GSI AUDIOSTAR PRO ™

CLINICAL TWO-CHANNEL AUDIOMETER

Setting The Clinical Standard


GSI AUDIOSTAR PRO ™

CLINICAL TWO-CHANNEL AUDIOMETER

Tradition of Excellence
The GSI AudioStar Pro™ continues
the tradition of excellence in clinical
audiometry by maintaining the
Grason-Stadler legacy of fast, Stand-Alone/
efficient, and familiar navigation. PC Enabled
Audiologists appreciate the flexibility
of a stand-alone audiometer that offers
GSI Suite
seamless data transfer to a computer. GSI Suite captures, saves and
In the event of a network failure or shares patient information improving
computer lock-up, you will not lose reporting to support the needs of the
patient data or the ability to test. The contemporary clinic. GSI Suite is
stand-alone configuration is optimized available as a Noah module for
with direct connection to a wireless seamless integration to hearing aid
keyboard and mouse making it fast and fittings.
easy to enter patient demographics,
report comments, and expedite test
administration. In addition, direct
connection to a printer and the inte-
grated print button make it possible to
print a complete report for immediate
review with the patient or physician.

The one-button, one-function


front panel of the AudioStar Pro is EMR/EHR Ready
recognized worldwide as the Gold
User login and password controls
Standard of user-friendly design,
provide security for patient data in
GSI Suite Features
allowing audiologists to test with
confidence. AudioStar Pro has every compliance with HIPAA. Complete • Powerful custom report designer
desired feature such as: an extra large audiometric results can be transferred • Select from a variety of provided
display that reduces eye strain, an to software such as report templates
ergonomic housing that maximizes GSI Suite and
hand and wrist comfort, and light Noah® 4, or EMR/EHR • Test results in EMR/EHR compatible
formats PDF, JPEG, XPS, TIFF
pipes around selected test buttons integrated
with your
Compatible • Retrieve test results for review
allowing concentrated focus on the
patient. facility’s • Print reports from PC compliant
EMR/EHR system. printers
• Network results from multiple
instruments by combining GSI
Suite with OtoAccess™ or Noah 4

Setting The Clinical Standard


Technical Features
Panel Navigation
Gold standard ergonomic design with
familiar one-button, one-function front
Custom Configuration panel navigation.
Configurable tests and preferences to improve workflow
and optimize speed of testing.

Special Tests
Pre-configured special tests such as QuickSIN, BKB-SIN,
TEN (HL), ABLB, SISI and Tone Decay addresses the
research trends in hearing evaluation.

Assistant
Monitor
Allows direct
communication
between operator
and assistant
eliminating
the need for
an external
intercom system.

Word Lists
Integrated internal digital word lists provide effortless
access to commonly used word lists providing consistent
and reliable recorded speech testing.

Two-Channel Control
Audiologist can mix signals and route them
to either one or both ears making hearing
evaluations easier than ever.
Direct Print Data Transfer
Expedite the data entry and test administration. Electronic data is transferred with a single button press.
Use the external keyboard and mouse to enter
patient demographics, session comments and
print a complete report to a connected printer.

VRA Controls
The built-in VRA controls allow for fast and simple activation
of VRA systems.

On-Board Navigation
Access integrated word lists
and menu selections using
the front panel navigation.
audiostarpro ™

Fast Efficient
Features such as speech auto play, immediate connection to both the Additional features integrated into
auto advance and auto frequency display and intuitive control panel while the AudioStar Pro makes testing
selection allow the examiner to being surrounded by new innovative every patient even more seamless
improve productivity and patient features. This revolutionary audiometer by configuring pure tone and speech
flow. Standard features, such as the maintains maximum flexibility in a preferences. The built-in auxiliary
headset microphone, allows the two-channel audiometer allowing full intercom, active microphone during
operator to communicate directly control in testing procedures for every tone presentation as well as remote
with patients while the built-in patient population. testing capabilities provide a multitude
monitor speaker makes it easy for of options for testing difficult to test
third parties to participate in the
hearing evaluations. Sound field
Customizable patients. The integrated intelligent
functionality of activating sound field
options are available from the Test Type Buttons allow access to speakers during hearing aid or
standard 90 dB HL output to the protocols that are customized to cochlear implant testing improves
high performance speakers with facility preferences. The AudioStar both efficiency and accuracy.
96 dB HL - both without any external Pro uses a Configuration Application
software to customize preferences
amplification needed. The maximum
output of 102 dB HL is available with for efficient and consistent testing. Compatible
an external amplifier - achieving the Auto Play, Auto Score, and Word The AudioStar Pro addresses the
highest output in the industry. Four List "Favorites" ensure seamless challenges of data management
speaker calibration is available for integration of the 50+ commonly such as patient security and HIPAA
advanced sound field testing. One used word lists included with every compliance through our electronic
button, one function incorporated with AudioStar Pro. The integrated Special record solutions: (E-Record Solutions).
a large display maintains the GSI Tests including Quick-SIN, BKB-SIN These solutions safely and easily
legacy of fast and efficient navigation. and TEN (HL) address research integrate audiometric data into every
trends in hearing evaluation. The facility’s data management system
AudioStar Pro combined with from VA, ENT, Private Practice, Hearing
Familiar GSI Suite, brings a new level of Aid, Cochlear Implant and Pediatric
The AudioStar Pro maintains the user customized reporting options that facilities.
friendly interface that has defined GSI addresses the requirements of both
since 1949. Audiologists will feel an the facility and referral sources.
GSI AudioStar Pro ™ PC Enabled/Stand Alone
• Transfer data to connected PC with an E-Record
Solution Software
• Print complete report directly to a compatible
CLINICAL TWO-CHANNEL AUDIOMETER USB printer

Special Tests (user defined)


• MLB
Product • Pulsed/FM: Pulsed and Modulated
• Pediatric Noise • Lombard test

Specifications • Pediatric Noise Pulsed • Pure Tone Stenger


• Speech Stenger
Speech – Channels 1 and 2
Dimensions and Weight • Microphone: For live voice testing and
• SAL
• W x D x H: 20.1 inches x 14.6 inches x • Doerfler - Stewart Test
communications
13.2 inches (LCD raised)
• INT/EXT A & INT/EXT B: Can be utilized for Communications and Monitoring
51.0 cm x 37.0 cm x 33.5 cm internal wave files or recorded speech material • Talk Forward: Permits the tester to speak
• Height with LCD lowered – 5.5 inches (14.0 cm) from an external digital device through the examiner microphone into the
• Weight: 17 lb (7.7 kg) Intensity Range: selected transducer.
• Shipping Weight: 27 lb (12.25 kg) • Air Conduction: -10 dB HL to 100 dB HL for • Talk Back: Allows the examiner to listen to
• Power Consumption: 90 Watts TDH 50P (Linear Type 4) comments from the patient in the testing booth.
• Bone Conduction • Monitor: The monitor headset or monitor speaker
Channels  Mastoid: -10 dB HL to 55 dB HL built into the instrument housing may be used
• Two Independent Channels by the examiner to listen to Channel 1, Channel
 Forehead: -10 dB HL to 35 dB HL
2, Aux intercom, and/or Talk Back signals.
Pure Tone – Channels 1 and 2 • Sound Field:* -10 dB HL to 80 dB HL
• Aux Intercom: The built-in Auxiliary Intercom and
Frequency Range: • Paired Inserts:* -10 dB HL to 95 dB HL assistant headset allows the examiner to speak
• Air Conduction: 125 Hz to 12,000*** Hz Masking Intensity Range: directly to an Assistant and allows the assistant
• High Frequency:* 8,000 Hz to 20,000 Hz • Speech Noise: to hear what is being presented to the patient.
(8 kHz, 9 kHz, 10 kHz, 11.2 kHz, 12.5 kHz,  Air Conduction: -10 dB HL to 95 dB HL (TDH 50P) • On-Board VRA Control: The built-in VRA controls
14 kHz, 16 kHz, 18 kHz*** and 20 kHz***) facilitate fast and simple activation of VRA systems.
 Bone Conduction:
• Full Frequency Range:* 125 Hz to 20,000 Hz -10 dB HL to 50 dB HL (mastoid) Optional Accessories
• Bone Conduction: 250 Hz to 8,000 Hz -10 dB HL to 35 dB HL (forehead) • Wireless keyboard and mouse
• Sound Field:* 125 Hz to 8,000 Hz  Sound Field: -10 dB HL to 80 dB HL • Gooseneck microphone
• Paired Inserts:* 125 Hz to 8,000 Hz • White Noise:
• Frequency Accuracy: ±1%  Air Conduction: -10 dB HL to 95 dB HL (TDH 50P)
Environmental Requirements
• Total Harmonic Distortion: • Temperature: +15ºC to 40ºC (59 to 104ºF)
 Bone Conduction:
 < 2% (earphones and paired insert phones*) -10 dB HL to 50 dB HL (mastoid) • Relative Humidity: 5% to 90% (non-condensing)
 < 5% (bone vibrator) -10 dB HL to 35 dB HL (forehead) • Ambient Pressure Range: 98 kPa to 104 kPa
Intensity Range:**  Sound Field: -10 dB HL to 80 dB HL • Background Sound Level: <35 dB(A)
• Air Conduction (TDH 50P): -10 dB HL to 120 dB HL • Frequency of Use: Once a year to multiple
Special Tests times per day
• High Frequency:* -20 dB HL to 100 dB HL
• ABLB or Fowler: Tone alternating between • Storage Temperature: -20ºC to + 60ºC (-4ºF to 140ºF)
(with Sennheiser HDA 200 Phones)
Channel 1 and Channel 2: Channel 1 is 400
• Bone Conduction msec ON, 400 msec OFF followed by Channel Quality System
 Mastoid: -10 dB HL to 75 dB HL 2, 400 msec ON, 400 msec OFF. • Manufactured, designed, developed and marketed
 Forehead: -10 dB HL to 65 dB HL • SISI: An intensity increment is added to a tone under ISO 13485 certified quality systems
• Sound Field:* in the selected channel for 200 msec, every 5
seconds. The HL increments are in 1, 2 or 5 dB. Compliance/Regulatory Standards
 -10 dB HL to 90 dB HL (basic speakers)
• High Frequency:* Pure tone testing in the Designed, tested and manufactured to meet
 -10 dB HL to 96 dB HL (high performance the following domestic (USA), Canadian,
frequency range of 8,000 Hz to 20,000 Hz using
speakers) European and International Standards:
circum-aural headphones.
 -10 dB HL to 102 dB HL (high performance • ANSI S3.6, ANSI S3.43, IEC 60645-1, IEC
• TEN: TEN masking noise will be presented to
speakers and external booster amplifier) 60645-2, ISO 389
the test ear. Pure tone stimuli between 500 Hz
• Paired Inserts:* -10 dB HL to 110 dB HL and 4000 Hz may be used at 1, 2, or 5 dB • UL 60601-1 American Standards for Medical
• Masking Intensity Range increments to obtain TEN thresholds. Electrical Equipment
(Calibrated in effective masking) • QuickSIN: Six (6) sentences with five (5) key • IEC/EN 60601-1 International Standards for
 Narrow Band Noise: words per sentence are presented in four-talker Medical Electrical Equipment
Maximum dB HL is 15 dB below tone babble noise. The sentences are presented at • CSA C22.2 # 601-1-M90
pre-recorded signal-to-noise ratios. The SNR’s
 White Noise: used are 25, 20, 15, 10, 5, and 0. • Medical Device Directive (MDD) to comply with
Maximum dB HL is 30 dB below tone “EC Directive” 93/42/EEC
• BKB-SIN: Contains 18 List Paris. The sentences
Signal Format: are presented at prerecorded signal-to-noise
• Steady: Tone continuously present. ratios that decrease in 3-dB steps. Each list in
* Optional configuration
the pair is individually scored, and the results of
501C-1 Rev. D

• Pulsed: Tone pulsed 200 msec ON, 200 msec OFF ** The maximum HL values are applicable to
the two lists are averages to obtain the List Pair
• FM: Modulation Rate: 5 Hz score. Results are compared to normative data the middle frequencies only
Modulation Depth: +/- 5% to obtain the SNR Loss. *** Optional configuration

www.grason-stadler.com