CALIBRATION PROCEDURE
FOR
2128 PATIENT SIMULATOR
weeeeenee06/04/1997 29:80 7574730655-13 SO erence PEPRPerEE Peet
SPECIFICATION
2.10
MODEL 2128 TEST PROCEDURE
INTRODUCTION
SCOPE, THIS PROCEDURE COVERS THE FINAL CALIBRATION OF
THE 2120.
IN THIS PROCEDURE THE SECTIONS ARE AS FOLLOWS:
SECTION 2 - POWER SUPPLIES AND REFERENCES
SECTION 3 = OSCILLATOR FREQUENCIES
SECTION 4 — ECG COMPONENT WAVES, OTHER WAVEFORMS
SECTION 5 - LOW LEVEL ECG OUTPUT
SECTION 6 ~ RESPIRATION
SECTION 7 — RLOOD PRESSURE
EQUIPMENT NEEDED:
DIGITAL VOLTMETER = 4 1/2 DIGIT, 10vV RESOLUTION.
FREQUENCY COUNTER - 6 DIGIT, PERIOD MEASURING WITH AT
LEAST 100vS RESULUTION, +01 ACCURACY.
osciLtoscore - oc -~ MHZ, exT noRrzoNTAL, Z00pV/DIV
SENSITIVITY, DIFFERENTIAL INPUTS.
SPECIAL EOUIPMENT ~ MODEL TR21 TEST SYSTEM, ExTENDER =
CABLE FOR 212 POWER PLUB. .
REFERENCE ORAWINGS:
SCHEMATIC ~ 1026218
UPPER PCB ASSEMBLY - 1026258
LOWER PCA ASSEMBLY - 1026264
NOTE +
BEFORE GEGINNING CALIBRATION ALL POTS SHOULD BE SET
FO MID-RANGE ON UNIT.
POWER SUPPLY CALIBRATION *
2126 TOP PANELs*'RESPIRATION'', **sLOOD PRESSURE'' AND
‘FLV UTES POWER-SWITCRES~ALL—!'ON't.
“s-Leap gca'*
SWITCH IN ‘*NORM"*=ROSITION , RESPIRATION ' ‘RATE @PM'S IN
sort, Seusmwiry Su ¥ Gn)
Steg!" POSITION. "TEGO RATE BPHT' IN *Peete PPT eee eer ree
2.30
3-15
212e FRONT PANEL) POWER OFF¢ ALL KNOBS FULLY COUNTER”
CLOCKWISE.
WITH UNIT OPEN ON BENCH, CONNEGT POWER EXTENDER BETWEEN
JS AND PS ON THE LOWER BOARD.
CONNECT BATTERY.
press ‘BATTERY TEST!" POSITION OF FRONT PANEL OCKER
wire. YELLOW LED SHOULD GLOW STEADILY.
PLUG BATTERY ELININATOR INTO JACK ON REAR PANEL -
PRESS FRONT PANEL ROCKER SWITCH TO ‘TON POWER'’ POSITION.
YeLLOw LED SHOULD BEGIN BLINKING.
ye ent Tea 10 = .20N
CONNECT DVM BETWEEN TPE AND TRIO% p= 102 ~3-D%0—3,109
3.072 : P- 9 +3.100
ADJUST R75 TO READ €r26s-VOLTS * AMV.
PER ene
MOVE :OVM LEAD -ON.TR10—T0-2RHt p-—READ“3rOO—TOB+2O-YOUTS =
Te
MOVE OVM LEAD ON AP9 TO TRS. TPIl Liscomven: scobhe.
RESPIRATION CALIBRATION
R21 SHOULD BE CALIBRATED IN THE SuV PRESSURE MODE.
THEN SWITCH TO '*RESPIRATION''.
ON UNIT TOP PANEL, "RESPIRATION LEAD TO RL" AT
ON UNIT FRONT PANEL ''RESPIRATION SR@ Stat 1t3'"
Seco COMPLEX'T AT TTRESPIT, +
CONNECT THE TR21 RESPIRATION LEADS TO THE UNIT ECG
SNAPS ON THE TOP PANELs GLACK COR GREEN) CLIP LEAD TO
srALtt, REO COR YELLOW) CLIP LEAD TO TTLL'T.
THE EXTERNAL DVM SHOULD READ LESS THAN 4.0 VOLTS. tf KT
READS HORE THAN 4.0 VOLTS. ECG 10009 15 >2x ERROR.
2.0%
ADJUST RS3 FOR 3.00 ON THE TR21 DISPLAY.
SWITCH UNIT dRQ t* TO a ADJUST R58 FOR A 1.00
READING.
SWITCH UNIT ‘''aRa *' TO ‘'O.5'", ADJUST R59 FOR A x050
READING.
SWITCH UNIT “"TaRa '* TO ''0.2'", ADJUST R60 FOR A 0.20
READING.86/04/1997 89:88 7574730505-13 TRANSCAT/ETL He
Peper eee
6.18 SWITCH UNIT TARAS TO TTO.tT Ts ApsustERsaPrOR A 0.10
READING.
ecz0 SwITcH UNIT ‘TARGIT TO TTAPNEA'', OBSERVE 0,00 50 os
rcaoinc. CHECK Aet AGA
| gea2: SWITCH UNTT *THRAt! TO''3,0'', DASERVE 3.00
READING.
6.26 MOVE THE CLIP ON TLL'T, SNAP TO
LAT! SNAP, SWITCH
“RESPIRATION LEAD Seneett TO
Atty OBSFRVE 3.00 40.10.
6.26 MOVE THE CLIP FROM
LAT! TO THE 'TRAT'SNAP, SWITCH
RESPIRATION LEAD Pammmet! TO ''RAT'; OBSERVE 3,00 40.10-
6.30 THIS COMPLETES RESPIRATION CALIORATION, REMOVE ALL
LEADS.
NOTE: THE RESPIRATION WAVE MAY BE OBSERVED ON THE
‘SCOPE CONNECTED TO THE TR2L."'FILYER''OUTPUT. AT
3.088 , IT WILL BE A 300 MY P-P SINE WITH 4.5 TO 44.0
oc OFFSET. :
7.00 2129 BLOOD PRESSURE CALIBRATION
7.01 STATIC PRESSURE ADJUSTMENT -
poz connect pv Sara
: w
EXT METER’ POSTS, DVM ON 10V
RANGE. TR 21 IN-''212"" LOAD MODEr
PLUG TR2t B-P. CABLE INTO THE 2126 STATIC BP JACK.
TURN TR21 FUNCTION SWITCH TO 'TSTATIC 10V
Apuust xv
SeSTATIC 10V"! POTENTIOMETER FOR 10-00 2 SHV.
Switch TR21 FUNCTION TO ''STATIC OUT
+ “switch over
FOR 10uV RESOLUTION.
as 1oVrer ep eo -
SET yovcPl 1 PD ZS
5 DYM te ew)dy
ipee eet
7.06
reat
rete
Pere eeeeeeeeereeee
SwivcH 2128 TO ''40uV TRANSDUCER SENSITIVITY’!
SWITCH FRONT PANGL BP seLEcTaR TO’! RY (100)"', O7HGRTo epee
ADJUST R230 FOR 40.00 HV ON THE OVH.
SwrtcH 2126 TO ''5uV'" TRANSDUCER SENSITIVITY.
ADJUST R234 FOR 5.O00HV.
SNITCH 2128 BACK TO 4OnV, CHECK FOR 40.00HV: READJUST
R230 IF NECESSARY. SWITCH 2128 TO SuV, CHECK FOR
5.000°HVs READJUST R234 IF NECESSARY.
ROTATE THE 2128 OP SELECTOR AND CHECK voLTacEs: Siy
weose (200) = OOo p.15mv
Rv (100) = |5,000¢.01mV
PA (a0) = 4,9008.06MV
Ly (40) = 2,000. 03RV :
AORTA (20) = 1,0002,02mV 6
SWITCH THE 2128 TO 40uV, ROTATE BP SELECTOR ANO CHECK
VOLTAGE
its,
WEDGE (200) = —«5U.60 1.50
RV (300) = 40,00 +-08mV :
PA (80) 32-00 £,60MV
Ly (409 = 16.00 £.30HV
AORTA (209 = 8.00 tetsmy
THIS COMPLETES THE STATIC BP ADJUSTMENT OF THE 2128,
NOTE: THIS SECTION IS PASSIVE: THE 2128 POWER
MAY BE OFF.86/84/1997 89:88 = 7674738655-13.
aerererre ree
2120 DYNAMIC BLOOD PRESSURE CALIGRATION
guITch TR21 TO TTRRESSURES' MaDe, - TTAONVES SENSITIVITY,
212"* LOAD, "'SYSTOLIC'* OISPLAY ON BUILT IN PVM. os?
a
EXTERNAL OVM WILL INDICATE THE DIASTOLIC PRESSURE
civeroom. Reconoect DYm ro TR-21-
CONNECT THE OSCILLOSCOPE VERTICAL INPUT AT 0.2V/DIV
to THE TR21 "IFILTER' OUTPUT JACK. HORIZONTAL SWEEP
40am G7.
ZERO AND CALT@RATE TR21 TO 100MM USING FFSSBUIUEENe,
SHOULD BE SET AT 0-5 SEC/OIV.
CALIBRATOR.
PLUG TR2i @F CABLE INTO THE 2128 M'DYNAMTC’* BP JACK.
SET TOP PANEL SWITCHES ON 2128 FOR 4OvV TRANSOUCER
SENSITIVITY AND BO BPM ECG RATE. SWITCH FRONT PANEL
ep SELECTOR To ''ATM (OPF)'", TURN 2128 MAIN POWER ON.
ZERO THE TR21 TO THE 2128 USING TR21 ‘*ZERO'* CONTROL.
TURN THE 2128 BP SELECTOR TO ‘'RV''. BOTH SECTIONS
1 AND 2 OF THE INTERIOR DIP SWITCH, S13. SHOULD BE ON.
BOTH RZ06 AND RZ0E SHOULO BE SET AT MIDRANGE POSITION.
ADJUST R211 FOR 25.0HM ON THE TR21.
. mv
SwITcH THE 2126 BP SELECTOR "LVS".
Sat 1203 70 120-2
ADJUST RX7S FOR 12g,.09Hm ON THE TAL.
aY
DISCONNECT THE TR21 BP CABLE FROM THE 2128.
OFFSET CURVE: ADJUSTMENT PROCEDURE
IN THE PREVIOUS STEPS R211 WAS SET TO ITS FINAL VALUE.
IF R211 1S READJUSTED THE OFFSET CURVE MUST BE RE-
AosusTED ALSO.Pere rer Preece
i
//
Tae
7.43
7-80
WITH THE SCOPE GRATICULE CENTERLINE. SWITCH OFF GROUND |
OC COUPLING. = |
ALTERNATE R208 AND R206 ADJUSTMENT To GET THE FLATTEST
TRACE POSSIBLE. THE ENTIRE Tract MUST LIE WITHIN 475MV
REMOVE THE SCOPE PROBES AND SWITCH S13 #1 BACK TO 'ON"
RECONNECT THE OSCILLOSCOPE TO THE *
Oui
OF THE TR21. SCOPE CONTROLS AT 0.2V/D1V,0.5 SEC/DIV~
chece TRAM CAL YOuY Box
PLUG THE TR21 GP CABLE INTO THE 212
FILTER’? GUTPUT
19 | DYNAMIC?! BP JACK.
SWITCH THE 2128 TO '*ATM (OFF)''- ZERO THE TR21~
SWITCH THE 2128 TO '*RY! CHECK FOR 25, 0Nme 1mm, (EF MOT, STEP TST
10
OBSERVE PROPER WAVEFORM ON SCOPE, (WAVES ON PAGE >
Oe
SWITCH THE 2128 TO **AOREA'. ADJUST R453 FOR 80.0MML.
E-IMM DIASTOLIC. ADJUST RISL FOR 120, OMME IMM SYSTOLIC.
FEI
OASERVE THE PROPER WAVEFORM ON SCOPE.@6/ed/1997 69:80 — 7574730655-13
eet
qees gwiten THE d128 To "*Ly''. AUST IRA76#ron 120.00miL. LM
SYSTOLIC. DIASTOLIC SHOULD BE LESS THAN 100MM.
OBSERVE PROPER WAVEFORM.
7.66 SWITCH THE 2128 TO 'TPAT'. ADJUST BLE FOR 10.00MML, 11M
DIASTOLIC. ADJUST RISP FOR 25.0 mM. 1NM SYSTOLIC.
oaseRve PROPER WAVEFORM.”
7167 SWITCH THE 2128 TO ''WEDGE'' . ADJUST_AL99 (OFFSET) AND
R125 CAMPLIYUDE) FOR A 10.00MM SYSTOLIC ANO A Z.ctitse, LH
DIASTOLIC. GE
7-68 DISCONNEGT THE TR21 BP CABLE FROM THE 2220. SWITCH
THE TR21 TO SUV SENSTIVITY; ZERO AND CALIQRATE TO
1OONM USING THE BUILT-IN TR21 CALIBRATOR.
7.69 SWITCH THE 2128 TO ''SuV SENSITIVITY, 8P SELECTOR
( SWITCHED TO 'TATM COFFD**. 7
7.70 RECONNECT THE TR21 BP CABLE TO THE 2128 DYNAMIC BP JACK.
ZERO THE TRa1.
Te7L SWITCH THE 24128 TO “AORTA''. ADJUST RZ1S FOR
120,0MM SYSTOLIC, 80-0MM DIASTOLIC BOTH +0.2MM,
7.72 CHECK AORTA WAVE ON SGOPE, THEN CHECK LV, PA, RV AND
WEDGE FOR PROPER WAVEFORMS AND PRESSURES £0-3MM.
7.73 THIS COMPLETES THE 2128 BLOOD PRESSURE CALIBRATION.ere ee