Overview
The following is a short summary of the tour:
1. Preparing for a measurement
Turning ON the 4395A
Connecting the DUT
2. Setting up the 4395A
Selecting the analyzer type
Setting the active channel
Setting the input port
Setting the frequency range
Performing the automatic scaling
3. Making a calibration
4. Reading a measurement result
Reading a measured value by using marker
5. Printing out the measurement result
Conguring and connecting a printer
Making a hardcopy of the display
After you nish this tour, you will understand how to make a basic measurement in the
network analyzer mode of operation.
The 4395A performs a power on self-test. About 20 seconds later, the model name, revision
number, and other information should appear on the LCD to indicate that the 4395A has
normally started up.
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
In the MEASUREMENT block, press 4Meas5. Press ANALYZER TYPE .
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
In the ACTIVE CHANNEL block, press Verify the Chan 1 active channel indicator
4Chan 1 5. lights.
Note Changing the analyzer type presets the 4395A for the active channel. If you
want to keep the current measurement settings when changing the analyzer
type, rst set the other channel to active.
FFFFFFFFF FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
FFFFFFFFFFFFFFFFFFFFFFFFFF
In the MEASUREMENT block, Press Press AUTO SCALE .
4Scale Ref5.
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
FFFFFFFFFFFFFFFFFFFFF
Press RESPONSE . Disconnect the DUT then, connect the THRU.
FFFFFFFFFFFF
Press THRU . WAIT - MEASURING CAL STANDARD is
displayed.
FFFFFFFFFFFF FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
The THRU softkey label is underlined when Press DONE: RESPONSE .
the measurement is completed.
\Cor" is displayed on the left side of the display to show that the frequency response error is
corrected.
The measured value is now corrected for the frequency response error.
Note If the trace is changed, it requires an adjustment of the scale. Perform the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
In the MARKER block, press 4Marker5. Verify a marker appears on the trace.
Turn the knob to the right to move the Read the values at the right top of the display.
marker toward the right.
The marker has a search function that makes it easier and faster to evaluate the trace results.
For example, to search for the maximum value and its frequency on the trace:
FFFFFFFFF
The marker immediately moves to the Read the frequency and response values
maximum point on the displayed trace. displayed at the upper right of the display.
Caution Do not connect a printer to \TEST SET - I/O INTERCONNECT". Doing so could
damage the printer.
Note For more information about printer, see the chapter 12.
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
Overview
The following is a short summary of the tour:
1. Preparing for a measurement
Turning ON the 4395A
Connecting the test signal source
2. Setting up the 4395A
Selecting the analyzer type
Setting the active channel
Selecting the input
Setting the frequency range
3. Making a Measurement
Reading the peak level using the marker
Setting the resolution bandwidth to see low level signals
Searching for harmonics using the search function
4. Saving and recalling the 4395A settings
Preparing the disk
Saving 4395A settings
Entering the le name
Recalling the 4395A settings
After you nish this tour, you will understand how to make a basic measurement in the
spectrum analyzer mode of operation.
The 4395A performs a power on self-test. About 20 seconds later, the model name, revision
number, and other information should appear on the LCD to indicate that the 4395A has
normally started up.
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
Press SPECTRUM ANALYZER .
Note Changing the analyzer type presets the 4395A for the active channel. If you
want to keep the current measurement settings when changing the analyzer
type, rst set the other channel to active.
In the ACTIVE CHANNEL block, press Verify the Chan 2 active channel indicator
4Chan 25. lights.
Note All selected settings are stored separately for each channel. You must select an
active channel (1 or 2) before you can change the measurement setup for that
channel.
FFFF FFFFFFFFFFFFFFFFFFFFFFFFFFFFF
FFFFFFFFFFFFFFFFFFFFFFFFFFFFF
Press SEARCH:PEAK . Marker moves to the top of the CAL OUT
signal.
The trace moves upward to place the tip of the maximum peak at the top line of the grid.
Now, with the noise
oor level lowered by narrowing the resolution bandwidth, the second and
third harmonics can be seen as shown below:
FFFFFFFFFFFFFFFFFFFFFFFFFFFFF
FFFFFFFFFFFFFFFFFFFFFFFF
Press NEXT PEAK . The marker moves to the third (or second)
harmonic.
Verify the disk is not write protected. Insert the disk into the disk drive
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
Toggle STOR DEV [MEMORY] to [DISK] , and Toggle FORMAT [DOS] to [LIF] .
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
Press INIT DISK: YES . The message, \INITIALIZE DISK In Progress,"
is displayed. After the disk is initialized, this
message is turned o.
Note The 4395A can use either a LIF (Logical Interchange Format) or a MS-DOS (Disk
Operating System) format disk.
Note The 4395A can initialize a 1.44 MB 3.5 inch exible disk only.
FFFFFFFFFFFFFF
Press 4Save5. Press STATE .
The 4395A requests the le name you want to use for the saved settings.
Entering the File Name
Note If a keyboard is connected, you can use it for le name entry. If not, use the
front-panel controls as described in the following steps.
FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF
Turn the rotary knob to move the arrow Press SELECT LETTER .
below the rst character, S.
Keep entering characters until SATOUR is If you enter a wrong character, press
FFFFFFFFFFFFFFFFFFFFFFFFFF
entered. BACK SPACE to erase the character.
FFFFFFFFFFFF
To complete the le name entry, press DONE . Verify the disk access indicator lights (this
shows that the 4395A is saving the settings to
the disk).
Note The le name for a LIF format can be up to 10 characters long. However,
with the 4395A, the last 2 characters are reserved for a sux. Therefore,
you can enter a le name of up to 8 characters. Either upper or lower case is
recognized in the LIF format.
A le name for a MS-DOS(DOS) format consists of a le name and an extension.
The le name can be up to 8 characters long and the extension contains up to 3
characters. A period separates the extension from the le name. The extension
Press 4Preset5.
The 4395A is set to the preset conditions. However, the 4395A settings from the previous
examples are stored in the SATOUR le on the disk.
Note Sux, \_S," means the 4395A settings are saved. If you save the 4395A
settings in a DOS format, an extension, \.sta," is appended to the le name.
After the disk access lamp goes out, all 4395A settings that you set are recalled. You can verify
them on the display. If you want to know what settings are saved, see chapter 8.
Note Your 4395A must be equipped with option 010 to serve as an impedance
analyzer. Otherwise, impedance analyzer mode is not available.
The 4395A performs a power on self-test. About 20 seconds later, the model name, revision
number, and other information should appear on the LCD to indicate that the 4395A has
normally started up.
Available analyzer modes are listed in the menu under the ANALYZER TYPE softkey. You can
use the 4395A as an impedance analyzer by selecting the corresponding analyzer mode from
that menu. Follow these steps:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press the 4Chan 15 key in the Make sure that the indicator lamp
ACTIVE CHANNEL block. beside the 4Chan 15 key is ON.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNN
To specify the frequency at which Press the 4k/m5 key to indicate that
to start the sweep, enter 415 405 405 the unit is kHz.
using the numeric keys.
NNNNNNNNNNNNNNNNN
Enter 405 4.5 455 using the numeric Press the 4x15 key.
keys.
NNNNNNNNNNNNNNNNN
Enter 435 405 405 using the numeric Press the 4215 key.
key.
Note A smaller IF bandwidth reduces trace noise, but increases measuring time.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Enter 485 using the numeric key. Press the 4215 key.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ON off .
Note When you perform impedance measurement with the 43961A, you must set IF
bandwidth equal to or less than 300 Hz and averaging factor equal to or greater
than 8.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
With the calibration menu displayed on the LCD screen, calibrate the 4395A for the OPEN,
SHORT, and LOAD states in order.
Note Be sure to use the calibration kit included in the 43961A package.
OPEN Calibration
Follow these steps:
NNNNNNNNNNNNNN
SHORT Calibration
Follow these steps:
NNNNNNNNNNNNNNNNN
Connect the 0
termination to the Press SHORT .
43961A's OUTPUT port. NNNNNNNNNNNNNNNNN
Remove the 0
termination.
LOAD Calibration
Follow these steps:
NNNNNNNNNNNNNN
Connect the 50
termination to Press LOAD .
the 43961A's OUTPUT port. NNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Remove the 50
termination.
NNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNN
Wait until the SHORT softkey's the circuit into the OPEN state.
label is underlined to indicate that
the SHORT compensation is
complete.
NNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
These settings are applied to Channel 1, which has been the active channel in the scenario of
this tour. Note that, in Steps 6 and 7, you will switch from Channel 1 to Channel 2, and assign
dierent settings to Channel 2.
Connecting the DUT
How to connect a DUT diers depending on which test xture you use. For more information,
refer to the documentation that comes with your test xture.
As soon as you have connected the DUT, the 4395A will measure and display the impedance
characteristics.
Performing the Automatic Scaling
Often, the trace obtained after specifying the frequency range is too large or too small
vertically for the grid. However, by using the automatic scaling function, you can obtain the
optimum vertical setting automatically.
Follow these steps:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Press the 4Chan 25 key in the Make sure that the indicator lamp
ACTIVE CHANNEL block. beside the 4Chan 25 key is ON.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Enter 485 using the numeric key. Press the 4215 key.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ON off .
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Now both Channels 1 and 2 are assigned specic settings. You can not only have one of the
two channels displayed at a time, but also have both channels displayed in parallel, as you will
learn in the next step.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
The screen is split into upper and lower halves. The upper half shows the absolute impedance
value while the lower half shows the phase.
NNNNNNNNNNNNNN NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN