Professional Documents
Culture Documents
ReleaseNotes V22.16 066064106
ReleaseNotes V22.16 066064106
Version: [22.16]
Last revised: [Nov 23, 2011]
Product: TDS1000B and TDS2000B
Introduction
-----------------------
Version v22.16 is a firmware update for the TDS1000B and TDS2000B
Series oscilloscope.
Installation Guide
-----------------------
This firmware update is intended to update TDS1000B and TDS2000B
Series oscilloscopes to version v22.16. Only install this firmware
update if your oscilloscope is running a firmware version older than v22.16 (such
as v22.15).
Please Note: The flash drive used to update the firmware must be 2 GB or smaller
for version lower than v22.11.
History
-----------------------
V22.16 Nov 23, 2011
In version 22.16, the firmware addresses the following issues:
-Support the 200X attenuation probe(CHn->probe->Attenuation)
Description
-----------------------
Version V22.13 only for internal test process improve, did not
add new features compare to v22.01.
************
Users who would like to disable Autoset and Autorange, read the
following:
************
Users who program their own custom application software to
communicate with the TDS2000B or TDS1000B should read the following:
(1) Be familiar with the programmers guide Status and Events section.
(4) Connect the TDS1000B / TDS2000B to a USB 2.0 port for reliable
communication.
(5) Programs should check for a timeout error after each read.
Why?: In rare circumstances, the read following a query may time out.
When a timeout error is observed, it is necessary to re-send the
query and re-read the query response.
*********************************************************************
The following only apply if using a TEK-USB-488 adapter
*********************************************************************
(6) When instrument is communicating via a TEK-USB-488 GPIB to USB
adapter, *WAI command, the BUSY? query, and the *OPC? query are
preferred methods of achieving synchronization. See the programmer
manual for examples of how to use these methods.
(8) Refer to the firmware release notes for the TEK-USB-488 adapter,
for programming recommendations generic to the adapter.