'PRESSURE - ',
below set Low pressure is defined
TIME+ ', or
point as pressure that is OR
TIME- '.
15mmHg below the
pressure set point Press and hold 'OEFLATE' for 2
continuously for more seconds to deflate the cuff.
than 1 second .
Alarm will stop automatically
whenever the unit can reguiate
the cuff to within 15 mmHg of the
pressure set point.
I
Message Action to silence
Alarm Pressure Time
Condition Description display alarm tone for 30 Action required to correct
tone display display
(steady) seconds alarm condition
Tourniquet The unit's elapsed TIME LIMIT Constant Normal Current Momentarily Press Press and hold T IME + ' to
time timer, which advances tourniquet 'PRESSURE+ ', increase time limit set point,
exceeded when the cuff is time, 'PRESSURE - ',
inflated, has reached flashing TIME+', or
the time limit set point TIME-'. OR
and caused the time
alarm to activate.
Press and hold 'DEF LATF for 2
seconds to deflate the cuff.
Battery The battery voltage has BATT LOW Cycle 1s Normal Normal Momentarily Press Plug in AC power
Low dropped to a level on/ 10s 'PRESSURE+',
indicating that the unit off 'PRESSURE-',
2 segments OR
should be plugged into TIME+', or
(stage 1) darkened in
AC power for continued TIME-'.
battery
use, and to charge the
indicator Press 'ON/ STANDBY' to shut unit I
battery. down.
symbol.
_,.
~
Battery The battery voltage has BATT LOW Constant Normal Normal Momentarily Press Plug in AC power immediatety
Low dropped to a level 'PRESSURE+',
indicating that the unit 1 segment 'PRESSURE-',
OR
may trigger a "BATT darkened in TIME+', or
(stage 2) TIME-'.
ERROR" alarm (see battery
below) and shut down indicator Press 'ON/STANDBY' to shut unit
at any time. symbol. down.
Cuff The unit has detected PRESSURE Constant Current cuff Normal Momentarily Press Momentarily press
pressuri- an approximate pressure, TIME+' or 'PRESSURE + ' or
zed during pressure of 50mmHg or flashing TIME-'. 'PRESSURE - ' button to display
power up more in the cuff when pressure set point
_.. the unit was powered
u, up. The unit assumes a
OR
procedure is in
progress, changes the
pressure set point to Adjust pressure set point by
the detected pressure, pressing and holding
immediately regulates 'PRESSURE + ' or
the cuff to this 'PRESSURE - '.
pressure, and activates
alarm to alert the user
of the condition.
Cuff not The unit has detected CUFF NOT Constant Current cuff Normal Momentarily Press Disconnect hose to cuff to
fully pressure in the cuff DEFLATED pressure, 'PRESSURE+', exhaust pressure.
deflated greater than 1SmmHg, flashing 'PRESSURE - ',
60 seconds after the TIME+', or
cuff has been deflated . TIME-'.
I
...
Message
Condition Description display Alarm Pressure Time Action to silence
(steady) tone display display alarm tone for 30 Action required to col'Teet
seconds alarm condition
Leak in cuff The unit has detected a CUFF LEAK Constant Current cuff Normal
or hose large leak in a cuff Momentarily Press Fix leak
pressure, 'PRESSURE+',
defined as the cuff (may also flashing 'PRESSURE-',
failing to reach the display
'TIME+', or OR
pressure set point in a PRESSURE
reasonable time, or the 'TIME- '.
LOW)
pump running Reduce pressure set point until
excessively while unit can maintain pressure
regulating in 'Inflating'
or 'Regulating' mode OR
and the pressure set
point is not being
adjusted. Press and hold 'DEFLATE' for 2
seconds to deflate the cuff.
Calibration The unit has detected ERROR Constant "CAL" "Err" None Disconnect the hose and press
failure that the calibration in
'ON/STANDBY' twice to shut unit
the pressure transducer down and restart
is invalid. The unit has
shut down in a safe
state*. OR
Internal The unit has detected ERROR Constant "Err" Error code None Press 'ON/STANDBY' twice to
electronic an internal error and shut unit down and restart.
failure has shut down in a
safe state*. When cuff deflation is required,
disconnect hose to cuff.
'
* In the 'safe state' mode, the current pressure in the cuff is held (in the absence of leaks). The 'safe state' mode helps prevent unexpected loss
of occlusion during a procedure if a sudden failure occurs. See Section 2.8.2 for more detail.
a ,
Although i! is vefl( unlikely, internal hardware failures may also cause erratic operation and/or unintelligible
displays with or without alarms, and may or may not put the P.T.S. in the 'safe state' mode.
If either type of error occurs:
a) Set the unit to standby by pressing the 'ON/STANDBY' button. This removes power from the internal
instrument circuitry and all instrument functions, causing the cuff to hold pressure (in the absence of leaks).
b) If required, attempt to restart the unit by pressing the 'ON/STANDBY' again to restart the unit.
c) If required to continue the procedure, clamp the hose with a hemostat to maintain cuff pressure, then
immediately discor,mect the faulty P.T.S. unit and connect a replacement unit.
·,,
d) If cuff deflation i~ required, disconnect the cuff from the P.T.S. unit.
WARNING: J~:~!J
cases of internal hardware failure, erratic operation, or unintelligible displays, it is
possible that" ihe pressure in the cuff is not accurately displayed on the P.T.S. unit and that cuff
pressure may be present even when the P.T.S. unit appears to be shut down. The user must
immediately determine if the cuff is inflated or deflated and continually monitor the cuff to ensure
patient safety. If deflation of the cuff is necessary, disconnect the hose from the P.T.S. unit or from
the cuff and confirm that the cuff deflates completely.
~7
SECTION 3.0: P.T.S. MAINTENANCE INSTRUCTIONS
3.1 GENERAL MAINTENANCE INFORMATION
While th p T s has been designed and manufactured to high industry standards, it is recomm
nd
'od ' ~nspec
. · tlon testing and calibration ('maintenance') be performed as described in this e e~ that
pen ,c · · , , . Th. r e .d section to
ensure continual safe and effective operation. . is sec i~n aIso s . rves as a gu1 e to troubleshootin
iting unscheduled maintenance. The maintenance intervals hsted below are provided as a . 9 ~nd
exped . . f It . t . t QU1dehne·
refer also to the policies in your practice setting or genera ourrnque main enance procedures and intervals. ,
CAUTION: Do not attempt to disassemble_ or open the enclosure of your P.T.S. unit. The P.T.S. n
11
designed to be disassembled and serviced by anyone other than Delfi staff. Disassembl ot
attempted service by anyone other than Delfi staff poses a risk of electric shock, damage to th Y and
and Injury to the patient and will void all warranties. Internal parts in the P.T.S. can only be servt ~nit,
the factory by Delfi staff. Please contact Delfi if you have problems with your P.T.S. unit that c: at
be resolved by following the maintenance and troubleshooting procedures described below. nnot
3.2.1 CLEANING
The exterior of the unit may be cleaned as often as required with a cloth that has been dampened (not
dripping) with a mild detergent The exterior of the cuff hose may be cleaned using a mild detergent solution
or alcohol. Tourniquet cuffs should be cleaned in accordance with the manufacturer's instructions.
CAUTION: Do not attempt to clean or flush out the interior of the hose assembly. Do not allow fluids
or debris to enter the hose connectors on the P.T.S. unit or the hose assembly.
CAUTION: Avoid exposing the AC power supply to liquids. Do not immerse in fluid. Do not allow the
AC power supply to lie on the floor where pooling of liquids may occur. Clean by damp cloth (alcohol
or mild detergent wipe) only. The AC power supply is resistant to occasional splashing or dripping of
fluids, but is not fluid--tight tf immersed in or exposed to excessive amounts of liquids, the AC power
supply may faiJ and may pose an electricaJ shock hazard.
3.2.2 INSPECTION
The unit should be externally inspected as folJows at least once every three months:
• Obvious external damage.
• Missing or illegible labels and warnings.
• Kinks or damage in the power corct
• Secure connection and locking of the power cord plug to the receptacle on the P.T.S. unit
• Secure connection and Jocking of the hose connectors on the P.T.S. unit to the hose assembly.
• Kinks or damage in the hose assembly.
3.3 CAUBRATION
Caibration should be pedormed ever, six months_ or after any unscheduled maintenance.
Caibration a1ows the output SV- from the presstn transtu:er to be ~ againSt a
of the P.T.S.
cabated presstft S0Wte. The dirtbl!UCe between 1he known presswe a,d the pressll'8 ,neasll8d by t,le
18
•
. recorded at each of four set points, and these four calibration factors are used to correct the
transducerst~: pressure transducers during normal operation . The calibration factors are stored in memory.
signal from
UIPMENT REQUIRED:
EO • Calibrated o to 500 f!lmHg pressure gauge.
Ad 'ustable oto 500 mmHg pressure source.
: Su~table pneumatic hoses and connectors.
UTION: The following steps must be taken In the exact order to calibrate the unit. Failure to do 80
CA ult In incorrect pressure readings while the unit Is in operation.
may res
a) To enter the calibration r:node, pre~s and hold the 'INf'.'LA~E' an? 'DEFLATE' buttons simultaneously, then
press 'ON/STANDBY'd~o slw1t:~}~et~mt on . Wh~~ t~e umtdfimlsthes i~s s~art-uf sequence the unit will enter the
calibration mode and isp ay m e_p~essure isp ay an ~ ernatmg CAL and "O" in the time display. This
. to indicate to the user that the umt 1s now ready to calibrate. The time display indicates the expected
~:ference pressure to be applied to the cuff. The pressure display indicates the sensed pressure.
b) With the hose connector on _the P.:.S. unit op~n ~o at~osphere, press the 'PRESSURE + • button to
indicate zero reference pressure Is applied. The umt will adjust the transducer output corresponding to zero
pressure. The unit will then sound a tone to indicate that the reference pressure was taken.
c) once the zero point is calibrated, press the 'TIME+ ' button to advance the unit to the next pressure level.
The time display will now be alternating between "CAL" and "50". Connect the P.T.S. to the reference
pressure source and gauge. Apply a calibrated reference pressure of 50 +/- 1 mmHg to the cuff port. Once
the pressure has stabilized, press the 'PRESSURE + ' button to indicate the reference pressure is applied.
The unit calibrates the 50mmHg point and sounds a tone when sucpessfully completed. If the reference
pressure is more than 15 mmHg different from the sensed pressure, the unit shows "CAL" and "Err" messages
in the pressure .and time displays respectively and halts the calibration routine. If this happens, power off the
unit and start over.
d) Repeat the preceding step for reference pressures of 250 +/- 1 mmHg and 475 +/- 1 mmHg. The unit will
not allow calibration points to be skipped.
e) Remove the reference pressure source and press the 'TIME- ' button three times to cycle the reference
pressure back to zero. The time display will alternate between "CAL" and "O". Press the 'INFLATE' and
'DEFLATE' buttons simultaneously to exit the calibration . The unit saves the new calibration values in
memory and displays "CAL" in both displays. Power off the unit by pressing the 'ON/STANDBY' button.
19
n uthe new tot point 18 lcaa than 400 mmHg, thore la a algnlf,cant leak In either In the tett cuff and hvse or ,n
the P.T s. lntornol pnoumotlca. Ropoat tho toat with a variety of different cuffe and hose as&embhct. If the
unit contlnuoa to fall, the leak lo Internal and the P.T S. unit mu8t be Bervlced by Delfi.
20
1-ifVO
I~
. b ttery compartment surface. The connector cannot be installed the wrong way; if it will not slide
with_th e thae way down, rotate it one-half turn and try again.
easily aII
_ TROUBLESHOOTING ~UIDE
36
lists a number of possible malfunctions and their most likely causes. While it is not possible to list
31
Tab 1e ·onceivable malfunction and possible causes, the table will help the user solve the most common
every c
problems.
Table 3.1: Troubleshooting
Cuff does not deflate. 'DEFLATE' button not Press and hold the 'DEFLATE' button for at least 2
held for 2 seconds. seconds.
Green AC indicator light AC power supply Ensure wall socket is working, of the correct
does not illuminate when assembly not plugged voltage, and that the plug is all the way in.
unit is plugged in to AC in to suitable wall
power. outlet.
21
L
MALFUNCTION POSSIBLE CAUSES CORRECTIVE ACTIONS
Red Alarm Indicator light Alarm Indicator llght Confirm that alarm Indicator light illuminate,- - -
does not illuminate during not working. during self-check upon power-up. Contact Delfi.
alarm conditions.
No cuff pressure and/or Faulty pressure/time Confirm that all segments of the display illumi-
tourniquet time reading. display. during self-check upon power-up. nate
BATT ERROR alarm. Fully discharged battery Connect to AC power and allow battery to
recharge for 8 hours.
Faulty or dead battery Replace battery pack (see Section 3.5 above).
pack.
Battery does not charge Faulty or dead battery Replace battery pack (see Section 3.5 above).
when unit is plugged in to pack.
AC power.
Unit does not turn off Pressure in cuff Deflate cuff and ensure it deflates fully
(cannot be set to standby) (disconnect hose if required) to clear "CUFF NOT
DEF LATED" alarm.
22
3.7 INTERNAL HARDWARE SERVICING
The P.T.S. is designed with self-test and self-monitoring features to warn .
O
it is very unlikely, modes of failure may also occur that cause t' f fa!lures (see Table 2.1 ). Although
may or may not trigger alarms. I~ the maintenance, calibratio~.rr=~~ ~ioeration an~/or illegible displays and
above do not restore normal operation, contact Delfi for service advice. ubleshooting procedures described
-
_J_
Alternating current.
Earth ground.
Direct current.
@. cTUVus: Medical Equipment with respect to electrical shock, fire and mechanical
hazards only in accordance with UL 2601 (AC power supply only).
& ATTENTION:
/,\. ATTENTION: FOR USE BY TRAINED
ill PERSONNEL ONLY.
DANGER: EXPLOSION HAZARD. DO
~~~TigR~~&f
DO NOT IMMERSE.
THIS UNIT MUST BE CHARGED AT LEAST 8
HOURS BEFORE INITIAL USE, CALIBRATION,
NOT USE IN THE PRESENCE OF FLAM- OR FUNCTIONAL CHECK. REFER 10
MABLE ANESTHETICS OR GASES. OPERATOR'S fv1ANUAL FOR CHARGING
1 CAUTION: RISK OF ELECTRIC SHOCK.
DO NOT DISASSEMBLE. TO BE SERVI-
SERIAL #XXXXX
INSTRUCTIONS. REMOVE AND DISCARD
THIS LABEL WHEN COMPLETE.
CED BY QUALIFIED PERSONNEL ONL
USE DELFI BATTERY AND POWER SUP-
PLY ONLY.
*Delfi
P.T.S. Portable Tourniquet System
DANGER : FIRE HAZARD. USE
& GENUINE DELFI REPLACEMENT
BATTERY ONLY.
*Delfi
Battery #4-2100-017
Only for use with Delfi P.T.S. Portable
9-2100-001 SERIAL #XXXXX Tourniquet Systems (#9-2100 series).
U.S. PATENTS: 6,213,939 ; 5,931,853 ; 5,649,954
4,469 ,099 ; PAT. PEND . and FOREIGN PATENTS
*Delfi
Delfi Medical Innovations Inc .
Store at room temperature or cooler.
Do not Incinerate.
Recycle/dispose at proper facility,
Delfl Medical Innovations Inc. Vancouver BC Canada Manufac tured In .- - ~ f o r
Vancouver BC Canada Delft Medical Innovations Inc.
800-933-3022
800-933-3022 Vancouver BC Canada 800-933-3022
MADE IN CANADA www.delflmedlcal.com www.delfimedlcal.com
23