CS series firmware upgrading guidance

Product Type: Chemistry Analyzer


Model: CS series Alarm Code: None
Error Message: None
Editor: Auditor:
Released Date:2015.9.29
Reversion History
Content Date

Method:
Please follow guidance as below strictly.

1) First of all please power on the instrument and close the operation software, then enter “\\Dirui\CS-400

Auto-Chemistry Analyzer” open “Tool” and click to run “upgrade software”.

2) Please input “DiRuiWB” as password in the pop-out interface, then click “Login”

3) After login the interface as picture 2-1 will pop out, first please click “Connect” to connect upgrading
software (if “Connect” button is gray, please check whether the instrument is open or the serial port cable
is connected well), if the upgrading software is on-line successfully there will be a hint “connect ok!” as
picture 2-2 showed.

Picture 2-1 Picture 2-2


4) Please click “file” ,then select the firmware file which you want to upgrade ( or the file we have sent to
you).
5) Click “program button” to start upgrading, you need to wait till the program download completely. When
the interface show “subboard program over” and the data no longer change as picture 2-3 showed that
means the download progress is complete.

Picture 2-3

6) After upgrading complete in order to accomplish upgrading process please power off the instrument and
reboot it after 10 seconds for every upgraded firmware.