Report No: AC - 4022-34187
Check Source Reading [eaeeeceee
Rate Calibration
[swe [Pica [ar [ ee [om [emo | ae | ere]
20Ci Cs-137 | 912.45 (O10 500 | uR/hr 185.5 3.06 - Pass
20 Ci Cs-137 O10 500 2.44 - Pass
20Ci Cs-137 0.61 - Pass
20 Ci Cs-137 0.42 - Pass
20 Ci Cs-137 1.83 - Pass
20 Gi Cs-137 0.75 - Pass
2000 Ci Cs-137 -1.28 - Pass
2000 Ci Cs-137 “3.56 - Pass
2000 Ci Cs-137 -169 - Pass,
Dose Calibration
ae
[norcrertar oss ows] me | 0 | ss | 2e0-Pas |
Calibration Procedure: CAL-450-451 paf
Calibration Description: The 451P-RYR has an operating range of 0 to 5 Rihr. The unit is exposed
through the side of the detector and calibrated on all ranges. Allrreadings
were corrected for background. The % Error was calculated using
Equation 1
Environmental Constraints: The 451P-RYR survey meter is designed to read accurately from -20 to
50C. The unit is pressurized, therefore, requires no air density
corrections.
Calibration Uncertainty: 3.6% with 2 4% associated with the uncertainty of the source
Accuracy Requirement: 10% of Reading
100° (UUT — Re ference)
Equation 1 me, ee
Re ference
Fluke Biomedical sponge 2 of 3 sei
‘s4oz4e 200 4403802907 wn auksbiemedia om