¢ Report No: AC - 4022-34187 Check Source Reading [eaeeeceee Rate Calibration [swe [Pica [ar [ ee [om [emo | ae | ere] 20Ci Cs-137 | 912.45 (O10 500 | uR/hr 185.5 3.06 - Pass 20 Ci Cs-137 O10 500 2.44 - Pass 20Ci Cs-137 0.61 - Pass 20 Ci Cs-137 0.42 - Pass 20 Ci Cs-137 1.83 - Pass 20 Gi Cs-137 0.75 - Pass 2000 Ci Cs-137 -1.28 - Pass 2000 Ci Cs-137 “3.56 - Pass 2000 Ci Cs-137 -169 - Pass, Dose Calibration ae [norcrertar oss ows] me | 0 | ss | 2e0-Pas | Calibration Procedure: CAL-450-451 paf Calibration Description: The 451P-RYR has an operating range of 0 to 5 Rihr. The unit is exposed through the side of the detector and calibrated on all ranges. Allrreadings were corrected for background. The % Error was calculated using Equation 1 Environmental Constraints: The 451P-RYR survey meter is designed to read accurately from -20 to 50C. The unit is pressurized, therefore, requires no air density corrections. Calibration Uncertainty: 3.6% with 2 4% associated with the uncertainty of the source Accuracy Requirement: 10% of Reading 100° (UUT — Re ference) Equation 1 me, ee Re ference Fluke Biomedical sponge 2 of 3 sei ‘s4oz4e 200 4403802907 wn auksbiemedia om